← Product Code [KWQ](/productcode/KWQ) · K141632

# ZEVO (TM) ANTERIOR CERVICAL PLATE SYSTEM (K141632)

_Medtronic Sofamor Danek USA, Inc. · KWQ · Dec 4, 2014 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K141632

## Device Facts

- **Applicant:** Medtronic Sofamor Danek USA, Inc.
- **Product Code:** [KWQ](/productcode/KWQ.md)
- **Decision Date:** Dec 4, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The ZEVO™ Anterior Cervical Plate System is interbody screw fixation from C2 to T1. The system is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusions in patients with: 1) degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), 2) trauma (including fractures), 3) tumors, 4) defined as kyphosis, lordosis, or scoliosis, 5) pseudoathrosis, and/or 6) failed previous fusions.

## Device Story

ZEVO™ Anterior Cervical Plate System provides temporary anterior spinal stabilization during cervical fusion; consists of titanium alloy plates and bone screws with nitinol subcomponents; implanted via anterior approach into cervical vertebral bodies (C2-T1); used by surgeons in clinical settings; provides mechanical fixation to support bone healing; benefits patients by stabilizing spine during fusion development.

## Clinical Evidence

Bench testing only. Testing performed per ASTM F1717-13 (spinal implant constructs in vertebrectomy model) and ASTM F2129-08 (corrosion susceptibility). Device met all pre-determined acceptance criteria.

## Technological Characteristics

Materials: Titanium alloy and nitinol (NiTi). Components: Plates and bone screws. Standards: ASTM F1717-13, ASTM F2129-08. Fixation: Anterior screw fixation into vertebral body. Form factor: Various sizes. Sterilization: Supplied sterile and non-sterile.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- Anterior Cervical Spine Locking Plate (CSLP) System ([K030866](/device/K030866.md))
- ATLANTIS® Anterior Cervical Plate System ([K130640](/device/K130640.md), [K081038](/device/K081038.md))
- ZEPHIR® Anterior Cervical System ([K030327](/device/K030327.md), [K994239](/device/K994239.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image contains the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized graphic. The graphic depicts three overlapping human profiles facing to the right, creating a sense of depth and connection. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 4, 2014

Medtronic Sofamor Danek USA, Incorporated Ms. Kristi Frisch Senior Regulatory Affairs Specialist 1800 Pyramid Place Memphis, Tennessee 38132

Re: K141632

Trade/Device Name: ZEVO™ Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: KWQ Dated: October 16, 2014 Received: October 17, 2014

Dear Ms. Frisch:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Lori A. Wiggins -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

### K141632

Device Name ZEVOTM Anterior Cervical Plate System

## Indications for Use (Describe)

The ZEVO™ Anterior Cervical Plate System is interbody screw fixation from C2 to T1. The system is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusions in patients with: 1) degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), 2) trauma (including fractures), 3) tumors, 4) defined as kyphosis, lordosis, or scoliosis, 5) pseudoathrosis, and/or 6) failed previous fusions.

### Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

## FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## ZEVO™ Anterior Cervical Plate System 510(k) Summary December, 2014

| I.   | Company:                | Medtronic Sofamor Danek USA, Inc.<br>1800 Pyramid Place<br>Memphis, TN 38132<br>(901) 396-3133 |
|------|-------------------------|------------------------------------------------------------------------------------------------|
| II.  | Contact:                | Kristi Frisch<br>Senior Regulatory Affairs Specialist                                          |
| III. | Proprietary Trade Name: | ZEVO™ Anterior Cervical Plate System                                                           |
| IV.  | Common Name:            | Spinal Intervertebral Body Fixation<br>Appliance                                               |
| V.   | Classification Name:    | 21 CFR 888.3060 - Spinal Intervertebral<br>Body Fixation Orthosis                              |
|      | Classification:         | Class II                                                                                       |
|      | Product Codes:          | KWO                                                                                            |

### VI. Product Description:

The ZEVO™ Anterior Cervical Plate System consists of temporary implants (plates and bone screws) intended for anterior screw fixation of the cervical spine during the development of a cervical spinal fusion. Fixation is provided by bone screws inserted into the vertebral body of the cervical spine using an anterior approach. The device is intended for the anterior screw fixation from C2- T1 in the cervical spine.

{4}------------------------------------------------

The ZEVO™ Anterior Cervical Plate System implants are available in a broad range of size offerings, and are supplied in both sterile and non-sterile form. The implant components are made from titanium alloy, with plates having subcomponents manufacturing from nitinol (NiTi).

#### Indications for Use: VII.

The ZEVO™ Anterior Cervical Plate System is intended for anterior interbody screw fixation from C2 to T1. The system is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusions in patients with: 1) degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), 2) trauma (including fractures), 3) tumors, 4) deformity (defined as kyphosis, lordosis, or scoliosis, 5) pseudoathrosis, and/or 6) failed previous fusions.

### Summary of Technological Characteristics VIII.

The subject ZEVO™ Anterior Cervical Plate System components have the same fundamental scientific technology as the predicate Cervical Spine Locking Plate (CSLP), ATLANTIS® Anterior Cervical System and ZEPHIR® Anterior Cervical System devices. The subject devices are manufactured from the same titanium alloy and nitinol material as the predicate ATLANTIS® devices. The predicate and subject devices are also anterior screw fixation devices where fixation is provided by bone screws inserted into the vertebral body of the cervical spine.

### IX. Identification of Legally Marketed Predicate Devices Used to Claim Substantial Equivalence

The subject plate and bone screws are substantially equivalent to the predicates:

Primary Predicate:

- Anterior Cervical Spine Locking Plate (CSLP) System, (K030866, SE . 04/18/2003) [SYNTHES® Spine]

{5}------------------------------------------------

Additional predicates:

- ATLANTIS® Anterior Cervical Plate System (K130640, SE . 06/04/2013; K081038, SE 08/15/2008) [Medtronic]
- ZEPHIR® Anterior Cervical System (K030327, SE 02/26/2003; . K994239, SE 06/19/2000) [Medtronic]

To the best of Medtronic's knowledge, none of the predicates listed here have been subject to a design-related recall.

### Brief Discussion of the Non-Clinical Tests Submitted X.

The subject ZEVO™ Anterior Cervical Plate devices were tested in accordance to ASTM F1717-13 "Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model" and ASTM F2129-08 "Standard Test Method for Conducting Cyclic Potentiodynamic Polarization Measurements to Determine the Corrosion Susceptibility of Small Implant Devices." The subject device met the pre-determined acceptance criteria for all tests. The test results are provided to demonstrate that the subject devices are substantially equivalent to the predicate devices.

### Conclusions Drawn from the Non-Clinical Tests XI.

Based on the non-clinical test results and additional supporting documentation provided in this pre-market notification, the subject devices demonstrated substantial equivalence to the listed predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K141632](https://fda.innolitics.com/device/K141632)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
