RT ELEMENTS, DOSE REVIEW, BRAIN METASTASES, ADAPTIVE HYBRID SURGERY ANALYSIS
K142108 · Brainlab AG · MUJ · Dec 16, 2014 · Radiology
Device Facts
Record ID
K142108
Device Name
RT ELEMENTS, DOSE REVIEW, BRAIN METASTASES, ADAPTIVE HYBRID SURGERY ANALYSIS
Applicant
Brainlab AG
Product Code
MUJ · Radiology
Decision Date
Dec 16, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The RT Elements are applications for radiation treatment planning for use in stereotactic, conformal, computer planned, Linac based radiation treatment of cranial, head and neck, and extracranial lesions
Device Story
RT Elements is a suite of radiation treatment planning software applications; includes Dose Review, Brain Metastases, and Adaptive Hybrid Surgery Analysis. Inputs include DICOM RT data. Dose Review enables visualization of isodose lines, dose-volume histograms (DVHs), dose comparison, and summation. Brain Metastases provides optimized planning and display for cranial multi-metastases. Adaptive Hybrid Surgery Analysis simulates automated template-based radiation treatment plans. Used in hospitals, clinics, and operating rooms by radiation oncologists, medical physicists, and neurosurgeons. Output is used for treatment evaluation in tumor board meetings or surgical settings to assist clinical decision-making and optimize radiation delivery.
Clinical Evidence
No clinical data. Verification and validation were performed via bench testing, including design specification verification and usability testing to ensure workflow and user interface effectiveness. All tests met acceptance criteria.
Technological Characteristics
Software-based radiation treatment planning system. Features DICOM RT data processing, isodose line visualization, DVH calculation, and automated template-based planning. Operates as a standalone software application suite. No specific hardware materials or energy sources defined as it is a software-only device.
Indications for Use
Indicated for radiation treatment planning for stereotactic, conformal, Linac-based radiation therapy of cranial, head and neck, and extracranial lesions. No specific demographic, regional, or cultural limitations for patients; clinical decision for use rests with the medical professional.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or a group of people.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 16, 2014
Brainlab AG % Mr. Alexander Schwiersch Regulatory Affairs Manager Kapellenstrasse 12 85622 Feldkirchen GERMANY
Re: K142108 Trade/Device Name: RT Elements Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: MUJ Dated: November 17, 2014 Received: November 21, 2014
Dear Mr. Schwiersch:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert A. Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K142108
Device Name RT Elements
Indications for Use (Describe)
The RT Elements are applications for radiation treatment planning for use in stereotactic, conformal, Linac based radiation treatment of cranial, head and neck, and extracranial lesions.
The "Dose Review" application as one RT Element contains features for review of isodose lines, review of DVHs. dose comparison and dose summation.
The "Brain Metastases" application as one RT Element provides optimized planning and display for cranial multimetastases radiation treatment planning.
The "Adaptive Hybrid Surgery Analysis" application as one RT Element simulates an automated template-based radiation treatment plan. The simulated plan is intended for treatment evaluation for example in tumor board meetings or operating rooms.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510 (K) SUMMARY RT ELEMENTS
IN ACCORDANCE WITH REQUIREMENTS OF 21 CFR PART 807.92
| Manufacturer: | Brainlab AG<br>Kapellenstrasse 12<br>85622 Feldkirchen<br>Germany<br><br>Phone: +49 89 99 15 68 0<br>Fax: +49 89 99 15 68 33 |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Rainer Birkenbach |
| Contact person: | Alexander Schwiersch |
| Summary date: | 12/5/2014 |
| Device: | RT Elements |
| Trade name: | Elements Dose Review, Elements Brain Metastases, Elements Adaptive<br>Hybrid Surgery Analysis |
| Common/Classification<br>Name: | System,Planning,Radiation Therapy Treatment |
| Main Predicate Device: | iPlan RT (K103246) |
| Secondary Predicate<br>Device: | Not applicable |
| Device classification<br>name: | Medical charged-particle radiation therapy system |
| Regulatory Class: | Class II |
| Regulation Number: | 21 CFR 892.5050 |
| Product Code: | MUJ |
| Intended use: | The RT Elements are applications for radiation treatment planning for use in<br>stereotactic, conformal, computer planned, Linac based radiation treatment of<br>cranial, head and neck, and extracranial lesions |
| Indications for use: | The RT Elements are applications for radiation treatment planning for use in<br>stereotactic, conformal, computer planned, Linac based radiation treatment of<br>cranial, head and neck, and extracranial lesions<br><br>The "Dose Review" application as one RT Element contains features for<br>review of isodose lines, review of DVHs, dose comparison and dose<br>summation. |
| | The "Brain Metastases" application as one RT Element provides optimized<br>planning and display for cranial multi-metastases radiation treatment planning. |
| | The "Adaptive Hybrid Surgery Analysis" application as one RT Element<br>simulates an automated template-based radiation treatment plan. The<br>simulated plan is intended for treatment evaluation for example in tumor board |
{5}------------------------------------------------
meetings or operating rooms.
Device description: The "Dose Review" application as one RT Element contains features for review of isodose lines, review of DVHs, dose comparison and dose summation.
> The "Brain Metastases" application as one RT Element provides optimized planning and display for cranial multi-metastases radiation treatment planning.
> The "Adaptive Hybrid Surgery Analysis" application as one RT Element simulates an automated template-based radiation treatment plan. The simulated plan is intended for treatment evaluation for example in tumor board meetings or operating rooms.
#### Operator Profile
Typical users are medical professionals, including but not limited to radiation oncologists, medical physicists or physicians as well as neurosurgeons (for Adaptive Hybrid Surgery Analysis)
#### Patient Population
There are no demographic, regional or cultural limitations for patients. It is up to the user to decide if the system shall be used to assist a certain procedure.
#### Conditions of use
The system can be used in a hospital environment, in a doctor's office, at the operating theater in case of Adaptive Hybrid Surgery Analysis
- Substantial The RT Elements are part of a new software generation at Brainlab. The equivalence: features provided by the RT Elements including viewing of DICOM RT data, radiation treatment planning for stereotactic radiotherapy treatments is not a new technology for Brainlab. The intended use is considered substantially equivalent to what is possible with iPlan RT as predicate device. The single Elements to not provide the complete features of the predicate device but allow a subset of functionality of the predicate device.
Conclusion: The RT Elements and iPlan RT have identical functionalities in DICOM viewing and identical technical characteristic. The RT functionalities are equivalent to those of iPlan RT.
#### Verification/validation Verification
The verification of the RT Elements has been carried out thoroughly both at the top level and on the underlying subsystems. The verification was done according to verification plan to demonstrate that the design specifications are met.
summary:
{6}------------------------------------------------
#### Non-clinical validation
The validation was done in accordance with the validation plan containing usability tests which should ensure that workflows or user interface result in a useful interface.
All test reports were finally rated as successful according to their acceptance criteria. The non-clinical validation has been performed with software and units that are considered equivalent to the final version of the product, as warranted by 21 CFR 820.30 (g) and which have the UI as planned for the release.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.