← Product Code [LPH](/productcode/LPH) · K141960

# Exactech® Novation® Crown Cup® (K141960)

_Exactech, Inc. · LPH · Sep 3, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K141960

## Device Facts

- **Applicant:** Exactech, Inc.
- **Product Code:** [LPH](/productcode/LPH.md)
- **Decision Date:** Sep 3, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3358
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also potentially indicated for ankylosing spondylitis, congenital hip dysplasia, revision of failed previous reconstructions where sufficient bone stock is present, and to restore mobility resulting from previous fusion. Cemented femoral stems and cemented acetabular cups are intended for cemented fixation only. Press-fit femoral stems and acetabular cups are intended for press-fit fixation. Femoral heads and endoprostheses are intended for use in cemented and press-fit applications.

## Device Story

The Exactech Novation Crown Cup is an acetabular component used in hip arthroplasty. This submission covers a manufacturing process change regarding the application of a titanium plasma spray coating to the device surface. The device is intended for surgical implantation by orthopedic surgeons in clinical settings. It functions as a prosthetic hip component to restore joint mobility and function. The modification involves only the coating application process; the fundamental design, materials, and intended use remain identical to the predicate devices. Clinical benefit is derived from the restoration of hip joint function in patients with degenerative or traumatic hip conditions. The device is evaluated through engineering analyses, including stereological evaluation, tensile/shear testing, and abrasion resistance testing of the porous coating to ensure performance equivalence.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing only, specifically stereological evaluation, tensile and shear testing, and abrasion resistance testing of the porous coating.

## Technological Characteristics

Acetabular component for hip arthroplasty. Features a titanium plasma spray coating on the porous surface. Designed for press-fit or cemented fixation. The device is a semi-constrained metal/polymer prosthesis.

## Regulatory Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

## Predicate Devices

- Exactech Novation Crown Cup and Liners ([K070479](/device/K070479.md))
- Exactech Novation Crown Cup and Liners ([K100269](/device/K100269.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol representing the human form, depicted as three stylized profiles facing to the right. The profiles are connected and flow together, creating a sense of unity and collaboration.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 3, 2015

Exactech Incorporated Mr. Thomas McNamara Regulatory Affairs Specialist 2320 North West 66th Court Gainesville, Florida 32653

Re: K141960 Trade/Device Name: Exactech® Novation® Crown Cup® Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, LZO, JDI, LWJ, MBL Dated: July 22, 2015 Received: July 27, 2015

Dear Mr. McNamara:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{1}------------------------------------------------

Page 2 - Mr. Thomas McNamara

(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

K141960 510(k) Number:

#### Device Name: Exactech® Novation® Crown Cup®

#### INDICATIONS

All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also potentially indicated for ankylosing spondylitis, congenital hip dysplasia, revision of failed previous reconstructions where sufficient bone stock is present, and to restore mobility resulting from previous fusion.

- . Cemented femoral stems and cemented acetabular cups are intended for cemented fixation only.
- Press-fit femoral stems and acetabular cups are intended for press-fit fixation.
- . Femoral heads and endoprostheses are intended for use in cemented and press-fit applications.

Prescription Use_ x (Part 21 CFR 801 Subpart D)

and/or

Over-The-Counter Use_ (21 CFR 807 Subpart C)

Please do not write below this line - use another page if needed.

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

# Exactech® Novation® Crown Cup® Plasma Spray Coating Manufacturing Change Traditional 510(k)

# 510(k) Summary

| Sponsor:          | Exactech®, Inc<br>2320 NW 66th Court<br>Gainesville FL, 32653                                        |
|-------------------|------------------------------------------------------------------------------------------------------|
|                   | Phone: (352) 377-1140<br>Fax: (352) 378-2617                                                         |
|                   | FDA Establishment Number 1038671                                                                     |
| Date:             | August 27, 2015                                                                                      |
| Contact Person:   | Thomas McNamara<br>Regulatory Affairs Specialist<br>Telephone: (352) 377-1140<br>Fax: (352) 378-2617 |
| Proprietary Name: | Exactech® Novation® Crown Cup®                                                                       |

Common Name: Acetabular Component

### Classification Name:

- Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or ● Non-Porous, Uncemented, 21 CFR 888.3353, Class II, Product Code LZO
- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented, 21 CFR ● 888.3350 Class II, Product Code JDI
- Prosthesis, Hip, Semi-constrained, Meta/Polymer, Porous Uncemented, 21 ● CFR 888.3358, Class II, Product Code LPH
- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented, 21 CFR 888.3360, Class II, Product Code LWJ
- Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Porous, 21 ● CFR 888.3358, Class II, Product Code MBL

#### Legally Marketed Device to Which Substantial Equivalence Is Claimed:

| Name                                      | Manufacturer  | 510(k) Number    |
|-------------------------------------------|---------------|------------------|
| Exactech Novation Crown Cup and<br>Liners | Exactech, Inc | K070479, K100269 |

### Indication for Use:

All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also

{4}------------------------------------------------

# Exactech® Novation® Crown Cup® Plasma Spray Coating Manufacturing Change Traditional 510(k)

potentially indicated for ankylosing spondylitis, congenital hip dysplasia, revision of failed previous reconstructions where sufficient bone stock is present, and to restore mobility resulting from previous fusion.

- Cemented femoral stems and cemented acetabular cups are intended for cemented fixation only.
- Press-fit femoral stems and acetabular cups are intended for press-fit fixation. .
- Femoral heads and endoprostheses are intended for use in cemented and pressfit applications.

### Device Description

The proposed Novation Crown Cup devices represent a modification to the predicate Novation Crown Cup cleared in K070479 and K100269. Both the predicate and proposed devices have the same intended use, general design features, and basic fundamental scientific technology. The only difference between the predicate and the proposed devices is the process by which the titanium plasma spray coating is applied.

# Testing:

The following engineering analyses were conducted to demonstrate substantial equivalence of the proposed Novation Crown Cup devices to the predicate Novation Crown Cups:

- . Stereological Evaluation of Porous Coating
- Tensile and Shear Testing of Porous Coating
- Abrasion Resistance Testing .

# Substantial Equivalence Conclusion:

Results of engineering studies referenced in this 510(k) submission demonstrate the proposed Novation Crown Cup are substantially equivalent to cleared Novation Crown Cup devices.

---

**Source:** [https://fda.innolitics.com/device/K141960](https://fda.innolitics.com/device/K141960)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
