The PONTIS™ Nonabsorbable Ultra-High Molecular Weight Polyethylene Braided Sutures are indicated to secure soft tissue to soft tissue reattachment in the hand: Collateral Ligaments around the PIP, DIP and MCP Joints Flexor and Extensor Tendons The PONTIS™ 3mm Nonabsorbable Suture Anchors with Ultra-High Molecular Weight Polyethylene Braided Suture are indicated to secure soft tissue to bone reattachment in the hand: Collateral Ligaments around the PIP, DIP and MCP Joints Flexor and Extensor Tendons
Device Story
Device consists of 316L stainless steel 3mm suture anchors preloaded with UHMWPe braided sutures; also includes standalone UHMWPe sutures. Used by surgeons to reapproximate and secure soft tissue to bone or soft tissue to soft tissue in the hand. Anchor delivered into bone via single-use driver and handpiece; sutures secured via manual knot tying. Provides mechanical fixation for ligament and tendon repairs. Benefits include secure tissue reattachment to facilitate healing. Device is sterile, single-use, and intended for clinical/surgical settings.
Clinical Evidence
Bench testing only. Performance evaluated in standardized foam bone models comparing anchor insertion torque, suture knot pull strength, pull-out failure force, and cyclic pull-out performance. Biocompatibility data from predicate K101126 remains applicable.
Technological Characteristics
316L stainless steel anchor; UHMWPe braided suture. Single-use, implantable. Sterilized by ethylene oxide. Mechanical fixation via knot-tying. No software or electronic components.
Indications for Use
Indicated for soft tissue-to-soft tissue or soft tissue-to-bone reattachment in the hand, specifically for collateral ligaments (PIP, DIP, MCP joints) and flexor/extensor tendons.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
PONTIS Sutures and Suture Anchors with Optional Crimps (K101126)
K020159 — SMITH & NEPHEW TWINFIX TI QUICK T · Smith & Nephew, Inc. · Mar 26, 2002
K023021 — MODIFICATION TO SMITH & NEPHEW TWINFIX TI QUICK T · Smith & Nephew, Inc. · Oct 4, 2002
K141011 — G-FORCE GEN 2 SUTURE ANCHOR SYSTEM · Wrightmedicaltechnologyinc · Jul 30, 2014
K993261 — TALON ANCHOR SNAP-PAK · Mitek Products · Dec 2, 1999
K243203 — OC JuggerKnotless Soft Anchor; OC JuggerLoop Soft Anchor · Riverpoint Medical · Oct 29, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
11
Image /page/0/Picture/2 description: The image shows the logo for Pontis Orthopaedics. The word "PONTIS" is written in large, bold letters, with the word "ORTHOPAEDICS" written in smaller letters below it. There is a curved line above the "S" in "PONTIS".
. ---- .
1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 -
| Owner's Name | Leonard Gordon |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Address | 2299 Post Street Suite 107<br>San Francisco, CA 94115 |
| Phone Number | (415) 567-8935 |
| Fax Number | (415) 567-8934 |
| 510(k) Owner | PONTIS Orthopaedics, LLC |
| Contact Person | Leonard Gordon<br>President |
| Date Prepared | June 24, 2014 |
| Trade Name | PONTISTM Ultra-High Molecular Weight Polyethylene Suture and<br>PONTISTM 3mm Nonabsorbable Suture Anchors with Ultra-High<br>Molecular Weight Polyethylene Braided Suture |
| Common name | 1. Suture, nonabsorbable, synthetic, polyethylene<br>2. Suture, Anchor |
| Classification Name | Class II 21 CFR § 888.3040 Smooth or threaded metallic bone<br>fixation fastener<br>Class II 21 CFR § 878.5000 Nonabsorbable poly(ethylene<br>terephthalate) surgical suture |
| Section | §C.F.R. Section 878.4495 |
| Product Code | PONTISTM Ultra-High Molecular Weight Polyethylene Suture<br>• GAT (suture, nonabsorbable, synthetic, polyethylene)<br>PONTISTM 3mm Nonabsorbable Suture Anchors with Ultra-<br>High Molecular Weight Polyethylene Braided Suture<br>• MBI (fastener, fixation, nondegradable, soft tissue)<br>• GAT (suture, nonabsorbable, synthetic, polyethylene) |
| Predicate Device | PONTISTM Sutures and Suture Anchors with Optional Crimps<br>K101126 |
| Device Description | Brief Description of the Device, Sutures<br>The PONTISTM 3mm anchor is made of 316L stainless steel and<br>incorporates non-absorbable UHMWPe sutures which are |
| | available in United States Pharmacopoeia (USP) sizes 4-0 to 2-0 in various lengths and cleared to market under K094028. UHMWPe sutures and 316L Stainless steel are used in a wide variety of medical devices including previously approved implants of this type. The UHMWPe Sutures will be secured by knot tying. The UHMWPe sutures are supplied sterile and armed with cutting needles. A single use driver and hand piece will hold the excess suture, which delivers the preloaded anchor into the bone. The suture strands are used to reapproximate and secure the soft tissue to bone. The anchor is the same as was cleared under K101126, PONTIS Sutures and Suture Anchor with Optional Crimps. |
| Reason for 510(k) | Modified Device (suture change only) |
| Indications for Use | The PONTIS™ Nonabsorbable Ultra-High Molecular Weight<br>Polyethylene Braided Suture are indicated to secure soft tissue to<br>soft tissue reattachment in the hand:<br>Collateral Ligaments around the PIP, DIP and MCP<br>Joints Flexor and Extensor Tendons The PONTIS™ 3mm Nonabsorbable Suture Anchors with Ultra-<br>High Molecular Weight Polyethylene Braided Suture are indicated<br>to secure soft tissue to bone reattachment in the hand:<br>Collateral Ligaments around the PIP, DIP and MCP<br>Joints Flexor and Extensor Tendons |
| Technological<br>Characteristics | The PONTIS 3mm Suture Anchor is the same anchor cleared<br>under K101126. All biocompatibility testing performed, which<br>was conducted under the predicate submissions K101126 remain<br>applicable for this submission as is contains the same materials.<br>The suture material has changed to Ultra High Molecular Weight<br>Polyethylene Suture. UHMWPe sutures, which was cleared under<br>K094028. This indicates that the components are safe for their<br>intended use.<br><br>The PONTIS 3mm Anchor with UHMWPe as well as the<br>predicate is a single use, implantable device, sterilized by ethylene<br>oxide. |
## 510(K) Summary
:
{1}------------------------------------------------
:
:
:
'
.
:
: :
:
:
:
·
.
:
.
.
.
·
{2}------------------------------------------------
## Substantial Equivalence The substantial equivalence of the PONTiSTM Nonabsorbable Ultra-High Molecular Weight Polyethylene Braided Suture and PONTISTM 3mm Nonabsorbable Suture Anchors with Ultra-High Molecular Weight Polyethylene Braided Suture is based on the equivalence in intended use, operational principals, and indications for use to the predicate PONTISTM Sutures and Suture Anchors with Optional Crimp (K101126). The only difference is the suture type, which is being changed to Ultra High Molecular Weight Polyethylene Suture (UHMWPe) instead of stainless steel. The sutures will be secured by knots instead of an alternate crimp. The UHMWPe sutures were cleared to market under K094028. PONTiSTM UHMWPe Sutures and PONTiSTM 3mm Suture Anchors with UHMWPe suture are substantially equivalent to other sutures currently marketed in conjunction with suture anchors and present no substantial differences in design, intended use and function to previously approved products. Nonclinical Tests A collection of tests was conducted to determine safety and Performed performance on the proposed device. Additionally, the subject device labeling is consistent both with FDA's guidance as well as current medical practice. The PONTiSTM 3mm Suture Anchors with UHMWPe was compared in standardized foam bone model materials. The Anchor insertion torque, suture strength (knot pull), pull-out failure mechanism (pull-out or suture break strength) and pullout failure force were documented. In addition cyclic pull-out performance was characterized. In all performance tests the PONTiSTM UHMWPe Sutures and 3mm Suture Anchors with UHMWPe Suture secured by knots are equivalent to products currently marketed for the same indications as confirmed by a comparison to the literature documenting the comparative characteristics of suture anchors. The performance studies and biocompatibility reviews indicate that the PONTiSTM UHMWPe sutures are safe for their intended use. Conclusions Drawn The testing demonstrated that the products are equivalent and did not affect safety and efficacy of the device or raise any new questions of safety or efficacy.
{3}------------------------------------------------
:
: : :
. . . . . . . . . . .
.
:
.
.
:
:
: :
. :
.
. .
: : :
| | Based on the indications for use, technological characteristics and<br>performance test results, the PONTiS™ UHMWPe Sutures and<br>3mm Suture Anchors with UHMWPe suture is substantially<br>equivalent to the predicate PONTiS™ Sutures and Suture Anchors<br>with Optional Crimp K101126 and present no substantial<br>differences in design, intended use and function. |
|--|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
:
# END OF 510(k) SUMMARY
:
:
-
:
and the comments of the comments of the comments of
: 上一篇:
.
: 上一篇:
:
:
. . . . . .
11 11 11 11
.
.
And Children Children Children
.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20903-0002
July 24, 2014
PONTIS Orthopaedics, LLC Dr. Leonard Gordon President 2299 Post Street, Suite 103 San Francisco, California 94115
Re: K141711
Trade/Device Name: PONTiS Nonabsorbable Ultra-High Molecular Weight Polyethylene Braided Sutures and PONTiS'100 3mm Nonabsorbable Suture Anchors with Ultra-High Molecular Weight Polyethylene Braided Suture Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI, GAT Dated: June 24, 2014 Received: June 25, 2014
Dear Dr. Gordon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{5}------------------------------------------------
Page 2 - Dr. Leonard Gordon
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### Statement of Indications for Use
510(k) Number (if known): K141711
Manufacturer: PONTiS Orthopaedics, LLC
Device Name: PONTiSTM Nonabsorbable Ultra-High Molecular Weight Polyethylene Braided Sutures and PONTIS™ 3mm Nonabsorbable Suture Anchors with Ultra-High Molecular Weight Polyethylene Braided Suture
Indications for Use:
The PONTiSTM Nonabsorbable Ultra-High Molecular Weight Polyethylene Braided Sutures are indicated to secure soft tissue to soft tissue reattachment in the hand:
- Collateral Ligaments around the PIP, DIP and MCP Joints .
- . Flexor and Extensor Tendons
The PONTiSTM 3mm Nonabsorbable Suture Anchors with Ultra-High Molecular Weight Polyethylene Braided Suture are indicated to secure soft tissue to bone reattachment in the hand:
- Collateral Ligaments around the PIP, DIP and MCP Joints .
- Flexor and Extensor Tendons ●
| Prescription Use<br>(21 CFR Part 801 Subpart D) | XX | AND/OR | Over the Counter Use<br>(21 CFR Part 801 Subpart C) |
|-------------------------------------------------|----|--------|-----------------------------------------------------|
|-------------------------------------------------|----|--------|-----------------------------------------------------|
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
# Elizabeth 詞&aank -S
Division of Orthopedic Devices
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.