The Honeywell HomMed Genesis Touch™ Retrospective Physiological Monitoring System is designed to retrospectively monitor vital signs include noninvasive blood pressure. pulse oximetry, pulse rate, weight, temperature, blood sugar, prothrombin time/international normalized ratio (PT/INR), peak expiratory flow (PEF) and forced expiratory volume (FEV1). The Genesis Touch Retrospective Physiological Monitoring System collects; displays and transmits vital signs measurements from commercially available medical devices designed for home use. Collected measurement data from the Genesis Touch System can then be transmitted via a communication module to a central viewing station where the data can be viewed and analyzed by a healthcare professional. The Genesis Touch Retrospective Physiological Monitoring System is intended for home use by patients and caregivers or in a healthcare related environment by healthcare providers. The Genesis Touch Retrospective Physiological Monitoring System is not intended for emergency use or real-time monitoring and does not have auditory or visual alarms for out-of-limit parameters.
Device Story
Genesis Touch System is a tablet-based vital signs monitoring station; collects data from paired Bluetooth medical peripherals or manual entry; supports BP, pulse oximetry, pulse rate, weight, temperature, blood sugar, PT/INR, PEF, FEV1. Device uses text/voice prompts and animated video instructions to guide users. Encrypted data transmitted to Honeywell HomMed LifeStream Management Suite; healthcare providers use LifeStream for remote configuration, data triaging, and trending. Used in homes by patients/caregivers or clinical settings by providers. Facilitates preventative care through retrospective review of clinical information. System is 'locked down' to prevent non-medical use. Benefits include improved patient management and clinical oversight without real-time alarm requirements.
Clinical Evidence
Bench testing only. Risk-based verification and validation performed per FDA 'General Principles of Software Validation' guidance. Testing covered usability, hardware/software interoperability, data transfer, accuracy, display/transmission, and system reliability. No clinical data required.
Technological Characteristics
Proprietary software application on dedicated commercially available tablet platform. Features touch screen, Wi-Fi, Bluetooth, and video capability. 'Locked down' OS prevents unauthorized software installation. Data transmission via communication module to central server. No electrical/EMC testing required as it is a software application on existing hardware.
Indications for Use
Indicated for patients and caregivers in home settings or healthcare providers in clinical environments for retrospective monitoring of vital signs (BP, pulse oximetry, pulse rate, weight, temperature, blood sugar, PT/INR, PEF, FEV1). Not for emergency use or real-time monitoring.
Regulatory Classification
Identification
A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring three stylized human profiles facing to the right, representing health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 10, 2014
Honeywell Hommed Claudia Jackson Regulatory Affairs Specialist 3400 West Intertech Drive Brookfield, Wisconsin 53045
Re: K141792 Trade/Device Name: Genesis Touch System Regulation Number: 21 CFR 870.2910 Regulation Name: Radiofrequency Physiological Signal Transmitter and Receiver Regulatory Class: Class II Product Code: DRG Dated: October 21, 2014 Received: October 22, 2014
Dear Claudia Jackson,
This letter corrects our undated substantially equivalent letter.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Melissa A. Torres -S
- Bram D. Zuckerman, M.D. For Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known):
Device Name: Honeywell HomMed Genesis Touch System
Indications For Use:
The Honeywell HomMed Genesis Touch™ Retrospective Physiological Monitoring System is designed to retrospectively monitor vital signs include noninvasive blood pressure. pulse oximetry, pulse rate, weight, temperature, blood sugar, prothrombin time/international normalized ratio (PT/INR), peak expiratory flow (PEF) and forced expiratory volume (FEV1). The Genesis Touch Retrospective Physiological Monitoring System collects; displays and transmits vital signs measurements from commercially available medical devices designed for home use. Collected measurement data from the Genesis Touch System can then be transmitted via a communication module to a central viewing station where the data can be viewed and analyzed by a healthcare professional.
The Genesis Touch Retrospective Physiological Monitoring System is intended for home use by patients and caregivers or in a healthcare related environment by healthcare providers.
The Genesis Touch Retrospective Physiological Monitoring System is not intended for emergency use or real-time monitoring and does not have auditory or visual alarms for out-oflimit parameters.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
# Honeywell HomMed
### 510(k) Summary as required by section 807.92(c) Honeywell HomMed Genesis Touch™ System
| Applicant Name: | Honeywell HomMed, LLC. |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 3400 Intertech Drive, Suite 200<br>Brookfield, Wisconsin 53045<br>Ph: (262) 252-783-5440<br>Establishment Registration #3004183721 |
| Summary Date: | 06/25/2014 |
| Contact Person: | Claudia Jackson, Regulatory Affairs Specialist<br>Honeywell HomMed, LLC<br>3400 Intertech Drive, Suite 200<br>Brookfield, Wisconsin 53045<br>Ph: (262) 252-5838<br>Fax: (262) 252-6119 |
| Corporate Contact: | Greg Lillegard, Director of Quality<br>Honeywell HomMed, LLC<br>Ph: (262) 252-262-6062<br>Fax: (262) 252-262-6119 |
| Trade Name: | Honeywell HomMed Genesis Touch System |
| Common Name: | Patient Vital Signs Monitor |
### Classification Name:
| Regulation<br>Number | Product<br>Code | Classification Name | Device<br>Class |
|------------------------------------------------------------------------------------------------------|-----------------|-----------------------------------------------------------------|-----------------|
| 870.2910 | DRG | Radiofrequency Physiological Signal<br>Transmitter and Receiver | II |
| Medical device product codes also supported by Genesis Touch by means<br>of separate medical devices | | | |
| 870.1130 | DXN | Noninvasive Blood Pressure Measurement<br>System | II |
| 880.2700 | FRI | Patient Weight Scale | I |
| 870.2700 | DQA | Oximeter | II |
| 862.1345 | NBW | Glucose Test System | II |
| 868.1860 | BZH | Meter, Peak Flow, Spirometry | II |
| 864.7750 | GJS | Test, Time, Prothrombin | II |
Predicate Device(s): Honeywell HomMed Genesis DM Pro BP, K101242; Honeywell HomMed Genesis Touch System, K112858
510(k) Summary
{4}------------------------------------------------
## onevw HomMe
Device Description:
The modified Honeywell HomMed Genesis Touch™ Retrospective Physiological Monitoring System (Genesis Touch 10 System) is a vital signs monitoring station.
The modified Genesis Touch System is a proprietary software application running on a dedicated software/commercially available tablet platform with touch screen; Wi-Fi, Bluetooth and video capability.
The modified Genesis Touch System uses text and voice prompts to collect users' vital signs information from paired Bluetooth noninvasive medical device peripherals, such as blood pressure, pulse oximetry, pulse rate, weight, temperature, blood sugar, and allows for manual entry of blood sugar, prothrombin time/international normalized ratio (PT/INR), peak expiratory flow (PEF) and forced expiratory volume (FEV1). The encrypted data is automatically or at the user's discretion, forwarded to a remote central monitoring station (Honeywell HomMed LifeStream Management Suite) for retrospective review by healthcare providers.
LifeStream Management Suite, a Windows-based application interface is used by the healthcare provider for the remote configuration of all the Genesis monitors, triaging of patient data, trending of clinical information, and the opportunity to provide preventative patient care.
The modified Genesis Touch System also asks the user disease specific symptom management questions and provides reminder information, just like the predicates Genesis DM and Genesis Touch System. Both the modified Genesis Touch System and predicate Genesis Touch System, offer animated video instructions and optional one-touch, video-based communication in addition to traditional telemonitoring.
The modified Genesis Touch monitor expands the user population to include patients and caregivers willing and capable of managing its use. The modified Genesis Touch System is a touch screen tablet with a consumer-based Android platform like the predicate Genesis Touch System. This type of user interface is easily usable by a wider age and acuity range. The user population of the predicate monitors, Genesis Touch System and Genesis DM Monitor are designated for use with adult and pediatric patients over twelve years of age. The expanded user population of the modified Genesis Touch System is presented in order to compliment the user population of interfaced peripheral medical devices which do not limit the user population to being over twelve years of age. This modification does not change the intended use of the modified Genesis Touch System as a communication hub that collects; displays and transmits vital signs data from noninvasive medical device peripherals via a communication module to a central viewing station for retrospective review by a healthcare professional. This modification does not affect the safety and effectiveness of the Genesis Touch System when used as labeled.
{5}------------------------------------------------
### Honevw HomM
Indications for Use:
The Honeywell HomMed Genesis Touch™ Retrospective Physiological Monitoring System is designed to retrospectively monitor vital signs. Vital signs include noninvasive blood pressure, pulse oximetry, pulse rate, weight, temperature, blood sugar, prothrombin time/international normalized ratio (PT/INR), peak expiratory flow (PEF) and forced expiratory volume (FEV1). The Genesis Touch Retrospective Physiological Monitoring System collects; displays and transmits vital signs measurements from commercially available medical devices designed for home use. Collected measurement data from the Genesis Touch System can then be transmitted via a communication module to a central viewing station where the data can be viewed and analyzed by a healthcare professional.
The Genesis Touch Retrospective Physiological Monitoring System is intended for home use by patients and caregivers or in a healthcare related environment by healthcare providers.
The Genesis Touch Retrospective Physiological Monitoring System is not intended for emergency use or real-time monitoring and does not have auditory or visual alarms for out-oflimit parameters.
Technological Characteristics:
The modified Genesis Touch Retrospective Physiological Monitoring System (Genesis Touch System) is substantially equivalent to the predicate devices, Honeywell HomMed Genesis DM and Honeywell HomMed Genesis Touch System as a proprietary software application running on a dedicated software/hardware platform, with the function of data collection, data display, data transmission, and data communication from noninvasive peripheral medical device types to a central server. The predicate Genesis DM uses hard button push keys to advance menu selection, selection of vitals collection, and provide responses while the modified Genesis Touch System and predicate Genesis Touch System both use graphic icon touch keys to accomplish the same tasks.
The modified Genesis Touch System and the predicate Genesis Touch System are proprietary software applications operating on a dedicated commercially available tablet platform with the minimum performance specifications consistent with typical tablet computer hardware and equipment specifications. The modified Genesis Touch System and the predicate Genesis Touch offers a touch screen user interface; Wi-Fi, Bluetooth and video capability.
The predicate Genesis DM is a monitoring system with proprietary firmware which operates in a dedicated hardware design.
The modified Genesis Touch System application and both predicate devices, Genesis DM, and Genesis Touch System, restrict, or "locks down", the system's functionality such that the user cannot install other programs, nor use the device for any other purpose other than the intended use.
510(k) Summary
{6}------------------------------------------------
## Honevw HomMe
The modified Genesis Touch System and the predicate, Genesis Touch System both have the ability to collect vital signs data via wireless technology (Bluetooth) or manually, while the predicate Genesis DM Monitor collects vital signs data using both wireless (Bluetooth) and wired technology. The predicate Genesis DM Monitor collects; displays and transmits the vital signs for noninvasive blood pressure, pulse oximetry, pulse rate, weight and temperature with the option to connect to peripheral medical devices and collect vital signs data for blood sugar, prothrombin time/international normalized ratio (PT/INR), peak expiratory flow (PEF) and forced expiratory volume (FEV1). The predicate Genesis Touch System collects; displays and transmits the vital signs for noninvasive blood pressure, pulse oximetry, pulse rate, weight and manually entered temperature. The modified Genesis Touch System combines the physiological parameters collected, displayed and transmitted by both of the predicates to include, noninvasive blood pressure, pulse oximetry, pulse rate, weight, temperature, blood sugar, prothrombin time/international normalized ratio (PT/INR), peak expiratory flow (PEF) and forced expiratory volume (FEV1).
The functionality of the modified Genesis Touch System is substantially equivalent to the predicates Genesis Touch System and Honeywell HomMed Genesis DM Monitor such that all of the monitors use text and voice prompts to collect users' vital signs or upload data from peripheral medical devices, and then forward that data to a remote central monitoring station for retrospective review by a healthcare professional. All of the monitors are designed for home use or in a healthcare related environment. The monitors are not intended for emergency use or realtime monitoring and do not have auditory or visual alarms for out-of-limit parameters.
#### Performance (Non-clinical) Data:
The modified Genesis Touch System is a software application loaded onto a commercially available tablet, therefore no electrical safety or electromagnetic testing is required. Risk based verification and validation testing was performed in accordance with the FDA guidance's, "General Principles of Software Validation". Verification and regression testing of the existing and new feature presented for the modified Genesis Touch System, ensured that the software met performance and functional specifications for usability, hardware and software interoperability, data transfer, accuracy, display and transmission, system reliability and serviceability.
#### Conclusion
The verification, validation, and regression testing performed on the modified Genesis Touch System did not raise any new issues of safety or efficacy and demonstrated that the modified Genesis Touch System performs safely, effectively, and correctly in accordance with specifications and labeling claims as intended and was found to perform as well as or better than the predicate devices, Genesis Touch System and Genesis DM Monitor.
Based on the intended use and performance results, the modified Genesis Touch System has been shown to be substantially equivalent to the predicate devices.
#### 510(k) Summary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.