← Product Code [OCY](/productcode/OCY) · K141820

# JAGWIRE HIGH PERFORMANCE GUIDWIRE (K141820)

_Boston Scientific Corp · OCY · Oct 21, 2014 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K141820

## Device Facts

- **Applicant:** Boston Scientific Corp
- **Product Code:** [OCY](/productcode/OCY.md)
- **Decision Date:** Oct 21, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

Indicated for use in selective cannulation of the biliary ducts including, but not limited to the common bile, cystic, right and left hepatic ducts. The guidewires are designed to be used during endoscopic biliary procedures for catheter introduction and exchanges. The Jagtail™ High Performance Guidewire Extension attached to the extendable Jagwire™ 260 cm Guidewire is designed for use during endoscopic biliary procedures for non-rapid exchange catheter introduction and exchanges. The attachment of the Jagtail High Performance Guidewire Extension to the extendable Jagwire 260 cm guidewire creates an extended guidewire that can be used to exchange out a non-rapid exchange biliary catheter without removing the original guidewire from the associated duct. When the exchange is completed, the Jagtail High Performance Guidewire Extension can be detached and the original, extendable Jagwire 260 cm guidewire can be used in a conventional manner.

## Device Story

Guidewires and extension facilitate biliary duct cannulation and catheter exchange during endoscopic procedures. Metal alloy core guidewires feature striped PTFE jackets and hydrophilic radiopaque distal tips for visualization. Extension device (stainless steel core, PTFE sleeve) attaches to 260 cm extendable guidewire via alignment tool to allow catheter exchange without removing the primary guidewire from the duct. Used by clinicians in endoscopic settings. Benefits include maintaining guidewire position during catheter exchanges, reducing procedural complexity.

## Clinical Evidence

Bench testing only. Biocompatibility testing (cytotoxicity, sensitization, intracutaneous reactivity, latex, USP physiochemical) performed per ISO 10993-1. Performance testing included tip adhesion shear strength, tip tensile strength, column strength, dimensional testing, cantilever strength, and atraumatic tip assessment.

## Technological Characteristics

Metal alloy core (guidewire) or stainless steel core (extension); striped PTFE jacket; hydrophilic radiopaque distal tip (Tungsten/Pebax). Tip attachment via adhesive primer or coextruded resin. Dimensions: 200 cm (extension). Connectivity: None. Sterilization: Not specified. Software: None.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Jagwire™ High Performance Guidewire ([K960186](/device/K960186.md))
- Sensor Guidewire (510(k) Exempt)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing right, with flowing lines connecting them. The profiles are meant to represent the department's focus on health and human well-being.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 14, 2014

Boston Scientific Corp. Allison Baillie Regulatory Affairs Specialist II 100 Boston Scientific Way Marlborough, MA 01752

K141820 Re:

> Trade/Device Name: Jagwire™ High Performance Guidewire Jagwire™ High Performance Extendable Guidewire Jagwire™ High Performance Guidewire Extension Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: OCY Dated: July 3. 2014 Received: July 7, 2014

Dear Allison Baillie,

This letter corrects our substantially equivalent letter of October 21, 2014.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Joyce M. Whang -S

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K141820

Device Name

Jagwire™ High Performance Guidewire, Jagwire™ High Performance Extendable Guidewire

Indications for Use (Describe)

Indicated for use in selective cannulation of the biliary ducts including, but not limited to the common bile, cystic, right and left hepatic ducts. The guidewires are designed to be used during endoscopic biliary procedures for catheter introduction and exchanges.

Type of Use (Select one or both, as applicable):

| <div> <span> <svg height="16" width="16"> <polygon points="0,0 0,16 16,16 16,0" style="fill:none;stroke:black;stroke-width:1"></polygon> <polyline points="2,8 6,12 14,4" style="fill:none;stroke:black;stroke-width:2"></polyline> </svg> </span>     Prescription Use (Part 21 CFR 801 Subpart D)    </div> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg height="16" width="16"> <polygon points="0,0 0,16 16,16 16,0" style="fill:none;stroke:black;stroke-width:1"></polygon> </svg> </span>     Over-The-Counter Use (21 CFR 801 Subpart C)    </div>                                                                                             |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K141820

#### Device Name

Jagtail™ High Performance Guidewire Extension

#### Indications for Use (Describe)

The Jagtail™ High Performance Guidewire Extension attached to the extendable Jagwire™ 260 cm Guidewire is designed for use during endoscopic biliary procedures for non-rapid exchange catheter introduction and exchanges. The attachment of the Jagtail High Performance Guidewire Extendable Jagwire 260 cm guidewire creates an extended guidewire that can be used to exchange out a non-rapid exchange biliary catheter without removing the original guidewire from the associated duct. When the exchange is completed, the Jagail High Performance Guidewire Extension can be detached and the original, extendable Jagwire 260 cm guidewire can be used in a conventional manner.

| Type of Use (Select one or both, as applicable)                               |                                                                              |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span>    Prescription Use (Part 21 CFR 801 Subpart D)   </div> | <div> <span> </span>    Over-The-Counter Use (21 CFR 801 Subpart C)   </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K141820 Pg. 1 of 3

## SECTION 5 510(k) SUMMARY

#### 1. Submitter

Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752 Telephone: 508-683-4942 Fax: 508-683-5939

Contact: Allison Baillie Regulatory Affairs Specialist II Date Prepared: October 17, 2014

#### 2. Device

| Trade Names:         | Jagwire™ High Performance Guidewire,<br>Jagwire™ High Performance Extendable Guidewire,<br>Jagtail™ High Performance Guidewire Extension |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name:         | Endoscopic Guidewire                                                                                                                     |
| Classification Name: | Endoscope and accessories                                                                                                                |
| Regulation Number:   | 876.1500                                                                                                                                 |
| Product Code:        | OCY                                                                                                                                      |
| Classification:      | Class II                                                                                                                                 |

## 3. Predicate Devices

| Trade Name:                           | Jagwire™ High Performance Guidewire,<br>Jagwire™ High Performance Extendable Guidewire,<br>Jagtail™ High Performance Guidewire Extension |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer and<br>Clearance Number: | Boston Scientific Corporation, K960186                                                                                                   |
| Common Name:                          | Endoscopic Guidewire                                                                                                                     |
| Classification Name:                  | Endoscope and accessories                                                                                                                |
| Regulation Number:                    | 876.1500                                                                                                                                 |
| Product Code:                         | EZB, OCY                                                                                                                                 |
| Classification:                       | Class II (previously 510(k) exempt under Procode EZB)                                                                                    |
| Trade Name:                           | Sensor Guidewire                                                                                                                         |
| Manufacturer and<br>Clearance Number: | Boston Scientific Corporation, 510(k) Exempt                                                                                             |
| Common Name:                          | Endoscopic Guidewire                                                                                                                     |
| Classification Name:                  | Endoscope and accessories                                                                                                                |
| Regulation Number:                    | 876.1500                                                                                                                                 |
| Product Code:                         | EZB                                                                                                                                      |
| Classification:                       | Class II (previously 510(k) exempt under Procode EZB)                                                                                    |

{5}------------------------------------------------

## 4. Device Description

Jagwire™ High Performance Guidewire, and Jagwire™ High Performance Extendable Guidewire:

The Boston Scientific Jagwire™ High Performance Guidewire is constructed of a metal alloy core, which is encapsulated in a striped PTFE jacket with a 5 cm hydrophilic radiopaque distal tip. The striped jacket provides endoscopically visible movement markings.

## Jagtail™ High Performance Guidewire Extension:

The Boston Scientific Jagtail™ High Performance Guidewire Extension is a 200 cm long guidewire extension constructed of a stainless steel core with a connector at the distal end. The shaft is coated with a PTFE sleeve. The Jagtail High Performance Guidewire Extension is exclusively compatible with the Jagwire High Performance Extendable Guidewire, 260 cm. The distal end of the Jagtail High Performance Guidewire Extension is inserted into the proximal end of the extendable Jagwire 260 cm guidewire with the use of the alignment tool. The alignment tool is then removed from the Jagtail High Performance Guidewire Extension.

## 5. Indication for Use:

## Jagwire™ High Performance Guidewire, and Jagwire™ High Performance Extendable Guidewire:

Indicated for use in selective cannulation of the biliary ducts including, but not limited to the common bile, cystic, right and left hepatic ducts. The guidewires are designed to be used during endoscopic biliary procedures for catheter introduction and exchanges.

#### Jagtail™ High Performance Guidewire Extension:

The Jagtail™ High Performance Guidewire Extension attached to the extendable Jagwire™ 260 cm Guidewire is designed for use during endoscopic biliary procedures for non-rapid exchange catheter introduction and exchanges. The attachment of the Jagtail High Performance Guidewire Extension to the extendable Jagwire 260 cm guidewire creates an extended guidewire that can be used to exchange out a non-rapid exchange biliary catheter without removing the original guidewire from the associated duct. When the exchange is completed, the Jagtail High Performance Guidewire Extension can be detached and the original, extendable Jagwire 260 cm guidewire can be used in a conventional manner.

{6}------------------------------------------------

## 6. Technological Characteristics:

The materials used to attach the Tungsten/Pebax tip of the proposed Jagwire™ Guidewires have changed slightly. The current adhesive primer material used to attach the current Jagwire guidewire tip has been discontinued by the supplier.

Two options for replacement a material for the discontinued adhesive primer material are proposed. One option is a direct replacement of the discontinued adhesive primer material. The second option is a material that is coextruded with the guidewire tip, which has also simplified the tip attachment process.

With the exception of the changes in the tip attachment materials, the proposed Jagwire™ guidewires are identical in design, materials and intended use to the currently commercialized Jagwire™M guidewires. There were no changes made to the Jagtail™ High Performance Guidewire Extension.

#### 7. Performance Data:

Non-clinical testing was successfully performed on the proposed Jagwire™ Guidewires as described below.

## Biocompatibility Testing Summary:

The proposed Jagwire Guidewires were assessed in accordance with ISO 10993-1, and are classified as surface, limited contact duration (<24hours), mucous membrane contact devices. All required testing was completed, which included testing for: Cytotoxicity, Sensitization, Intracutaneous Reactivity, Latex and USP Physiochemical testing was completed with acceptable results.

#### Performance Testing Summary:

Non-clinical comparative performance bench testing was successfully completed to demonstrate that the new adhesive primer, and new resin (coextruded with the tip of the proposed Jagwire) meets the Product Specifications of the currently marketed Jagwire Guidewires. This testing included: Tip Adhesion Shear Strength, End Product Tip Tensile, Tip Column Strength, Dimensional Test for Tip Outer Diameter, Tip Cantilever Strength, Atraumatic Tip, and Appearance and Cleanliness.

#### 8. Conclusion:

Boston Scientific has demonstrated that the proposed Jagwire™ Guidewires are substantially equivalent to the currently marketed Jagwire™ Guidewires.

---

**Source:** [https://fda.innolitics.com/device/K141820](https://fda.innolitics.com/device/K141820)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
