K141811 · Mortara Instrument, Inc. · MHX · Nov 19, 2014 · Cardiovascular
Device Facts
Record ID
K141811
Device Name
MORTARA MONITORING WAVEFORM VIEWER
Applicant
Mortara Instrument, Inc.
Product Code
MHX · Cardiovascular
Decision Date
Nov 19, 2014
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Monitoring Waveform Viewer is a software component intended to be utilized within another software application. The software component can continuously receive and display data generated by multiparameter patient monitoring systems in near real time, display reported alarm states, display static waveforms for review of existing data, and perform measurements with the support of a caliper tool. The Monitoring Waveform Viewer software component is not intended to be used as an alarm device. The data and analysis provided by the Monitoring Waveform Viewer is reviewed, confirmed, and used by trained medical personnel in the diagnosis of patients with various conditions.
Device Story
Monitoring Waveform Viewer is a software component integrated into host medical applications; receives waveform and parameter data (ECG, ART, PA, ICP, IABP, RR, EtCO2, SpO2) from multi-parameter patient monitoring systems. Displays data in near real-time; provides visual indicators for alarm states (High/Medium/Low priority). Features include magnification tool for signal inspection and manual caliper tool for horizontal/time and vertical/signal amplitude measurements. Used by trained medical personnel in clinical settings to review patient data, confirm findings, and support diagnostic decision-making. Does not function as an alarm system; no alarm silencing or notification capabilities. Output used for printing or inclusion in patient records.
Clinical Evidence
No clinical data. Bench testing only. Software verification and validation were performed to confirm compliance with predetermined functional and performance requirements.
Technological Characteristics
Software-only component; no physical materials, sterilization, or electrical/EMC requirements. Operates as a drawing surface with parameter and waveform display areas. Supports manual caliper and magnification tools. Connectivity via integration into host software applications. Algorithm type: rule-based data visualization and measurement processing.
Indications for Use
Indicated for use by trained medical personnel as a software component within a host application to display waveforms, parameters, and alarm states from multi-parameter patient monitoring systems. Supports review of static waveforms, signal magnification, and manual caliper measurements for amplitude and interval analysis. Not for use as a primary or secondary alarm device; does not provide alarm acknowledgement, silencing, or audible/visual alarm notifications.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract image of three faces in profile, stacked on top of each other, representing the department's focus on people and their well-being.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 19, 2014
Mortara Instrument, Inc. Ms. Amy Yang Sr. Regulatory Affairs Engineer 7865 N. 86th Street Milwaukee, WI 53224
Re: K141811
Trade/Device Name: Mortara Monitoring Waveform Viewer Regulation Number: 21 CFR 870.1025 Regulation Name: Monitor, physiological, patient (with arrhythmia detection or alarms) Regulatory Class: Class II Product Code: MHX Dated: October 2, 2014 Received: October 6, 2014
Dear Ms. Yang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{2}------------------------------------------------
Page 2 – Ms. Amy Yang
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Melissa A. Torres -S
For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141811
Device Name Monitoring Waveform Viewer
Indications for Use (Describe)
The Monitoring Waveform Viewer is indicated for use as follows:
The Monitoring Waveform Viewer is a software component utilized within another software application.
The Monitoring Waveform Viewer performs the following:
· Display of data generated by multi-parameter patient monitoring systems continuously producing waveforms and parameters.
· Display of reported alarm states produced by such monitoring systems.
· Display of static waveforms for review of existing data, with the possibility to activate a magnification tool. The magnification tool will enlarge the signal located under the magnification icon.
· Perform measurements with the support of a caliper tool (providing horizontal/time measurements and vertical/signal amplitude measurements). The caliper tool can be activated when the Monitoring Waver displays still waveforms.
· Provide selected waveform episodes to the host application for printing or inclusion in the patient record.
The Monitoring Waveform Viewer:
- · Is not a primary or secondary alarm device.
· Does not provide a user interface for acknowledging or silencing active alarms.
· Does not generate audible or visual (flashing) notifications of active alarms.
The Monitoring Waveform Viewer does not provide any support for implementing an application that is part of an alarm system.
Type of Use (Select one or both, as applicable)
| Description Use (Part 91 CFR 901 Subpart D) | Over-The-Counter Use (91 CFR 901 Subpart C) |
|---------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: overline;">☒</span> | ☐ |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Abbreviated 510(k) Notification
## 510(k): Mortara Monitoring Waveform Viewer Summary K141811
# Submitter:
Date: November 18, 2014
| Amy Yang, Sr. Regulatory Affairs Engineer | |
|-------------------------------------------|----------------------------------|
| Mortara Instrument, Inc. | |
| 7865 N. 86th Street | |
| Milwaukee, WI 53224 | |
| FAX: | (414) 354-4760 |
| Phone: | (414) 354-1600 |
| Contact: | Amy Yang (see above) |
| Trade Name: | Mortara Monitoring Waveform View |
| Trade Name: | Mortara Monitoring Waveform Viewer |
|-----------------------------|----------------------------------------------------------------------------|
| Common Name: | Monitoring Waveform Viewer |
| Classification Name: | Monitor, Physiological, Patient (with Arrhythmia Detection or Alarms) |
| Classification Description: | Arrhythmia detector and alarm (including ST-segment measurement and alarm) |
| Classification Regulation: | 21 CFR §870.1025 |
| Product Code: | MHX |
# Legally marketed devices to which S.E. is claimed:
| Mortara Monitoring Waveform<br>Viewer | Predicate 510(k) Number | Predicate Manufacturer / Model |
|---------------------------------------|-------------------------|--------------------------------------------------------|
| Surveyor Central Station | K131929 | Mortara Instrument, Inc. / Surveyor<br>Central Station |
# Description:
The Monitoring Waveform Viewer is a software component that includes a system for reviewing snapshots of monitoring data and offers a measurement module with user controlled calipers and magnifier glass.
Medical device data or information system applications may utilize the Monitoring Waveform Viewer software component that includes a stand-alone executable module that functions as a host application. The Monitoring Waveform Viewer is organized as a drawing surface with three areas:
- a parameters area located at the bottom of the window, .
- a parameter waveform area located at the right and
- . a main waveform area.
The parameters area is dedicated to report the values of parameters not associated with displayed waveforms. The parameter waveform area is dedicated to report the values of parameters associated to the displayed waveforms.
The application supports several numerical parameters such as: ECG (electrocardiogram), ART (arterial pressure), PA (pulmonary artery invasive blood pressure), ICP (intracranial pressure), IABP (intra-aotic balloon pump pressure), RR (respiration rate), EtCO2 (end tidal carbon dioxide), SpO2 (oxygen saturation), etc.
The Monitoring Waveform Viewer receives data parameters and waveforms supplied from multi-parameter patient monitoring systems. The Monitoring Waver is able to display alarm states by highlighting the corresponding parameter box. The Monitoring Waveform Viewer operating in near real time mode supports alarm priorities of High Priority, Medium Priority, and Low Priority.
{5}------------------------------------------------
# Abbreviated 510(k) Notification
The Monitoring Waveform Viewer allows the user to right-click over any ECG lead or the ECG parameter box to activate a context menu for selecting the ECG gain and displayed leads. The user is able to select a list of active (with data) ECG leads received from the host application. It also allows the user to select gains of 5, 10, 20 and 40mm/mV. The Monitoring Waveform Viewer provides a caliper tool for performance amplitude and interval measurements on the displayed traces. The caliper can be placed anywhere on the waveform display area.
Further description and details concerning use of the Monitoring Waveform Viewer can be found in:
- Tab 2 Device Description, Monitoring Waveform Viewer Block Diagram .
- . Tab 12 Proposed Labeling documents, Monitoring Waveform Viewer product brochure, Monitoring Waveform Viewer User Manual.
## Technology Comparison:
The Mortara Monitoring Waver utilizes the same or similar technology for each parameter as utilized by the predicate devices.
## Intended Use:
The Monitoring Waveform Viewer is a software component intended to be utilized within another software application. The software component can continuously receive and display data generated by multiparameter patient monitoring systems in near real time, display reported alarm states, display static waveforms for review of existing data, and perform measurements with the support of a caliper tool.
The Monitoring Waveform Viewer software component is not intended to be used as an alarm device. The data and analysis provided by the Monitoring Waveform Viewer is reviewed, confirmed, and used by trained medical personnel in the diagnosis of patients with various conditions.
## Indications for Use:
The Monitoring Waveform Viewer is indicated for use as follows: The Monitoring Waveform Viewer is a software component utilized within another software application.
The Monitoring Waveform Viewer performs the following:
- Display of data generated by multi-parameter patient monitoring systems continuously producing waveforms and parameters.
- . Display of reported alarm states produced by such monitoring systems.
- . Display of static waveforms for review of existing data, with the possibility to activate a magnification tool. The magnification tool will enlarge the signal located under the magnification icon.
- . Perform measurements with the support of a caliper tool (providing horizontal/time measurements and vertical/signal amplitude measurements). The caliper tool can be activated when the Monitoring Waveform Viewer displays still waveforms.
- . Provide selected waveform episodes to the host application for printing or inclusion in the patient record.
The Monitoring Waveform Viewer:
- Is not a primary or secondary alarm device. .
- . Does not provide a user interface for acknowledging or silencing active alarms.
- . Does not generate audible or visual (flashing) notifications of active alarms.
The Monitoring Waveform Viewer does not provide any support for implementing an application that is part of an alarm system.
{6}------------------------------------------------
# Abbreviated 510(k) Notification
## Performance Testing:
Sterilization Validation - The Mortara Monitoring Waveform Viewer is not sterilized or sterilizable, and therefore this section does not apply to the Waveform Viewer itself.
Shelf Life Testing – The Mortara Monitoring Waveform Viewer is not sterilizable, and therefore this section does not apply to the Waveform Viewer itself.
Biocompatibility Testing - The Moritoring Waveform Viewer is software only, and therefore this section does not apply to the Waveform Viewer itself.
Software Testing - Software for the Mortara Monitoring Waveform Viewer was designed and developed according to a robust software development process, and was rigorously verified and validated. Test results indicated that the Mortara Monitoring Waver complies with its predetermined specification.
Electrical Safety - The Mortara Monitoring Waver is software only, and therefore this section does not apply to the Waveform Viewer itself.
Electromagnetic Compatibility Testing - The Mortara Monitoring Waveform Viewer is software only, and therefore this section does not apply to the Waveform Viewer itself.
Performance Testing – Bench – The Mortara Monitoring Waveform Viewer was tested in accordance with internal requirements and procedures, and test results indicated that the device complies with the predetermined requirements. This testing includes performance and functional.
Performance Testing – Animal – Animal performance testing was not performed and is not necessary to demonstrate safety and effectiveness of the Mortara Monitoring Waveform Viewer.
Performance Testing – Clinical performance testing was not performed and is not necessary to demonstrate safety and effectiveness of the Mortara Monitoring Waveform Viewer.
Conclusion - The results of these activities demonstrate that the Mortara Monitoring Waver is as safe, as effective, and performs as well as or better than the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.