← Product Code [FRN](/productcode/FRN) · K141789

# PLUM 360 INFUSION SYSTEM WITH HOSPIRA MEDNET, SMART CARD PLUG N PLAY CE 3.0 MODULE FOR PLUM A+ INFUSION SYSTEM UPGRADE (K141789)

_Hospira, Inc. · FRN · Jan 16, 2015 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K141789

## Device Facts

- **Applicant:** Hospira, Inc.
- **Product Code:** [FRN](/productcode/FRN.md)
- **Decision Date:** Jan 16, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5725
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

Plum 360™ Infusion System with Hospira MedNet™ and Plum A+™ (360 Enabled™) Infusion System are indicated for use in parenteral, enteral and epidural therapies and the administration of whole blood and blood products.

## Device Story

Volumetric infusion system; delivers parenteral, enteral, epidural therapies and blood products via dedicated administration sets. User interface includes LCD, numeric keypad, audible/visual indicators, nurse call interface, and drug library. Hospira MedNet™ software enables communication between pump, server, and facility systems. Operated by clinicians in healthcare settings. Enhancements include CPU software v15.0 (drug library, autoprogramming, alarm management), connectivity engine CE 3.0 (improved Wi-Fi/throughput), and ergonomic hardware (enclosure, tubing guides, ratcheting pole clamp). Smart Card Plug 'n' Play module allows Plum A+™ upgrades. Output provides controlled fluid delivery; assists clinical decision-making via drug library safety limits and automated programming. Benefits include improved infusion safety, connectivity, and workflow efficiency.

## Clinical Evidence

No clinical data. Bench testing only. System verification and validation confirmed design inputs met. Human factors studies conducted to validate use-related error mitigations with passing results.

## Technological Characteristics

Volumetric infusion pump. Materials: Biocompatible per ISO 10993-1. Connectivity: Wi-Fi, network interface. Software: v15.0 (drug library, autoprogramming, alarm management). Hardware: LCD, keypad, ratcheting pole clamp, proximal tubing guides. Standards: ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-24.

## Regulatory Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

## Predicate Devices

- Plum A+™ Infusion System with Hospira Mednet™ ([K042081](/device/K042081.md))
- Abbott Plum A+ Pump and Sets ([K982159](/device/K982159.md))
- Lifeshield Latex-Free Microbore Extension Set, Model 14949 and Others ([K052722](/device/K052722.md))
- Hospira Enteral Feeding Set, Model 20640 ([K061432](/device/K061432.md))
- Hospira Infusion Blood Sets ([K101677](/device/K101677.md))
- Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe Tubing ([K103224](/device/K103224.md))
- Infusion Sets ([K103344](/device/K103344.md))
- Lifeshield Vision Infusion Set with Pre-pierced Reseal ([K113683](/device/K113683.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a human figure, represented by three overlapping profiles, with the head tilted upwards.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 16, 2015

Hospira, Incorporated Yuliva Matlin. M.S., M.B.A. Associate Director of Global Regulatory Affairs 275 North Field Drive Lake Forest, Illinois 60045

Re: K141789

Trade/Device Name: Plum 360™ Infusion System with Hospira Mednet™, Smart Card Plug 'N Play CE 3.0 Upgrade Module for Plum A+TM Infusion System, and Plum Administration Sets Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II

Product Code: FRN, PHC, FPA Dated: December 15, 2014 Received: December 17, 2014

Dear Ms. Malin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

# Tina Kiang -S

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K141789

Device Name

Plum 360 Infusion System with Hospira Mednet Smart Card Plug 'N Play CE 3.0 Upgrade Module for Plum A+ Infusion System Plum Administration Sets

Indications for Use (Describe)

Plum 360™ Infusion System with Hospira MedNet™ and Plum A+™ (360 Enabled™) Infusion System are indicated for use in parenteral, enteral and epidural therapies and the administration of whole blood products.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Hospira logo. The logo features the company name "Hospira" in a sans-serif font, with the "H" slightly larger than the other letters. Above the name is an abstract graphic consisting of three curved lines that converge at a point, creating a sense of movement or flow.

# 510(k) Summary of Safety and Effectiveness

## K141789

| Submitter Information                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|---------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name                                                                                        | Hospira, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Address                                                                                     | 275 N Field Drive, Lake Forest, IL 60045 USA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Phone Number                                                                                | 224-212-4857                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Fax Number                                                                                  | 224-212-5402                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Establishment Registration Number                                                           | 3005579246                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Name of Contact Person                                                                      | Yuliya Matlin, M.S., M.BA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Date Prepared                                                                               | June 30, 2014                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Name of Device                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Trade or Proprietary Name                                                                   | Plum 360TM Infusion System with Hospira MedNetTM                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                                                                             | Smart Card Plug'n' Play Module with CE 3.0 for Plum A+TM Infusion System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                                                                             | PlumTM Administration Sets                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Common or Usual Name                                                                        | Infusion Pumps, Infusion safety management software &<br>Intravascular Administration Sets                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Classification Name                                                                         | Infusion Pumps, Infusion safety management software &<br>Intravascular Administration Sets                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Classification Panel                                                                        | 80                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Regulation(s)                                                                               | 21 CFR 880.5725 & 21 CFR 880.5440                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Product Code(s)                                                                             | FRN, PHC, FPA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Legally Marketed Device(s) To Which<br>Equivalence is Claimed                               | Infusion System:<br>Plum A+TM Infusion System with Hospira MednetTM (K042081)<br>Administration Sets:<br>Abbott Plum A+ Pump and Sets (K982159)<br>Lifeshield Latex-Free Microbore Extentsion Set, Model 14949 and<br>Others (K052722)<br>Hospira Enteral Feeding Set, Model 20640 (K061432)<br>Hospira Infusion Blood Sets (K101677)<br>Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe<br>Tubing (K103224)                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                                                             | Infusion Sets (K103344)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                                                             | Lifeshield Vision Infusion Set with Pre-pierced Reseal (K113683)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Reason for 510(k) Submission                                                                | Introduction of the next generation of Plum™ Infusion System-Plum<br>360™ Infusion System with Hospira MedNet™ with enhanced<br>software, hardware and connectivity features.<br><br>Introduction of an optional Smart Card Plug 'n' Play module with<br>CE 3.0 for Plum A+TM Infusion System upgrade.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                                                                             | Changes to secure lock component of Plum™ Administration Sets.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Device Description                                                                          | The Plum 360TM Infusion System with Hospira MedNet™ remains a<br>volumentric Infusion System intended for use in parenteral, enteral<br>and epidural therapies and the administration of whole blood and<br>blood products. Dedicated administration sets are used for delivery of<br>infusion therapy to the patient.<br><br>The device has a user interface consisting of LCD display, front<br>panel numeric keypads and keys, audible and visual indicators,<br>audible indicator volume control, nurse call interface, network<br>interface connectors, the drug library, and software that provides pre-<br>defined display screens for therapy programming.<br><br>The continued use of the Hospira MedNet™ Software allows<br>communication between the Infusion Pump, Hospira MedNet™<br>server and the facility's communication systems. |
| Intended Use/Indication For Use                                                             | Plum 360™ Infusion System with Hospira Mednet™ and Plum<br>A+TM (360 EnabledTM) Infusion System are indicated for use in<br>parenteral, enteral and epidural therapies and the administration of<br>whole blood and blood products.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Summary of the Technological Characteristics of the Device Compared to the Predicate Device |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                                                             | Plum 360TM Infusion System with Hospira MedNet™ employs the same fundamental scientific technology, principles of<br>operation and intended/indications for use. In addition, the subject device includes the following notable characteristics:<br>• Latest CPU software version 15.0 featuring drug library enhancements, autoprogramming and alarm management<br>enhancments.<br>• Enhanced connectivity engine CE 3.0 with imporved Wi-Fi Performance, network connectivity and throughput to<br>Hospira MedNet™<br>• Enhanced hardware including egronomic enclosure, proximal tubing guides, ratcheting pole clamp and keypad.<br>• Enhancements to dedicated administration sets to include a modified secure lock.<br>• An optional Smart Card Plug 'n' Play with CE 3.0 Module allows Plum A+TM Infusion System upgrades to the           |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Hospira logo. The logo features the company name "Hospira" in a simple, sans-serif font. Above the name is an abstract graphic consisting of three curved lines that converge at the bottom, creating a stylized, arch-like shape.

### Performance Data

### Summary of Non-Clinical Tests Conducted For Determination of Substantial Equivalence

Performance Test Summary - Subject Device

configuration.

System verification and validation activities for subject devices confirmed that the system meets and design inputs. All the testing met the acceptance criteria.

Risk management activities are incorporated in to the design and development process and safety assurance cases has been generated to demonstrate the safety of the subject devcies.

Risk management as well as verification activities incorporate the principles of applicable FDA guidance such as "Total Product Life Cycle: infusion Pump –Premarket Notification [510(k)] Submissions (DRAFT GUIDANCE)"

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Hospira logo. The logo consists of the word "Hospira" in a sans-serif font, with a stylized graphic above it. The graphic is composed of three curved lines that converge at the top, resembling a stylized arch or a simplified representation of a medical device. The logo is in black and white.

### Plum 360™ Infusion system with Hospira MedNet™/ Smart Card Plug 'n' Play CE 3.0 Module Traditional 510(k) Section 5: 510(k) Summary

issued in April 23, 2010 and "General Principles of Software Validation; Final Guidance for Industry and FDA Staff," issued January 11, 2002.

Human Factors studies have been conducted to validate the effectiveness of use related error mitigations.

| Characteristics                                                                                                                                                                                                                                                                                                                                                                            | Standard/Test/FDA Guidance                                                                                                                                                                                                       | Results Summary |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|
| Electrical and Mechanical<br>Safety                                                                                                                                                                                                                                                                                                                                                        | ANSI/AAMI ES60601-1 Medical Electrical Equipment - Part<br>1: General Requirements for Basic Safety and Essential<br>Performance, 2005 / (R)2012 / C1:2009 / (R)2012, including<br>Amendment 1:2012                              | PASS            |
| Electromagnetic<br>Compatibility                                                                                                                                                                                                                                                                                                                                                           | IEC 60601-1-2 Medical electrical equipment - Part 1-2:<br>General requirements for basic safety and essential<br>performance - Collateral standard: Electromagnetic<br>compatibility - Requirements and tests, Edition 3:2007-03 | PASS            |
| Alarms System                                                                                                                                                                                                                                                                                                                                                                              | IEC 60601-1-8 Collateral Standard: General requirements,<br>tests and guidance for alarm systems in medical electrical<br>equipment and medical electrical systems, Edition 2.0 2006-10                                          | PASS            |
| Infusion Pump Basic Safety<br>and Essential Performance                                                                                                                                                                                                                                                                                                                                    | IEC 60601-2-24 Medical Electrical Equipment - Part-24:<br>Particular Requirements for the Basic Safety and Essential<br>Performance of Infusion Pump and Controllers, Edition 2.0<br>2012-10                                     | PASS            |
| Biocompatibility                                                                                                                                                                                                                                                                                                                                                                           | ISO 10993-1 Biological evaluation of medical devices - Part 1:<br>Evaluation and testing within a risk management process,<br>Fourth edition 2009-10-15                                                                          | PASS            |
| Summary of Clinical Tests Conducted for Determination of Substantial Equivalence and/or of Clinical Information                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                  |                 |
| Clinical evaluation is not required for this submission to support substantial equivalence. Human factors studies have been<br>conducted on subject devcies demonstrating passing results.                                                                                                                                                                                                 |                                                                                                                                                                                                                                  |                 |
| Conclusion                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                  |                 |
| Hospira, Inc. considers Plum 360™ Infusion System Hospira MedNet™ and Plum A+™ (360 Enabled™) Infusion System<br>with Hospira MedNet™ to be as safe and effective as its predicate devices. Therefore, Plum 360™ Infusion System<br>compatible with Hospira MedNet™ and Plum A+™ (360 Enabled™) Infusion System with Hospira MedNet™ are<br>substantially equivalent to predicate devices. |                                                                                                                                                                                                                                  |                 |

---

**Source:** [https://fda.innolitics.com/device/K141789](https://fda.innolitics.com/device/K141789)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
