← Product Code [IYN](/productcode/IYN) · K141768

# LOGIQ V5/LOGIQ V3 (K141768)

_GE Healthcare · IYN · Aug 4, 2014 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K141768

## Device Facts

- **Applicant:** GE Healthcare
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** Aug 4, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

LOGIQ V5/LOGIQ V3 is a general-purpose ultrasound system used for acquiring and storing ultrasound images for applications such as: Obstetrics; Abdominal, Obstetrical, Gynecological, Small parts, Vascular/Peripheral Vascular, Adult Cephalic, Pediatric, Musculoskeletal, Transcranial, Neonatal Cephalic, Urological and Cardiac. The system used for performing measurements, annotating, printing/reporting on the acquired/stored images.

## Device Story

General-purpose entry-level ultrasound scanner for private clinics; focuses on OB/GYN. Takes acoustic signals via transducers; transforms inputs into real-time digital images/measurements; displays on system monitor. Used by clinicians for general radiology, cardiology, and vascular evaluation. Features include image freeze, cine loop, spatial/temporal filters, digital harmonics, raw data access, LOGIQ View, CrossXBeam, SRI-HD, B-Flow, TVI, Anatomical-M mode, and Auto Optimization. Output assists healthcare providers in diagnostic imaging and fluid flow analysis; aids clinical decision-making by providing visual data for patient assessment.

## Clinical Evidence

No clinical studies were required to support substantial equivalence. Bench testing only, including acoustic output, biocompatibility, cleaning/disinfection effectiveness, thermal, electrical, electromagnetic, and mechanical safety testing.

## Technological Characteristics

Ultrasonic pulsed doppler imaging system. Materials evaluated per ISO 10993-1. Complies with AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, NEMA UD 2, and NEMA UD 3. Supports DICOM connectivity. Imaging modes: B, M, PW/CW Doppler, Color, Power, Color M, Harmonic, and Combined modes.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- GE LOGIQ F Series ([K133034](/device/K133034.md))
- Voluson E series ([K131267](/device/K131267.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 4, 2014

GE HEALTHCARE BRYAN BEHN REGULATORY AFFAIRS MANAGER 9900 INNOVATION DRIVE RP-2138 WAUWATOSA WI 53226

Re: K141768 Trade/Device Name: Logiq V5 / Logiq V3 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: June 30, 2014 Received: July 1, 2014

Dear Mr. Behn:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

This determination of substantial equivalence applies to the following transducers intended for use with the Logiq V5 and Logiq V3, as described in your premarket notification:

| Transducer Model Number |
|-------------------------|
| 4C-RS                   |
| E8C-RS                  |
| 3Sc-RS                  |
| L6-12-RS                |

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Smh.7)

for

Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

K141768

Device Name

LOGIQ V5/LOGIQ V3 Diagnostic Ultrasound System

Indications for Use (Describe)

The LOGIQ V5/LOGIQ V3 is general purposed ultrasound imaging and analysis systems providing digital acquisition, processing and display capability and clinical applications including: Abdominal, Obstetrical, Gynecological, Small parts, Vascular/Peripheral Vascular, Adult Cephalic, Pediatric, Musculoskeletal, Transcranial, Neonatal Cephalic, Transvaginal, Urological and Cardiac.

Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

## FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881(1/14)

Page 1 of 7

{3}------------------------------------------------

# Indications for Use Forms

The following Indications for Use forms are appended.

System: LOGIQ V5/LOGIQ V3 Transducer: 4C-RS Transducer: 3Sc-RS Transducer: L6-12-RS Transducer: E8C-RS

The following forms represent indications with clinical applications and exam types along with the modes of operation for the LOGIQ V5 and LOGIQ V3. Combinations identified "P" for the transducers represents those previously cleared with another GE Ultrasound system. Combinations identified as "N" are new.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo, which consists of the letters 'G' and 'E' intertwined within a circular frame. The logo is rendered in a light blue color, and the design has a vintage or classic feel. The circular frame around the letters has a decorative, swirling pattern.

#### Diagnostic Ultrasound Indications for Use Form

## LOGIQ V5 Ultrasound System

Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:

| Clinical Application          | Mode of Operation |   |    |    |               |            |       |   |                     |                     |                |       |
|-------------------------------|-------------------|---|----|----|---------------|------------|-------|---|---------------------|---------------------|----------------|-------|
|                               |                   |   |    |    | Doppler Modes |            |       |   | Combined<br>Modes * | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Anatomy/Region of Interest    | B                 | M | PW | CW | Color         | Color<br>M | Power |   |                     |                     |                |       |
| Ophthalmic                    |                   |   |    |    |               |            |       |   |                     |                     |                |       |
| Fetal/OB                      | N                 | N | N  | N  | N             | N          | N     | N | N                   | N                   | N              |       |
| Abdominal[1]                  | N                 | N | N  | N  | N             | N          | N     | N | N                   | N                   | N              |       |
| Pediatric                     | N                 | N | N  | N  | N             | N          | N     | N | N                   | N                   | N              |       |
| Small Organ (specify)[2]      | N                 | N | N  | N  | N             | N          | N     | N | N                   | N                   | N              |       |
| Neonatal Cephalic             | N                 | N | N  | N  | N             | N          | N     | N | N                   | N                   | N              |       |
| Adult Cephalic                | N                 | N | N  | N  | N             | N          | N     | N | N                   | N                   | N              |       |
| Cardiac [3]                   | N                 | N | N  | N  | N             | N          | N     | N | N                   | N                   | N              |       |
| Peripheral Vascular           | N                 | N | N  | N  | N             | N          | N     | N | N                   | N                   | N              |       |
| Musculo-skeletal Conventional | N                 | N | N  | N  | N             | N          | N     | N | N                   | N                   | N              |       |
| Musculo-skeletal Superficial  | N                 | N | N  | N  | N             | N          | N     | N | N                   | N                   | N              |       |
| Thoracic/Pleural (specify)    |                   |   |    |    |               |            |       |   |                     |                     |                |       |
| Other (specify)               |                   |   |    |    |               |            |       |   |                     |                     |                |       |
| Exam Type, Means of Access    |                   |   |    |    |               |            |       |   |                     |                     |                |       |
| Transcranial                  | N                 | N | N  | N  | N             | N          | N     | N | N                   | N                   | N              |       |
| Transorbital                  |                   |   |    |    |               |            |       |   |                     |                     |                |       |
| Transesophageal               |                   |   |    |    |               |            |       |   |                     |                     |                |       |
| Transrectal                   |                   |   |    |    |               |            |       |   |                     |                     |                |       |
| Transvaginal                  | N                 | N | N  | N  | N             | N          | N     | N | N                   | N                   | N              |       |
| Intraoperative (specify)      |                   |   |    |    |               |            |       |   |                     |                     |                |       |
| Intraoperative Neurological   |                   |   |    |    |               |            |       |   |                     |                     |                |       |
| Intravascular/Intraluminal    |                   |   |    |    |               |            |       |   |                     |                     |                |       |
| Intracardiac                  |                   |   |    |    |               |            |       |   |                     |                     |                |       |
| Laparoscopic                  |                   |   |    |    |               |            |       |   |                     |                     |                |       |

[1] Abdominal includes GYN and Urological; Notes:

[2] Small Organ includes breast, testes, and thyroid;

[3] Cardiac is Adult and Pediatric;

[*] Combined modes are color/power Doppler with B-mode

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Prescription Use (Per 21 CFR 801.109)

Page 2 of 7

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo, which consists of the letters 'G' and 'E' intertwined within a circular frame. The logo is rendered in a light blue color, providing a clean and recognizable design. The intertwined letters are stylized with elegant curves, giving the logo a classic and timeless appearance. The circular frame adds a sense of unity and completeness to the overall design.

#### Diagnostic Ultrasound Indications for Use Form

## LOGIQ V3 Ultrasound System

Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:

| Clinical Application          | Mode of Operation |   |    |    |               |            |       |   |                     |                     |                |       |
|-------------------------------|-------------------|---|----|----|---------------|------------|-------|---|---------------------|---------------------|----------------|-------|
|                               |                   |   |    |    | Doppler Modes |            |       |   | Combined<br>Modes * | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Anatomy/Region of Interest    | B                 | M | PW | CW | Color         | Color<br>M | Power |   |                     |                     |                |       |
| Ophthalmic                    |                   |   |    |    |               |            |       |   |                     |                     |                |       |
| Fetal/OB                      | N                 | N | N  | N  | N             | N          | N     | N | N                   | N                   | N              |       |
| Abdominal[1]                  | N                 | N | N  | N  | N             | N          | N     | N | N                   | N                   | N              |       |
| Pediatric                     | N                 | N | N  | N  | N             | N          | N     | N | N                   | N                   | N              |       |
| Small Organ (specify)[2]      | N                 | N | N  | N  | N             | N          | N     | N | N                   | N                   | N              |       |
| Neonatal Cephalic             | N                 | N | N  | N  | N             | N          | N     | N | N                   | N                   | N              |       |
| Adult Cephalic                | N                 | N | N  | N  | N             | N          | N     | N | N                   | N                   | N              |       |
| Cardiac[3]                    | N                 | N | N  | N  | N             | N          | N     | N | N                   | N                   | N              |       |
| Peripheral Vascular           | N                 | N | N  | N  | N             | N          | N     | N | N                   | N                   | N              |       |
| Musculo-skeletal Conventional | N                 | N | N  | N  | N             | N          | N     | N | N                   | N                   | N              |       |
| Musculo-skeletal Superficial  | N                 | N | N  | N  | N             | N          | N     | N | N                   | N                   | N              |       |
| Thoracic/Pleural (specify)    |                   |   |    |    |               |            |       |   |                     |                     |                |       |
| Other (specify)               |                   |   |    |    |               |            |       |   |                     |                     |                |       |
| Exam Type, Means of Access    |                   |   |    |    |               |            |       |   |                     |                     |                |       |
| Transcranial                  | N                 | N | N  | N  | N             | N          | N     | N | N                   | N                   | N              |       |
| Transorbital                  |                   |   |    |    |               |            |       |   |                     |                     |                |       |
| Transesophageal               |                   |   |    |    |               |            |       |   |                     |                     |                |       |
| Transrectal                   |                   |   |    |    |               |            |       |   |                     |                     |                |       |
| Transvaginal                  | N                 | N | N  | N  | N             | N          | N     | N | N                   | N                   | N              |       |
| Intraoperative (specify)      |                   |   |    |    |               |            |       |   |                     |                     |                |       |
| Intraoperative Neurological   |                   |   |    |    |               |            |       |   |                     |                     |                |       |
| Intravascular/Intraluminal    |                   |   |    |    |               |            |       |   |                     |                     |                |       |
| Intracardiac                  |                   |   |    |    |               |            |       |   |                     |                     |                |       |
| Laparoscopic                  |                   |   |    |    |               |            |       |   |                     |                     |                |       |

[1] Abdominal includes GYN and Urological; Notes:

[2] Small Organ includes breast, testes, and thyroid;

[3] Cardiac is Adult and Pediatric;

[*] Combined modes are color/power Doppler with B-mode

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Prescription Use (Per 21 CFR 801.109)

Page 3 of 7

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circle. The circle has a double-line border, with the inner line being slightly thicker than the outer line.

GE Healthcare 510(k) Premarket Notification Submission

## Diagnostic Ultrasound Indications for Use Form LOGIQ V5/LOGIQ V3 with 4C-RS Transducer

Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:

| Clinical Application          | Mode of Operation |   |    |    |               |            |       |   |                    |                     |                |  |
|-------------------------------|-------------------|---|----|----|---------------|------------|-------|---|--------------------|---------------------|----------------|--|
|                               |                   |   |    |    | Doppler Modes |            |       |   | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse |  |
| Anatomy/Region of Interest    | B                 | M | PW | CW | Color         | Color<br>M | Power |   |                    |                     | Other          |  |
| Ophthalmic                    |                   |   |    |    |               |            |       |   |                    |                     |                |  |
| Fetal/OB                      | P                 | P | P  | P  | P             | P          | P     | P | P                  | P                   | P              |  |
| Abdominal [1]                 | P                 | N | P  | P  | P             | N          | P     | P | P                  | P                   | P              |  |
| Pediatric                     |                   |   |    |    |               |            |       |   |                    |                     |                |  |
| Small Organ (specify) [2]     | N                 | N | N  | N  | N             | N          | N     | N | N                  | N                   | N              |  |
| Neonatal Cephalic             |                   |   |    |    |               |            |       |   |                    |                     |                |  |
| Adult Cephalic                |                   |   |    |    |               |            |       |   |                    |                     |                |  |
| Cardiac [3]                   |                   |   |    |    |               |            |       |   |                    |                     |                |  |
| Peripheral Vascular           | N                 | N | N  | N  | N             | N          | N     | N | N                  | N                   | N              |  |
| Musculo-skeletal Conventional |                   |   |    |    |               |            |       |   |                    |                     |                |  |
| Musculo-skeletal Superficial  |                   |   |    |    |               |            |       |   |                    |                     |                |  |
| Thoracic/Pleural (specify)    |                   |   |    |    |               |            |       |   |                    |                     |                |  |
| Other (specify)               |                   |   |    |    |               |            |       |   |                    |                     |                |  |
| Exam Type, Means of Access    |                   |   |    |    |               |            |       |   |                    |                     |                |  |
| Transcranial                  |                   |   |    |    |               |            |       |   |                    |                     |                |  |
| Transorbital                  |                   |   |    |    |               |            |       |   |                    |                     |                |  |
| Transesophageal               |                   |   |    |    |               |            |       |   |                    |                     |                |  |
| Transrectal                   |                   |   |    |    |               |            |       |   |                    |                     |                |  |
| Transvaginal                  |                   |   |    |    |               |            |       |   |                    |                     |                |  |
| Intraoperative (specify)      |                   |   |    |    |               |            |       |   |                    |                     |                |  |
| Intraoperative Neurological   |                   |   |    |    |               |            |       |   |                    |                     |                |  |
| Intravascular/Intraluminal    |                   |   |    |    |               |            |       |   |                    |                     |                |  |
| Intracardiac                  |                   |   |    |    |               |            |       |   |                    |                     |                |  |
| Laparoscopic                  |                   |   |    |    |               |            |       |   |                    |                     |                |  |

N = new indication; P = previously cleared by FDA (K133034)

[1] Abdominal includes GYN and Urological; Notes:

[2] Small Organ includes breast, testes, and thyroid;

[3] Cardiac is Adult and Pediatric;

[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PDI

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Prescription Use (Per 21 CFR 801.109)

Page 4 of 7

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a blue circle. The circle has decorative swirls around the perimeter, giving it a classic and recognizable appearance.

GE Healthcare 510(k) Premarket Notification Submission

## Diagnostic Ultrasound Indications for Use Form LOGIQ V5/LOGIQ V3 with E8C-RS Transducer

Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:

| Clinical Application              | Mode of Operation |   |    |    |               |   |                    |                     |                |       |
|-----------------------------------|-------------------|---|----|----|---------------|---|--------------------|---------------------|----------------|-------|
| <i>Anatomy/Region of Interest</i> | B                 | M | PW | CW | Doppler Modes |   | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic                        |                   |   |    |    |               |   |                    |                     |                |       |
| Fetal/OB                          | P                 | P | P  |    | P             | P | P                  | P                   | N              |       |
| Abdominal [1]                     | N                 | N | N  |    | N             | N | N                  | N                   | N              |       |
| Pediatric                         |                   |   |    |    |               |   |                    |                     |                |       |
| Small Organ (specify)[2]          |                   |   |    |    |               |   |                    |                     |                |       |
| Neonatal Cephalic                 |                   |   |    |    |               |   |                    |                     |                |       |
| Adult Cephalic                    |                   |   |    |    |               |   |                    |                     |                |       |
| Cardiac [3]                       |                   |   |    |    |               |   |                    |                     |                |       |
| Peripheral Vascular               |                   |   |    |    |               |   |                    |                     |                |       |
| Musculo-skeletal Conventional     |                   |   |    |    |               |   |                    |                     |                |       |
| Musculo-skeletal Superficial      |                   |   |    |    |               |   |                    |                     |                |       |
| Thoracic/Pleural (specify)        |                   |   |    |    |               |   |                    |                     |                |       |
| Other (specify)                   |                   |   |    |    |               |   |                    |                     |                |       |
| Exam Type, Means of Access        |                   |   |    |    |               |   |                    |                     |                |       |
| Transcranial                      |                   |   |    |    |               |   |                    |                     |                |       |
| Transorbital                      |                   |   |    |    |               |   |                    |                     |                |       |
| Transesophageal                   |                   |   |    |    |               |   |                    |                     |                |       |
| Transrectal                       |                   |   |    |    |               |   |                    |                     |                |       |
| Transvaginal                      | P                 | P | P  |    | P             | P | P                  | P                   | N              |       |
| Intraoperative (specify)          |                   |   |    |    |               |   |                    |                     |                |       |
| Intraoperative Neurological       |                   |   |    |    |               |   |                    |                     |                |       |
| Intravascular/Intraluminal        |                   |   |    |    |               |   |                    |                     |                |       |
| Intracardiac                      |                   |   |    |    |               |   |                    |                     |                |       |
| Laparoscopic                      |                   |   |    |    |               |   |                    |                     |                |       |

N = new indication; P = previously cleared by FDA (K133034)

 [1] Abdominal includes GYN and Urological; Notes:

[2] Small Organ includes breast, testes, and thyroid

[3] Cardiac is Adult and Pediatric;

[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PDI

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Prescription Use (Per 21 CFR 801.109)

Page 5 of 7

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined in a stylized script, enclosed within a circular frame. The logo is colored in a light blue hue, and the background is white.

GE Healthcare 510(k) Premarket Notification Submission

## Diagnostic Ultrasound Indications for Use Form LOGIQ V5/LOGIQ V3 with 3Sc-RS Transducer

| Clinical Application<br>Anatomy/Region of Interest | B | M | Doppler Modes |   |   |   | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other |
|----------------------------------------------------|---|---|---------------|---|---|---|--------------------|---------------------|----------------|-------|
| Ophthalmic                                         |   |   |               |   |   |   |                    |                     |                |       |
| Fetal/OB                                           |   |   |               |   |   |   |                    |                     |                |       |
| Abdominal[1]                                       | P | P | P             | N | P | P | P                  | P                   | P              | N     |
| Pediatric                                          | P | P | P             | N | P | P | P                  | P                   | P              | N     |
| Small Organ (specify)[2]                           |   |   |               |   |   |   |                    |                     |                |       |
| Neonatal Cephalic                                  |   |   |               |   |   |   |                    |                     |                |       |
| Adult Cephalic                                     | N | N | N             | N | N | N | N                  | N                   | N              | N     |
| Cardiac[3]                                         | P | P | P             | N | P | P | P                  | P                   | P              | N     |
| Peripheral Vascular                                |   |   |               |   |   |   |                    |                     |                |       |
| Musculo-skeletal Conventional                      |   |   |               |   |   |   |                    |                     |                |       |
| Musculo-skeletal Superficial                       |   |   |               |   |   |   |                    |                     |                |       |
| Thoracic/Pleural (specify)                         |   |   |               |   |   |   |                    |                     |                |       |
| Other (specify)                                    |   |   |               |   |   |   |                    |                     |                |       |
| Exam Type, Means of Access                         |   |   |               |   |   |   |                    |                     |                |       |
| Transcranial                                       | P | P | P             | N | P | P | P                  | P                   | P              | N     |
| Transorbital                                       |   |   |               |   |   |   |                    |                     |                |       |
| Transesophageal                                    |   |   |               |   |   |   |                    |                     |                |       |
| Transrectal                                        |   |   |               |   |   |   |                    |                     |                |       |
| Transvaginal                                       |   |   |               |   |   |   |                    |                     |                |       |
| Intraoperative (specify)                           |   |   |               |   |   |   |                    |                     |                |       |
| Intraoperative Neurological                        |   |   |               |   |   |   |                    |                     |                |       |
| Intravascular/Intraluminal                         |   |   |               |   |   |   |                    |                     |                |       |
| Intracardiac                                       |   |   |               |   |   |   |                    |                     |                |       |
| Laparoscopic                                       |   |   |               |   |   |   |                    |                     |                |       |

Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:

N = new indication; P = previously cleared by FDA (K133034)

[1] Abdominal includes GYN and Urological Notes:

[2] Small Organ includes breast, testes, and thyroid

[3] Cardiac is Adult and Pediatric

[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PDI

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Prescription Use (Per 21 CFR 801.109)

Page 6 of 7

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined in a stylized script, enclosed within a circular frame. The logo is colored in a light blue hue, and the background is white.

GE Healthcare 510(k) Premarket Notification Submission

## Diagnostic Ultrasound Indications for Use Form LOGIQ V5/LOGIQ V3 with L6-12-RS Transducer

| Clinical Application          | Mode of Operation |   |    |    |       |         |       |                    |                     |                |       |
|-------------------------------|-------------------|---|----|----|-------|---------|-------|--------------------|---------------------|----------------|-------|
| Anatomy/Region of Interest    | B                 | M | PW | CW | Color | Color M | Power | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic                    |                   |   |    |    |       |         |       |                    |                     |                |       |
| Fetal/OB                      |                   |   |    |    |       |         |       |                    |                     |                |       |
| Abdominal[1]                  | N                 | N | N  | N  | N     | N       | N     | N                  | N                   | N              | N     |
| Pediatric                     | P                 | N | P  |    | P     | N       | P     | P                  | P                   | P              | N     |
| Small Organ (specify)[2]      | P                 | N | P  |    | P     | N       | P     | P                  | P                   | P              | N     |
| Neonatal Cephalic             | N                 | N | N  |    | N     | N       | N     | N                  | N                   | N              | N     |
| Adult Cephalic                |                   |   |    |    |       |         |       |                    |                     |                |       |
| Cardiac[3]                    |                   |   |    |    |       |         |       |                    |                     |                |       |
| Peripheral Vascular           | P                 | N | P  |    | P     | N       | P     | P                  | P                   | P              | N     |
| Musculo-skeletal Conventional | P                 | N | P  |    | P     | N       | P     | P                  | P                   | P              | N     |
| Musculo-skeletal Superficial  | P                 | N | P  |    | P     | N       | P     | P                  | P                   | P              | N     |
| Thoracic/Pleural (specify)    |                   |   |    |    |       |         |       |                    |                     |                |       |
| Other (specify)               |                   |   |    |    |       |         |       |                    |                     |                |       |
| Exam Type, Means of Access    |                   |   |    |    |       |         |       |                    |                     |                |       |
| Transcranial                  |                   |   |    |    |       |         |       |                    |                     |                |       |
| Transorbital                  |                   |   |    |    |       |         |       |                    |                     |                |       |
| Transesophageal               |                   |   |    |    |       |         |       |                    |                     |                |       |
| Transrectal                   |                   |   |    |    |       |         |       |                    |                     |                |       |
| Transvaginal                  |                   |   |    |    |       |         |       |                    |                     |                |       |
| Intraoperative (specify)      |                   |   |    |    |       |         |       |                    |                     |                |       |
| Intraoperative Neurological   |                   |   |    |    |       |         |       |                    |                     |                |       |
| Intravascular/Intraluminal    |                   |   |    |    |       |         |       |                    |                     |                |       |
| Intracardiac                  |                   |   |    |    |       |         |       |                    |                     |                |       |
| Laparoscopic                  |                   |   |    |    |       |         |       |                    |                     |                |       |

Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:

N = new indication; P = previously cleared by FDA (K133034)

Notes: [1] Abdominal includes GYN and Urological;

[2] Small Organ includes breast, testes, and thyroid;

[3] Cardiac is Adult and Pediatric;

[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PDI

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Prescription Use (Per 21 CFR 801.109)

{10}------------------------------------------------

## 510(k) Summary

| In accordance with 21 CFR 807.92 the following summary of information is provided: |                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date:                                                                              | June 30, 2014                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Submitter:                                                                         | GE Healthcare<br>9900 Innovation Dr<br>Wauwatosa, WI 53226                                                                                                                                                                                                                                                                                                                                                                                  |
| Primary Contact Person:                                                            | Bryan Behn<br>Regulatory Affairs Manager<br>GE Healthcare<br>T:(414)721-4214<br>F:(414)918-8275                                                                                                                                                                                                                                                                                                                                             |
| Secondary Contact Person:                                                          | Jiawei ZHANG<br>Regulatory Affairs<br>GE Healthcare<br>T: +86 510 8527 8259<br>F: +86 510 8522 7347                                                                                                                                                                                                                                                                                                                                         |
| Device: Trade Name:                                                                | LOGIQ V5, LOGIQ V3                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Common/Usual Name:                                                                 | LOGIQ V5, LOGIQ V3                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Classification Names:                                                              | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                                                    | Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550                                                                                                                                                                                                                                                                                                                                                                                    |
| Product Code:                                                                      | IYN Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560,<br>90-IYO Diagnostic Ultrasound Transducer, 21 CFR 892.1570,<br>90-ITX                                                                                                                                                                                                                                                                                                           |
| Predicate Device(s):                                                               | K133034 GE LOGIQ F Series<br>K131267 Voluson E series                                                                                                                                                                                                                                                                                                                                                                                       |
| Device Description:                                                                | The LOGIQ V5/LOGIQ V3 is a general-purpose entry level<br>ultrasound scanner from LOGIQ family for private clinics<br>focusing on OB/GYN. The device is a general-purpose imaging<br>and analysis system providing real-time digital acquisition,<br>processing and display capability intended for general radiology<br>imaging and evaluation with some cardiology and vascular<br>applications.                                          |
| Intended Use:                                                                      | LOGIQ V5/LOGIQ V3 is a general-purpose ultrasound system<br>used for acquiring and storing ultrasound images for applications<br>such as: Obstetrics; Abdominal, Obstetrical, Gynecological,<br>Small parts, Vascular/Peripheral Vascular, Adult Cephalic,<br>Pediatric, Musculoskeletal, Transcranial, Neonatal Cephalic,<br>Urological and Cardiac. The system used for performing<br>measurements, annotating, printing/reporting on the |

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized script, enclosed within a blue circle. The circle has decorative swirls around the letters, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, and is widely recognized as the symbol of the General Electric company.

acquired/stored images.

LOGIO V5 and LOGIO V3 employ the same fundamental Technology: scientific technology as its predicate devices.

Comparison to Predicate Devices Determination of Substantial Equivalence: The LOGIQ V5 and LOGIQ V3 systems are substantially equivalent to the predicate devices with regard to intended use, imaging capabilities, technological characteristics and safety and effectiveness.

- o The systems are all intended for diagnostic ultrasound imaging and fluid flow analysis.
- The LOGIQ V5/LOGIQ V3 has a subset of clinical indications for use available on the predicate LOGIQ F series.
- o The LOGIQ V5/LOGIQ V3 has a subset of imaging modes on the predicate LOGIQ F series.
- The LOGIQ V5/LOGIQ V3 has a subset of transducers on ● the predicate LOGIQ F series.
- The systems are manufactured with materials which have been evaluated and found to be safe for the intended use of the device.
- . The systems have acoustic power levels which are below the applicable FDA limits.
- . The LOGIO V5/LOGIO V3 and the predicate LOGIO F series systems have similar capability in terms of performing measurements, capturing digital images, reviewing and reporting studies.
- The LOGIQ V5/LOGIQ V3 and predicate systems have been designed in compliance with approved electrical and physical safety standards.

|                                                                                                                                                                                                                                                                                                      | Proposed Device<br>LOGIQ V5/LOGIQ<br>V3 | Predicate Device<br>LOGIQ F Series<br>(K133034) |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|-------------------------------------------------|
| Indications and Clinical Applications:                                                                                                                                                                                                                                                               |                                         |                                                 |
| • Fetal/Obstetrics;                                                                                                                                                                                                                                                                                  | ✓                                       | ✓                                               |
| • Abdominal/<br>GYN and Urological                                                                                                                                                                                                                                                                   | ✓                                       | ✓                                               |
| • Small Organ (breast, testes,<br>thyroid);                                                                                                                                                                                                                                                          | ✓                                       | ✓                                               |
| • Pediatric                                                                                                                                                                                                                                                                                          | ✓                                       | ✓                                               |
| • Neonatal Cephalic;                                                                                                                                                                                                                                                                                 | ✓                                       | ✓                                               |
| • Adult Cephalic;                                                                                                                                                                                                                                                                                    | ✓                                       | ✓                                               |
| • Cardiac (adult and pediatric);                                                                                                                                                                                                                                                                     | ✓                                       | ✓                                               |
| • Peripheral Vascular;                                                                                                                                                                                                                                                                               | ✓                                       | ✓                                               |
| • Musculo-skeletal Conventional and Superficial;                                                                                                                                                                                                                                                     | ✓                                       | ✓                                               |
| • Transcranial                                                                                                                                                                                                                                                                                       | ✓                                       | ✓                                               |
| • Transvaginal;                                                                                                                                                                                                                                                                                      | ✓                                       | ✓                                               |
| Contact Type<br>• Surface, Cavitary , TEE                                                                                                                                                                                                                                                            | ✓                                       | ✓                                               |
| Image modes:<br>• B; M; Color, Power, PW& CW Doppler modes, Color M-mode, Harmonic imaging, Combined modes                                                                                                                                                                                           | ✓                                       | ✓                                               |
| Transducers                                                                                                                                                                                                                                                                                          |                                         |                                                 |
| • 3Sc-RS                                                                                                                                                                                                                                                                                             | ✓                                       | ✓                                               |
| • 4C-RS                                                                                                                                                                                                                                                                                              | ✓                                       | ✓                                               |
| • E8C-RS                                                                                                                                                                                                                                                                                             | ✓                                       | ✓                                               |
| • L6-12-RS                                                                                                                                                                                                                                                                                           | ✓                                       | ✓                                               |
| • 8C-RS                                                                                                                                                                                                                                                                                              | ✓                                       | ✓                                               |
| • RAB2-6-RS                                                                                                                                                                                                                                                                                          | ✓                                       | ✓                                               |
| Processing &Display features:<br>Image freeze, Multiple images, Pan /Zoom, Image maps (color & gray), Cine loop, Spatial & temporal filters, L-R / T-B image rev., Digital harmonics, TGC, Raw data access, LOGIQ View, CrossXBeam, SRI-HD, B-Flow, TVI, Anatomical-M mode (AMM), Auto Optimization. | ✓                                       | ✓                                               |
| Tested to meet Electrical Safety, EMC and Biocompatibility Standards                                                                                                                                                                                                                                 | ✓                                       | ✓                                               |
| Track 3 (within FDA limits)                                                                                                                                                                                                                                                                          | ✓                                       | ✓                                               |

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a blue circle. There are decorative swirls around the outer edge of the circle, giving it a classic and recognizable appearance.

{13}------------------------------------------------

## Summary of Non-Clinical Tests:

LOGIQ V5/LOGIQ V3 has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic, and mechanical safety, and has been found to conform with applicable medical device safety standards. The LOGIO V5/LOGIQ V3 complies with voluntary standards:

- 1. AAMI/ANSI ES60601-1, Medical Electrical Equipment – Part 1: General Requirements for Safety
- 2. IEC60601-1-2, Medical Electrical Equipment -Part 1-2:General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility Requirements and Tests
- IEC60601-2-37, Medical Electrical Equipment -3. Part 2-37:Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment
- 4. NEMA UD 3. Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
- 5. ISO10993-1, Biological Evaluation of Medical Devices- Part 1: Evaluation and Testing- Third Edition
- 6. NEMA UD 2, Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment
- 7. ISO14971, Application of risk management to medical devices
- 8. NEMA, Digital Imaging and Communications in Medicine (DICOM) Set. (Radiology)

The following quality assurance measures are applied to the development of the system:

- Risk Analysis
- Requirements Reviews ●
- . Design Reviews
- Testing on unit level (Module verification)
- . Integration testing (System verification)
- Performance testing (Verification) ●
- . Safety testing (Verification)

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in white inside. There are some white swirls around the letters. The logo is simple and recognizable.

# GE Healthcare

510(k) Premarket Notification Submission

- Simulated use testing (Validation) ●
Transducer material and other patient contact materials such as needle guidance kits are biocompatible.

Summary of Clinical Tests:

The subject of this premarket submission, LOGIQ V5/LOGIQ V5, did not require clinical studies to support substantial equivalence.

- Conclusion: GE Healthcare considers the LOGIQ V5/LOGIQ V3 to be as safe, as effective, and performance is substantially equivalent to the predicate device(s).

---

**Source:** [https://fda.innolitics.com/device/K141768](https://fda.innolitics.com/device/K141768)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
