K141617 · Rex Medical, L.P. · QEW · Nov 14, 2014 · Cardiovascular
Device Facts
Record ID
K141617
Device Name
CLEANER ROTATIONAL THROMBECTOMY SYSTEM
Applicant
Rex Medical, L.P.
Product Code
QEW · Cardiovascular
Decision Date
Nov 14, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cleaner™ rotational thrombectomy system is indicated for mechanical declotting of native vessel dialysis fistulae and synthetic dialysis access grafts. The Cleaner™ Rotational Thrombectomy System is indicated for mechanical declotting and controlled and selective infusion of physician-specified fluids, including thrombolytics, in the peripheral vasculature.
Device Story
Percutaneous infusion system; utilizes rotating sinusoidal wire at 4000 RPM to create fluid vortex; facilitates mechanical declotting and dispersion of physician-specified fluids (e.g., thrombolytics) in peripheral vasculature; contrast media/solutions infused through catheter lumen to distal side hole; wire rotation enhances solution penetration into clot; residual clot aspirated via introducer sheath; sterile, single-use disposable; used by physicians in clinical settings; provides mechanical declotting and targeted drug delivery to restore vessel patency.
Clinical Evidence
No human clinical data. Evidence based on bench testing (infusion performance, electrical safety, EMC) and a GLP animal study assessing device performance in restoring vessel patency in peripheral vasculature. Animal study confirmed successful restoration of patency in occluded vessels.
Technological Characteristics
Percutaneous catheter with rotating sinusoidal wire; 4000 RPM rotation; radiopaque a-traumatic distal tip; sterile, single-use disposable; infusion lumen with distal side hole; mechanical energy source for rotation.
Indications for Use
Indicated for mechanical declotting of native vessel dialysis fistulae and synthetic dialysis access grafts, and mechanical declotting and controlled/selective infusion of physician-specified fluids (including thrombolytics) in the peripheral vasculature.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a circular seal with an eagle emblem and text around it, representing the Department of Health & Human Services - USA. To the right, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" in a bold, sans-serif font, followed by "ADMINISTRATION" in a smaller font.
July 29, 2020
Rex Medical, L.P. % MDI Consultants, Inc. Susan Goldstein-Falk 55 Northern Boulbevard Great Neck, New York 11021
Re: K141617
Trade/Device Name: Cleaner Rotational Thrombectomy System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: QEW, KRA
Dear Susan Goldstein-Falk:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated November 14, 2014. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Gregory O'Connell, OHT2: Office of Cardiovascular Devices, (301) 796-6075, Gregory.Oconnell(@FDA.HHS.gov.
Sincerely,
Digitally signed by Gregory Gregory W. W. O'connell -S O'connell -S Date: 2020.07.29 19:20:18 -04'00
Gregory O'Connell Assistant Director Plaque Modification Devices Team DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features an emblem with three stylized human profiles facing right, overlaid and slightly offset from each other, representing the department's focus on people. The emblem is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 14, 2014
Rex Medical, L.P. % Susan Goldstein-Falk MDI Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, NY 11021
Re: K141617
> Trade/Device Name: Cleaner Rotational Thrombectomy System Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II Product Code: KRA Dated: October 6, 2014 Received: October 7, 2014
Dear Ms. Goldstein-Falk,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{2}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K141617
Device Name
Cleaner™ Rotational Thrombectomy System
| Indications for Use (Describe) | |
|--------------------------------|--|
|--------------------------------|--|
The Cleaner™ Rotational Thrombectomy System is indicated for mechanical declotting of native vessel dialysis fistulae and synthetic dialysis access grafts.
The Cleaner™ Rotational Thrombectomy System is indicated for mechanical declotting and controlled and selective infusion of physician-specified fluids, including thrombolytics, in the peripheral vascular.
### Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
The assigned 510(k) number is:_____K141617
| Submitter: | Rex Medical, L.P.<br>555 East North Lane, Suite 5035<br>Conshohocken, PA 19428 |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Colin Valentis<br>Development Engineer<br>Phone: (610) 940-0665 x106<br>Fax: (610) 940-1590<br>Email: cvalentis@rexmedical.com |
| Date Prepared: | September 19th, 2014 |
| Trade Name: | Cleaner™ Rotational Thrombectomy System |
| Common Name: | Thrombectomy catheter |
Classification Name: Continuous Flush Catheter (21 CFR 870.1210, Product Code KRA) Peripheral Atherectomy Catheter (21 CFR 870.4875, Product Code MCW)
# Predicate Device(s):
| K013635 | Trellis Peripheral Infusion System |
|---------|-----------------------------------------|
| K950907 | Pulse*Spray Infusion System |
| K091029 | Cleaner™ Rotational Thrombectomy System |
| K120346 | Cleaner™ Rotational Thrombectomy System |
# Device Description:
The Cleaner™ Rotational Thrombectomy System is a percutaneous infusion system that utilizes a rotating sinusoidal wire to increase the dispersion of delivered solution. The devices employ mechanical rotation of a flexible "S" shaped dispersion wire at 4000 RPM which creates a fluid vortex within the treatment site. Contrast media and physician specified solutions, including thrombolytics, may be infused through the catheter lumen to a side hole at the distal end. The dispersion wire uses mechanical rotation to allow the infused solution to penetrate the clot increasing the effectiveness of the treatment. Any residual clot can be aspirated through an introducer sheath prior to restoration of flow. The distal soft tip is a radiopaque a-traumatic tip on the distal end of the dispersion wire. The devices are sterile and single-use disposable.
{5}------------------------------------------------
# Intended Use:
The Cleaner™ rotational thrombectomy system is indicated for mechanical declotting of native vessel dialysis fistulae and synthetic dialysis access grafts.
The Cleaner™ Rotational Thrombectomy System is indicated for mechanical declotting and controlled and selective infusion of physician-specified fluids, including thrombolytics, in the peripheral vasculature.
# Technological Characteristics:
The Cleaner™ Rotational Thrombectomy System is similar with regard to materials, intended use, principles of operation and technological characteristics to the predicate device. Any differences that may exist do not significantly affect the safety and efficacy of the device. Results of bench testing and animal studies demonstrate Cleaner™ Rotational Thrombectomy System is as safe and effective as the legally marketed predicate device.
# Non-Clinical Performance Testing:
The expanded indications for use to include controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature. are based on the bench testing and animal data presented in the performance section of the submission. A GLP animal study was conducted utilizing the Cleaner Rotational Thrombectomy System to assess device performance at restoring vessel patency in the peripheral vasculature. The animal testing show the Cleaner Rotational Thrombectomy System successfully restores patency to occluded vessels. Bench testing included: infusion testing of the predicate device to the proposed device, Electrical Safety, and EMC testing.
# Conclusions:
Rex Medical considers the Cleaner™ Rotational Thrombectomy System to be substantially equivalent to the predicate devices listed above. The conclusions are based on performance testing and similarities in indications for use, materials, technological characteristics, principle of operation and design features.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.