K142076 · Dickson & Dickson Healthcare (Us), Inc. · FSY · Sep 30, 2014 · General, Plastic Surgery
Device Facts
Record ID
K142076
Device Name
CLAVEGUARD FREEDOM SURGICAL LIGHT
Applicant
Dickson & Dickson Healthcare (Us), Inc.
Product Code
FSY · General, Plastic Surgery
Decision Date
Sep 30, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4580
Device Class
Class 2
Indications for Use
The Claveguard Freedom Surgical Light is designed to illuminate the surgical site in Operating theatres
Device Story
ClaveGuard Freedom Surgical Light provides illumination for surgical sites in operating rooms. System utilizes Light Emitting Diodes (LEDs) with adaptive focus and shadow-free illumination capabilities. Operates at distances of 70-140cm from surgical site. Configuration includes 1 to 3 light heads mounted on a ceiling-pivoting suspension system; optional components include monitor extension arms, monitor mounts, and in-light video camera. Operated by surgical staff to provide clear visualization of the operative field, facilitating surgical decision-making and procedural accuracy.
Clinical Evidence
Bench testing only. Device conforms to IEC 60601-2-41:2001, IEC 60601-1:2005, IEC 60601-2-41:2009, IEC 60601-1-2:2007, and IEC 60601-1-4:2000. No clinical data utilized.
Technological Characteristics
LED-based surgical luminaire; ceiling-mounted pivoting suspension system; adaptive focus; shadow-free illumination. Operates at 70-140cm distance. Conforms to IEC 60601-1, IEC 60601-2-41, IEC 60601-1-2, and IEC 60601-1-4 standards.
Indications for Use
Indicated for illumination of surgical sites in operating theatres for patients undergoing surgical procedures.
Regulatory Classification
Identification
A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 30, 2014
Dickson & Dickson Healthcare (US) Incorporated % Ragunath Muniandy Black Diamond Video Incorporated 503 Canal, Boulevard Richmond, California 94804
Re: K142076
Trade/Device Name: ClaveGuard Freedom Surgical Light Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical lamp Regulatory Class: Class II Product Code: FSY Dated: May 30, 2014 Received: August 4, 2014
Dear Muniandy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{2}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Binita S. Ashar -S 2014.09.30 17:27:22 -04'00'
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142076
Device Name
ClaveGuard Freedom Surgical Light
#### Indications for Use (Describe)
The Claveguard Freedom Surgical Light is designed to illuminate the surgical site in Operating theatres
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for ClaveGuard. The logo features a stylized blue letter 'G' with an arrow incorporated into its design. Below the 'G' is the word "CLAVEGUARD" in a larger, bolder font, also in blue. Underneath "CLAVEGUARD" is the tagline "SAFE | STERILE | SUSTAINABLE" in a smaller font.
Image /page/3/Picture/1 description: The image shows three different logos. The first logo is a red and white logo with the text "AS/NZS ISO 9001:2008 Quality Management System Certified Registration Number 159". The second logo is a blue and white logo with the text "JAS-ANZ". The third logo is a green and yellow logo with a kangaroo and the text "AUSTRALIAN OWNED".
ClaveGuard Pty Ltd www.claveguard.com Unit 3, 40 Brodie St. Rydalmere NSW 2116 Australia Phone (+61 2) 8845 3800 Fax (+61 2) 9999 1441
# 510(K) Summary
Date Prepared: 05/30/2014
| 510K Owner: | ClaveGuard Pty, Ltd<br>Unit 3, 40 Brodie St, Rydalmere,<br>NSW 2116, Australia. |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Ragunath Muniandy<br>Regulatory Affairs Specialist<br>503 Canal Blvd.,<br>Richmond, CA 94804 |
| Trade Name: | ClaveGuard Freedom surgical light |
| Common Name: | Surgical Light |
| Classification Name: | Light, Surgical Accessories |
| Classification Panel: | General and Plastic Surgery |
| CFR section: | 21 CFR 878.4580 |
| Class: | 2 |
| Product Code: | FSY |
| Predicate Device: | Berchtold Chromophare E668 (K090378) |
| Device Description: | The ClaveGuard Freedom surgical lights are suitable for all<br>types of surgical procedures in operating rooms. The light is<br>based on Light Emitting Diodes (LEDs) in combination with<br>adaptive focus and shadow free illumination. It operates at a<br>distance of 70-140cm from the surgical site.<br>A light system may consist of 1 to 3 light heads supported by<br>pivoting suspension system that is mounted to Healthcare<br>facility's ceiling. The system may include optional monitor<br>extension arms, monitor mounts, and in-light video camera. |
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for ClaveGuard. The logo features a stylized letter "G" in blue, with an arrow incorporated into its design. Below the "G" is the word "CLAVEGUARD" in a larger, bolder blue font. Underneath "CLAVEGUARD" is the tagline "SAFE | STERILE | SUSTAINABLE" in a smaller font.
Image /page/4/Picture/1 description: The image shows three different logos. The first logo is a red and white logo with the text "GLOBAL CERTIFICATION PTY LTD" at the top and "AS/NZS ISO 9001:2008 Quality Management System Certified Registration Number 159" at the bottom. The second logo is a blue and white logo with the text "JAS-ANZ" at the top. The third logo is a green and yellow logo with a kangaroo in the center and the text "AUSTRALIAN OWNED" around the edge.
ClaveGuard Pty Ltd www.claveguard.com Unit 3, 40 Brodie St. Rydalmere NSW 2116 Australia Phone (+61 2) 8845 3800 Fax (+61 2) 9999 1441
Indications for use: The ClaveGuard Freedom surgical light is designed to illuminate the surgical site in Operating theatres.
### Substantial Equivalence
| (SE) Rational: | The ClaveGuard Freedom surgical light is substantially equivalent to predicate devices since intended use, operational principle, basic technology and design are similar. The minor differences between the ClaveGuard Freedom surgical lights and predicate devices raise no new issues of safety and effectiveness, as these differences have no effect on the performance, function or general intended of use of the device. Therefore based on the applicable testing and the equivalence information presented in this submission, Claveguard Pty Ltd. believes that ClaveGuard Freedom surgical light does not raise any new safety or efficacy issues. |
|----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
### Summary of Nonclinical
| Testing: | - This device conforms to IEC 60601-2-41:2001 specifications<br>for performance of surgical lamps. Testing was conducted in<br>accordance with this standard to support substantial<br>equivalence. The ClaveGuard Freedom surgical light meets all<br>requirements of the performance standard and achieved<br>specified desired values per the product design specifications.<br>Clinical testing was not required or utilized to support<br>substantial equivalence |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of Safety<br>and Effectiveness: | Testing and evaluation indicate that the system meets the needs<br>of the users of the device and does not raise any new safety and<br>efficacy of the predicate device.<br>The ClaveGuard Freedom surgical light conforms to the<br>IEC 60601-1:2005 + CORR. 1:2006 + CORR. 2:2007 +<br>AM1:2012 “Medical Electrical Equipment- Part 1 General<br>requirement for basic safety and essential performance. IEC 60601-2-41:2009 Medical Electrical Equipment- Part<br>2-41: Particular requirement for basic safety and essential<br>performance of surgical luminaires for diagnosis. IEC 60601-1-2:2007 “Medical Electrical Equipment- Part<br>1-2 General requirement for basic safety and essential<br>performance- Collateral Standard: Electromagnetic<br>compatibility- requirement and test. |
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for ClaveGuard. The logo features a stylized letter 'G' in blue, with an arrow incorporated into its design. Below the 'G', the word 'ClaveGuard' is written in a bold, sans-serif font, also in blue. Underneath 'ClaveGuard', the words 'SAFE STERILE SUSTAINABLE' are written in a smaller, thinner font, separated by vertical lines.
Image /page/5/Picture/1 description: The image shows three different logos. The first logo is a red and white logo with the text "communication pty ltd" at the top and "AS/NZS ISO 9001:2008 Quality Management System Certified Accreditation Number 155" at the bottom. The second logo is a blue and white logo with the text "JAS-ANZ" at the top. The third logo is a green and yellow logo with a kangaroo in the center and the text "AUSTRALIAN OWNED" around the edge.
ClaveGuard Pty Ltd www.claveguard.com Unit 3, 40 Brodie St. Rydalmere NSW 2116 Australia Phone (+61 2) 8845 3800 Fax (+61 2) 9999 1441
- 4. IEC 60601-1-4:2000 "Medical Electrical Equipment- Part 1-4 General requirement for safety- Collateral Standard: Programmable Electrical Medical System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.