AS20 COMPOSITE GRAFT

K141746 · Wrightmedicaltechnologyinc · MQV · Aug 4, 2014 · Orthopedic

Device Facts

Record IDK141746
Device NameAS20 COMPOSITE GRAFT
ApplicantWrightmedicaltechnologyinc
Product CodeMQV · Orthopedic
Decision DateAug 4, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3045
Device ClassClass 2
AttributesTherapeutic, Pediatric

Intended Use

AS20 resultant paste is intended for use as a bone graft substitute to be injected or digitally packed into open bone voids/gaps that are not intrinsic to the stability of bony structure of the skeletal system (i.e., the extremities and pelvis) to cure in situ. These open bone voids may be the result of benign bone cysts and tumors (in adults and pediatric patients > 6 years old), surgically created osseous defects or osseous defects created from traumatic injury to the bone. The paste provides a bone graft substitute that resorbs and is replaced with bone during the healing process. The AS20 paste cured in situ provides an open void/gap filler that can augment provisional hardware (e.g. K Wires) to help support bone fragments during the surgical procedure. The cured paste acts only as a temporary support media and is not intended to provide structural support during the healing process. AS20 is provided sterile for single use only. The AS20 Core Decompression Procedure Kit, consisting of a bone void filler and manual surgical instruments, is intended to be used during core decompression procedures. The bone void filler component resorbs and is replaced with bone during the healing process. The bone void filler included in the AS20 Core Decompression Procedure Kit is not intended to be used as a load-bearing device.

Device Story

Bone graft substitute; injectable/packable paste; cures in-situ. Components: calcium sulfate (primary osteoconductive filler; dissolves for vascular infiltration) and calcium phosphate (secondary porous scaffold; osteoclastic resorption). Supplied in syringe mix package (dual syringes, powder, liquid vial). Used by surgeons in clinical settings for bone void filling or core decompression procedures. Augments provisional hardware (e.g., K-wires) for temporary support of bone fragments; not for structural/load-bearing support. Resorbs and is replaced by bone during healing. Benefits: biocompatible scaffold; facilitates bone regeneration.

Clinical Evidence

No clinical data provided. Substantial equivalence supported by non-clinical bench testing including dissolution, porosity, Vicat set time, Gillmore set time, and diametral tensile strength.

Technological Characteristics

Composite bone graft consisting of calcium sulfate and calcium phosphate. Supplied as powder and liquid for syringe-based mixing. Cures in-situ via hydration reactions. Sterile, single-use. No electronic components or software.

Indications for Use

Indicated for adults and pediatric patients > 6 years old with open bone voids/gaps in extremities and pelvis not intrinsic to skeletal stability; voids resulting from benign bone cysts/tumors, surgical osseous defects, or traumatic injury. Not for load-bearing use.

Regulatory Classification

Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 4, 2014 Wright Medical Technology, Incorporated % Leslie Fitch, Ph.D. Senior Regulatory Affairs Specialist 1023 Cherry Road Memphis, Tennessee 38117 Re: K141746 Trade/Device Name: AS20 Composite Bone Graft Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV Dated: June 27, 2014 Received: June 30, 2014 Dear Dr. Fitch: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Lori A. Wiggins for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. 510(k) Number (if known) K141746 Device Name AS20 Composite Bone Graft #### Indications for Use (Describe) AS20 resultant paste is intended for use as a bone graft substitute to be injected or digitally packed into open bone voids/ gaps that are not intrinsic to the stability of bony structure of the skeletal system (i.e., the extremities and pelvis) to cure in situ. These open bone voids may be the result of benign bone cysts and tumors (in adults and pediatric patients > 6 years old), surgically created osseous defects or osseous defects created from traumatic injury to the bone. The paste provides a bone graft substitute that resorbs and is replaced with bone during the healing process. The AS20 paste cured in situ provides an open void/gap filler that can augment provisional hardware (e.g. K Wires) to help support bone fragments during the surgical procedure. The cured paste acts only as a temporary support media and is not intended to provide structural support during the healing process. AS20 is provided sterile for single use only. The AS20 Core Decompression Procedure Kit, consisting of a bone void filler and manual surgical instruments, is intended to be used during core decompression procedures. The bone void filler component resorbs and is replaced with bone during the healing process. The bone void filler included in the AS20 Core Decompression Procedure Kit is not intended to be used as a load-bearing device. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) - Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. | FOR FDA USE ONLY | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)<br>Laurence D. Coyne -A<br>(Division Sign-Off)<br>Division of Orthopedic Devices<br>510(k) Number: K141746 | This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Headquarters Wright Medical Technology, Inc 023 Cherry Road Memphis, TN 38117 1 867 9971 mt.com Image /page/3/Picture/3 description: The image shows the Wright logo. The logo consists of two overlapping trapezoids, one red and one orange, to the left of the word "WRIGHT" in red, sans-serif font. Below the word "WRIGHT" is the phrase "FOCUSED EXCELLENCE" in a smaller, sans-serif font. #### 510(K) SUMMARY In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the use of the AS20 Composite Bone Graft. | (a)(1). Submitted By: | Wright Medical Technology, Inc.<br>1023 Cherry Road<br>Memphis, TN 38117 | |-------------------------------------------|------------------------------------------------------------------------------------------------------------| | Date: | June 27, 2014 | | Contact Person: | Leslie Fitch, PhD<br>Senior Regulatory Affairs Specialist<br>Office: (901) 867-4120<br>Fax: (901) 867-4190 | | (a)(2). Proprietary Name:<br>Common Name: | AS20 Composite Bone Graft<br>Bone Void Filler | | Classification Name and Reference: | 21 CFR 888.3045 - Class II | | Device Product Code, Device Panel: | MQV, Orthopedic | | (a)(3). Predicate Devices: | K132656: PRODENSE® Bone Graft<br>Substitute | ### (a)(4). Device Description AS20 is a bone graft substitute to be injected and/or digitally packed into open bone voids/gaps that are not intrinsic to the structural stability of the skeletal system and cure in-situ. It is supplied in a syringe mix package (dual syringes of powder and liguid vial) for mixing it into a paste and delivering it to the defect site. When the components are mixed according to directions, an injectable paste forms and subsequently hardens via hydration reactions. The benefits of this composite include: - Calcium Sulfate . - Primary osteoconductive filler l - Resorbs first primarily through simple dissolution to allow early vascular infiltration ■ - 트 Excellent clinical history - Calcium Phosphate ● - Osteoclastic resorption 트 - Secondary porous scaffold that is resorbed after primary filler - I TCP granules are resorbed in the third and final phase {4}------------------------------------------------ ### (a)(5). Intended Use AS20 resultant paste is intended for use as a bone graft substitute to be injected or digitally packed into open bone voids/gaps that are not intrinsic to the stability of bony structure of the skeletal system (i.e., the extremities and pelvis) to cure in situ. These open bone voids may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), surgically created osseous defects or osseous defects created from traumatic injury to the bone. The paste provides a bone graft substitute that resorbs and is replaced with bone during the healing process. AS20 paste cured in situ provides an open void/gap filler that can augment provisional hardware (e.g. K Wires) to help support bone fragments during the surgical procedure. The cured paste acts only as a temporary support media and is not intended to provide structural support during the healing process. AS20 is provided sterile for single use only. ### AS20 Core Decompression Procedure Kit The AS20 Core Decompression Procedure Kit, consisting of a bone void filler and manual surgical instruments, is intended to be used during core decompression procedures. The bone void filler component resorbs and is replaced with bone during the healing process. The bone void filler included in the AS20 Core Decompression Procedure Kit is not intended to be used as a load-bearing device. ### (a)(6). Technological Characteristics Comparison The subject shares many technical characteristics with the predicate PRODENSE® (K132656). The implant material is identical. The only differences are the addition of a smaller volume kit and the addition of syringe mixing accessories for the smaller volume kits (2-10 cc). ### (b)(1). Substantial Equivalence - Non-Clinical Evidence Performance testing shows that the subject device is substantially equivalent to the predicate. Testing of the subject implant mixed with the subject or predicate mixing system included dissolution, porosity, Vicat set time, Gillmore set time, and diametral tensile strength. #### (b)(2). Substantial Equivalence - Clinical Evidence N/A {5}------------------------------------------------ ## (b)(3). Substantial Equivalence – Conclusions The design characteristics of the subject system do not raise any new types of questions of safety or effectiveness. From the evidence submitted in this 510(k), the subject device can be expected to perform at least as well as the predicate device. E2-3
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%