← Product Code [OLO](/productcode/OLO) · K141601

# IASSIST KNEE SYSTEM (K141601)

_Zimmer Cas · OLO · Sep 11, 2014 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K141601

## Device Facts

- **Applicant:** Zimmer Cas
- **Product Code:** [OLO](/productcode/OLO.md)
- **Decision Date:** Sep 11, 2014
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The iASSIST Knee System is a computer assisted stereotaxic surgical instrument system to assist the surgeon in the positioning of orthopedic implant system components intra-operatively. It involves surgical instruments and position sensors to determine alignment axes in relation to anatomical landmarks and to precisely position alignments and implant components relative to these axes. Example orthopedic surgical procedures include but are not limited to: Total Knee Arthroplasty.

## Device Story

Computer-assisted stereotaxic surgical system; assists surgeons during Total Knee Arthroplasty. Inputs: positional data from wireless tracking sensors ('pods') attached to surgical instruments and patient anatomy. Processing: system calculates alignment axes relative to anatomical landmarks; guides positioning of femoral and tibial cutting planes. Output: real-time positional feedback displayed to surgeon. Used in OR; operated by orthopedic surgeons. Benefits: precise implant component placement; improved surgical workflow. This iteration supports an alternate surgical flow using 2 pods (vs 4 in predicate); includes software updates for bone registration robustness, wireless communication, and ESD resistance.

## Clinical Evidence

Bench testing only. No clinical data. Performance verified via software system tests (functionality, communication, fault conditions), bench tests (bone registration, ESD resistance, battery life), and full-use simulations using sawbones to validate surgical flow and ergonomics. Electrical safety verified per IEC 60601-1:2005.

## Technological Characteristics

System comprises tracking sensors (pods), computer, software, and surgical instruments. Connectivity: wireless communication between pods and system. Electrical safety: IEC 60601-1:2005. Software: includes algorithms for bone registration and wireless communication. Sterilization: not specified.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- iASSIST Knee System ([K122326](/device/K122326.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 11, 2014

ZIMMER CAS Mr. Christopher Mclean Quality and regulatory Affairs Associate Director 75 QUEEN STREET, SUITE 3300 MONTREAL, Canada H3C 2N6

Re: K141601

Trade/Device Name: iASSIST™ Knee System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: Class II Product Code: OLO Dated: August 11, 2014 Received: August 12, 2014

Dear Mr. Mclean:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

K141601

Device Name iASSIST Knee System

Indications for Use (Describe)

The iASSIST Knee System is a computer assisted stereotaxic surgical instrument system to assist the surgeon in the positioning of orthopedic implant system components intra-operatively. It involves surgical instruments and position sensors to determine alignment axes in relation to anatomical landmarks and to precisely position alignments and implant components relative to these axes.

Example orthopedic surgical procedures include but are not limited to: Total Knee Arthroplasty.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

## FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS iASSIST™ KNEE SYSTEM

Applicant: Zimmer CAS 75 Queen Street, suite 3300 Montreal, Quebec Canada, H3C 2N6 Tel.: 514 861 4074 Fax: 514 866 2197

Contact Person: Christopher McLean

Date Summary Prepared: June 12, 2014

Device Trade Name: iASSIST™ Knee System

Device Classification Name: Orthopedic Stereotaxic Instrument (product code OLO): 21 CFR & 882.4560

#### Predicate Device:

i ASSIST Knee System, from Zimmer CAS, 510(k) # K122326, cleared Nov. 30th, 2012

#### Device Description:

As in the predicate, the iASSIST Knee System consists of tracking sensors ('pods'), a computer system, software, and surgical instruments designed to assist the surgeon in the placement of Total Knee Replacement components. The pods combined with the surgical instruments provide positional information to help orient and locate the main femoral and tibial cutting planes as required in knee replacement surgery. This includes means for the surgeon to determine and thereafter track each of the bones' alignment axes relative to which the cutting planes are set.

#### Indications for Use / Intended Use:

This is unchanged from the predicate:

The iASSIST Knee System is a computer assisted stereotaxic surgical instrument system to assist the surgeon in the positioning of orthopedic implant system components intra-operatively. It involves surgical instruments and position sensors to determine alignment axes in relation to anatomical landmarks and to precisely position alignments and implant components relative to these axes.

Example orthopedic surgical procedures include but are not limited to: Total Knee Arthroplasty.

#### Technological Comparisons to the Predicates:

The fundamental scientific technology of the predicate is unchanged including the operating principle and control mechanism.

The changes involved altering the system to allow for an alternate surgical flow requiring only 2 pods as compared to the predicate 4 pod method. This involved the introduction of a new 2 pod kit along with software adjustments to allow for the altered step sequence.

In addition, general system improvements and updates have been implemented in the predicate. These involved secondary engineering changes and improvements. They did not involve

{4}------------------------------------------------

changes to the intended and indication for use and did not raise any new issues of safety and effectiveness. These included the following:

- -Update of the pod electrical certification to latest IEC 60601-1:2005 electrical safety standard
- Removal of no longer used or redundant pod and instrument graphical symbols and mechanical interfaces
- The modification or addition instruments and methods to facilitate use and improve the compatibility with implant systems
- Improvements to the system software components to improve system usability, to increase the robustness of bone registration and wireless communication algorithms, to further increase the resistance to electro-static discharges, and to further ensure the pod's battery lifetime to accommodate the system changes.

#### Performance Data:

Non-clinical tests were performed to assess that no new safety and efficacy issues were raised in the device. The main tests included the following:

- 1. Software system tests were performed to ensure that the required functionalities were maintained or correctly updated per the changes without the introduction of hazardous anomalies. They included the hardware and user interfaces, the communication protocols. the use of pods and instruments as applicable, the response to temperature effects, and the verification of fault conditions.
- 2. Performance tests were performed under simulated bench test conditions and analyses to verify the implementation of the performance of the bone registration related functionalities.
- 3. Bench test and analyses were performed to verify the robustness and compatibility of the added or modified instruments, to verify the resistance of the pods to electro-static discharges, and to verify the sufficiency the pod's battery expected lifetime as in the predicate.
- 4. Full use simulations tests using sawbones were performed to verify and validate the overall system performance in terms of the system usage, surgical flow, and instrument ergonomics.
- 5. Electrical certification test related to the certification update.

#### Conclusion:

The information and data provided in this 510(k) Premarket Notification established that the modified iASSIST Knee System is substantially equivalent to its predicate version.

---

**Source:** [https://fda.innolitics.com/device/K141601](https://fda.innolitics.com/device/K141601)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
