UNICON (ETAFILCON A) SOFT (HYDROPHILIC) CONTACT LENS WITH UV BLOCKER FOR DAILY WEAR
Applicant
Unicon Optical Co., Ltd.
Product Code
LPL · Ophthalmic
Decision Date
Oct 17, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
The UNICON (Etafilcon A) Soft (Hydrophilic) Contact Lens with UV Blocker for Daily Wear is indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic or non-aphakic persons with non-diseased eyes who may have 1.00D or less astigmatism. The UNICON (Etafilcon A) Soft (Hydrophilic) Contact Lens with UV Blocker for Daily Wear are indicated for single-use disposable wear and are to be discarded after each removal. The UNICON (Etafilcon A) Soft (Hydrophilic) Contact Lens with UV Blocker for Daily Wear helps protect against transmission of harmful UV radiation to the cornea and into the eye.
Device Story
Aspherical soft contact lens; manufactured via cast-molding; composed of etafilcon A (HEMA/MAA copolymer) with UV absorber and Blue 15:3 tint. Designed for daily wear; single-use disposable. Prescribed by eye care professionals for patients with myopia/hyperopia and low astigmatism. Lens provides refractive correction and UV radiation protection for the cornea. Device is non-electronic; no software or algorithms involved.
Clinical Evidence
No new clinical data provided. Safety and effectiveness were established based on previous clinical performance testing of etafilcon A lenses and comprehensive non-clinical bench testing, including biocompatibility, sterilization, oxygen permeability, and mechanical properties.
Technological Characteristics
Material: Etafilcon A (HEMA/MAA copolymer cross-linked with TMPTMA and EGDMA). Water content: 58%. Manufacturing: Cast-molding. UV protection: Benzotriazole absorber. Tint: Blue 15:3. Sterilization: Moist heat (overkill method per EN/ISO 17665:2006). Dimensions: 7.85-10.0mm base curve, 12.0-15.0mm diameter. Non-electronic; no software.
Indications for Use
Indicated for daily wear correction of myopia and hyperopia in aphakic or non-aphakic persons with non-diseased eyes and <= 1.00D astigmatism. Single-use disposable.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
VISTAKON (Etafilcon A) Soft (hydrophilic) Contact Lens (spherical), Clear and Visibility Tinted with UV Blocker for Daily Disposable Wear (K051900)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized human figure in profile, with three overlapping heads suggesting a sense of community and interconnectedness.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 17, 2014
Unicon Optical Co., Ltd. % Mr. Michael Lee President No. 45, Mingsheng Rd., Danshui District New Taipei, 251 TW
Re: K141917
> Trade/Device Name: Unicon (Etafilcon A) Soft (hydrophilic) Contact Lens with UV Blocker for Daily Wear Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: July 8, 2014 Received: July 15, 2014
Dear Mr. Lee.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141917
#### Device Name
Unicon (etafilcon A) Soft (hydrophilic) Contact Lens with UV Blocker for Daily Wear
#### Indications for Use (Describe)
The UNICON (etafilcon A) Soft (hydrophilic) Contact Lens with UV Blocker for Daily Wear is indicated for the correction of ametropia (myopia and hyperopia) in aphakic persons with non-diseased eye who may have 1.00 D or less astigmatism. The UNICON (etafilcon A) Soft (hydrophilic) Contact Lens with UV Blocker for Daily Wear are single disposable wear and are to be discarded after each removal. UNICON (etaffilcon A) Soft (hydrophilic) Contact Lens with UV Blocker for Daily Wear helps protect against transmission of harmful UV radiation to the cornea and into the eye.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|--------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
UNICON (Etafilcon A) Soft (Hydrophilic) Contact Lens with UV Blocker for Daily Wear
#### 510(k) Summary
| 5.1 | Type of Submission: | Traditional |
|-----|---------------------|------------------------------------------------------------------------------------------------------|
| 5.2 | Preparation Date: | 7th July, 2014 |
| 5.3 | Submitter: | UNICON Optical Co., LTD. |
| | Address: | No 16, Gongye E. 9th Rd., Hsinchu Science Park,<br>Baoshan Township, Hsinchu County 30075,<br>Taiwan |
| | Phone: | +886-3-5775586 |
| | Fax: | +886-3-5777868 |
| | Contact: | Robert |
#### 5.4 Identification of the Device:
| Proprietary/Trade name: | UNICON (Etafilcon A) Soft (Hydrophilic) Contact Lens with UV Blocker for Daily Wear |
|-------------------------|-------------------------------------------------------------------------------------|
| Common Name: | Soft (hydrophilic) contact lenses for daily wear |
| Classification Name: | LENS, CONTACT, (DISPOSABLE) |
| Device Classification: | II |
| Regulation Number: | 886.5925 |
| Panel: | Ophthalmic |
| Product Code: | MVN ; LPL |
#### 5.5 Identification of the Predicate Device:
| Predicate Device<br>Name: | VISTAKON (Etafilcon A) Soft (hydrophilic)<br>Contact Lens (spherical), Clear and Visibility<br>Tinted with UV Blocker for Daily Disposable<br>Wear |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | VISTAKON |
| Product Code: | MVN ; LPL |
| 510(k) Number: | K051900 |
{4}------------------------------------------------
#### 5.6 Intended Use and Indications for Use of the subject device.
The UNICON (Etafilcon A) Soft (Hydrophilic) Contact Lens with UV Blocker for Daily Wear is indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic or non-aphakic persons with non-diseased eyes who may have 1.00D or less astigmatism. The UNICON (Etafilcon A) Soft (Hydrophilic) Contact Lens with UV Blocker for Daily Wear are indicated for single-use disposable wear and are to be discarded after each removal. The UNICON (Etafilcon A) Soft (Hydrophilic) Contact Lens with UV Blocker for Daily Wear helps protect against transmission of harmful UV radiation to the cornea and into the eye.
#### 5.7 Device Description
UNICON (Etafilcon A) Soft (Hydrophilic) Contact Lens with UV Blocker for Daily Wear is available as aspherical lenses manufactured by case-molding method. The model illuminated with high water content (58 %). The hydrogel lens' material is a random copolymer composed of 2-hydroxyethyl methacrylate (HEMA) and methacrylic acid (MAA), which was cross-linked with Trimethylolpropane trimethacrylate (TMPTMA) and Ethylene Glycol Dimethacrylate (EGDMA) via UV photo-polymerization. The UNICON (Etafilcon A) Soft (Hydrophilic) Contact Lens with UV Blocker for Daily Wear with visible tint is light tinted blue using Blue 15:3 to make the lens more visible for handling. The lenses also contain a UV absorber { { 2-[ 3-( 2 -( 2 -( 2 -( ) - 4 - h dr ox y p hen y l which is included during the manufacturing process as a monomer to block UV radiation. The average transmittance characteristics are less than 5% in the UVB range of 280 to 315 nm and less than 30% in the UVA range of 316 to 380nm.
With more specific explanation, the "UNICON (Etafilcon A) Soft (Hydrophilic)
{5}------------------------------------------------
Contact Lens with UV Blocker for Daily Wear' is 58% water and 42% HEMA/MAA cast-molding aspherical soft contact lens. It is blister packaged and produced from +4.00D to -6.00D in 0.25D per step, from +4.50D to +6.00D in 0.50D per step, and from -6.50D to -13.00D in 0.50D per step.
#### 5.8 Non-clinical Testing
A series of non-clinical studies were completed on this product. The tests and studies were performed in UNICON QC lab and subcontracted test services which were according to the FDA guidance (Premarket Notification (510(k)) Guidance Document for Daily Wear Contact Lenses, February 27, 1997) and related recognized consensus standards. All the test results were met the requirements of products specification.
The following tests were conducted on "UNICON (Etafilcon A) Soft (Hydrophilic) Contact Lens with UV Blocker for Daily Wear'':
- . Sterilization Test
- o Shelf Life Test
- Biocompatibility
- -Cytotoxicity
- Irritation -
- -Acute Systemic toxicity
- Water Content
- . Mechanical Properties of Materials
- . Refractive Index
- Oxygen Permeability
- Transmittance
- Extractables Test
{6}------------------------------------------------
- . Specific Gravity
The Sterilization Method used on "UNICON (Etafilcon A) Soft (Hydrophilic) Contact Lens with UV Blocker for Daily Wear" is moist heat sterilization and the study for sterilization is based on the EN/ISO 17665 : 2006 Sterilization of health care product-Moist heat-Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices. The overkill method is used as validation method for sterilization and the results demonstrates achievement of the required sterility assurance limit (SAL) of 10 °.
#### 5.9 Clinical Testing
The safety and effectiveness of etafilcon A lenses have been established through previous clinical performance testing. And the pre-clinical testing results are sufficient in establishing substantial equivalence.
#### 5.10 EMC and Electrical safety
The devices do not require EMC/Electrical Safety evaluation.
#### 5.11 Substantial Equivalence Determination
The UNICON (Etafilcon A) Soft (Hydrophilic) Contact Lens with UV Blocker for Daily Wear submitted in this 510(k) file is substantially equivalent in intended use, main materials, design, safety and performance claims to the cleared VISTAKON (Etafilcon A) Soft (hydrophilic) Contact Lens (spherical), Clear and Visibility Tinted with UV Blocker for Daily Disposable Wear which is the subject of K051900. Differences between the devices cited in this section do not raise any new issues of safety or effectiveness.
{7}------------------------------------------------
UNICON (Etafilcon A) Soft (Hydrophilic) Contact Lens with UV Blocker for Daily Wear
| | Proposed Device | Predicate Device |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name | UNICON (Etafilcon A) Soft<br>(Hydrophilic) Contact Lens<br>with UV Blocker for Daily<br>Wear | VISTAKON (Etafilcon A) Soft<br>(hydrophilic) Contact Lens<br>(spherical), Clear and Visibility<br>Tinted with UV Blocker for<br>Daily Disposable Wear |
| Regulatory<br>Number | 886.5925 | 886.5925 |
| Classification | II | II |
| Intended Use | The UNICON (Etafilcon A)<br>Soft (Hydrophilic) Contact Lens<br>with UV Blocker for Daily<br>Wear is indicated for daily wear<br>for the correction of ametropia<br>(myopia and hyperopia) in<br>aphakic or non-aphakic persons<br>with non-diseased eyes who<br>may have 1.00D or less<br>astigmatism. The UNICON<br>(Etafilcon A) Soft (Hydrophilic)<br>Contact Lens with UV Blocker<br>for Daily Wear are indicated for<br>single-use disposable wear and<br>are to be discarded after each<br>removal. The UNICON<br>(Etafilcon A) Soft (Hydrophilic)<br>Contact Lens with UV Blocker<br>for Daily Wear helps protect<br>against transmission of harmful<br>UV radiation to the cornea and<br>into the eye. | The VISTAKON (Etafilcon A)<br>Soft (hydrophilic) Contact Lens<br>(spherical), Clear and Visibility<br>Tinted with UV Blocker for Daily<br>Disposable Wear for the<br>correction of refractive ametropia<br>(myopia and hyperopia) in<br>aphakic or not-aphakic persons<br>with not diseased eyes who may<br>have 1.00D or less of<br>astigmatism.<br>VISTAKON (Etafilcon A) Soft<br>(hydrophilic) Contact Lens, Clear<br>and Visibility Tinted with UV<br>Blocker for Daily Disposable<br>Wear help protect against<br>transmission of harmful UV<br>radiation to the cornea and into<br>the eye.<br>The VISTAKON (Etafilcon A)<br>Soft (hydrophilic) Contact Lens,<br>Clear and Visibility Tinted with<br>UV Blocker for Daily Disposable<br>Wear is to be prescribed for Daily<br>Disposable Wear and is to be<br>discarded after each removal. |
| Prescription Use | Yes | Yes |
| Single Use | Yes | Yes |
{8}------------------------------------------------
| Material | Etafilcon A | Etafilcon A |
|-------------------------------------------------------|---------------------------------------------|--------------------------------------------------------------|
| Base Curve | 7.85mm to 10.0mm | 7.85mm to 10.0mm |
| Diameter | 12.0mm to 15.0mm | 12.0mm to 15.0mm |
| Water Content | 58% | 58% |
| UV<br>transmittance<br>@280~315 nm<br>@316~380 nm | Avg < 5%<br>Avg < 30% | Avg < 5%<br>Avg < 30% |
| UV absorbing | Benzotriazol | Benzotriazole |
| Refractive Index | 1.399 | 1.40 |
| Manufacturing<br>Method | Cast Molded | Cast Molded |
| Sterilization<br>method | Traditional moist heat | Traditional moist heat |
| Primary<br>packaging | PP container with aluminum<br>foil sealing | PP container with aluminum<br>foil sealing |
| Power | +6.00D to -13.00D | +20.00D to -20.00D |
| Specific Gravity | 1.14 | 0.98-1.12 |
| Tensile Strength | 1.66 Mpa | 2.00 Mpa |
| Break Elongation | 393% | 189% |
| Secant Modulus | 0.43 Mpa | 1.06 Mpa |
| Oxygen<br>permeability<br>(edged corrected)<br>@ 35°C | 30.8 x 10-11 (cm2/sec)<br>(mlO2/ml-mmHg) | 28 x 10-11 (cm2/sec)<br>(mlO2/ml-mmHg) |
| Visible Light<br>Transmittance | Minimum 90% | Minimum 85% |
| Color & Tint | Blue 15:3 | Reactive Blue Dye #4 |
| FDA 21 CFR<br>(Color additive) | 74.3045<br>[Phthalocyaninato(2-)]<br>copper | 73.3121<br>Poly(hydroxyethyl<br>methacrylate)-dye copolymers |
{9}------------------------------------------------
## 5.12 Similarity and differences
The difference between the proposed device and predicate device is color additive. The color additive used on proposed device is Blue 15:3 ([Phthalocyaninato (2-)]copper) which has already been approved by US FDA. According to FDA 21 CFR 874.3045, this color additive may be safely used for coloring contact lenses. Therefore, the difference of proposed device and predicate device did not raise any problems of safety or effectiveness. The proposed device is substantially equivalent to the predicate device in intended use, main materials, design, safety and performance claims.
### 5.13 Conclusion
After analyzing bench tests and clinical test, it can be concluded that UNICON (Etafilcon A) Soft (Hydrophilic) Contact Lens with UV Blocker for Daily Wear is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.