← Product Code [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS) · K141796

# DEPUY SYNTHES TOMOFIX OSTEOTOMY SYSTEM (K141796)

_Synthes (USA) Products, LLC · HRS · Aug 29, 2014 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K141796

## Device Facts

- **Applicant:** Synthes (USA) Products, LLC
- **Product Code:** [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS.md)
- **Decision Date:** Aug 29, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The DePuy Synthes TOMOFIX Osteotomy System is intended for osteotomies, treatment of bone and joint deformities, fixation of fractures, and malalignment caused by injury or disease, such as osteoarthritis, of the distal femur and proximal tibia. Specifically, – The TOMOFIX Medial Proximal Tibia Plates are indicated for open- and closed-wedge osteotomies fixation of fractures, and malalignment caused by injury or disease, such as osteoarthritis, of the medial proximal tibia – The TOMOFIX Lateral Proximal Tibia Plates are indicated for open- and closed-wedge osteotomies fixation of fractures, and malalignment caused by injury or disease, such as osteoarthritis, of the lateral proximal tibia – The TOMOFIX Lateral Distal Femur Plates are indicated for open- and closed-wedge osteotomies fixation of fractures, and malalignment caused by injury or disease, such as osteoarthritis, of the lateral distal femur – The TOMOFIX Medial Distal Femur Plates are indicated for closed-wedge osteotomies fixation of fractures, and malalignment caused by injury or disease, such as osteoarthritis, of the medial distal femur

## Device Story

System comprises five titanium plate families with locking and combination locking/compression holes; designed for stable fixation of osteotomies, fractures, and malalignments of distal femur and proximal tibia. Plates are anatomically contoured, available in right/left versions, and provided sterile or non-sterile. Used by orthopedic surgeons in clinical/surgical settings. Device provides mechanical stabilization of bone segments to facilitate healing and correct alignment. Benefits include restoration of joint function and reduction of deformity-related pain.

## Clinical Evidence

Bench testing only. Mechanical performance demonstrated via axial testing, torsion testing, bending stiffness analysis (using ProEngineer CAD), and dynamic fatigue testing on plate-screw constructs mounted on sawbone models and POM-C test blocks.

## Technological Characteristics

Titanium bone fixation plates with locking and combination locking/compression holes. Anatomically contoured. Mechanical performance validated via axial, torsion, and fatigue testing. No software or electronic components.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Synthes TOMOFIX Medial Distal Femur Plates ([K081353](/device/K081353.md))
- Synthes TOMOFIX Osteotomy System ([K023941](/device/K023941.md))
- Synthes Large Fragment Dynamic Compression Locking System ([K000682](/device/K000682.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of three human profiles facing right, arranged in a way that suggests unity and connection. The profiles are black and are set against a white background. Encircling the profiles are the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 29, 2014

Synthes (USA) Products LLC Ms. Kara A. R. Elkin Regulatory Affairs Specialist 1301 Goshen Parkway West Chester, Pennsylvania 19380

Re: K141796

Trade/Device Name: DePuy Synthes TOMOFIX Osteotomy System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliance and accessories Regulatory Class: Class II Product Code: HRS Dated: August 1, 2014 Received: August 4, 2014

Dear Ms. Elkin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

Page 2 – Ms. Kara A.R. Elkin

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Lori A. Wiggins -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

See PRA Statement below.

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

# Indications for Use

510(k) Number (if known)

## K141796

Device Name

DePuy Synthes TOMOFIX Osteotomy System

# Indications for Use (Describe)

The DePuy Synthes TOMOFIX Osteotomy System is intended for osteotomies, treatment of bone and joint deformities, fixation of fractures, and malalignment caused by injury or disease, such as osteoarthitis, of the distal femur and proximal tibia.

Specifically.

- The TOMOFIX Medial Proximal Tibia Plates are indicated for open- and closed-wedge osteotomies fixation of fractures, and malalignment caused by injury or disease, such as osteoarthritis, of the medial proximal tibia
- The TOMOFIX Lateral Proximal Tibia Plates are indicated for open- and closed-wedge osteotomies, fixation of fractures and malalignment caused by injury or disease. such as osteoarthritis, of the lateral proximal tibia
- The TOMOFIX Lateral Distal Femur Plates are indicated for open-and closed-wedge osteotomies, fixation of fractures and malalignment caused by injury or disease, such as osteoarthritis, of the lateral distal femur
- The TOMOFIX Medial Distal Femur Plates are indicated for closed-wedge osteotomies, fixation of fractures and malalignment caused by injury or disease, such as osteoarthritis, of the medial distal femur

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "SYNTHES" in a bold, sans-serif font. To the left of the word is a circular logo with a stylized design. A registered trademark symbol is located to the right of the word.

______________________________________________________________________________________________________________________________________________________________________________

| 5.0                 | 510(k) Summary                                                                                                                                                                                                                                                                                                                                                                    | Page 1 of 3 |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Sponsor:            | DePuy Synthes<br>1301 Goshen Parkway<br>West Chester, PA 19380                                                                                                                                                                                                                                                                                                                    |             |
|                     | Phone: (610) 719-1239<br>Fax: (484) 356-9682                                                                                                                                                                                                                                                                                                                                      |             |
|                     | Contact Person: Kara A. R. Elkin<br>Date Prepared: August 1, 2014                                                                                                                                                                                                                                                                                                                 |             |
| Device Name:        | DePuy Synthes TOMOFIX Osteotomy System                                                                                                                                                                                                                                                                                                                                            |             |
| Classification:     | Product Code - HRS, Plate, Fixation, Bone, Class II, §888.3030 –<br>Single / multiple component metallic bone fixation appliance and<br>accessories.                                                                                                                                                                                                                              |             |
| Predicate Devices:  | Synthes TOMOFIX Medial Distal Femur Plates (K081353)                                                                                                                                                                                                                                                                                                                              |             |
|                     | Synthes TOMOFIX Osteotomy System (K023941)                                                                                                                                                                                                                                                                                                                                        |             |
|                     | Synthes Large Fragment Dynamic Compression Locking System<br>(K000682)                                                                                                                                                                                                                                                                                                            |             |
| Device Description: | The DePuy Synthes TOMOFIX Osteotomy System consists of five<br>different titanium plate families with locking and combination<br>locking/compression holes. This system features plates designed to<br>provide stable fixation of osteotomies of the distal femur and<br>proximal tibia.                                                                                          |             |
|                     | The DePuy Synthes TOMOFIX Medial Distal Femur Plates are part<br>of the DePuy Synthes TOMOFIX Osteotomy System used for closed<br>wedge femoral osteotomies. The subject plates are anatomically<br>contoured to fit the medial distal femur, are available in right and left<br>versions, as well as sterile and non-sterile.                                                    |             |
| Intended Use:       | The DePuy Synthes TOMOFIX Osteotomy System is intended for<br>osteotomies, treatment of bone and joint deformities, fixation of<br>fractures, and malalignment caused by injury or disease, such as<br>osteoarthritis, of the distal femur and proximal tibia.                                                                                                                    |             |
|                     | Specifically,<br>– The TOMOFIX Medial Proximal Tibia Plates are indicated<br>for open- and closed-wedge osteotomies fixation of fractures,<br>and malalignment caused by injury or disease, such as<br>osteoarthritis, of the medial proximal tibia<br>– The TOMOFIX Lateral Proximal Tibia Plates are indicated<br>for open- and closed-wedge osteotomies fixation of fractures, |             |

{4}------------------------------------------------

and malalignment caused by injury or disease, such as osteoarthritis, of the lateral proximal tibia

- The TOMOFIX Lateral Distal Femur Plates are indicated for open- and closed-wedge osteotomies fixation of fractures, and malalignment caused by injury or disease, such as osteoarthritis, of the lateral distal femur
- The TOMOFIX Medial Distal Femur Plates are indicated for closed-wedge osteotomies fixation of fractures, and malalignment caused by injury or disease, such as osteoarthritis, of the medial distal femur

Substantial The intent of this submission is to update the indications for use Equivalence: statement of the DePuy Synthes TOMOFIX Osteotomy System (cleared under K023941) by specifying osteotomy type (open or closed) and specific anatomical region indicated for each plate family in the system. Information has been provided to support the use of the DePuy Synthes TOMOFIX Osteotomy System for the proposed indications to support a substantially equivalent decision in comparison to the predicate Synthes TOMOFIX Osteotomy System as cleared under K023941.

> It is also the intent of this submission to modify the design of the DePuy Synthes TOMOFIX Medial Distal Femur Plate. Information presented supports substantial equivalence of the DePuy Synthes TOMOFIX Medial Distal Femur Plate to the predicate devices, Synthes TOMOFIX Medial Distal Femur Plates (K081353), Synthes TOMOFIX Osteotomy System (K023941) and Synthes Large Fragment Dynamic Compression Locking System (K000682). The proposed plates have similar indications for use, design characteristics, materials, and incorporate the same fundamental technology.

Performance Data

Static and dynamic construct testing was completed for the plates included in the DePuy Synthes TOMOFIX Osteotomy System in order to demonstrate comparable mechanical performance to the predicates. The mechanical testing was designed to assess the stiffness and fatigue strength of the subject devices. The following tests and analyses were performed:

- Axial Testing conducted by mounting the plates on an ● osteotomized sawbone femur model and subjecting the plates to an axial load
- Torsion Testing - conducted by mounting the plates on an osteotomized sawbone femur model and subjecting the plates to a torsional load
- Bending Stiffness - ProEngineer CAD software was used to calculate and compare stiffness properties in the critical cross section of the plates

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "SYNTHES" in bold, black font. To the left of the word is a circular logo with a stylized image inside. To the right of the word is a small, superscripted "R" in a circle, indicating a registered trademark.

- . Dynamic (Fatigue) Testing – conducted by mounting the plate-screw constructs on POM-C test blocks with a milled surface to match the bottom of the plate and introducing a cyclical axial force
The results of the mechanical testing and engineering analyses demonstrate substantial equivalence.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K141796](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K141796)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
