The 3B Luna CPAP and Auto CPAP Systems are intended to deliver positive pressure for the treatment of Obstructive Sleep Apnea. The optional integrated humidifier is indicated for the humidification and warming of air from the flow generator. These devices are intended for single patient use by prescription in the home or hospital / institutional environment on adult patients.
Device Story
Microprocessor-controlled, blower-based CPAP/Auto-CPAP system; delivers positive airway pressure (4-20 cm H2O) to patient via mask interface. Includes optional integrated heated humidifier for air warming/humidification. Used in home or hospital environments by adult patients under prescription. Device features color LCD, menu-driven user interface, and redesigned enclosure. Healthcare providers use device output (pressure/flow monitoring) to manage sleep apnea therapy; patient benefits from improved airway patency during sleep.
Clinical Evidence
Bench testing only. Performance testing conducted per ISO 17510-1:2007. Biocompatibility testing (ISO 10993-3, -5, -6, -10) and particulate/volatile testing leveraged from predicate K132967. Electrical safety and EMC testing performed per IEC 60601-1 and IEC 60601-1-2.
Technological Characteristics
Microprocessor-controlled blower system; flame retardant engineering thermoplastic housing. Pressure range 4-20 cm H2O. Class II equipment, Type BF applied part. IP22 ingress protection. Connectivity: standalone. Power: 100-240VAC. Integrated heated humidifier (1-5 settings).
Indications for Use
Indicated for adult patients with Obstructive Sleep Apnea requiring positive airway pressure therapy. Intended for single-patient use in home or hospital/institutional settings.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a symbol consisting of three stylized human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 22, 2015
3B™ Medical, Inc Mr. Alex Lucio Vice President 21301 Highway 27 N. Lake Wales, FL 33859
Re: K141770
Trade/Device Name: Luna CPAP and Auto-CPAP System Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: BZD Dated: March 6, 2015 Received: March 23, 2015
Dear Mr. Lucio:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Lucio
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Teiashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number: K141770 Luna CPAP and Auto CPAP System
Indications for Use:
The 3B Luna CPAP and Auto CPAP Systems are intended to deliver positive pressure for the treatment of Obstructive Sleep Apnea. The optional integrated humidifier is indicated for the humidification and warming of air from the flow generator. These devices are intended for single patient use by prescription in the home or hospital / institutional environment on adult patients.
| Prescription Use: <span style="text-decoration: underline;">X</span> | AND/OR | Over-The-Counter Use: <span style="text-decoration: underline;"></span> |
|----------------------------------------------------------------------|--------|-------------------------------------------------------------------------|
| (Part 21 CFR 801 Subpar D) | | (21 CFR 801 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
## 510(k) Summary
| Device Trade Name | Luna CPAP and Auto-CPAP |
|--------------------------|-------------------------------------------|
| | Systems with Integrated Heated Humidifier |
| | |
| Common/Usual Name | CPAP System, Auto-CPAP system |
| | |
| Date Prepared | April 16, 2015 |
| | |
| Sponsor Identification | 3B Medical, Inc. |
| | |
| | 21301 Highway 27 N. |
| | Lake Wales, FL 33859 |
| Phone | 863-226-6285 |
| Fax | 863-226-6284 |
| Email | alucio@3bproducts.com |
| | |
| Submission Correspondent | Alex Lucio |
| | 3B Medical, Inc. |
| | 21301 Highway 27 N. |
| | Lake Wales, FL 33859 |
| Phone | 863-226-6285 |
| Fax | 863-226-6284 |
| Email | |
{4}------------------------------------------------
| Establishment Registration # | 3008566132 |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | BMC Medical CO., LTD<br>5/f Main Building No.19<br>Gucheng Street West, Shinjingshan<br>Beijing, CHINA 100043 |
| Classification | Class II Device |
| Classification Name | Non-continuous ventilator |
| Classification Panel | Medical Device |
| Classification Reference | 21 CFR 878.5905 |
| Products Code | BZD |
| Medical Specialties | Anesthesiology |
| Predicate Device(s) | RESmart® CPAP and Auto-CPAP Systems<br>(K132967) |
| Reason for Submission: | Device Modification |
| Intended Use | The 3B Luna CPAP and Auto CPAP Systems<br>are intended to deliver positive pressure for the<br>treatment of Obstructive Sleep Apnea. The<br>optional integrated humidifier is indicated for the<br>humidification and warming of air from the flow<br>generator. These devices are intended for single<br>patient use by prescription in the home or<br>hospital / institutional environment on adult<br>patients. |
| Device Description | The Luna CPAP and Auto CPAP System is a<br>microprocessor-controlled, blower-based system<br>that generates positive airway pressure from 4 to<br>20 cm H20. The device is intended for |
2
{5}------------------------------------------------
use with a patient interface (mask). The device has been modified to include a color LCD, menu driven user interface, and a redesigned enclosure. The electrical circuit was redesigned to incorporate the color LCD. The basic functionality and performance characteristics of the Luna CPAP and Auto CPAP are unchanged from the predicate device RESmart CPAP and Auto CPAP (K132967).
Non-Clinical Testing Extensive non-clinical testing was conducted according to ISO 17510-1:2007, Sleep Apnea Breathing Therapy-Part I: Sleep Apnea Breathing Therapy Equipment. Side by side Performance Testing Bench equivalence with the predicated device.
- Biocompatibility The materials in the predicate device are identical and manufactured with the same manufacturing processes as the predicate device K132967; hence biocompatibility testing is not required.
Materials used in the construction of components that contact the heated humidified gas pathway permanent are classified communicating devices" (with tissue/bone/dentin).
The appropriate biological tests conducted and passed for these components, in accordance with FDA guidance #G95-1- were performed in K132967:
- ISO 10993-3 Genotoxicity, .
- ISO 10993-5 Cytotoxicity ●
- ISO 10993-6 Implantation and ●
- ISO 10993-10 Sensitization and Irritation ●
{6}------------------------------------------------
Testing for particulate matter and volatiles also demonstrated compliance to EPA requirements in K132967.
The Luna CPAP has been tested to appropriate standards and other applicable requirements. The Luna CPAP with integrated heated humidifier was designed and tested according to:
- IEC equipment - Part 1: General Requirements for safety Medical electrical equipment - General requirements for basic safety and essential performance
- IEC 60601-1-2:2007, medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility – Requirements and tests
The proposed and predicate devices have identical materials, indications for use, and operating principles. Testing and validation of component part upgrades establish substantial equivalence between predicate and proposed devices.
Substantial Equivalence The proposed Luna CPAP and Auto CPAP System remain substantially equivalent to the RESmart II CPAP and AutoCPAP/Luna CPAP and AutoCPAP System (K132967) in that they have the same intended use, same operating principle, technology, identical materials, and manufacturing process. Designed validation and verification
{7}------------------------------------------------
tests were performed on the Luna CPAP and Auto-CPAP System because of the risk analysis and product requirements.
## Comparative Summary of Technological Characteristics
| Features/<br>Function | Proposed Device (K141770)<br>Luna CPAP and Auto CPAP Systems | Predicate Device (K132967)<br>RESmart CPAP/APAP<br>(Private labeled as 3B CPAP/APAP) |
|--------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Therapy<br>Delivered | CPAP,<br>Auto CPAP | CPAP,<br>Auto CPAP |
| Operation<br>Temperature | 5 to 35° C<br>(41 to | 5 to 30° C<br>(41 to 86 F) |
| Storage/Transport<br>Temperature | -25 to 70° C | -20 to 55° C |
| Humidity | 15% to 93% Non-condensing | ≤ 80% Non-condensing |
| Atmospheric<br>Pressure | 76 to 106 kPa | 86 to 106 kPa |
| Standards<br>Compliance | IEC 60601-1 General Requirements for Safety of<br>Medical Electrical Equipment<br>IEC60601-1-11 Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment<br>IEC 60601-1-2 Electromagnetic Compatibility | IEC 60601-1 General Requirements for Safety of<br>Medical Electrical Equipment<br>IEC 60601-1-2 Electromagnetic Compatibility<br>ISO 8185 General Requirements for Humidification Systems |
| Mode of<br>Operation | Continuous | Continuous |
| AC Power<br>Consumption | 100-240VAC, 50/60Hz,<br>2.0A max | 100-240VAC, 50/60Hz,<br>1.0A max |
| Software | Microprocessor controlled | Microprocessor controlled |
| System<br>Contents | Air blower, pressure-flow monitoring, pressure<br>controlling, user interface, heated humidifier,<br>power cord, carrying case, user manual. | Air blower, pressure-flow monitoring, pressure<br>controlling, user interface, heated humidifier,<br>power cord, carrying case, user manual. |
| Type of<br>Protection<br>Against Electric<br>Shock | Class II Equipment | Class II Equipment |
| Degree of<br>Protection<br>Against Electric<br>Shock | Type BF Vertical Applied Part | Type BF Vertical Applied Part |
| Degree of<br>Protection<br>Against Ingress<br>of Water | IP22 | IPX1-Drip- Proof, Vertical |
| Pressure Range | 4-20 cmH20 (in 0.5 cmH20 increments) | 4-20 cmH20 (in 0.5 cmH2) increments) |
| Sound Pressure<br>Level | <30 dB, when the device is working at the<br>pressure of 10 cmH20 | <30 dB, when the device is working at the pressure<br>of 10 cmH2O. |
| Housing | flame retardant Engineering thermoplastic | flame retardant Engineering thermoplastic |
| Pressure<br>Display<br>Accuracy<br>(cmH20) | $\pm (0.5+4\%)$ | 0.5 |
| Ramp (minutes) | 0-60 | 0-60 |
| Mask off alert | Yes | Yes |
| Integrated<br>Humidifier | Yes | Yes |
| Static and<br>dynamic<br>pressure<br>accuracies | 4 to 20 cmH20( $\pm$ 1 cmH20)<br>Measured in accordance with preen 17510 @ 6.6,<br>13.2, & 20 cm H2O @ 500 ml with BPM set to<br>10, 15, & 20 BPM performed at 23° C ( $\pm$ 2° C),<br>50% RH ( $\pm$ 5%), and an atmospheric pressure of<br>101.5 kPa. | 4 to 20 cmH20( $\pm$ 1 cmH20)<br>Measured in accordance with prEN 17510 @ 6.6,<br>13.2, & 20 cm H2O @ 500 ml with BPM set to 10,<br>15, & 20 BPM performed at 23° C ( $\pm$ 2° C), 50% RH<br>( $\pm$ 5%), and an atmospheric pressure of 101.5 kPa. |
| | Humidifier | |
| | Yes | Yes |
| | Water Capacity: 350 ml at recommended water level | Water Capacity :> 350 ml |
| | Heater Settings: 1 to 5 (95 to 167 °F) | Heater Settings: 1 to 5 (104 to 149 °F) |
| | Pressure Drop with Humidifier: < 0.4 cmH20 at 60 LPM flow | Pressure Drop with Humidifier :< 0.5 cmH20 at 60 LPM flow; |
| | Humidity Range: >10mg/L | Humidity Range: 10 to 40 mg H2O/L |
{8}------------------------------------------------
{9}------------------------------------------------
Conclusions There have been no changes in the material composition, intended use, or or operating principles. Performance bench testing and device validation and verification demonstrate that the proposed device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.