PRODIGY AUTOCODE EJECT BLOOD GLUCOSE MONITORING SYSTEM

K141914 · Ok Biotech Co., Ltd. · NBW · Apr 15, 2015 · Clinical Chemistry

Device Facts

Record IDK141914
Device NamePRODIGY AUTOCODE EJECT BLOOD GLUCOSE MONITORING SYSTEM
ApplicantOk Biotech Co., Ltd.
Product CodeNBW · Clinical Chemistry
Decision DateApr 15, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

Prodigy® AutoCode® Eject Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, palm, calf or thigh. The Prodigy® AutoCode® Eject Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The Prodigy® AutoCode® Eject Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Prodigy® AutoCode® Eject Blood Glucose Monitoring System should not be used for the diagnosis of, or screening for diabetes, or for neonatal use. The alternative site testing should be done only during steady-state times (when glucose is not changing rapidly). PRODIGY® No Coding Test Strips are intended for use with the PRODIGY® AutoCode® Eject blood glucose meter to measure the concentration of the blood glucose in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, palm, calf or thigh for self-testing at home. They are for testing outside the body (in vitro diagnostic use only). Do not use them for diagnosis of, or screening for diabetes or for testing on neonates. PRODIGY® No Coding Test Strips are used as an aid to monitor the effectiveness of diabetes control. This system contains a speaking function, but is not intended for use by the visually impaired.

Device Story

System measures capillary whole blood glucose via amperometric assay using glucose oxidase reagent on test strips; requires >0.7 µL sample. Meter processes electrical current proportional to glucose concentration; displays result; includes voice function; stores/averages results. Used by patients at home for diabetes management. Output informs patient of blood glucose level to monitor diabetes control. Not for diagnosis or neonatal use. Alternative site testing restricted to steady-state conditions. Single-patient use; includes disinfection protocols for infection control.

Clinical Evidence

Non-clinical testing performed per FDA guidance for In Vitro Diagnostic Test Systems. No specific clinical trial data or performance metrics (e.g., sensitivity, specificity) provided in the summary document.

Technological Characteristics

Electrochemical biosensor; glucose oxidase enzyme (Aspergillus niger); automatic temperature compensation via built-in thermistor; 0.7 μL sample volume; 20-600 mg/dL detection range; 20-60% HCT range; 6-second measurement time; 1.5V AAA battery powered; 450-test memory; speaking function; handheld form factor.

Indications for Use

Indicated for quantitative measurement of glucose in fresh capillary whole blood (fingertip, palm, forearm, upper arm, calf, thigh) for single-patient home use by people with diabetes to monitor glycemic control. Contraindicated for neonatal use, diabetes screening/diagnosis, and use on critically ill, shock, dehydrated, or hyper-osmolar patients.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized eagle or bird in flight, composed of three curved lines that suggest the shape of a human profile. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 15, 2015 OK BIOTECH CO., LTD. KE-MIN JEN OFFICIAL CORRESPONDENT NO. 91, SEC. 2, GONGDAOWU 5TH ROAD HSINCHU CITY 30070, CHINA (TAIWAN) Re: K141914 Trade/Device Name: Prodigy® Autocode Eject TM Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW. CGA Dated: March 1, 2015 Received: March 11, 2015 Dear Ke-min Jen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Stayce Beck -S For : Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K141914 ### Device Name PRODIGY® AutoCode Eject TM Blood Glucose Monitoring System ### Indications for Use (Describe) Prodigy® AutoCode® Eject Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, palm, calf or thigh. The Prodigy® AutoCode® Eject Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The Prodigy® AutoCode® Eject Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Prodigy® AutoCode® Eject Blood Glucose Monitoring System should not be used for the diagnosis of, or screening for diabetes, or for neonatal use. The alternative site testing should be done only during steady-state times (when glucose is not changing rapidly). PRODIGY® No Coding Test Strips are intended for use with the PRODIGY® AutoCode® Eject blood glucose meter to measure the concentration of the blood glucose in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, palm, calf or thigh for self-testing at home. They are for testing outside the body (in vitro diagnostic use only) . Do not use them for diagnosis of, or screening for dabetes or for testing on neonates. PRODIGY® No Coding Test Strips are used as an aid to monitor the effectiveness of diabetes control. This system contains a speaking function, but is not intended for use by the visually impaired. Type of Use (Select one or both, as applicable) _ Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. ### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows a logo with the letter K inside of an oval shape. Next to the K is the word "Kbiotech" in a stylized font. Below the logo is text in a non-English language, possibly Chinese or Taiwanese. The text appears to be the name of the company. TEL : 886-3-5160258 FAX : 886-3-5160028 Taiwan : China : TEL : 86-791-3899362 FAX : 86-791-3880131 > http://www.okbiotech.com E-mail:service@okbiotech.com # 5. 510(K) Summary of Safety and Effectiveness ## (Per 21 CFR 807.92) | Type Of 510(K) Submission | Traditional | |-------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Basis for the submission | A New Device | | Common Name Of The Proposed<br>Device | Blood Glucose Monitoring System | | Trade name | PRODIGY AutoCode Eject™ Blood Glucose<br>Monitoring System | | 510(K) Submitter | OK BIOTECH CO., LTD.<br>No. 91, Sec. 2, Gongdao 5th Road, 30070,<br>Hsinchu City, Taiwan<br>Telephone: +886-3-516-0258<br>Fax:+886-3-516-0028<br>Email: service@okbiotech.com | | Owner Number | 9090860 | | Date prepared | April 14, 2015 | | Official Correspondent | Dr. JEN, KE-MIN<br>TEL: 886-3-5208829 FAX: 886-3-5209783<br>Email: ceirs.jen@msa.hinet.net | | Preference For Continued<br>Confidentiality (21 CFR 807.95) | 510(k) Summary | | Classification Regulation | SYSTEM, TEST, BLOOD GLUCOSE, OVER<br>THE COUNTER<br>(21 CFR 862.1345) | | Class | II | | Panel | Clinical Chemistry | | Product Code | NBW | | Predicate Device | PRODIGY Preferred® Blood Glucose Monitoring<br>System | #### ● Intended Use: Prodigy® AutoCode® Eject Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, palm, calf or thigh. The Prodigy® AutoCode® Eject Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The Prodigy® AutoCode® Eject Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Prodigy® AutoCode® Eject Blood Glucose Monitoring System should not be used for the diagnosis of, {4}------------------------------------------------ Image /page/4/Picture/0 description: The image contains a logo with the letters 'K' and 'b' intertwined, followed by the word 'biotech' in a bold, sans-serif font. Below the word 'biotech', there is a line of text in a different language, possibly Chinese or Japanese, indicating the company's name or description. The logo appears to be for a biotechnology company, given the use of the term 'biotech'. E-mail:service@okbiotech.com or screening for diabetes, or for neonatal use. The alternative site testing should be done only during steady-state times (when glucose is not changing rapidly). PRODIGY® No Coding Test Strips are intended for use with the PRODIGY® AutoCode® Eject blood glucose meter to measure concentration the of blood glucose in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, palm, calf or thigh for self-testing at home. They are for testing outside the body (in vitro diagnostic use only) . Do not use them for diagnosis of, or screening for diabetes or for testing on neonates. PRODIGY® No Coding Test Strips are used as an aid to monitor the effectiveness of diabetes control. This system contains a speaking function, but is not intended for use by the visually impaired. #### ● Device Description: The Prodigy® AutoCode® Eject Blood Glucose Monitoring System consists of a meter and Prodigy No Coding Test Strips. The system utilizes an electrochemical method-based meter and dry reagent biosensor (test strips) for blood glucose testing. The size of the current is proportional to the amount of glucose present in the sample, providing a quantitative measurement of glucose in fresh whole blood and control solutions. The Prodigy® AutoCode® Eject Blood Glucose Monitoring System is marketed as a meter only with a carrying case, batteries, Owner's Manual, Quick Reference Guide, Logbook, and Warranty Card. The Prodigy® AutoCode® Eject Blood Glucose Monitoring System is also marketed as a meter kit with a carrying case, batteries, Owner's Manual, Quick Reference Guide, Logbook, and Warranty Card, Prodigy Lancing Device, Prodigy Lancets, Prodigy No Coding Test Strips, and Control Solution. The Prodigy No Coding Test Strips utilizes the active enzyme is Glucose Oxidase, derived from Aspergillus niger. The Prodigy® AutoCode® Eject Blood Glucose Monitoring System has a speaking function. #### ● Test Principle The Blood glucose test is based on the measurement of electrical current generated by the reaction of capillary whole blood glucose with glucose oxidase on the test strip. The meter measures the strength of the current which is proportional to the concentration of glucose present and displays the corresponding blood glucose level. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Kbiotech. The logo features a stylized letter K enclosed in a circular design, followed by the word "biotech" in a bold, sans-serif font. Below the word "biotech" is text in a different language, possibly Chinese or Japanese. http://www.okbiotech.com E-mail:service@okbiotech.com #### ● Comparison Table | Comparison Items | Subject device | Predicate device | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | MANUFACTURER | OK Biotech Co., Ltd. | Prodigy Diabetes Care, LLC | | BRAND NAME | Prodigy | Prodigy | | Model Number | AutoCode EjectTM | Preferred® | | Trade Name | Prodigy® AutoCode EjectTM<br>Blood Glucose Monitoring<br>System | Prodigy Preferred® Blood<br>Glucose Monitoring System | | Product Code | NBW | NBW | | 510K NO | K141914 | K122338 | | Similarities | | | | Indications for use | Prodigy® AutoCode® Eject<br>Blood Glucose Monitoring<br>System is intended to be used<br>for the quantitative<br>measurement of glucose (sugar)<br>in fresh capillary whole blood<br>samples drawn from the<br>fingertips, forearm, upper arm, | The Prodigy Preferred Blood<br>Glucose Monitoring System is<br>intended to be used for the<br>quantitative measurement of<br>glucose (sugar) in fresh<br>capillary whole blood samples<br>drawn from the fingertips,<br>forearm, upper arm, palm, calf | | | Prodigy® AutoCode® Eject<br>Blood Glucose Monitoring<br>System is intended to be used<br>by a single person and should<br>not be shared. The Prodigy®<br>AutoCode® Eject Blood<br>Glucose Monitoring System is<br>intended for self-testing outside<br>the body (in vitro diagnostic<br>use) by people with diabetes at<br>home as an aid to monitor the<br>effectiveness of diabetes<br>control. The Prodigy®<br>AutoCode® Eject Blood<br>Glucose Monitoring System<br>should not be used for the<br>diagnosis of, or screening for<br>diabetes, or for neonatal use.<br>The alternative site testing<br>should be done only during<br>steady-state times (when<br>glucose is not changing<br>rapidly).<br>PRODIGY® No Coding Test<br>Strips are intended for use with<br>the PRODIGY® AutoCode®<br>Eject blood glucose meter to<br>measure concentration the of<br>blood glucose in fresh capillary<br>whole blood samples drawn<br>from the fingertips, forearm,<br>upper arm, palm, calf or thigh<br>for self-testing at home. They<br>are for testing outside the body<br>(in vitro diagnostic use only) .<br>Do not use them for diagnosis<br>of, or screening for diabetes or<br>for testing on neonates.<br>PRODIGY® No Coding Test<br>Strips are used as an aid to<br>monitor the effectiveness of<br>diabetes control.<br>This system contains a<br>speaking function, but is not<br>intended for use by the visually<br>impaired. | E-mail:service@okbiotech.com<br>Blood Glucose Monitoring<br>System is intended to be used<br>by a single person and should<br>not be shared.<br>The Prodigy Preferred Blond<br>Glucose Monitoring System is<br>intended for self testing outside<br>the body (in vitro diagnostic<br>use) by people with diabetes at<br>home as an aid to monitor the<br>effectiveness of diabetes<br>control. The Prodigy Preferred<br>Blood Glucose Monitoring<br>System should not be used for<br>the diagnosis of or screening of<br>diabetes or for neonatal use.<br>Alternative site testing should<br>be done only during steady<br>-state times (when glucose is<br>not changing rapidly). | | Test Principle | The test is based on the | Same | | current generated by the<br>reaction of capillary whole<br>blood glucose with glucose<br>oxidase on the test strip. The<br>meter measures the strength<br>of the current which is<br>proportional to the<br>concentration of glucose<br>present and displays the<br>corresponding blood glucose<br>level. | | E-mail:service@okbiotech.com | | Enzyme | Glucose oxidase | Same | | Specimen Type | Capillary whole blood from<br>fingertip and alternative sites<br>(palm, forearm, upper-arm, calf<br>and thigh) | Same | | Test Strip | PRODIGY® No-Coding Test<br>Strips | Same | | Control solution | PRODIGY® Control Solution<br>(Level 1 & Level 2) | Same | | Sample Volume | 0.7 μL | Same | | Operating<br>Temperature | 50 °F - 104 °F<br>10~85% R. H. | Same | | Strip Storage<br>Temperature | 39.2 - 104 °F<br>10~85% R. H. | Same | | HCT Range | 20 ~ 60 % | Same | | Detecting range | 20~600 mg/dL | same | | code-checking<br>mechanism | Code number checking | Same | | temperature<br>compensation<br>mechanism | Automatic compensation with<br>built-in thermistor | Same | | Differences | | | | Measuring Time | 6 seconds | 7 seconds | | Meter size | 100 mm (L) × 56 mm (W) × 23 | 71 mm (L) × 60 mm (W) × 19 | | | mm (H)…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...