K141848 · Iogyn, Inc. · PGT · Aug 29, 2014 · Obstetrics/Gynecology
Device Facts
Record ID
K141848
Device Name
IOGYN SYSTEM
Applicant
Iogyn, Inc.
Product Code
PGT · Obstetrics/Gynecology
Decision Date
Aug 29, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.1710
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The IOGYN System is intended to distend the uterus by filling it with saline to facilitate viewing with a hysteroscope during diagnostic and operative hysteroscopy and provide fluid management through the closed-loop recirculation of filtered distension fluid. It is also intended for cutting and coagulation of uterine tissue such as intrauterine polyps and myomas using a bipolar resecting device.
Device Story
IOGYN System is a closed-loop hysteroscopic insufflator and cutter-coagulator. It uses a controller with two peristaltic pumps to manage saline infusion and aspiration, maintaining intra-uterine pressure (30-125 mmHg) via a pressure sensor feedback loop. A bipolar radiofrequency (RF) resecting device, operated via footswitch, performs tissue resection and coagulation. Resected tissue is aspirated through the device, filtered (removing particles, cellular material, proteins, and pathogens), and recirculated into a 3-liter saline bag. Used in clinical settings by physicians, the system provides real-time fluid management and visualization. The controller features a GUI for pressure settings and RF output control. Benefits include continuous, optically clear visualization and efficient tissue removal during hysteroscopic procedures.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993-1), sterility (ANSI/AAMI/ISO 11137-2), shelf-life (ASTM F1980), transit (ASTM D4169), and performance testing (simulated use, durability, hi-pot, tensile strength, flow rate, pressure testing, and dead volume). Results demonstrated that modified components meet design specifications and are suitable for intended use.
Technological Characteristics
System includes controller, footswitch, fluid management accessories (tubing/filters), and bipolar RF resecting device. Materials: Polyester and Parylene C (insulation), PVC (tubing). Energy: Bipolar RF. Connectivity: Electrical pressure sensor monitoring. Sterilization: Sterile single-use disposables. Software: Pressure control algorithm for peristaltic pumps.
Indications for Use
Indicated for patients undergoing diagnostic or operative hysteroscopy requiring uterine distension and resection/coagulation of intrauterine tissue (e.g., polyps, myomas).
Regulatory Classification
Identification
A closed loop hysteroscopic insufflator with cutter-coagulator is a prescription device configured for hysteroscopic insufflation, resection, and coagulation. It is used to perform diagnostic and surgical procedures (i.e., resection and coagulation). This device type contains a closed-loop recirculating fluid management system for the controlled delivery of filtered distension fluid. This device type also contains a bipolar radiofrequency device used in conjunction with a hysteroscope for resection and coagulation of intrauterine tissues.
Special Controls
In combination with the general controls of the Food, Drug & Cosmetic Act, the Closed Loop Hysteroscopic Insufflator with Cutter-coagulator is subject to the following special controls:
*Classification.* Class II (special controls). The special control(s) for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Software validation, verification, and hazard analysis must be provided.
(3) Electrical equipment safety, including appropriate thermal and mechanical safety and electromagnetic compatibility (EMC) testing must be performed.
(4) Device components that are labeled sterile must be validated to a sterility assurance level of 10
−6 .(5) Shelf-life testing that demonstrates the device packaging maintains sterility and the functionality of the device is maintained following simulated shipping and handling must be provided to support the proposed shelf life.
(6) Non-clinical testing data must demonstrate the performance characteristics of the device. Detailed protocols and the test reports must be provided for each test.
(i) The following tests must be performed for the resection portion of the device:
(A) Mechanical testing to assess critical joint strength.
(B) Device electrode temperature testing.
(C) Coagulation depth testing.
(D) Simulated use testing.
(E) Device durability testing.
(ii) The following tests must be performed for the fluid management portion of the device:
(A) Mechanical testing to assess tensile strength of connections.
(B) Pressure testing that demonstrates the following parameters, including accuracy of the pressure displayed; appropriate detection and response to overpressure conditions; activation of a secondary overpressure relief valve at the maximum safe level; and all accessories within the fluid path meet the pressure requirements.
(C) Fluid delivery volume testing that demonstrates that the maximum fluid volume delivered is below a predefined level.
(D) Flow rate testing.
(E) Simulated use testing.
(F) Filtration testing.
(G) Blood filtration capacity testing.
(H) Tissue collection capacity testing.
(I) Filtrate characterization and testing that demonstrates that the continuous reintroduction of filtrate into the uterus does not pose a safety risk.
(7) Clinician labeling must include:
(i) Specific instructions and the clinical training needed for the safe use of the device.
(ii) Appropriate warnings, precautions, and information related to overpressurization.
(iii) Appropriate EMC information.
(iv) An expiration date/shelf life.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are rendered in a dark color, contrasting with the white background of the seal.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 29, 2014
Iogyn, Inc. Rich Christensen Chief Operating Officer 20195 Stevens Creek Boulevard, Suite 120 Cupertino, CA 95014
Re: K141848 Trade/Device Name: Iogyn System Regulation Number: 21 CFR§ 884.1710 Regulation Name: Closed Loop Hysteroscope Insufflator with Cutter-coagulator Regulatory Class: II Product Code: PGT Dated: August 15, 2014 Received: August 18, 2014
Dear Rich Christensen,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Herbert P. Lerner -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### 7.0 INDICATIONS FOR USE STATEMENT
# Indications for Use Statement
| 510(k)<br>Number | To be determined. |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | IOGYN System |
| Indications<br>For Use | The IOGYN System is intended to distend the uterus by filling it<br>with saline to facilitate viewing with a hysteroscope during<br>diagnostic and operative hysteroscopy and provide fluid<br>management through the closed-loop recirculation of filtered<br>distension fluid. It is also intended for cutting and coagulation of<br>uterine tissue such as intrauterine polyps and myomas using a<br>bipolar resecting device. |
Prescription Use _____________________________________________________________________________________________________________________________________________________________
AND/OR Over-The-Counter
Use
(21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{4}------------------------------------------------
# 510(k) SUMMARY
## 510(k) Summary for IOGYN System
#### A. Sponsor
IOGYN, Inc.
#### B. Contact
Rich Christensen Chief Operating Officer IOGYN, Inc 20195 Stevens Creek Blvd., Suite 120 Cupertino, CA 95014
Date Prepared: July 16, 2014
#### C. Device Name
| Trade name: | IOGYN System |
|----------------------|------------------------------------------------------------------|
| Common/usual Name: | Closed Loop Hysteroscopic Insufflator with Cutter-<br>coagulator |
| Classification Name: | Closed Loop Hysteroscopic Insufflator with Cutter-<br>coagulator |
#### D. Predicate Device(s)
| Trade name: | IOGYN System |
|-------------------------|------------------------------------------------------------------|
| Common/usual Name: | Closed Loop Hysteroscopic Insufflator with Cutter-<br>coagulator |
| Classification Name: | Closed Loop Hysteroscopic Insufflator with Cutter-<br>coagulator |
| Premarket Notification: | K132695 |
#### E. Device Description
Device Name: IOGYN System
The IOGYN System is comprised of the following:
- IOGYN Controller with Integrated Fluid Management .
- Footswitch O
- Fluid Management Accessories o
- IOGYN Resecting Device ●
{5}------------------------------------------------
The IOGYN System provides an integrated control system with bipolar radiofrequency outputs (cut and coagulation) and fluid management through the use of two integrated peristaltic pumps. The Resecting Device is a bipolar radiofrequency device configured for the resection and aspiration of uterine pathology. Fluid infusion and aspiration of the uterine cavity are controlled by the IOGYN Controller's peristaltic pumps, in conjunction with the Fluid Management Accessories; these components form a closed-loop recirculating system. These integrated peristaltic pumps are operated by a software pressure control algorithm that measures and controls intra-uterine cavity pressure by varying saline infusion and aspiration rates in response to pressure changes during aspiration and tissue resection using the Resecting Device. The IOGYN Controller Graphical User Interface (GUI) has fluid control settings which allow the user to toggle infusion ON/OFF and to set the cavity pressure from 30-125mmHg.
The IOGYN System uses a closed-loop fluid management system with a single 3liter saline bag that continuously recirculates filtered distension fluid throughout the procedure. The fluid absorption is limited by the volume within the 3-liter saline bag minus the dead volume within the system, which limits the deliverable volume to less than 2.5L. The 2.5L of recirculated fluid is passed through two levels of filtration - the first level of filtration removes resected tissue and bulk particles; the second level of filtration removes cellular materials, hemoglobin, plasma proteins, cytokines, coagulation factors, bacteria and viruses. These two levels of filtration generate optically clear, sterile, filtered distension fluid which is returned to the 3 liter saline bag and recirculated.
## Components Description
The IOGYN System consists of the following components:
- o Component 1: IOGYN controller with Integrated Fluid Management
- Component 1a: Footswitch 0
- Component 1b: Fluid Management Accessories 0
- Component 2: Resecting Device ●
The IOGYN System is for use with the IOGYN Endoscope. Refer to IOGYN Endoscope Instructions.
{6}------------------------------------------------
Image /page/5/Figure/1 description: This image is labeled Figure 1: IOGYN System Components and IOGYN Endoscope. The image shows the IOGYN Controller with Integrated Fluid Management Component 1, Fluid Management Accessories Component 1b, FootSwitch Component 1a, Resecting Device Component 2, and the IOGYN Endoscope. All of these components are laid out on a white surface.
## The IOGYN Controller includes:
- A touch screen with an intuitive Graphical User Interface (GUI) including . user-adjustable cavity pressure control;
- RF cut and coagulation output for the IOGYN Resecting Device;
- o Two pumps for the controlled infusion and aspiration of fluids and tissue;
- o An electrical connection for pressure sensor monitoring of the uterine cavity pressure;
A footswitch connection with the IOGYN Controller for user activation of RF energy and aspiration; and Accessories (footswitch, Fluid Management Accessories):
Image /page/5/Figure/8 description: This image is titled "Figure 2: IOGYN Controller with Integrated Fluid Management". The image is split into two sections, the left side shows the IOGYN Controller, which includes a touch screen, infusion pump, aspiration pump, footswitch connector, resecting device connector, and pressure sensor connector. The right side of the image shows the footswitch, which includes the CUT pedal, COAG pedal, and aspiration button.
Component 1: The IOGYN Controller is provided non-sterile and is intended to be cleaned prior to use pursuant to provided instructions. (Figure 2).
{7}------------------------------------------------
Accessories to the IOGYN Controller consist of the (1) Footswitch and (2) the Fluid Management Accessories.
Component 1a: The Footswitch has a three button arrangement with individual pedals for cut, coagulation, and aspiration. (Figure 2). When acted upon by the user, it communicates with the IOGYN Controller to perform the designated function. The footswitch connects to the foot switch connector on the front of the IOGYN Controller.
Component 1b: The Fluid Management Accessories to the IOGYN Controller consist of infusion and aspiration tubing sets. (Figure 3). The infusion tubing set consists of a pressure sensor, an infusion tube and a disposable introducer seal for the endoscope. The infusion tube in conjunction with the infusion pump is responsible for insufflation of the uterine cavity via pressure feedback control from the pressure sensor to the IOGYN Controller. The aspiration tubing set consists of an aspiration tube, a container to collect tissue, a molecular filter and filter tubing. The aspiration tubing set in conjunction with the aspiration pump is responsible for the aspiration and filtration of uterine outflow and the subsequent recirculation into the 3-liter saline bag. The Fluid Management Accessories are provided sterile, as a single-use disposable device.
Image /page/6/Figure/4 description: This image is titled "Figure 3: Fluid Management Accessories" and shows two different fluid management accessories. On the left is the Fluid Management Accessories Infusion Tubing Set, which includes a pressure sensor, pressure sensor connector, saline spike with integrated float valve, introducer, luer connection for endoscope, and infusion tube. On the right is the Fluid Management Accessories Aspiration Tubing Set, which includes an aspiration tube, molecular filter, tissue catch tube, aspiration valve, tissue catch, filter tube, and saline spike.
Component 2: The Resecting Device is a sterile, single use disposable bipolar radiofrequency device for the resection and coagulation of tissue in a salineinsufflated environment. (Figure 4). The Resecting Device features an outer tube and an internal reciprocating electrode. The IOGYN Controller provides bipolar radiofrequency output to the Resecting Device sufficient for the cutting and
{8}------------------------------------------------
coagulation of tissue within the endoscopic environment. An internal full loop electrode reciprocates within the outer tube window, which is positioned at the distal tip of the device. When the RF Cut pedal is activated, the bipolar RF fullloop electrode moves distally to electro-surgically cut the tissue in the outer tube window. The resected tissue is simultaneously aspirated from the treatment site through the inner diameter of the full loop electrode, through the shaft and device handle area, and through the aspiration tube of the Fluid Management Accessories via the aspiration pump on the IOGYN Controller. When the RF Coagulation is activated, the bipolar RF full-loop electrode electro-surgically coagulates the tissue adjacent to the full loop electrode.
The Resecting Device is provided sterile, as a single-use disposable device.
Image /page/7/Figure/3 description: This image is labeled Figure 4: Resecting Device and shows two different views of the device. The left side of the image shows the resecting device with labels pointing to the device cable, device handle, and shaft. The right side of the image shows the resecting device distal tip with labels pointing to the outer tube, insulator, and full loop electrode.
The device description for the modified device is identical to the device description of the Predicate Device cleared per K132695 with the exception of the following modifications:
- 1. Resecting Device:
- · The Inner Tube Electrical Insulation material is being changed from FEP (Fluorinated Ethylene Propylene) to Polyester. The material change is being made to improve manufacturing reliability by use of a material that is stiffer and shrinks at a lower minimum temperature than FEP.
- · The Outer Tube Electrical Insulation material is being changed from Aesno Med and Trogamid (Nylon/Polyamide) to Parylene C. The material change is being made to increase manufacturing reliability by reducing a multi-component assembly of two insulation materials to a single vapor deposited conformal coating.
- 2. Fluid Management Accessories:
- · The Infusion Tube material and dimensions are being changed from AdvantPure AdvantaFlex (OD: 5/16" ID: 3/16") to polyvinyl chloride (PVC) (OD: .355" ID: 1/4") and the Aspiration Tube material and dimensions are being changed from PureWeld XL Thermoplastic Elastomer (TPE) (OD: 5/16" ID: 3/16") to polyvinyl chloride (PVC) (OD: .355" ID: 1/4"). The PVC tubing also contains two colorants: Ultramarine
{9}------------------------------------------------
Blue and Ultramarine Violet. The material changes are being made to reduce supply chain risk associated with use of multiple vendors and to optimize cost structure.
#### F. Intended Use
The IOGYN System is intended to distend the uterus by filling it with saline to facilitate viewing with a hysteroscope during diagnostic and operative hysteroscopy and provide fluid management through the closed-loop recirculation of distention fluid. It is also intended for cutting and coagulation of uterine tissue such as intrauterine polyps and myomas using a bipolar resecting device.
#### G. Technological Characteristics
The proposed Resecting Device and the Fluid Management Accessories (components of the IOGYN System) have the same fundamental design, operating principles and the intended/indications for use as the predicate device (K132695). The proposed device has differences in materials and dimensions when compared to the cleared device (K132695).
#### H. Substantial Equivalence
The indications for use, technology and principles of operation of the IOGYN System remain unchanged when compared to the IOGYN System cleared per K132695. Modifications described above were confirmed to be acceptable by means of biocompatibility evaluation and appropriate bench/performance testing. The modified device is substantially equivalent to the predicate device.
#### I. Performance Testing (Bench Evaluation)
As a part of the design control and material change process, Fluid Management Accessories and Resecting Device test samples assembled with the material changes were subjected to extensive testing at the system, component, and subassembly levels to ensure that the material changes did not alter device, component or system performance and that the IOGYN System met its performance specifications. The following testing was complete to evaluate the effects of the design change:
- o Biocompatibility testing per ISO 10993-1:2009 and FDA #G95-1
- Sterility testing per ANSI/AAMI/ISO 11137-2:2006(R)2010 and AAMI . TIR33:2005.
- Shelf Life Testing ASTMF1980-07(2011)
- Transit Testing per ASTM D4169(2009) ●
- Performance Testing ●
- Simulated Use O
- Durability Testing O
{10}------------------------------------------------
- Hi Pot Testing O
- Tensile Strength O
- Flow Rate O
- Pressure Testing O
- O System undeliverable volume (dead volume)
#### J. Conclusion
Biocompatibility test results were deemed acceptable for the intended use and the devices were considered biocompatible.
All device bench test results were acceptable. Data demonstrate that the Resecting Device and the Fluid Management Accessories meet design specifications and are suitable for the intended use and labeled shelf-life.
IOGYN, Inc. has demonstrated that the proposed modifications to the IOGYN System components are substantially equivalent to their predicate in IOGYN System (K132695).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.