K141649 · Reflow Medical · DQY · Nov 5, 2014 · Cardiovascular
Device Facts
Record ID
K141649
Device Name
SPEX SUPPORT CATHETER
Applicant
Reflow Medical
Product Code
DQY · Cardiovascular
Decision Date
Nov 5, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The speX Support Catheter is intended to be used in conjunction with steerable guidewires to access discreet regions of the peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices and provide a conduit for the delivery of saline solutions or diagnostic/therapeutic agents
Device Story
Over-the-wire, single-lumen support catheter; used to provide additional support to steerable guidewires during peripheral vascular access. Features proximal female luer hub, flexible shaft with varying stiffness, and radiopaque tapered distal tip. Distal end includes lubricious coating. Operated manually by physician during percutaneous interventional procedures. Device facilitates guidewire/device exchange and serves as conduit for saline or diagnostic/therapeutic agents. Benefits patient by enabling navigation through complex peripheral anatomy and assisting in crossing occlusions via controlled forward advancement of tapered tip.
Indicated for use in peripheral vasculature to access discrete regions, facilitate guidewire/interventional device placement and exchange, and deliver saline or diagnostic/therapeutic agents. Compatible with 0.035" guidewires.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, composed of three intertwined human profiles facing right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 5, 2014
ReFlow Medical Ms. Rebecca Pine VP. Regulatory/Clinical Affairs & Quality Assurance 1003 Calle Sombra San Clemente, CA 92673
Re: K141649 Trade/Device Name: speX Support Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: October 2, 2014 Received: October 2, 2014
Dear Ms. Pine,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{2}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman -S
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
### 510(k) Number (if known)
K141649
#### Device Name
speX Support Catheter, 50cm, 300006; speX Support Catheter, 90cm, 300007; speX Support Catheter, 135cm, 300008; speX Support Catheter. 150cm. 300009
### Indications for Use (Describe)
The speX Support Catheter is intended to be used in conjunction with steerable guidewires to access discreet regions of the peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices and provide a conduit for the delivery of saline solutions or diagnostic/therapeutic agents
Type of Use (Select one or both, as applicable)
스 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/2/Picture/15 description: The image shows the FDA logo. The logo is in a light gray color. The letters FDA are large and stylized.
# Bram D. Zuckerman -S 2014.11.05 06:17:25 -05'00'
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| APPLICANT: | ReFlow Medical |
|------------------------|---------------------------------------------------------------------------------------------------------------------|
| DATE PREPARED: | October 1, 2014 |
| CONTACT PERSON: | Rebecca K Pine<br>ReFlow Medical<br>1003 Calle Sombra<br>San Clemente, CA 92673<br>Phone: (760) 809.5178 |
| TRADE NAME: | speX Support Catheter |
| COMMON NAME: | Guide Catheter |
| CLASSIFICATION NAME: | Percutaneous Catheter |
| DEVICE CLASSIFICATION: | Class 2, per 21 CFR 870.1250 |
| PRODUCT CODE | DQY |
| PREDICATE DEVICES: | Volcano Valet Microcatheter (K112035)<br>Quickcross Support Catheter (K072750)<br>Minnie Support Catheter (K082337) |
## Substantially Equivalent To:
The speX Support Catheter is substantially equivalent in intended use, principal of operation and technological characteristics to the Volcano Valet Microcatheter (K112035), QuickCross Support Catheter (K072750), and the Minnie Support Catheter (K082337).
## Description of the Device Subject to Premarket Notification:
The speX Support Catheter is a device intended to provide additional support to a steerable guidewire when accessing discrete regions of the peripheral vasculature. The device is an over the wire single lumen catheter with a proximal female hub and a distal flexible part with a radiopaque tapered tip. The distal end of the catheter has a lubricious coating. The speX Support Catheter is 0.035" guidewire compatible and has working lengths of: 50cm, 90c, 135cm and 150cm.
## Indication for Use:
The speX Support Catheter is intended to be used in conjunction with steerable guidewires to access discreet regions of the peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices and
{5}------------------------------------------------
provide a conduit for the delivery of saline solutions or diagnostic/therapeutic agents.
# Technological Characteristics:
The speX Support Catheter has similar physical and technical characteristics to the predicate device. All devices are highly flexible with a stabilizing radiopaque distal tip. In the case of the speX Support Catheter and the Volcano Catheter, both tips are shapeable. All devices utilize a luer hub handle. All devices are fabricated from medical-grade polymers with varying stiffness shafts. All devices are provided sterile and are intended for single-use. A summary of the similarities and differences between the speX Support Catheter and the identified predicates are shown in the table below:
| | speX Support<br>Catheter | QuickCross<br>Support Catheter<br>(K072750) | Valet Mirocatheter<br>(K112035) | Minnie Support<br>Catheter<br>(K082337) |
|----------------------------|--------------------------------------------------------------------------|---------------------------------------------------------------------------------|--------------------------------------------------------------------------|--------------------------------------------------------------------------|
| Function | Support catheter | Support catheter | Support catheter | Support catheter |
| Principle of<br>Operation | Controlled penetration<br>to create a channel<br>across the occlusion | Controlled<br>penetration to create<br>a channel across the<br>occlusion | Controlled<br>penetration to create a<br>channel across the<br>occlusion | Controlled<br>penetration to create a<br>channel across the<br>occlusion |
| Anatomical<br>Site | Peripheral | Peripheral<br>Coronary | Peripheral<br>Coronary | Peripheral<br>Coronary |
| Delivery to<br>Site | Percutaneous access;<br>over the wire | Percutaneous access;<br>over the wire | Percutaneous access;<br>over the wire | Percutaneous access;<br>over the wire |
| Visualization | Radiographic imaging | Radiographic<br>imaging | Radiographic imaging | Radiographic imaging |
| Flexibility | Highly flexible | Highly flexible | Highly flexible | Highly flexible |
| Working<br>length | 50cm<br>90cm<br>135cm<br>150cm | 65cm-150cm | 100cm-150cm | 90cm-150cm |
| Crossing<br>Mechanism | Distal Tip;<br>Radiopaque;<br>tapered | Distal Tip;<br>Radiopaque marker<br>Tapered | Distal Tip;<br>Radiopaque<br>tapered | Distal Tip;<br>Radiopaque<br>tapered |
| Crossing<br>method | Forward (advancement)<br>penetration of tapered<br>tip | Forward<br>(advancement)<br>penetration of<br>tapered tip | Forward<br>(advancement)<br>penetration of tapered<br>tip | Forward<br>(advancement)<br>penetration of tapered<br>tip |
| Use interface | Luer as handle | Luer as handle | Luer as handle | Luer as handle |
| Luer<br>connector | Yes | Yes | Yes | Yes |
| Manually<br>operated | Yes | Yes | Yes | Yes |
| Tip Material | Polymer | Polymer | Polymer | Polymer |
| Distal Tip OD | .022" | 0.02" - 0.041" | 0.02" | .0205"-0.0415" |
| Shapeable tip | Yes | No, available in<br>straight or angled | Yes | No, available straight |
| Catheter<br>Material | Polymer;<br>varying stiffness shaft;<br>braid reinforced; PTFE<br>lining | Polymer:<br>varying stiffness<br>shaft: stainless steel<br>braid (some models); | Polymer; varying<br>stiffness shaft, metal<br>braid | Polymer; metal braid<br>incorporated into<br>shaft |
| Hydrophilic<br>coating | Yes, distal 40cm | Yes | Yes, distal 30cm | Yes, distal 40cm |
| | speX Support<br>Catheter | QuickCross<br>Support Catheter<br>(K072750) | Valet Mirocatheter<br>(K112035) | Minnie Support<br>Catheter<br>(K082337) |
| Catheter OD | 4.6F | 3.0-4.8F | 1.8-3.5F | 2.9-4.8F |
| Single use | Yes | Yes | Yes | Yes |
| Guidewire<br>compatibility | 0.035" | 0.014" -0.035" | 0.014" -0.035" | 0.014" -0.035" |
| Targeted<br>injection | Yes- saline solutions,<br>diagnostic/therapeutic<br>agents | Yes- saline solutions,<br>diagnostic agents | Yes- saline solutions,<br>contrast agents | Yes- saline solutions,<br>diagnostic/therapeutic<br>agents |
{6}------------------------------------------------
# Performance Data:
All necessary testing has been performed for the speX Support Catheter to assure substantial equivalence to the predicate device and demonstrate the device performs as intended. All testing was performed on test units representative of finished devices.
The device design was qualified through the following tests:
- Simulated Use (baseline, aged) ●
- Tensile Strength (baseline, aged) ●
- Torque (baseline, aged) ●
- Pressure Test (baseline, aged)
- Coating verification (baseline, aged)
- Corrosion Test (baseline, aged)
- Luer Verification (baseline, aged)
- Dimensional verification and visual inspections (baseline, aged) ●
- Sterilization Validation ●
- Biocompatibility ●
- Cytotoxicity .
- l Irritation
- Sensitization
- Materials Mediated Pyrogenicity
- I Acute Systemic Tox
- Hemolysis (direct, indirect)
- PTT
- Complement Activation
- I Thrombosis
The speX Support Catheter met all specified criteria and did not raise new safety or performance questions.
# Basis for Determination of Substantial Equivalence:
The data and information presented in this submission and the similarities and differences between the subject and predicate devices support a determination of substantial equivalence.. The speX Support Catheter is determined by ReFlow Medical, to be substantially equivalent to the identified predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.