K141834 · Q Core Medical , Ltd. · FPA · Aug 5, 2014 · General Hospital
Device Facts
Record ID
K141834
Device Name
SAPPHIRE ADMINISTRATION SET
Applicant
Q Core Medical , Ltd.
Product Code
FPA · General Hospital
Decision Date
Aug 5, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The dedicated Q Core administration sets for the Sapphire pump are intended for single-patient use and single-use only.
Device Story
Dedicated administration set for Sapphire infusion pump; facilitates fluid delivery; includes standard components (Luer locks, PVC tubing, Y-connector, clamps); features Anti-Free Flow Valve (AFFV) and pressure-activated back check valve (PAV) to prevent free flow; provided sterile and non-pyrogenic; used in hospital or home settings under physician order; ensures safe fluid administration by maintaining flow control and preventing backflow.
Clinical Evidence
Bench testing only. Testing included visual inspection, dimensional testing, physical requirements, and flow accuracy verification.
Technological Characteristics
Components: Luer locks, PVC tubing, Y-connector, tubing clamp. Materials: Biocompatible, non-DEHP, latex-free. Safety features: Anti-Free Flow Valve (AFFV) and pressure-activated back check valve (PAV). Sterilization: Sterile, non-pyrogenic. Connectivity: Dedicated for use with Sapphire infusion pumps.
Indications for Use
Indicated for single-patient, single-use administration of fluids via the Sapphire infusion pump. Use restricted to order of a licensed medical practitioner in hospital or home settings.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
K142036 — INFUSOMAT SPACE VOLUMETRIC INFUSION PUMP ADMINISTRATION SETS · B.Braun Medical, Inc. · Aug 25, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized representation of a human figure with three faces in profile, stacked on top of each other. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 5, 2014
Q Core Medical Limited C/O Ms. Rhona Shanker Director, Regulatory Consulting 12154 Darnestown Road, #236 Gaithersburg, MD 20878
Re: K141834
Trade/Device Name: Q Core Administration Set Regulation Number: 21 CFR 880.5440 Regulation Name: Administration Set Regulatory Class: II Product Code: FPA Dated: July 7, 2014 Received: July 7, 2014
Dear Ms. Shanker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Ms. Shanker
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mary S. Runner -S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K141834
Device Name
Administration Set
Indications for Use (Describe)
The dedicated Q Core administration sets for the Sapphire pump are intended for single-patient use and single-use only.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Mary E. Brooks -A 2014.08.05 13:52:51 -04'00' This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data souces, gather and maintain the data needed and complete and review the collection of inferently data sources, gather and mailitain the data needed and complete of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
| Owner/Submitter | Q Core Medical Ltd.<br>29 Yad Haruzim St.<br>Netanya 4250529<br>ISRAEL<br>Ph: +972-73-2388888<br>Fax: +972-73-2388800 |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Rhona Shanker<br>FDA Regulatory Consultant to Q Core Medical Ltd<br>Ph: 301251-9570<br>Fax: 301-251-9571 |
| Trade Name | Q Core Administration Set |
| Common Name | Administration Set |
| Classification Name | Intravascular administration set 21 CFR 880.5440<br>Product Code:<br>FPA - Administration Sets<br>Class II |
| Predicate Device<br>Administration Sets | Q Core Administration Sets cleared under K123049<br>Bodyguard Infusion System sets with back check valve cleared<br>under K031749 |
#### Device Description
The dedicated Q Core Administration Sets for the Sapphire infusion pump are provided sterile and are for single-use and single-patient use only. Components from the previously cleared administration sets were used to make a set with new configuration, all within the parameters (e.g. length, diameters, materials) of the sets cleared with the Sapphire Infusion Pump. This new configuration is substantially equivalent to the administration sets that were cleared under K123049, having identical indications for use and technological characteristics. In addition, this set includes a pressure activated back check valve ("PAV") that has the same cracking pressure as the back check valve that was cleared with administration sets under K031749 (Bodyguard
Image /page/3/Picture/4 description: The image shows the logo for Q Core Medical. The logo features a blue circle with a speech bubble-like shape inside, with a small droplet extending from the bottom of the bubble. Below the circle, the text "Q CORE" is displayed in a larger, bolder blue font, and below that, the word "medical" is written in a smaller, lighter blue font.
{4}------------------------------------------------
Infusion System). The PAV valve has been added to make it consistent with other sets on the market.
#### Indications for Use
The dedicated Q Core administration sets for the Sapphire pump are intended for single-patient use and single-use only.
These are the same indications for use as the predicate Administration Sets cleared under K123049.
### Technological Characteristics
The proposed O Core Administration Set configuration to be used with the Sapphire Infusion Pump is substantially equivalent to the primary predicate devices (K123049) in the following respects:
- All sets are dedicated for use with Q Core infusion pumps. 1.
- All sets can be used only by or under the order of a licensed medical practitioner. 2.
- All sets consist of standard, conventional components such as Luer locks, PVC tubing, Y-3. connector, tubing clamp.
- 4. All sets use materials with the same characteristics (biocompatible, non-DEHP, latex free).
- 5. All sets have a cassette with an Anti-Free Flow Valve [AFFV] and other means to protect against free flow. The new set includes an additional means to protect against free flow, which is the inclusion of a pressure activated back check valve (PAV) that has the same cracking pressure as the back check valve that was cleared under K031749 (Bodyguard Infusion System)
- All sets are provided sterile, non-pyrogenic, intended for single patient use and single 6. use.
- All sets are intended for either hospital or home use. 7.
# Pre-Clinical Testing
Preclinical testing included in the submission to demonstrate that the new administration set is safe and performs as intended involved the following:
- Visual tests .
- Dimensional testing ●
- Physical requirements ●
- Flow accuracy ●
Image /page/4/Picture/19 description: The image shows the logo for Q Core Medical. The logo consists of a blue circle with a speech bubble inside of it. Below the circle, the text "Q CORE" is written in blue, with the word "medical" written in a smaller font size below it.
{5}------------------------------------------------
## Conclusion
The new Administration Set having the new configuration and an additional pressure activated back check valve, that is to be used with the Sapphire Infusion Pump, is substantially equivalent to sets cleared under K123049 (primary predicate) with respect to the indications for use, technological characteristics and materials and to K031749 (Bodyguard Infusion System) with respect to the cracking pressure of the back check valve.
Image /page/5/Picture/2 description: The image is a logo for Q Core Medical. The logo features a blue circle with a lighter blue speech bubble inside. Below the circle, the word "QCORE" is written in a sans-serif font, with "medical" written in a smaller font below it. The overall design is clean and modern.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.