← Product Code [CBK](/productcode/CBK) · K141943

# BREATHE TECHNOLOGIES LIFE2000 VENTILATION SYSTEM (K141943)

_Breathe Technologies · CBK · Jun 12, 2015 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K141943

## Device Facts

- **Applicant:** Breathe Technologies
- **Product Code:** [CBK](/productcode/CBK.md)
- **Decision Date:** Jun 12, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5895
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The Breathe Technologies Life2000 Ventilation System is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The ventilator is intended for use by qualified, trained personnel under the direction of a physician. Specifically, the ventilator is applicable for adult patients who require the following types of ventilatory support: - Positive Pressure Ventilation, delivered invasively (via ET tube) or non-invasively (via mask). - Assist/Control mode of ventilation. The ventilator is suitable for use in home and institutional settings.

## Device Story

Portable, battery-powered critical care ventilator; provides positive pressure ventilation via tracheostomy, endotracheal tube, or nasal mask. Input: external compressed gas supply (air/oxygen). Operation: delivers physician-prescribed volume using pneumatic active exhalation valve; supports Assist/Control mode. Output: mechanical ventilation to patient. Used in home/institutional settings by trained personnel/physicians. Small/lightweight (1.1 lbs) for belt/shoulder wear. Benefits: provides portable ventilatory support for adult patients; enables mobility compared to larger stationary ventilators.

## Clinical Evidence

Bench testing only. Compliance demonstrated with ISO 10993-1, IEC 60601-1, IEC 60601-1-2, ISO 80601-2-12, and ASTM F1246-91. Additional testing included ventilator cleaning validation, VOC risk characterization, static software analysis, comparative waveform testing, and human factors/usability testing.

## Technological Characteristics

Portable critical care ventilator; battery-powered; pneumatic active exhalation valve. Dimensions: 3.2 x 7.7 x 1.0 inches; Weight: 1.1 lbs. Gas supply: external compressed air/oxygen. Materials: polycarbonate (Universal Connector). Standards: ISO 10993-1, IEC 60601-1, IEC 60601-1-2, ISO 80601-2-12, ASTM F1246-91.

## Regulatory Identification

A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.

## Predicate Devices

- LTV-1200 ([K060647](/device/K060647.md))

## Reference Devices

- Breathe Technologies NIOV Ventilator ([K103345](/device/K103345.md))
- ResMed Ultra Mirage II Nasal Mask ([K050359](/device/K050359.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around the perimeter. In the center of the seal is a stylized image of three human profiles facing to the right, with flowing lines beneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 12, 2015

Breathe Technologies C/O Mr. Craig Coombs President Coombs Medical Device Consulting, Inc. 1193 Sherman St. Alameda, CA 94501

Re: K141943/S003

Trade/Device Name: Breathe Technologies Life2000 Ventilation System Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: June 8, 2015 Received: June 10, 2015

Dear Mr. Coombs:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Mr. Coombs

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K141943

Device Name

Breathe Technologies Life2000 Ventilation System

Indications for Use (Describe)

The Breathe Technologies Life2000 Ventilation System is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.

The ventilator is intended for use by qualified, trained personnel under the direction of a physician.

Specifically, the ventilator is applicable for adult patients who require the following types of ventilatory support:

- Positive Pressure Ventilation, delivered invasively (via ET tube) or non-invasively (via mask).

- Assist/Control mode of ventilation.

The ventilator is suitable for use in home and institutional settings.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| Device Information:                      |                                                                                                                                                          |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Category                                 | Comments                                                                                                                                                 |
| Sponsor:                                 | Breathe Technologies<br>175 Technology Drive, Suite 100<br>Irvine, CA 92618<br>Tel: 949-988-7700<br>Contact: Samir Ahmad, Ph.D.                          |
| Correspondent Contact<br>Information:    | Craig Coombs<br>President<br>Coombs Medical Device Consulting, Inc<br>1193 Sherman St.<br>Alameda, CA 94501<br>Office: 510.337.0140<br>Fax: 510.337.0416 |
| Device Common Name:                      | Mechanical Ventilator                                                                                                                                    |
| Device Classification & Name:            | 21 CFR 868.5895 Continuous Ventilator                                                                                                                    |
| Device Classification &<br>Product Code: | Class II<br>CBK                                                                                                                                          |
| Device Proprietary Name:                 | Breathe Technologies Life2000 Ventilation System                                                                                                         |

# Section 5: 510(k) Summary

#### Predicate Device Information:

| Predicate Device:                         | LTV-1200                 |
|-------------------------------------------|--------------------------|
| Predicate Device Manufacturer:            | Pulmonetic Systems, Inc. |
| Predicate Device Premarket Notification # | K060647                  |
| Predicate Device Common Name:             | Mechanical Ventilator    |
| Predicate Device Classification & Name:   | 21 CFR 868.5895          |
| Predicate Device Classification &         | Class II                 |
| Product Code:                             | CBK                      |

### b. Date Summary Prepared

12 June 2015

## c. Description of Device

The Breathe Technologies Life2000 Ventilation System is a portable, battery powered, critical care ventilator.

The Ventilator administers the physician-prescribed volume to the patient via the attached Breathe Technologies Patient Universal Connector which connects into the patient's tracheostomy tube, endotracheal tube, or any off the shelf non-invasive mask. It can also be used with the Breath Technologies NIOV Pillows Interface, a type of nasal mask.

The ventilator is small and light enough to be worn on a patient's belt, or slung over their shoulder. It is connected to a separate, third party, gas supply.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Breathe Technologies. The word "Breathe" is in a large, light blue font, with the word "TECHNOLOGIES" in a smaller font below it. There is a blue wave design below the word "TECHNOLOGIES".

The Ventilator is intended for Institutional or Home Use. It is not intended for use during emergency/medical transportation.

### d. Indications for Use

The Breathe Technologies Life2000 Ventilation System is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The ventilator is intended for use by qualified, trained personnel under the direction of a physician.

Specifically, the ventilator is applicable for adult patients who require the following types of ventilatory support:

- Positive Pressure Ventilation, delivered invasively (via ET tube) or noninvasively (via mask).

- Assist/Control mode of ventilation.

The ventilator is suitable for use in home and institutional settings.

#### e. Comparison to Predicate Device

The Life2000 Ventilation System is substantially equivalent in Intended Use, Indications for Use, technology, and performance to the LTV-1200 that was cleared under K060647.

The Life2000 serves a subset of the cohort (adult) that are indicated for the LTV-1200 (patients >5kg).

The Life2000 has all the ventilation modes that are necessary for a critical care ventilator, whereas the LTV-1200 includes modes like SIMV and CPAP.

The Life2000 has a pneumatic active exhalation valve. whereas the LTV-1200 has a mechanical/pneumatic active exhalation valve.

Any technological differences between the Breathe Technologies Life2000 and the LTV-1200 can be determined to be clinically insignificant when compared to the reference device, the Breathe Technologies NIOV Ventilator, cleared under K103345.

The following table presents a comparison of the features of the predicate and applications devices, along with an analysis of why the resultant differences do not negatively impact a conclusion of substantial equivalence.

{5}------------------------------------------------

|                                           | Predicate Device:<br>LTV-1200<br>K060647                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Application Device:<br>Life2000 Ventilation<br>System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Difference<br>Status                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use                    | The LTV 1200 ventilator is<br>intended to provide<br>continuous or intermittent<br>ventilatory support for<br>the care of individuals who<br>require mechanical ventilation.<br>The ventilator is a restricted<br>medical device intended for use<br>by qualified, trained personnel<br>under the direction of a<br>physician.<br>Specifically, the ventilator is<br>applicable for adult and pediatric<br>patients weighing<br>at least 5 kg (11 lbs.), who<br>require the following types of<br>ventilatory support:<br>- Positive Pressure Ventilation,<br>delivered invasively (via ET<br>tube) or non-invasively (via<br>mask).<br>- Assist/Control, SIMV, CPAP,<br>or NPPV modes of ventilation.<br>The ventilator is suitable for use<br>in institutional, home and<br>transport settings. | The Breathe Technologies<br>Life2000 Ventilation System is<br>intended to provide continuous<br>or intermittent ventilatory<br>support for the care of<br>individuals who require<br>mechanical ventilation.<br>The ventilator is intended for use<br>by qualified, trained personnel<br>under the direction of a<br>physician.<br>Specifically, the ventilator is<br>applicable for adult patients<br>who require the following types<br>of ventilatory support:<br>- Positive Pressure Ventilation,<br>delivered invasively (via ET<br>tube) or non-invasively (via<br>mask).<br>- Assist/Control mode of<br>ventilation.<br>The ventilator is suitable for use<br>in home and institutional<br>settings. | The Indications for<br>Use for the application<br>device is a subset of<br>the Indications for Use<br>of the predicate.<br>In particular, the<br>application device<br>treats only adult<br>patients, rather than<br>the predicate's >5kg<br>limit. The application<br>device does not have<br>all the ventilation<br>modes that the<br>predicate device<br>possesses. Finally, the<br>application device is<br>not indicated for use in<br>emergency/medical<br>transport settings. |
| Product<br>Classification<br>Code         | CBK                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | CBK                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| CFR Citation                              | 21 CFR 868.5895                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 21 CFR 868.5895                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Principal<br>Operator                     | Trained personnel under the<br>direction of a physician                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Trained personnel under the<br>direction of a physician                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Environment<br>of Use                     | Institution, Home, and<br>Transport                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Institution & Home                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Application device is not<br>validated to work in an<br>emergency/medical<br>transport environment.                                                                                                                                                                                                                                                                                                                                                                                  |
| Patient<br>Interface                      | Delivered invasively (via ET tube)<br>or non-invasively (via mask).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Delivered invasively (via ET tube)<br>or non-invasively (via mask).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Power Source                              | Battery Powered, can be run<br>while battery is charging                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Battery Powered, can be run<br>while battery is charging                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Operational<br>Modes                      | Volume Control<br>Volume Assist/Control<br>Volume Assist<br>Pressure Control<br>Pressure Support<br>SIMV<br>CPAP<br>NPPV                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Volume Control<br>Volume Assist/Control<br>Volume Assist                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | The operational modes<br>of the application<br>device are a subset of<br>the predicate.                                                                                                                                                                                                                                                                                                                                                                                              |
|                                           | Predicate Device:<br>LTV-1200<br>K060647                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Application Device:<br>Life2000 Ventilation<br>System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Difference<br>Status                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Active<br>Exhalation<br>Valve?            | Yes, Mechanical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Yes, Mechanical/Pneumatic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Clinically Equivalent                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Design<br>Designation                     | Portable Critical Care                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Portable Critical Care                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Size<br>WxLxH (in)                        | 3 x 10 x 12                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 3.2 x 7.7 x 1.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | The application device<br>is much smaller than<br>predicate device.<br>Potentially easier for<br>patient to handle.  No<br>new issues of S&E are<br>raised by the size<br>reduction                                                                                                                                                                                                                                                                                                  |
| Weight                                    | 13.4 lbs                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 1.1 lbs                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | The application device<br>is much smaller than<br>predicate device.<br>Potentially easier for<br>patient to handle.  No<br>new issues of S&E are<br>raised by the weight<br>reduction.                                                                                                                                                                                                                                                                                               |
| Volume<br>Setting<br>Range                | 50 - 2000 ml/breath                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 50 - 750 ml/breath                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | The volume range of<br>the application device<br>is a subset of the<br>predicate device                                                                                                                                                                                                                                                                                                                                                                                              |
| Resultant<br>Tidal Volume                 | 50 - 2000 ml/breath                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 50 - Up to 2000 ml/breath due<br>to venturi effect                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| PEEP Setting                              | 0 – 20 cmH2O                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 0 - 10 cmH2O                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | The PEEP range of the<br>application device is a<br>subset of the predicate<br>device                                                                                                                                                                                                                                                                                                                                                                                                |
| PIP Alarms &<br>Monitoring                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Adjustable<br>Inspiration<br>Time         | 0.3 - 9.9 seconds                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 0.15 to 3 seconds                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | I-time of application<br>device is a subset of<br>predicate. Only<br>clinically relevant<br>times are used                                                                                                                                                                                                                                                                                                                                                                           |
| Supply Gas                                | Oxygen, Air                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Oxygen, Air                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Method of<br>supply gas<br>pressurization | Internal turbine for Air<br>Compressed Source for O2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Compressed source for Air<br>Compressed source for O2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Clinically Equivalent<br>Result from Ventilator                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Sterilized?                               | Ventilator: No<br>Patient Circuit: No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Ventilator: No<br>Patient Circuit: No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |

# Tabular Comparison to Predicate Device

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Breathe Technologies. The word "Breathe" is written in a large, light blue font. Below that, the word "TECHNOLOGIES" is written in a smaller, light blue font. There is a curved, light blue line below the word "TECHNOLOGIES".

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Breathe Technologies. The word "Breathe" is written in a large, sans-serif font in blue. Below "Breathe" is the word "TECHNOLOGIES" in a smaller, sans-serif font, also in blue. Underneath the words is a blue wave design.

## f. Summary of Supporting Data

The Software Design and Validation process along with the bench testing of the device demonstrated that the Breathe Technologies Life2000 Ventilation System operates as intended.

In particular, testing demonstrated that Life2000 is compliant with the following Guidelines and Standards:

- ISO 10993-1 (2009): Biological evaluation of medical devices -- Part 1: Evaluation and Testing
- . FDA Draft Reviewer Guide for Ventilators (July 1995)
- ASTM F1246-91 (1991, Reapproved 2005); Standard Specification for Electrically Powered Home Care Ventilators, Part 1 - Positive Pressure Ventilators and Ventilator Circuits
- . IEC 60601 - 1 (2005): Medical electrical equipment - General Requirements for Safety
- . IEC 60601-1-2:2007 3rd Edition Medical Electrical Equipment- Part 1-2: General Requirements for Basic Safety and Essential Performance- Collateral Standard: Electromagnetic Compatibility- Requirements and Tests
- ISO 80601-2-12 (2011), First Edition 2011-04-15, Medical electrical equipment -Part 2-12: Particular requirements for basic safety and essential performance of critical care ventilators

The following additional testing was requested by the FDA to demonstrate Substantial Equivalence.

- Ventilator Cleaning validation .
- . Risk characterization of VOC's for most vulnerable patient population
- . Static Analysis Tool verification of software
- . Comparative Waveform Testing
- Human Factors and Usability Testing with all intended cohorts and users.

Breathe Technologies concludes from this comparison and the supporting data that the Breathe Technologies Life2000 Ventilation System is substantially equivalent to to the predicate Pulmonetic Systems LTV-1200 (K060647).

<sup>1</sup> Per Annex B.3, it was determined that all of the materials (and material processes) used in Life2000 are identical those in a reference device, the NIOV Ventilator and Accessories (K103345), except the polycarbonate material in the Universal Connector, which was shown to be identical to the polycarbonate used in the ResMed Ultra Mirage II Nasal Mask (K050359). This analysis was adopted for the Volatile Organic Compound and Particulate Matter testing.

---

**Source:** [https://fda.innolitics.com/device/K141943](https://fda.innolitics.com/device/K141943)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
