When intended to provide stabilization as an adjunct to fusion of the cervical spine and occipitocervical junction (occiput-T3), the plate/rod, plates, rod, hook and screw (3.2 mm cortex) components of the Synthes CerviFix, Axon, OC Fusion and Synapse Systems are indicated for skeletally mature patients using allograft and/or autograft for the following: · Degenerative Disc Disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies) · Spondylolisthesis · Spinal Stenosis · Fracture/dislocation · Atlantoaxial fracture with instability · Occipitocervical dislocation · Revision of previous cervical spine surgery · Tumor When used to treat these cervical and occipitocervical conditions, screws are limited to occipital fixation only. The rod and hook components are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1-T3) spine. The rods, clamps, screws, nuts, variable axis screws, and transverse bars are intended to promote fusion following reduction of fracture/dislocation or trauma in the upper thoracic spine (T1-T3). The use of these screws (3.5 mm, 4.0 mm and 4.5 mm cancellous, and 3.5 mm and 4.2 mm cortex) is limited to placement in T1-T3 in treating thoracic conditions only. They are not intended to be placed in or treat conditions involving the cervical spine. The Synthes CerviFix, Axon, and Synapse Systems can also be linked to the Synthes Universal Spinal System using the 3.5 mm/6.0 mm and 4.0 mm/6.0 mm parallel connectors from that system and via the CerviFix tapered rods using lamina hooks, transverse process hooks, 4.2 mm screws, and the 5.0 mm/6.0 mm parallel connector. Warning: This device is not intended for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic (T4-T12), or lumbar spine. The Synthes OC Fusion System is intended to provide stabilization as an adjunct to fusion of the occipital-cervical junction. A complete occipital-cervical-thoracic construct can be created by using hooks (C1-T3) and screws (T1-T3) that have been previously cleared within the Synthes CerviFix System, Synthes Axon System, and Synthes Synapse System. The OC Fusion System is indicated for skeletally mature patients using allograft and/or autograft for the following indications: · Degenerative disc disease of the cervical vertebrae (neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies) · Spondylolisthesis · Spinal Stenosis · Fracture/dislocation · Atlanto/axial fracture with instability · Occipital-cervical dislocation · Revision of previous cervical spine surgery · Tumors (primary and metastatic) The use of screws is limited to placement in the occiput. Screws are not intended to be placed into the cervical spine.
Device Story
Synapse and OC Fusion Systems are spinal fixation implants; components include hooks, pedicle screws, rods, plates, clamps, nuts, and transverse bars. Used by surgeons in clinical settings to provide stabilization as an adjunct to spinal fusion in skeletally mature patients. System allows construction of occipital-cervical-thoracic constructs to fit patient anatomy. Implants are secured using Class I surgical instruments. Output is a rigid construct providing mechanical stability to the spine; intended to promote fusion and alleviate symptoms associated with degenerative, traumatic, or neoplastic conditions. Benefits include stabilization of unstable spinal segments and support for bone graft healing.
Clinical Evidence
Bench testing only. Mechanical evaluation included static compression bending, dynamic compression bending, and dynamic torsion testing per ASTM F2706-08. No clinical data presented.
Technological Characteristics
Materials: Titanium aluminum niobium (Ti-6Al-7Nb) per ASTM F1295-11 and ISO 5832-11. Components: Hooks, screws, rods, plates, clamps, nuts. Mechanical fixation principle. No software or energy source.
Indications for Use
Indicated for skeletally mature patients requiring stabilization as an adjunct to fusion of the cervical spine, occipitocervical junction, or upper thoracic spine (T1-T3). Conditions include DDD, spondylolisthesis, spinal stenosis, fracture/dislocation, atlantoaxial instability, occipitocervical dislocation, revision surgery, and tumors. Contraindicated for pedicle fixation in cervical, thoracic (T4-T12), or lumbar spine.
Regulatory Classification
Identification
A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with the first profile being the most prominent and the other two receding into the background. The profiles are connected by a flowing line that resembles a ribbon or wave. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 25, 2014
Synthes USA Products, LLC Mr. Mitch Ohiwa Manager, Regulatory Affairs 325 Paramount Drive Raynham, Massachusetts 02767
Re: K141897
Trade/Device Name: Synapse System and OC Fusion System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: KWP, MNH, MNI Dated: July 11, 2014 Received: July 14, 2014
Dear Mr. Ohiwa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
Page 2 - Mr. Mitch Ohiwa
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Ronald#Alean -S for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141897
Device Name Synapse System
Indications for Use (Describe)
These Systems are intended for the following:
Hooks, Plate/Rods, Plates, Rods and Screws
When intended to provide stabilization as an adjunct to fusion of the cervical spine and occipitocervical junction (occiput-T3), the plate/rod, plates, rod, hook and screw (3.2 mm cortex) components of the Synthes CerviFix, Axon, OC Fusion and Synapse Systems are indicated for skeletally mature patients using allograft for the following: · Degenerative Disc Disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies)
· Spondylolisthesis
- · Spinal Stenosis
- · Fracture/dislocation
- Atlantoaxial fracture with instability
- Occipitocervical dislocation
- · Revision of previous cervical spine surgery
- Tumor
When used to treat these cervical and occipitocervical conditions, screws are limited to occipital fixation only.
#### Hooks and Rods
The rod and hook components are also intended to provide stabilization following reduction of fracture/ dislocation or trauma in the cervical/upper thoracic (C1-T3) spine.
Rods, Clamps, Screws, Nuts, Variable Axis Screws, Locking Screws, and Transverse Bars
The rods, clamps, screws, nuts, variable axis screws, and transverse bars are intended to promote fusion following reduction of fracture/dislocation or trauma in the upper thoracic spine (T1-T3).
The use of these screws (3.5 mm, 4.0 mm cancellous, and 3.5 mm and 4.2 mm cortex) is limited to placement in T1-T3 in treating thoracic conditions only. They are not intended to be placed in or treat conditions involving the cervical spine.
The Synthes CerviFix, Axon, and Synapse Systems can also be linked to the Synthes Universal Spinal System using the 3.5 mm/6.0 mm and 4.0 mm/6.0 mm parallel connectors from that system and via the CerviFix tapered rods using lamina hooks, transverse process hooks, 4.2 mm screws, and the 5.0 mm/6.0 mm parallel connector.
Warning: This device is not intended for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic (T4-T12), or lumbar spine.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141897
Device Name OC Fusion System
#### Indications for Use (Describe)
The Synthes OC Fusion System is intended to provide stabilization as an adjunct to fusion of the occipital-cervical junction. A complete occipital-cervical-thoracic construct can be created by using hooks (C1-T3) that have been previously cleared within the Synthes CerviFix System, and Synthes Synapse System.
The OC Fusion System is indicated for skeletally mature patients using allograft and/or autograft for the following indications:
· Degenerative disc disease of the cervical vertebrae (neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies)
- · Spondylolisthesis
- · Spinal Stenosis
- · Fracture/dislocation
- Atlanto/axial fracture with instability
- · Occipital-cervical dislocation
- · Revision of previous cervical spine surgery
- · Tumors (primary and metastatic)
The use of screws is limited to placement in the occiput. Screws are not intended to be placed into the cervical spine.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(K) SUMMARY
## A. Submitter Information
| Submitter: | Synthes USA Products, LLC<br>325 Paramount Drive<br>Raynham, Massachusetts 02767 |
|---------------------------------------------|-------------------------------------------------------------------------------------------|
| Contact Person: | Mitch Ohiwa<br>DePuy Synthes Spine<br>325 Paramount Drive<br>Raynham, Massachusetts 02767 |
| Telephone Number:<br>Fax Number:<br>E-mail: | 508-828-3225<br>508-828-3267<br>mohiwa@its.jnj.com |
### B. Date Prepared
September 2014
# C. Device Name
| Trade/Proprietary Name: | Synapse System<br>OC Fusion System |
|----------------------------------------|---------------------------------------|
| Common/Usual Name: | Spinal interlaminal fixation orthosis |
| Device Classification and Regulation: | Class II, per 21 CFR 888.3050 |
| Subsequent Regulation: | 21 CFR 888.3070 |
| Classification Product and Panel Code: | KWP (Orthopedic) |
| Subsequent Product and Panel Codes: | MNH (Orthopedic)<br>MNI (Orthopedic) |
### D. Predicate Device Nameu
Rtlo ct { "Rtgf lecyg<" Synthes Synapse System (most recently K133698)
Cfffkkqpcn'Rtgf kecygu" Synthes OC Fusion System (most recently K091689)."cpf Synthes CerviFix/Axon (K023675, most recently K030377 for CerviFix System)
{6}------------------------------------------------
## E. Submission Purpose
The purpose of this submission is to include additional components in the Synapse and OC Fusion Systems.
## F. Device Description
The Synapse System is a hook, pedicle screw, and rod spinal system which is designed to provide spinal fixation to allow for stabilization of spinal segments in skeletally mature patients as an adjunct to fusion. The Synapse System is composed of multiple components to allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. The Axon System and CerviFix System also included in the indications for use statement are additional hook, pedicle screw, and rod spinal systems. These systems consist of anchors including bone screws, pedicle screws and hooks; interconnection mechanisms including locking screws, set screws, parallel rod connectors, clamps, transverse bars, rod collars and nuts; longitudinal members including rods and rod/plates; and transverse connectors including transconnectors.
The OC Fusion System consists of occipital plates, occipital screws, and rods intended to provide stabilization as an adjunct to fusion of the occipital-cervical junction. The OC Fusion System can be used with components from the Synthes CerviFix System including Axon and Synapse components to create a complete occipital-cervical-thoracic spinal construct.
Fixation of the rods included in the Synapse and OC Fusion Systems is achieved by using hooks from one of the aforementioned cleared systems. System components in the Synapse and OC Fusion Systems are implanted with Class I general use surgical instruments.
# G. Indication for Use
### Synapse System
# Hooks, Plate/Rods, Plates, Rods and Screws
When intended to provide stabilization as an adjunct to fusion of the cervical spine and occipitocervical junction (occiput-T3), the plate/rod, plates, rod, hook and screw (3.2 mm cortex) components of the Synthes CerviFix, Axon, OC Fusion and Synapse Systems are indicated for skeletally mature patients using allograft and/or autograft for the following:
- . Degenerative Disc Disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies)
- Spondylolisthesis
- Spinal Stenosis ●
- . Fracture/dislocation
- Atlantoaxial fracture with instability ●
- Occipitocervical dislocation
- Revision of previous cervical spine surgery
- . Tumor
When used to treat these cervical and occipitocervical conditions, screws are limited to occipital fixation only.
{7}------------------------------------------------
### Hooks and Rods
The rod and hook components are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1-T3) spine.
#### Rods, Clamps, Screws, Nuts, Variable Axis Screws, Locking Screws, and Transverse Bars
The rods, clamps, screws, nuts, variable axis screws, and transverse bars are intended to promote fusion following reduction of fracture/dislocation or trauma in the upper thoracic spine (T1-T3). The use of these screws (3.5 mm, 4.0 mm and 4.5 mm cancellous, and 3.5 mm and 4.2 mm cortex) is limited to placement in T1-T3 in treating thoracic conditions only. They are not intended to be placed in or treat conditions involving the cervical spine.
The Synthes CerviFix, Axon, and Synapse Systems can also be linked to the Synthes Universal Spinal System using the 3.5 mm/6.0 mm and 4.0 mm/6.0 mm parallel connectors from that system and via the CerviFix tapered rods using lamina hooks, transverse process hooks, 4.2 mm screws, and the 5.0 mm/6.0 mm parallel connector.
Warning: This device is not intended for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic (T4-T12), or lumbar spine.
#### OC Fusion System
The Synthes OC Fusion System is intended to provide stabilization as an adjunct to fusion of the occipital-cervical junction. A complete occipital-cervical-thoracic construct can be created by using hooks (C1-T3) and screws (T1-T3) that have been previously cleared within the Synthes CerviFix System, Synthes Axon System, and Synthes Synapse System.
The OC Fusion System is indicated for skeletally mature patients using allograft and/or autograft for the following indications:
- Degenerative disc disease of the cervical vertebrae (neck pain of discogenic origin with ● degeneration of the disc as confirmed by patient history and radiographic studies)
- Spondylolisthesis ●
- . Spinal Stenosis
- . Fracture/dislocation
- . Atlanto/axial fracture with instability
- . Occipital-cervical dislocation
- . Revision of previous cervical spine surgerv
- Tumors (primary and metastatic)
The use of screws is limited to placement in the occiput. Screws are not intended to be placed into the cervical spine.
#### H. Summary of Similarities and Differences in Technological Characteristics, Performance, and Intended Use
The proposed additional components to the Synapse and OC Fusion Systems share the same intended use, indications for use, and technological characteristics as the previously cleared components in the Synapse, OC Fusion, and CerviFix/Axon Systems.
{8}------------------------------------------------
### I. Materials
The proposed additional components to the Synapse and OC Fusion Systems are manufactured from titanium aluminum niobium (Ti-6Al-7Nb (TAN)). The same material is used in the components of the previously cleared Synapse. OC Fusion, and CerviFix/Axon Systems. The material conforms to ASTM F1295-11 Standard Specification for Wrought Titanium-6Aluminum-7Niobium Alloy for Surgical Implant Applications and ISO 5832-11 Implants for surgery - Metallic materials - Part 11: Wrought titanium 6-aluminum 7-niobium allov.
### J. Performance Data
The proposed additional components to the Synapse System required no additional mechanical testing. The proposed additional components to the OC Fusion System were evaluated in mechanical testing including static compression bending testing, dynamic compression bending testing, and dynamic torsion testing. The mechanical tests performed followed ASTM F2706-08, Standard Test Methods for Occipital-Cervical and Occipital-Cervical-Thoracic Spinal Implant Constructs in a Vertebrectomy Model. Test results and analyses showed that the proposed components to the Synapse and OC Fusion Systems were substantially equivalent to or better than the predicate devices.
### K. Conclusion
The substantial equivalence justification demonstrates that the proposed additional components to the Synapse and OC Fusion Systems are as safe, as effective, and perform as well as the predicate components in the Synapse, OC Fusion, and Axon Systems.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.