K142097 · Biomedix, Inc. · FPA · Aug 29, 2014 · General Hospital
Device Facts
Record ID
K142097
Device Name
SELEC-3 I.V.ADMINISTRATION SET
Applicant
Biomedix, Inc.
Product Code
FPA · General Hospital
Decision Date
Aug 29, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The SELEC-3 I.V. Administration Set is used to provide a pathway to deliver fluid into the vascular system. This product is targeted for use by nurses and paramedics in pre-hospital, emergency room and field settings.
Device Story
Single-use, sterile IV administration set; provides fluid pathway to vascular system. Features selectable drop chamber (10, 15, or 60 drops/cc) and Y-sites for medication administration. Used by nurses and paramedics in pre-hospital, emergency room, and field settings. Device facilitates controlled fluid delivery to patients; manual operation by clinician.
Clinical Evidence
Bench testing only. Testing included flow rate verification, leakage, microbial ingress, and pull force testing. Biocompatibility assessed per ISO 10993-1:2009. Sterility validated per ANSI/AAMI/ISO 11137-2:2012.
Technological Characteristics
Constructed of biocompatible plastics. Features selectable drop chamber (10, 15, 60 drops/cc) and Y-sites. Gamma sterilized per ANSI/AAMI/ISO 11137-2:2012. Externally communicating, indirect blood path, prolonged duration device.
Indications for Use
Indicated for delivery of fluids and soluble drugs into the vascular system. Targeted for use by nurses and paramedics in pre-hospital, emergency room, and field settings.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
K050906 — VITALCARE I.V. ADMINISTRATION SET · Vitalcare Group, Inc. · Aug 23, 2005
K991932 — INTRAVASCULAR ADMINISTRATION SET · The Kipp Group · Aug 20, 1999
K191640 — TK Intravascular Administration Set · Anhui Tiankang Medical Technology Co., Ltd. · Aug 26, 2020
K223645 — I.V. Administration Set, I.V. Extension Set · Bq Plus Medical Co., Ltd. · May 18, 2023
K121511 — MACLIN POWER INTRAVASCULAR ADMINISTRATION SET · Maclin Power, Inc. · Mar 7, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of three human profiles facing right, stacked on top of each other. The profiles are stylized and appear to be connected, forming a single, unified shape.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 29, 2014
Biomedix, Inc. Ms. Myra J. Bender President 3895 West Vernal Pike Bloomington, IN 47404
Re: K142097
Trade/Device Name: Biomedix SELEC-3 I.V. Administration Set Regulation Number: 21 CFR 888.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: July 31, 2014 Received: August 1, 2014
Dear Ms. Bender:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Erin | Keith -S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K142097
Device Name
Biomedix SELEC-3 I.V. Administration Set
#### Indications for Use (Describe)
The SELEC-3 I.V. Administration Set is used to provide a pathway to deliver fluid into the vascular system. This product is targeted for use by nurses and paramedics in pre-hospital, emergency room and field settings.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for BIOMEDIX, Inc. The text is in a bold, sans-serif font with horizontal lines running through each letter. The word "BIOMEDIX" is stacked on top of the word "quality medical products".
# quality medical products
# SPECIAL 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92
| Submitter/Contact Information | |
|---------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Biomedix, Inc. |
| Address | 3895 West Vernal Pike<br>Bloomington, IN 47404 |
| Phone number | (812) 355-7000 |
| Fax number | (812) 355-4507 |
| Establishment Registration<br>Number | 1833470 |
| Name of contact person | Myra J. Bender |
| Date prepared | July 31, 2014 |
| Name of device | |
| Trade or proprietary<br>name | Biomedix SELEC-3 I.V. Administration Set |
| Common or usual name | Intravascular Administration Set |
| Classification name / Regulation<br>number | Set, Administration, Intravascular (21 CFR § 888.5440) |
| Classification panel | General Hospital |
| Product Code(s) | FPA |
| Legally marketed device(s) to<br>which equivalence is claimed | SELEC-3 I.V Administration Set (K925645)<br>Biomedix, Inc. |
| Reason for 510(k) submission | Device modification |
| Device description | The Biomedix SELEC-3 I.V. Administration Set is a single-<br>use, sterile device used for the administration of solutions<br>and fully soluble drugs into the vascular system of a patient. |
| Indications for use | The SELEC-3 I.V. Administration Set is used to provide a<br>pathway to deliver fluid into the body via the vascular<br>system. This product is targeted for use by nurses and<br>paramedics in pre-hospital, emergency room and field<br>settings. |
{4}------------------------------------------------
# Summary of Technological Characteristics
The SELEC-3 I.V. Administration Set is constructed of biocompatible plastics. It incorporates a Selectable Drop Chamber that allows the user to choose between a flow rate of 10, 15, or 60 drops/cc at any time. Also included are Y-sites that allow the user to add medication. The intended use, basic design, and function are identical to the predicate device.
## Testing Summary
The SELEC-3 I.V. Administration Set has met Biomedix testing and acceptance criteria related to flow rate verification, leakage, microbial ingress, and pull force testing.
## Biocompatibility
Biocompatibility has been assessed in accordance with ISO 10993-1:2009, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing. The SELEC-3 I.V. Administration Set has been categorized as an externally communicating device, indirect blood path, prolonged duration. Based on testing applicable to this categorization, biocompatibility has been demonstrated.
## Sterility
SELEC-3 I.V. Administration Sets are provided as sterile, single use devices that are individually packaged. The device is gamma sterilized through a validated process with the dose set in conformance with ANSI/AAMI/ISO 11137-2: 2012, "Sterilization of health care products-Radiation- Part 2: Establishing the Sterilization Dose," Method-VDmax25.
### Bacterial Endotoxins
The SELEC-3 I.V. Administration Set is monitored for bacterial endotoxins.
### Conclusion
The SELEC-3 I.V. Administration Set met all established acceptance criteria for design verification testing. Testing demonstrated that the device is safe and effective when used as intended.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.