K142103 · Teleflexmedical, Inc. · BSK · May 8, 2015 · Anesthesiology
Device Facts
Record ID
K142103
Device Name
CUFF PILOT, SURE SEAL
Applicant
Teleflexmedical, Inc.
Product Code
BSK · Anesthesiology
Decision Date
May 8, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5750
Device Class
Class 2
Attributes
Pediatric
Intended Use
To monitor intra-cuff pressures of supraglottic airways. The SureSeal™ supraglottic airway with attached Cuff Pilot™ is indicated for use in achieving and maintaining control of the airway during routine anesthetic procedures on fasted patients and emergency procedures using either spontaneous or Positive Pressure Ventilation (PPV). - An alternative to a face mask. - An airway device in routine anesthesia procedures. - Securing the immediate airway in anticipated or unexpected difficult airway situations. - Use in elective surgical procedures where tracheal intubation is not necessary. - Establishing an immediate, clear airway during cardiopulmonary resuscitation (CPR) in the profoundly unconscious patient requiring artificial ventilation when tracheal intubation is not possible.
Device Story
Cuff Pilot™ is a mechanical pressure monitoring accessory for supraglottic airways; SureSeal™ is a supraglottic airway with integrated Cuff Pilot™. Device uses internal bellows mechanism; bellows compress with increasing intra-cuff pressure and expand with decreasing pressure. O-ring indicator on bellows displays pressure status via color-coded zones (0-80 cm H2O). Used in hospitals, EMS, extended care, outpatient clinics, and MRI suites; operated by medical professionals. Provides visual pressure feedback to clinicians to assist in maintaining appropriate cuff pressure; helps prevent over-inflation or under-inflation. SureSeal™ is MR Safe; Cuff Pilot™ is disposable, single-patient use.
Clinical Evidence
Bench testing only. Testing included accuracy across pressure ranges, repeatability, real-time aging (3 years), temperature exposure (hot/cold), and drop/shipping tests. Biocompatibility testing (ISO 10993-1) for cytotoxicity, sensitization, irritation, and systemic toxicity confirmed non-reactive status.
Technological Characteristics
Mechanical bellows-based pressure indicator; color-coded visual scale. Materials: non-reactive per ISO 10993-1. MR Safe per ASTM F2052-06 (no magnetic/metal materials). Pressure range: 0-80 cm H2O. Accuracy: +/- 5 cm H2O. Disposable, single-patient use. No electronic components or software.
Indications for Use
Indicated for patients (pediatric to adult) requiring artificial airway management during routine anesthesia, emergency procedures, or CPR. Used for spontaneous or positive pressure ventilation. Contraindicated in patients where tracheal intubation is required but not possible, or in non-fasted patients for routine procedures.
Regulatory Classification
Identification
An inflatable tracheal tube cuff is a device used to provide an airtight seal between a tracheal tube and a patient's trachea.
K182548 — Accucuff Cuff Pressure Indicator · Tianjin Medis Medical Device Co., Ltd. · Jan 25, 2019
K173914 — VORTRAN Cuff Inflator (VCI) · Vortran Medical Technology 1, Inc. · May 25, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure in profile, with three faces overlapping each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 8, 2015
Teleflex Medical c/o Mr. Paul Dryden Consultant 3015 Carrington Mill Blvd. Morrisville, NC 27560
Re: K142103
Trade/Device Name: Cuff Pilot™, SureSeal™ with Cuff Pilot™ Regulation Number: 21 CFR 868.5750 Regulation Name: Inflatable Tracheal Tube Cuff Regulatory Class: II Product Code: BSK, CAE Dated: April 27, 2015 Received: April 28, 2015
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known)
#### K142103
Device Name
## Cuff Pilot™
Indications for Use (Describe)
To monitor intra-cuff pressures of supraglottic airways.
Patient Population:
Patients who have an artificial airway and for which the user would like to monitor cuff pressure, pediatric to adult.
Environments of Use:
To be used under medical supervision in hospital (EMS), extended care facilities and outpatient clinics, where a patient may have an artificial airway. It may also be used in MRI suites when attached to airways that are MR conditional or MR Safe.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
XX Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Expiration Date: January 31, 2017
See PRA Statement below.
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
# K142103
Device Name
# SureSeal™ with Cuff Pilot™
#### Indications for Use (Describe)
The SureSeal™ supraglottic airway with attached Cuff Pilot ™ is indicated for use in achieving and maintaining control of the airway during routine anesthetic procedures on fasted patients and emergency procedures using either spontaneous or Positive Pressure Ventilation (PPV).
- An alternative to a face mask.
- . An airway device in routine anesthesia procedures.
- Securing the immediate airway in anticipated or unexpected difficult airway situations.
- Use in elective surgical procedures where tracheal intubation is not necessary.
- Establishing an immediate, clear airway during cardiopulmonary resuscitation (CPR) in the profoundly ● unconscious patient requiring artificial ventilation when tracheal intubation is not possible.
Patient Population: Patients who need an artificial airway, pediatric to adult.
Environments of Use:
To be used under medical supervision in hospital (EMS), extended care facilities and outpatient clinics, where a patient needs an artificial airway. It may also be used in MRI suites and is considered MRI Safe.
Type of Use (Select one or both, as applicable)
XX Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary Page 1 of 7 8-May-15
| Teleflex Medical<br>3015 Carrington Mill Blvd<br>Morrisville, NC 27560 | Telephone: 919-361-3934<br>Fax: 919-433-4996 |
|------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact: | Justin Lovelace<br>Regulatory Affairs Specialist |
| Proprietary or Trade Name: | Cuff Pilot™ |
| Common/Usual Name: | Cuff Pilot™<br>SureSeal™ with Cuff Pilot™ |
| Classification Name: | Cuff, tracheal tube, inflatable (accessory<br>BSK – 21 CFR 868.5750, Class II<br>Airway, oropharyngeal<br>CAE – 21 CFR 868.5110, Class I |
| Predicate Devices: | Easy Cuff – K102704<br>LMA™ Classic – K130304 |
#### Device Description:
The Cuff Pilot™ is a simple device which allows the user to monitor the cuff pressure of supraglottic airways. It has been designed to display pressure ranges via color coded zones. The Cuff Pilot™ is designed with an internal bellows, is compressed as pressure increases and expands when intra-cuff pressure is less. There is an O-ring indicator that is on the bellows and this indicates the status of pressure within the define color coded pressure zones.
The SureSeal™ supraglottic airway is very similar in design, performance, indications for use, technology of operation, and materials to our LMA airways. It will have the Cuff Pilot™ attached and is considered MR Safe.
# Indications for Use - Cuff Pilot™
To monitor intra-cuff pressures of supraglottic airways.
Patient Population: Patients who have an artificial airway and for which the user would like to monitor cuff pressure, pediatric to adult.
Environments of Use: To be used under medical supervision in hospitals, pre-hospital (EMS), extended care facilities and outpatient clinics, where a patient may have an artificial airway. It may also be used in MRI suites when attached to airways that are MR conditional or MR Safe.
# Indications for Use - SureSeal™ with Cuff Pilot™
The SureSeal™ supraglottic airway with attached Cuff Pilot™ is indicated for use in achieving and
{5}------------------------------------------------
# 510(k) Summary Page 2 of 7 8-May-15
maintaining control of the airway during routine anesthetic procedures on fasted patients and emergency procedures using either spontaneous or Positive Pressure Ventilation (PPV).
- An alternative to a face mask. .
- An airway device in routine anesthesia procedures.
- Securing the immediate airway in anticipated or unexpected difficult airway situations. .
- . Use in elective surgical procedures where tracheal intubation is not necessary.
- . Establishing an immediate, clear airway during cardiopulmonary resuscitation (CPR) in the profoundly unconscious patient requiring artificial ventilation when tracheal intubation is not possible.
Patient Population: Patients who need an artificial airway
Environments of Use: To be used under medical supervision in hospitals, pre-hospital (EMS), extended care facilities and outpatient clinics, where a patient needs an artificial airway. It may also be used in MRI suites and is considered MRI Safe.
| Attribute | Predicate<br>Easy Cuff (K102704) | Proposed Device<br>Cuff Pilot™ |
|---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | To inflate cuffs and to measure and<br>monitor intra-cuff pressures of<br>endotracheal, supraglottic airways, or<br>tracheostomy tubes. | To monitor intra-cuff pressures of<br>supraglottic airways. |
| Environments of use | To be used under medical supervision in<br>hospitals, pre-hospital (EMS), extended<br>care facilities and outpatient clinics,<br>where a patient may be intubated. | To be used under medical supervision in<br>hospitals, pre-hospital (EMS), extended care<br>facilities and outpatient clinics, where a<br>patient may have an artificial airway. It may<br>also be used in MRI suites when attached to<br>airways that are MR conditional or MR Safe. |
| Patient population | Intubated patients | Patients who have an artificial airway and for<br>which the user would like to monitor cuff<br>pressure, pediatric to adult. |
| Technology | Bellows that move with changes in<br>pressure<br>Expands with higher pressures<br>Contracts with lower pressures | Bellows that move with changes in pressure<br>Contracts with higher pressures<br>Expands with lower pressures |
| Method of inflating cuff | Use the integrated syringe<br>Manual | Use an independent syringe<br>(normal practice)<br>Manual |
| Attaches to the Cuff Inflation<br>Pilot | Yes via a luer fitting | Yes via a luer fitting<br>May be permanently attached to an airway's<br>cuff inflation line |
| Types of airways to which it<br>can be used | Supraglottic airway<br>Endotracheal tube<br>Tracheostomy tube | Supraglottic airway |
| Single patient, disposable | Yes | Yes |
| Pressure Range of the device | 0 to 60 cm H₂O | 0 to 80 cm H₂O |
{6}------------------------------------------------
# 510(k) Summary Page 3 of 7 8-May-15
| Attribute | Predicate<br>Easy Cuff (K102704) | Proposed Device<br>Cuff Pilot™ |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Detection of "good range" | Color coded and a scale | Color coded zones |
| Packaged | Non-sterile | Non-sterile<br>Sterile |
| Accuracy | +/- 1 cmH2O up to 30 cmH2O<br>+/- 2 cmH2O at 40 cmH2O<br>+/- 5 cmH2O at 60 cmH2O | +/- 5 cmH2O up to 80 cmH2O |
| MRI Use | Not labeled | MR Safe based upon the lack of<br>materials which contain metal or are<br>magnetic per ASTM F2052-06,<br>rationale |
| Shelf-life | Not provided | 3 years |
| Performance Testing | Accuracy across the pressure range<br>Repeatability<br>Tested for accuracy after exposure to high<br>and low temperatures | Accuracy within the pressure zone<br>ranges<br>Repeatability<br>Subjected samples to aging and<br>exposure to cold and hot temperatures |
| Drop test | | Drop / Shipping test |
# Substantial Equivalence Discussion:
The Cuff Pilot™ is viewed as substantially equivalent to the predicate devices because:
#### Indications -
The Cuff Pilot™ is intended to monitor the intra-cuff pressure of Supraglottic airways. Discussion - The predicate, EasyCuff™ (K102704) is designed to inflate the cuff instead of requiring a separate syringe, which is the normal practice. This difference does not raise any new safety or effectiveness concerns, thus the proposed device is considered substantially equivalent. The proposed Cuff Pilot™ is intended to monitor changes in cuff pressure and display the pressure within a range or zone. Actual pressure will be set or checked with a separate cuff pressure gauge.
# Environment of Use -
The Cuff Pilot™ has the same environment of use as the predicate plus MRI suites. Discussion - The Cuff Pilot™ has no magnetic materials and thus can be considered MR Safe. It can be used with any artificial airway that is deemed to be MR Conditional or MR Safe. The predicate is likely MR Conditional but it was not an indication in the original submission. This difference does not raise any new safety or effectiveness concerns that have not been addressed in the rationale and thus the proposed device can be considered substantially equivalent.
#### Patient Population -
The Cuff Pilot™ has the same patient population as the predicate.
Discussion - Since patients who have a supraglottic airway are not technically considered "intubated"; the predicate, EasyCuff™ (K102704) term in not correct. This is no different than the predicate and thus the proposed device is considered substantially equivalent.
{7}------------------------------------------------
## 510(k) Summary Page 4 of 7 8-May-15
#### Technology -
The Cuff Pilot™ uses a bellows which moves in relationship to the intra-cuff pressure. As pressures increases the bellows collapses.
Discussion - The use of a bellows which moves in relationship to the intra-cuff pressure is identical to the predicate EasyCuff™ (K102704). The only differences is that for the proposed device the bellows compresses with increasing pressure while the predicate EasyCuff™ bellows expands with increasing pressure. Each device has color coded zones, which represent pressures, which informs the user of the relative cuff pressure within the zone. Any difference does not raise any new safety or effectiveness concerns, thus the proposed device is considered substantially equivalent.
#### Summary of Non-clinical Testing
The following is a summary of non-clinical bench testing.
#### Materials -
There are no materials of the Cuff Pilot™ in contact with the patient or in the gas pathway. Discussion - We have performed ISO 10993-1 cytotoxicity testing on the Cuff Pilot™. The results were non-reactive.
#### Performance -
We have performed a number of tests to demonstrate the accuracy and repeatability of the measured pressure for the Cuff Pilot™. The performance is similar to the predicate. EasyCuff™ (K102704).
The tests performed are:
- Real-time age testing - 3 years
- Accuracy across the pressure zone / range
- Repeatability
- Subjected samples to aging and exposure to cold and hot temperatures ●
- Drop / Shipping test ●
Discussion - The performance specifications of the Cuff Pilot™ and the predicate are substantially equivalent.
# Substantial Equivalence Conclusion -
The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicate have been found to be substantially equivalent.
# SureSeal™ with Cuff Pilot™
SureSeal™ is only offered with the Cuff Pilot™ is standard Supraglottic airway similar in design and function to the LMA Classic™. It is considered a Class 1 exempt device but we are seeking to have the Cuff Pilot™ attached as well as seeking MR Safe environment of use.
{8}------------------------------------------------
# 510(k) Summary
| Attribute | Predicate | Proposed Device |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | LMA Classic™ (K130304) | SureSeal™ with Cuff Pilot™ |
| Classification | Procode -CAE - airway, oropharyngeal<br>CFR – 868.5110<br>Class 1 - exempt | Procode -CAE – airway, oropharyngeal<br>CFR – 868.5110<br>Class 1 - exempt |
| Indications for Use | LMA Classic™ is indicated for use in<br>achieving and maintaining control of the<br>airway during routine anesthetic procedures<br>on fasted patients and emergency procedures<br>using either spontaneous or Positive Pressure<br>Ventilation (PPV).<br>• An alternative to a face mask.<br>• An airway device in routine anesthesia<br>procedures.<br>• Securing the immediate airway in<br>anticipated or unexpected difficult airway<br>situations.<br>• Use in elective surgical procedures where<br>tracheal intubation is not necessary.<br>• Establishing an immediate, clear airway<br>during cardiopulmonary resuscitation (CPR)<br>in the profoundly unconscious patient<br>requiring artificial ventilation when tracheal<br>intubation is not possible. | The SureSeal™ supraglottic airway is<br>indicated for use in achieving and<br>maintaining control of the airway during<br>routine anesthetic procedures on fasted<br>patients and emergency procedures using<br>either spontaneous or Positive Pressure<br>Ventilation (PPV).<br>• An alternative to a face mask.<br>• An airway device in routine anesthesia<br>procedures.<br>• Securing the immediate airway in<br>anticipated or unexpected difficult airway<br>situations.<br>• Use in elective surgical procedures where<br>tracheal intubation is not necessary.<br>• Establishing an immediate, clear airway<br>during cardiopulmonary resuscitation (CPR)<br>in the profoundly unconscious patient<br>requiring artificial ventilation when tracheal<br>intubation is not possible. |
| Patient population | Patients who need an artificial airway | Patients who need an artificial airway |
| Environments of use | To be used under medical supervision in<br>hospitals, pre-hospital (EMS), extended care<br>facilities and outpatient clinics, where a<br>patient needs an artificial airway. It may also<br>be used in MR suites as MR Conditional. | To be used under medical supervision in<br>hospitals, pre-hospital (EMS), extended care<br>facilities and outpatient clinics, where a<br>patient needs an artificial airway. It may<br>also be used in MRI suites when attached to<br>airways that are MR Safe. |
| Technology | Inflatable cuff and tube that rest above the<br>glottis opening | Inflatable cuff and tube that rest above the<br>glottis opening |
| Method of inflating cuff | Use integrated pilot check valve with a<br>separate syringe. Manual | Used with Cuff Pilot™ and<br>a separate syringe. Manual |
| Available in sizes | Sizes 1 to 6 | Sizes 1 to 6 |
| Single patient,<br>disposable | Yes | Yes |
| Attribute | Predicate | Proposed Device |
| | LMA ClassicTM (K130304) | SureSealTM with Cuff PilotTM |
| MRI Use | MR Conditional | MR Safe based upon the lack of materials<br>which contain metal or are magnetic per<br>ASTM F2052-06, rationale |
| Biocompatibility | Surface and Externally communicating<br>contact<br>(indirect gas pathway)<br>Tissue / Mucosa contact<br>Limited duration | Surface and Externally communicating<br>contact<br>(indirect gas pathway)<br>Tissue / Mucosa contact<br>Limited duration |
| Shelf-life | 3 years | 3 years |
| Performance Testing | | Age<br>Hot / Cold temperature<br>Drop / shipping |
{9}------------------------------------------------
#### 510(k) Summary Page 6 of 7 8-May-15
#### Substantial Equivalence Discussion:
The SureSeal™ is viewed as substantially equivalent to the predicate device because:
#### Indications -
The SureSeal™ is indicated for use in achieving and maintaining control of the airway during routine anesthetic procedures on fasted patients and emergency procedures using either spontaneous or Positive Pressure Ventilation (PPV).
Discussion - The predicate, LMA Classic™ (K130304) is also designed to achieve and maintain and airway. There are no differences, thus there are no new safety or effectiveness concerns raised, thus the proposed device is considered substantially equivalent.
#### Environment of Use -
The SureSeal™ has the same environment of use as the predicate including use in MRI suites. Discussion - The SureSeal™ and the Cuff Pilot™ have no magnetic materials and thus can be considered MR Safe. The predicate is MR Conditional because the inflation check valve does contain magnetic materials. This difference does not raise any new safety or effectiveness concerns that have not been addressed in the rationale and thus the proposed device can be considered substantially equivalent.
#### Patient Population -
The SureSeal™ has the same patient population as the predicate.
Discussion - There are no differences between the proposed device, thus it can be considered substantially equivalent.
#### Technology -
The SureSeal™M employs the same technology and principle of operation as the predicate. Discussion - The design and technology is similar between the devices and there are no differences which would raise any new safety or effectiveness concerns, thus the proposed device is considered substantially equivalent.
{10}------------------------------------------------
#### 510(k) Summary Page 7 of 7 8-May-15
# Materials -
The materials are similar between the devices.
Discussion - We have performed ISO 10993-1 testing for the materials for the SureSeal™. Based upon ISO 10993-1 and G95-1 the SureSeal™ would be considered as:
- . Surface Contact and Externally communicating (indirect gas pathway)
- Tissue / Mucosa contact
- Limited duration (<24 hours) .
This is identical to the predicate LMA Classic™ (K130304).
# Performance -
We have performed a number of tests on the SureSeal™ including age / shelf-life, hot and cold temperatures and drop /shipping testing.
Discussion - The SureSeal™ met it specifications after these tests.
#### Summary of Non-clinical Testing
The following is a summary of non-clinical bench testing.
#### Materials -
We have performed ISO 10993-1 testing for the materials of the SureSeal™.
- Cytotoxicity
- Sensitization ●
- Irritation ●
- Systemic Toxicity
# Performance -
We have performed a number of tests which included:
- Real-time age testing - 3 years
- Subjected samples to aging and exposure to cold and hot temperatures .
- Drop / Shipping test ●
Results - The SureSeal™ met its performance specifications.
# Substantial Equivalence Conclusion -
The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicate have been found to be substantially equivalent.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.