COBAS C ACETAMINOPHEN GEN.2 ASSAY, ACET2 CALIBRATOR
Applicant
Roche Diagnostics Operations (Rdo)
Product Code
DKB · Clinical Toxicology
Decision Date
Nov 25, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3200
Device Class
Class 2
Indications for Use
The cobas c Acetaminophen Gen.2 assay is an in vitro diagnostic test for the quantitative determination of acetaminophen of acetaminophen overdose in serum and plasma on Roche/Hitachi cobas c systems. The ACET2 calibrator is for use in the calibration of the Acetaminophen Gen.2 Roche assay.
Device Story
The cobas c Acetaminophen Gen.2 assay is a homogeneous enzyme immunoassay for quantitative analysis of acetaminophen in human serum and plasma. It uses a competitive binding principle: drug in the sample competes with enzyme-labeled drug (bacterial G6PDH) for anti-acetaminophen antibody binding sites. Enzyme activity is inversely proportional to drug concentration; activity is measured spectrophotometrically via absorbance change of NADH. The ACET2 calibrator is a liquid, ready-to-use reagent containing a known quantity of acetaminophen; the analyzer performs on-board dilution with NaCl to create a six-level calibration set. The device is used in clinical laboratories on Roche/Hitachi cobas c 501 analyzers. Healthcare providers use the quantitative results to assess overdose severity and guide clinical decision-making, potentially preventing liver/kidney damage.
Clinical Evidence
Bench testing only. Precision (CLSI EP5-A2) showed CVs 2.2-3.5%. Linearity (CLSI EP6-A) confirmed 5-200 µg/mL range. Method comparison (n=105) against predicate yielded Deming regression y = 1.02x - 0.699 (r=0.997). LoB/LoD/LoQ established at 1.5/3/5 µg/mL respectively. Interference testing confirmed no significant interference from endogenous substances (hemoglobin, lipids, bilirubin) or 24 common drugs.
Technological Characteristics
Quantitative enzyme immunoassay. Reagents: sheep polyclonal anti-acetaminophen antibody, G6PDH-labeled acetaminophen, NAD, glucose-6-phosphate. Energy source: spectrophotometric (340/415 nm). Form factor: cassette-based reagents for cobas c 501 analyzer. Connectivity: integrated with analyzer software. Standards: CLSI EP05-A2, EP06-A2, EP17-A2, EP09-A2.
Indications for Use
Indicated for quantitative determination of acetaminophen in serum and plasma to diagnose acetaminophen overdose. Acetaminophen is a widely used analgesic/antipyretic; overdose can cause severe liver/kidney damage or death. Quantitative measurement is the only reliable early diagnostic indicator.
Regulatory Classification
Identification
A clinical toxicology calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. A clinical toxicology calibrator can be a mixture of drugs or a specific material for a particular drug (e.g., ethanol, lidocaine, etc.). (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
{1}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k141928
B. Purpose for Submission:
New Device
C. Measurand:
Acetaminophen
D. Type of Test:
Quantitative enzyme immunoassay
E. Applicant:
Roche Diagnostics
F. Proprietary and Established Names:
cobas C Acetaminophen Gen.2 assay
ACET2 Calibrator
G. Regulatory Information:
1. Regulation section:
21 CFR Sec. 862.3030-Acetaminophen test system
21 CFR Sec. 862.3200-Clinical toxicology calibrator
2. Classification:
Class II
3. Product code:
LDP-Colorimetry, Acetaminophen
DKB-Calibrators, Drug Mixture
4. Panel:
Toxicology (91)
H. Intended Use:
1. Intended use(s):
See Indications for use below.
2. Indication(s) for use:
{2}
The cobas c Acetaminophen Gen.2 assay is an in vitro diagnostic test for the quantitative determination of acetaminophen of acetaminophen overdose in serum and plasma on Roche/Hitachi cobas c systems.
The ACET2 calibrator is for use in the calibration of the Acetaminophen Gen.2 Roche assay.
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
Roche Hitachi cobas c 501 clinical chemistry analyzer
I. Device Description:
The cobas c Acetaminophen Gen.2 assay consists of reagents packaged in a cassette labeled with their instrument positioning, R1 (Reagent 1) and R2 (Reagent 2).
- R1 contains anti-acetaminophen antibody (sheep polyclonal), Glucose-6-phosphate, Nicotinamide adenine dinucleotide (NAD), bovine serum albumin, preservatives (Proclin), and stabilizers (pepstatin and aprotinin).
- R2 contains acetaminophen labeled with bacterial Glucose-6-phosphate dehydrogenase, Tris buffer, preservatives (Sodium Azide and 2-methyl-2H-isothiazol-3-one (MIT)) bovine serum albumin, and stabilizers (pepstatin and aprotinin).
The ACET2 calibrator contains a known quantity of acetaminophen (~200 µg/mL), sodium phosphate, sodium EDTA, sodium chloride (NaCl), and preservatives (Proclin and MIT). The cobas c 501 analyzer dilutes the ACET2 calibrator on-board the analyzer with NaCl diluent, in order to create five concentration levels (level 2, 3, 4, 5, and 6) containing 10, 30.2, 75.0, 100, and 200 µg/mL of acetaminophen, respectively. Level 1 contains no acetaminophen.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Emit® tox™ Acetaminophen Assay
Emit® tox™ Acetaminophen Calibrators
2. Predicate 510(k) number(s):
K002974
3. Comparison with predicate:
{3}
| Assay Comparison - Similarities and Differences | | |
| --- | --- | --- |
| Feature | Predicate Device: Emit® tox™ Acetaminophen Assay | Candidate Device: Acetaminophen Gen.2 |
| Intended Use | The Emit® tox™ Acetaminophen Assay is a homogeneous enzyme immunoassay intended for in vitro diagnostic use in the quantitative analysis of acetaminophen in human serum or plasma. | Same |
| Test Principle | Homogeneous enzyme immunoassay | Same |
| Measuring Range | Up to 200 μg/mL (1324 μmol/L) | 5–200 μg/mL (33.1–1324 μmol/L) |
| Traceability | This method has been standardized against USP reference standards. | Same |
| Sample Types | Serum and EDTA, heparin, citrate and oxalate/fluoride plasma | Serum and K2- or K3-EDTA, or lithium heparinized plasma |
| Calibrator Comparison - Similarities and Differences | | |
| --- | --- | --- |
| Feature | Predicate Device: Emit® tox™ Acetaminophen Calibrators | Candidate Device: ACET2 Calibrator |
| Intended Use | The Emit® tox™ Acetaminophen Calibrators are intended for use with the Emit® tox™ Acetaminophen Assay. | ACET2 calibrator is for use in the calibration of the Acetaminophen Gen.2 Roche assay. |
| Format | Liquid ready-to-use | Same |
| Analyte | Acetaminophen | Same |
| Storage Conditions | 2-8°C | Same |
{4}
| Storage and Stability | When stored refrigerated at 2–8°C, the Emit® tox™ Acetaminophen Calibrators are stable until the expiration date printed on the dropper vial label. | Same |
| --- | --- | --- |
| Quantity | Emit® tox™ Acetaminophen Calibrators are a six-level set that contain the following acetaminophen concentrations: 0, 10, 25, 50, 100, 200 μg/mL. | The ACET2 calibrator has a nominal value of 200 μg/mL of acetaminophen. The analyzer dilutes the ACET2 calibrator on-board with NaCl diluent, in order to create five concentration levels. This results in a six- level calibrator set: 0.0, 10.0, 30.2, 75.0, 100, 200 μg/mL. |
K. Standard/Guidance Document Referenced (if applicable):
- Standard Title CLSI EP05-A2- Evaluation of Precision Performance of Quantitative Measurement Methods
- Standard Title CLSI EP06-A2- Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach
- Standard Title CLSI EP17-A2- Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures
- Standard Title CLSI EP09-A2- Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline-Second Edition
L. Test Principle:
The assay is based on a homogeneous enzyme immunoassay technique used for the quantitative analysis of acetaminophen in human serum or plasma. The assay is based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phoshate dehydrogenase (G6PDH) for antibody binding sites. Enzyme activity decreases upon binding to the antibody, so the drug concentration in the sample can be measured in terms of enzyme activity. Active enzyme converts oxidized nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that is measured spectrophotometrically (at 415 nm and 340 nm).
M. Performance Characteristics (if/when applicable):
All performance studies were conducted with the Roche Hitachi cobas c 501 clinical chemistry analyzer.
1. Analytical performance:
{5}
# a. Precision/Reproducibility:
Precision of the Acetaminophen Gen.2 assay was evaluated with a 21-day precision study according to CLSI EP05-A2, using control material as well as pooled human sera samples (from patients taking acetaminophen). The acetaminophen concentrations of all tested samples are noted in the table below. All samples were tested using 3 reagent kit lots, with 2 replicates per run and 2 runs per day. Data from all lots tested were similar. Data from a representative lot is presented below.
| Test Specimen | Mean Acetaminophen Concentration (μg/mL) | Within-Run (Repeatability) | | Between-Run | | Total Precision (within lab precision) | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD (μg/mL) | % CV | SD (μg/mL) | % CV | SD (μg/mL) | % CV |
| Human Sample 1 | 7.6 | 0.224 | 3.0 | 0.107 | 1.4 | 0.258 | 3.4 |
| Human Sample 2 | 73 | 1.66 | 2.3 | 0.314 | 0.4 | 1.95 | 2.7 |
| Human Sample 3 | 130 | 2.90 | 2.2 | 0 | 0 | 3.32 | 2.5 |
| Human Sample 4 | 169 | 3.69 | 2.2 | 0.377 | 0.2 | 4.66 | 2.8 |
| Human Sample 5 | 184 | 4.25 | 2.3 | 0 | 0 | 5.31 | 2.9 |
| Control 1 | 15.3 | 0.385 | 2.5 | 0.193 | 1.3 | 0.455 | 3.0 |
| Control 2 | 35.3 | 0.637 | 1.8 | 0.470 | 1.3 | 0.791 | 2.2 |
| Control 3 | 107 | 2.18 | 2.0 | 0.641 | 0.6 | 2.71 | 2.5 |
# b. Linearity/assay reportable range:
Linearity was assessed in a study using serum and Li-Heparin plasma samples containing acetaminophen concentrations across the measuring range of the assay (5-200 $\mu$ g/mL acetaminophen). Samples were prepared by 13 dilutions of a high concentration sample with an acetaminophen free sample. Expected concentrations were determined based on the known value of the commercially available standard material values and the dilution factors. Samples were analyzed in triplicate. Percent recoveries relative to the expected concentrations are shown below for serum and Li-Heparin plasma. Results support linearity across the measuring range.
{6}
Serum
| Sample | Expected (μg/mL) | C501 mean (μg/mL) (μg/mL) | Absolute deviation (μg/mL) | % recovery of mean values |
| --- | --- | --- | --- | --- |
| 1 | 0 | 0.00 | 0.0 | - |
| 2 | 3 | 2.40 | -0.6 | - |
| 3 | 6 | 5.40 | -0.6 | 90 |
| 4 | 24 | 23.1 | - | 96 |
| 5 | 48 | 45.7 | - | 95 |
| 6 | 72 | 68.4 | - | 95 |
| 7 | 96 | 92.1 | - | 96 |
| 8 | 12 | 113 | - | 94 |
| 9 | 14 | 139 | - | 97 |
| 10 | 168 | 160 | - | 95 |
| 11 | 192 | 188 | - | 98 |
| 12 | 216 | 206 | - | 95 |
| 13 | 240 | 232 | - | 97 |
Li-Heparin plasma
| Sample | Expected (μg/mL) | c 501 mean (μg/mL) | Absolute deviation (μg/mL) | % recovery of mean values |
| --- | --- | --- | --- | --- |
| 1 | 0 | 0 | 0 | - |
| 2 | 3 | 2.4 | -0.6 | - |
| 3 | 6 | 5.1 | -0.9 | 85 |
| 4 | 24 | 21.7 | - | 90 |
| 5 | 48 | 44.3 | - | 92 |
| 6 | 72 | 65.4 | - | 91 |
| 7 | 96 | 89.0 | - | 93 |
| 8 | 120 | 110 | - | 92 |
| 9 | 144 | 132 | - | 92 |
| 10 | 168 | 155 | - | 92 |
| 11 | 192 | 180 | - | 94 |
| 12 | 216 | 194 | - | 90 |
| 13 | 240 | 228 | - | 95 |
Dilution Accuracy:
The sponsor provided data supporting dilution accuracy (re-run function) by comparing manually diluted serum and Li-heparin plasma samples to expected values (5-fold dilution). Each sample was diluted with NaCl. Data is provided in the table below.
{7}
| Sample type | Replicate | Expected acetaminophen concentration (μg/mL) | Auto rerun acetaminophen concentration (μg/mL) | Mean Expected value | Mean autorerun value | % Recovery |
| --- | --- | --- | --- | --- | --- | --- |
| Serum | 1 | 234 | 232.0 | 224.7 | 230.7 | 102.7 |
| | 2 | 224 | 226.7 | | | |
| | 3 | 216 | 233.4 | | | |
| Serum | 1 | 510 | 566.8 | 535.7 | 561.3 | 104.8 |
| | 2 | 551 | 560.1 | | | |
| | 3 | 546 | 557.0 | | | |
| Serum | 1 | 839 | 871.3 | 839.0 | 887.1 | 105.7 |
| | 2 | 817 | 907.4 | | | |
| | 3 | 861 | 882.5 | | | |
| Plasma | 1 | 201 | 209.7 | 206.7 | 212.0 | 102.6 |
| | 2 | 213 | 213.7 | | | |
| | 3 | 206 | 212.6 | | | |
| Plasma | 1 | 614 | 629.2 | 611.7 | 625.7 | 102.3 |
| | 2 | 615 | 614.2 | | | |
| | 3 | 606 | 633.7 | | | |
| Plasma | 1 | 773 | 852.9 | 794.3 | 854.7 | 107.6 |
| | 2 | 785 | 873.2 | | | |
| | 3 | 825 | 838.1 | | | |
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The cobas c Acetaminophen Gen.2 assay is calibrated using a calibrator that was previously cleared under k002974. The Roche ACET2 calibrator is prepared gravimetrically from commercially available purified materials. See k002974 for traceability and stability information.
d. Detection limit:
LoB, LoD, and LoQ studies were performed based upon CLSI EP17-A2.
**LoB Protocol**
One analyte-free sample was measured in replicates of 5 for 2 runs on 2 analyzers for 3 days for each of 3 reagent lots (for total n=60 per lot). The LoB was determined to be 1.5 μg/mL.
**LoD Protocol**
Five low-analyte samples spiked with acetaminophen were measured with 1 determination per run (2 runs) on two analyzers with 3 reagent lots for six runs per day across three days. This study yielded a LoD of 3 μg/mL.
{8}
8
# LoQ Protocol
A set of six low-level samples was measured in two aliquots using 3 lots over at least 6 days (n=12 for each low-level sample). The total number of measurements made for each lot was 72.
Bias was determined by the difference between average and target values. The target values were determined by a chromatography system. Standard deviations were determined over all replicates for each lot. The following results were observed.
| Lot | Mean value (ug/mL) (C501) | SD (ug/mL) | Absolute bias compared to target value (ug/mL) |
| --- | --- | --- | --- |
| 1 | 4.73 | 0.123 | 0.367 |
| 2 | 4.63 | 0.129 | 0.475 |
| 3 | 4.88 | 0.221 | 0.217 |
The LoQ was determined to be 5 µg/mL.
## e. Analytical specificity:
### i. Interference from endogenous substances/cross reactants
Cross reactivity with compounds of similar structure were evaluated by using serum sample pools, with two acetaminophen target concentrations (~5.0 and ~30.0 µg/mL). Test samples included the acetaminophen plus potential cross-reactant (at the concentrations shown below). Control samples contained corresponding concentrations of acetaminophen without potential interferents. The table below provides the % cross reactivity for each of the compounds tested.
| Compound | Compound Concentration (µg/mL) | Acetaminophen (µg/mL) | % cross reactivity |
| --- | --- | --- | --- |
| Acetaminophen cysteine | 100 | 6.1 | 0.5 |
| Acetaminophen glucuronide | 1000 | 5.2 | n.d.* |
| Acetaminophen mercapturate | 300 | 5.4 | 0.2 |
| Acetaminophen sulfate | 200 | 6.1 | n.d.* |
| Phenacitin | 500 | 6.7 | 0.5 |
| Acetaminophen cysteine | 100 | 29.2 | -0.3 |
| Acetaminophen glucuronide | 1000 | 25.4 | -0.1 |
{9}
The n.d.* stands for 'not detectable'.
The assay was also evaluated with hemoglobin (hemolysis), lipids (lipemia), and bilirubin (Icterus) for potential interferences with the measurement of acetaminophen using serum sample pools. Evaluation of interference included normal levels of bilirubin and lipids (triglycerides). Recoveries of acetaminophen were within the sponsor's acceptance criteria of $\pm 10\%$ . No interference was noted when tested up to the concentrations shown in the table below. The H and I index values in the table are equivalent to $\sim 1\mathrm{mg / dL}$ .
| | Conc. Of Acetaminophen (μg/mL) | No interference up to |
| --- | --- | --- |
| Lipemia Level 1 | 6.3 | 1100 mg/dL |
| Lipemia Level 2 | 29.7 | 997 mg/dL |
| Lipemia Level 3 | 52.7 | 810 mg/dL |
| Lipemia Level 4 | 104 | 698 mg/dL |
| Lipemia Level 5 | 149 | 682 mg/dL |
| Lipemia Level 6 | 178 | 678 mg/dL |
| Hemolysis Level 1 | 5.8 | 926 H index |
| Hemolysis Level 2 | 28.2 | 936 H index |
| Hemolysis Level 3 | 46.1 | 1009 H index |
| Hemolysis Level 4 | 91.8 | 1008 H index |
| Hemolysis Level 5 | 135 | 1016 H index |
| Hemolysis Level 6 | 174 | 817 H index |
| Unconjugated bilirubin Level 1 | 6.2 | 47 I index |
| Unconjugated bilirubin Level 2 | 30.2 | 45 I index |
| Unconjugated bilirubin Level 3 | 51.1 | 63 I index |
| Unconjugated bilirubin Level 4 | 99.1 | 63 I index |
| Unconjugated bilirubin Level 5 | 150 | 63 I index |
| Unconjugated bilirubin Level 6 | 178 | 63 I index |
| Conjugated bilirubin Level 1 | 6.5 | 45 I index |
| Conjugated bilirubin | 30.3 | 46 I index |
{10}
# ii. Interference from common drugs
Interference from common drugs was also examined. A total of 24 commonly used drugs were examined for potential interferences. Testing was performed with serum sample pools at two levels of acetaminophen ( $\sim 7$ and $\sim 30~\mu \mathrm{g / mL}$ ). Recoveries of acetaminophen were within the sponsor's acceptance criteria of $\pm 10\%$ . The highest concentrations of common drugs shown not to interfere with the assay are provided in the table below.
| Drug | Highest Concentration Shown Not to Interfere with ACET2 |
| --- | --- |
| Acetylcysteine | 1663 μg/mL |
| Acetylsalicylic acid | 1000 mg/L |
| Ampicillin-sodium | 1000 mg/L |
| Ascorbic acid | 300 mg/L |
| Cefoxitin | 2500 mg/L |
| Cyclosporine | 5 mg/L |
| Doxycycline | 50 mg/L |
| Phenylbutazone | 400 mg/L |
| Rifampicin | 64 mg/L |
| Theophylline | 100 mg/L |
| Amitriptyline | 277 μg/mL |
| Caffeine | 1000 μg/mL |
| Codeine | 1.6 μg/mL |
| Diazepam | 5.1 μg/mL |
| Heparin | 5000 U/mL |
| Ibuprofen | 500 mg/L |
| Levodopa | 20 mg/L |
| Methyldopa | 20 mg/L |
| Metronidazole | 200 mg/L |
| Methionine | 1000 μg/mL |
| Phenylephrine | 20 μg/mL |
| Propoxyphene | 20 μg/mL |
| Salicylate | 1000 μg/mL |
{11}
| Secobarbital | 22 μg/mL |
| --- | --- |
| Imipramine | 280 μg/mL |
| Cysteine | 1300 μg/mL |
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
A method comparison test was performed on the Roche/Hitachi cobas c 501 analyzer by testing 117 human serum samples in singlicate on the Acetaminophen Gen.2 assay and the predicate device. The samples ranged in concentration from 5.2 to 198 $\mu \mathrm{g} / \mathrm{mL}$ . An additional set of eleven native serum samples from individual patients taking acetaminophen were spiked with a stock solution of $4\mathrm{mg / mL}$ acetaminophen in NaCl so concentrations were above the measuring range of the assay. These samples were spiked with a minimal volume of the stock solution, so that dilution effects were minimized. Deming regression analysis resulted in a slope of 1.02, a y-intercept of -0.699, and a correlation coefficient of 0.997. Ninety-five percent confidence intervals for the y-intercept and slope were -0.981 to -0.417 and 1.005 to 1.030, respectively.
b. Matrix comparison:
Matrix comparison studies were performed for Li-Heparin plasma vs serum, $\mathrm{K}_2$ -EDTA plasma vs serum, and $\mathrm{K}_3$ -EDTA plasma vs. serum. A total of 50 tubes were collected per anticoagulant and percent recovery was determined. All percent recoveries were within $\pm 10\%$ .
| anticoagulants | Sample concentration range tested (μg/mL) | Claimed Measuring Range (μg/mL) |
| --- | --- | --- |
| Li-Heparin | 2.6 - 187 | 5 – 200 |
| K2-EDTA (full) | 2.6 - 187 | |
| K3-EDTA (full) | 2.6 - 187 | |
In addition, the following linear regression statistics were obtained (Passing/Bablok):
| Matrix | slope | y-intercept | r value |
| --- | --- | --- | --- |
| Lithium-heparin vs serum | 0.989 | 0.0891 | 0.998 |
| K2-EDTA vs serum | 1.000 | 0.000 | 0.998 |
| K3-EDTA vs serum | 0.992 | -0.163 | 0.998 |
{12}
12
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
Normal therapeutic doses of acetaminophen result in serum concentrations of 10-30 µg/mL (66-199 µmol/L) in healthy adults. (From Jacobs DS, DeMott WR, Grad HI, et al. Laboratory Test Handbook 5th ed. Stow, Ohio: Lexi Comp Inc 2001;768).
Toxic manifestations have been observed at serum concentrations > 100 µg/mL, however the toxic range is generally reported at > 200 µg/mL. Toxic concentrations can be more effectively related to post dose interval; > 200, > 100, and > 50 µg/mL serum concentrations at 4, 8, and 12 hours post dose, respectively. (Rumack BH. Acetaminophen overdose. Arch Intern Med 1981;141:380)
Each laboratory should investigate the transferability of the expected values of its own patient population and if necessary determine its own reference ranges.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.