XN CAL

K141962 · Streck · KRX · Dec 5, 2014 · Hematology

Device Facts

Record IDK141962
Device NameXN CAL
ApplicantStreck
Product CodeKRX · Hematology
Decision DateDec 5, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.8150
Device ClassClass 2

Intended Use

XN CAL is used for the calibration and calibration verification of Sysmex XN series (XN-10, XN-11, XN-20, XN-21) analyzers. Assayed parameters include: WBC (10³/μL), RBC (10⁶/μL), HGB (g/dL), HCT (%), PLT (10³/μL), and RET (%).

Device Story

XN CAL is an in-vitro diagnostic calibrator designed to substitute for fresh whole blood in Sysmex XN series hematology analyzers. It consists of stabilized red blood cells, white blood cells, platelets, and nucleated red blood cells in a preservative medium. Packaged in polypropylene vials, it is used by laboratory personnel to calibrate and verify the accuracy of WBC, RBC, HGB, HCT, PLT, and RET parameters. The device ensures analyzer performance remains within specified limits, supporting accurate clinical hematology testing. It is stored at 2-8°C and has a 49-day closed-vial stability.

Clinical Evidence

Bench testing only. Studies included open-vial stability (4 hours), closed-vial stability (49 days), and precision performance. Data demonstrated consistent reproducibility and stability for the entire product dating, supporting substantial equivalence.

Technological Characteristics

Stabilized human blood components (RBC, WBC, PLT, nucleated RBC) in preservative medium. Polypropylene vials with screw caps. Calibrated to whole blood per CLSI H26-A2, H7-A3, H15-A3, and ICSH guidelines. Storage 2-8°C. No electronic or software components; purely a chemical/biological calibrator.

Indications for Use

Indicated for calibration and calibration verification of Sysmex XN series (XN-10, XN-11, XN-20, XN-21) hematology analyzers. Used by clinical laboratory professionals.

Regulatory Classification

Identification

A calibrator for cell indices is a device that approximates whole blood or certain blood cells and that is used to set an instrument intended to measure mean cell volume (MCV), mean corpuscular hemoglobin (MCH), and mean corpuscular hemoglobin concentration (MCHC), or other cell indices. It is a suspension of particles or cells whose size, shape, concentration, and other characteristics have been precisely and accurately determined.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. The logo is black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 5, 2014 STRECK INC C/O MS. DEBORAH KIPP REGULATORY AFFAIRS MANGER 7002 SOUTH 109TH STREET OMAHA, NE 68128 Re: K141962 Trade/Device Name: XN CALTM Regulation Number: 21 CFR 864.8150 Regulation Name: Calibrator for cell indices Regulatory Class: Class II Product Code: KRX Dated: October 27, 2014 Received: October 28, 2014 Dear Ms. Kipp: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. # Maria M. Chan -S Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) k141962 Device Name XN CAL Indications for Use (Describe) XN CAL is used for the calibration and calibration verification of Sysmex XN series (XN-10, XN-11, XN-20, XN-21) analyzers. Assayed parameters include: WBC (10^3/uL), RBC (10^6/uL), HGB (g/dL), HCT (%), PLT (10^3/ uL), and RET (%) Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(k) Summary | 510(k) Submitter: | Streck<br>7002 South 109th Street<br>La Vista, NE 68128 | |-------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------| | Official Correspondent:<br>Address:<br>Phone:<br>Fax:<br>Date Prepared: | Deborah Kipp, Regulatory Affairs Manager<br>7002 South 109th Street; La Vista, NE 68128<br>402-537-5215<br>402-537-5317<br>April 18, 2014 | | Names<br>Trade Name | XN-CALTM | Trade Name: Common Name: Classification Name: Product Code: Panel: XN CAL Assayed Hematology Calibrator Calibrator for Cell Indices (864.8150) KRX Hematology #### Predicate Device: XN CAL™ (K120745) #### Intended Use: XN CAL is used for the calibration and calibration verification of Sysmex XN series (XN-10, XN-11, XN-20, XN-21) analyzers. Assayed parameters include: WBC (103/pL), RBC (108/μL), HGB (g/dL), HCT (%), PLT (103/ μL), and RET (%) {4}------------------------------------------------ #### Description: Per the FDA guidance document, The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications (510(k)), a predicate device was selected in order to demonstrate substantial equivalence for XN CAL. The comparison to the "Comparison to Predicate Device" section. #### Comparison to Predicate Device | | XN CAL™ (K120745)-Predicate Device | XN CAL™ -Candidate Device | Same or Differences | |------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------| | Intended<br>Use<br>Statement | XN CAL is used for the calibration and<br>calibration verification of Sysmex<br>XN series (XN-10, XN-20) analyzers.<br>Assayed parameters include:<br>WBC ( $10^3$ /μL), RBC ( $10^6$ /μL),<br>HGB (g/dL), HCT (%), PLT ( $10^3$ / µL), and<br>RET (%) | XN CAL is used for the calibration and<br>calibration verification of Sysmex<br>XN series (XN-10, XN-11, XN-20, XN-21)<br>analyzers. Assayed parameters include:<br>WBC ( $10^3$ /μL), RBC ( $10^6$ /μL),<br>HGB (g/dL), HCT (%), PLT ( $10^3$ / µL), and<br>RET (%) | Addition of the XN-11 and<br>XN-21 analyzers. | | Open Vial<br>Stability | 4 hours | 4 hours | Same | | Closed Vial<br>Stability | 35 days | 49 days | Extension of Closed Vial<br>Stability Dating to 49<br>days | | Reagents | XN CAL contains the following: stabilized<br>red blood cell component(s), stabilized<br>white blood cell component(s), stabilized<br>platelet component(s), and stabilized<br>nucleated red blood cell component(s) in<br>a preservative medium. | XN CAL contains the following: stabilized red<br>blood cell component(s), stabilized white<br>blood cell component(s), stabilized platelet<br>component(s), and stabilized nucleated red<br>blood cell component(s) in a preservative<br>medium. | Same | | Storage<br>Conditions | 2 - 8°C | 2 - 8°C | Same | {5}------------------------------------------------ #### Discussion of Tests and Test Results: To substantiate the product performance claims for XN CAL, Streck collected product performance data for the following studies Open-Vial Stability, Closed-Vial Stability, and Precision Performance. The resultant data set established that XN CAL is safe and effective for its intended use and that the product is stable for the entire product dating. The product fulfills its intended use as instructed in the Instructions for Use. #### Conclusions Drawn From Tests: Study results show XN CAL to be consistently reproducible, substantially equivalent to the products, and stable for the entire product dating. XN CAL is a safe and effective product, which fulfills its intended use when used as instructed in the Instructions for Use.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...