The Stryker S2 Drill is intended for use with the Stryker Consolidated Operating Room Equipment (CORE) System. When used with a variety of attachments and cutting accessories, the drill is intended for use in cutting, reaming, decorticating, shaping, and smoothing of bone cement and teeth in a variety of surgical procedures, including but not limited to dental. ENT (ear, nose, and throat), neuro, spine, and endoscopic applications. It is also usable in the placement or cutting of screws, metal, wires, pins, and other fixation devices.
Device Story
Stryker S2 Drill is an electric-powered (40V DC) surgical handpiece designed for use with the Stryker Consolidated Operating Room Equipment (CORE) System. Device rotates various cutting accessories at speeds up to 75,000 RPM to perform bone, bone cement, and tooth manipulation (cutting, reaming, shaping, smoothing) and fixation device placement. Operated by surgeons in clinical settings (OR, dental, etc.) via footswitch control. Device features improved rotor driveshaft compared to predicate. Output is mechanical rotation of surgical attachments; healthcare providers use this to perform precise surgical bone/tissue work. Benefits include enhanced surgical access and efficiency in diverse specialties.
Clinical Evidence
No clinical testing was deemed necessary for this 510(k). Substantial equivalence is supported by bench testing, including verification of the improved rotor driveshaft, simulated use tests, and temperature testing across various torque-speed settings.
Technological Characteristics
Electric-powered (40V DC) surgical drill; 17mm motor diameter; 123.5mm length; 313g weight. Materials: Stainless steel and aluminum. Connectivity: Cable-connected to CORE console. Activation: Footswitch. Speed: 0-75,000 RPM. Attachment retention: Mechanical lock via rotation.
Indications for Use
Indicated for use in surgical procedures including dental, ENT, neuro, spine, and endoscopic applications for cutting, drilling, reaming, decorticating, shaping, and smoothing of bone, bone cement, and teeth, and for placement or cutting of fixation devices.
Regulatory Classification
Identification
An ear, nose, and throat electric or pneumatic surgical drill is a rotating drilling device, including the handpiece, that is intended to drive various accessories, such as an ear, nose, and throat bur (§ 874.4140), for the controlled incision or removal of bone in the ear, nose, and throat area.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is an abstract symbol that resembles an eagle or a bird in flight, composed of three stylized human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 26, 2014
Stryker Corporation Mr. Vishal Kanani Sr. Regulatory Affairs Representative 4100 E. Milham Avenue Kalamazoo, MI 49001
Re: K141935
Trade/Device Name: Stryker S2 Drill Regulation Number: 21 CFR 874.4250 Regulation Name: Ear, Nose, and Throat Electric or Pneumatic Surgical Drill Regulatory Class: Class II Product Code: ERL, HBE, DZJ Dated: August 25, 2014 Received: August 27, 2014
Dear Mr. Kanani:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
# K141935
Device Name Styker S2 Drill
#### Indications for Use (Describe)
The Stryker S2 Drill is intended for use with the Stryker Consolidated Operating Room Equipment (CORE) System. When used with a variety of attachments and cutting accessories, the drill is intended for use in cutting, reaming, decorticating, shaping, and smoothing of bone cement and teeth in a variety of surgical procedures, including but not limited to dental. ENT (ear, nose, and throat), neuro, spine, and endoscopic applications. It is also usable in the placement or cutting of screws, metal, wires, pins, and other fixation devices.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "Stryker" in a bold, sans-serif font. The letters are all capitalized except for the "t" and "y", which are lowercase. A registered trademark symbol is located to the right of the "r" in "Stryker". The word is in black and the background is white.
## 510(k) Summary
| 510(k) Owner: | Stryker Instruments<br>4100 E. Milham Avenue<br>Kalamazoo, MI 49001<br>(p) 269-323-7700<br>(f) 269-389-5412 |
|---------------------------|--------------------------------------------------------------------------------------------------------------|
| Contact Person: | Vishal Kanani<br>Sr. Regulatory Affairs Representative |
| Registration Number: | 1811755 |
| Date Summary<br>Prepared: | July 07, 2014 |
| Trade Name(s): | Stryker S2 Drill |
Common Name:
Classification Data:
Ear, nose, and throat electric or pneumatic surgical drill.
| | Product<br>Code | Device | Regulation<br>Number | Class | Review<br>Panel |
|--------------------|-----------------|------------------------------------------------------------------------------|----------------------|-------|-------------------------|
| Primary<br>Code | ERL | Drill, Surgical,<br>ENT (Electric or<br>Pneumatic)<br>Including<br>Handpiece | 21 CFR<br>874.4250 | II | Ear, Nose<br>and Throat |
| Secondary<br>Codes | HBE | Drills, burs,<br>trephines, and<br>accessories<br>(simple,<br>powered) | 21 CFR<br>872.4120 | II | Neurology |
| | DZJ | Driver, wire,<br>and bone drill,<br>manual | 21 CFR<br>872.4120 | II | Dental |
### Predicate Device:
| 510(k)<br>number | Product<br>code | Trade name | Manufacturer |
|------------------|-----------------|--------------------------------------------------------------------------|------------------------|
| K112593 | ERL | Stryker®<br>Consolidated<br>Operating Room<br>Equipment<br>(CORE) System | Stryker<br>Instruments |
{5}------------------------------------------------
4100 E. Milham Ave.
Kalamazoo, MI 49001
t: 269 323 7700 f: 269 389 5412 www.stryker.com
Image /page/4/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase except for the first letter, which is uppercase. There is a registered trademark symbol to the right of the word. The word is in black and the background is white.
| Indications for Use: | The Stryker S2 Drill is intended for use with the Stryker Consolidated Operating Room Equipment (CORE) System. When used with a variety of attachments and cutting accessories, the drill is intended for use in cutting, drilling, reaming, decorticating, shaping, and smoothing of bone, bone cement and teeth in a variety of surgical procedures, including but not limited to dental, ENT (ear, nose, and throat), neuro, spine, and endoscopic applications. It is also usable in the placement or cutting of screws, metal, wires, pins, and other fixation devices. |
|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Description: | The Stryker S2 Drill is an electric powered (40V DC) motor. When connected to the CORE console, it directly rotates cutting accessories up to speeds of 75,000 RPM. |
| Performance Data (Non Clinical Tests): | Following verification tests were performed which demonstrate that the design outputs of the modified device meet the design input requirements: <ul><li>Verification of the improved rotor driveshaft</li><li>Simulated use tests</li><li>Temperature testing of different torque-speed setting</li></ul> Results of these tests demonstrate that the functionality, integrity, and safety and effectiveness of the Stryker S2 Drill is sufficient for their intended use and support a determination of substantial equivalence. |
| Clinical Tests: | No clinical testing was deemed necessary for this 510(k). |
| Conclusion/Substantial Equivalence (SE) Rationale: | The Stryker S2 Drill is substantially equivalent in intended use, technological characteristics, safety, and effectiveness to the previously cleared Stryker CORE Sumex Drill. The products have the same fundamental scientific technology, basic design, functional characteristics and applications. The modifications introduced raise no new issues of safety and effectiveness. Therefore, the Stryker S2 Drill is substantially |
equivalent to the existing predicate device.
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "Stryker" in a bold, sans-serif font. The letters are all connected, and the word is in black. There is a registered trademark symbol to the right of the letter "r".
| Summary of Substantial Equivalence Table | | |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Description | Stryker CORE Sumex Drill (Predicate) | Stryker S2 Drill (Subject) |
| Intended Use | The Stryker Consolidated Operating<br>Room Equipment (CORE) System is<br>intended for use in cutting, drilling,<br>reaming, decorticating, shaping, and<br>smoothing of bone, bone cement and<br>teeth in a variety of surgical<br>procedures, including but not limited<br>to, dental, ENT (ear, nose, and<br>throat), neuro, spine, and endoscopic<br>applications. It is also usable in the<br>placement or cutting of screws,<br>metal, wires, pins, and other fixation<br>devices. | The S2 Drill is intended for use with<br>the Stryker Consolidated Operating<br>Room Equipment (CORE) System.<br>When used with a variety of<br>attachments and cutting accessories,<br>the drill is intended for use in<br>drilling, cutting, reaming,<br>decorticating, shaping and<br>smoothing of bone, bone cement<br>and teeth in a variety of surgical<br>procedures, including but not limited<br>to ENT, neuro, spine and Endoscopic<br>applications. It is also usable in the<br>placement or cutting of screws,<br>metal, wires, pins, and other fixation<br>devices. |
| Housing Material | Stainless Steel | Stainless Steel and Aluminum |
| Motor Diameter | 20mm | 17mm |
| Length of the drill | 105mm | 123.5mm |
| Weight of the drill | 399g | 313g |
| Power source | 40V DC Electric Motor connected via<br>cable to CORE console | 40V DC Electric Motor connected via<br>cable to CORE console |
| Speed | 0-75,000 rpm | 0-75,000 rpm |
| Attachment retention<br>method by drill | Mechanical lock activated by rotation<br>of attachment onto drill | Mechanical lock activated by<br>rotation of attachment onto drill |
| Mode of activation | Footswitch and Handswitch | Footswitch |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.