← Product Code [MQV](/productcode/MQV) · K141600

# NANOSS BIOACTIVE, NANOSS BIOACTIVE LOADED, AND NANOSS BIOACTIVE LOADED KIT (K141600)

_Pioneer Surgical Technology, Inc. (Rti Surgical, I · MQV · Oct 17, 2014 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K141600

## Device Facts

- **Applicant:** Pioneer Surgical Technology, Inc. (Rti Surgical, I
- **Product Code:** [MQV](/productcode/MQV.md)
- **Decision Date:** Oct 17, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3045
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

NanOss™ is intended for bony voids or gaps that are not intrinsic to the stability of bony structure. The product is indicated to be gently packed into bony voids or gaps in the skeletal system (i.e., extremities, posterolateral spine, and pelvis). NanOss must be mixed with autogenous blood or sterile saline for use in the extremities or pelvis. NanOss must be mixed with bone marrow aspirate and autograft bone as a bone graft extender in the posterolateral spine. These defects may be surgically created osseous defects or defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.

## Device Story

NanOss is a resorbable bone void filler composed of calcium phosphate and porcine gelatin; provided in granular form. Device acts as a scaffold for bone growth; resorbs over time and is replaced by natural bone. Used by surgeons in clinical settings to fill skeletal voids. Product supplied in mixing containers or syringe-based chambers. Clinician mixes material with autogenous blood, saline, or bone marrow aspirate/autograft depending on anatomical site. Packed into defects to facilitate healing. Benefits patient by providing osteoconductive filler for bone regeneration.

## Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including packaging and shelf-life validation (ASTM D4169-09, ASTM F1980), sterilization validation (ISO 11137-1), and instrument validation testing.

## Technological Characteristics

Resorbable bone void filler; calcium phosphate and porcine gelatin carrier. Granular form; radiopaque. Provided sterile for single-use. Packaging/shelf-life per ASTM D4169-09 and ASTM F1980. Sterilization per ISO 11137-1. No software or electronic components.

## Regulatory Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

## Predicate Devices

- FortrOss Bone Void Filler ([K110561](/device/K110561.md))
- NanOss™ BVF-E ([K100361](/device/K100361.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 17, 2014

Pioneer Surgical Technology, Incorporated (dba RTI Surgical, Incorporated) Ms. Sarah McIntyre Regulatory Affairs Associate II 375 River Park Circle Marquette, Michigan 49855

Re: K141600

Trade/Device Name: NanOss™, NanOss™ Loaded, NanOss™ Loaded Kit Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MOV Dated: September 16, 2014 Received: September 18, 2014

Dear Ms. McIntyre:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Ronald和Nean -S for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K141600

Device Name

NanOssTM, NanOss™ Loaded, NanOss™ Loaded Kit

#### Indications for Use (Describe)

NanOss™ is intended for bony voids or gaps that are not intrinsic to the stability of bony structure. The product is indicated to be gently packed into bony voids or gaps in the skeletal system (i.e., extremities, posterolateral spine, and pelvis). NanOss must be mixed with autogenous blood or sterile saline for use in the extremities or pelvis. NanOss must be mixed with bone marrow aspirate and autograft bone as a bone graft extender in the posterolateral spine. These defects may be surgically created osseous defects or defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.

Type of Use (Select one or both, as applicable)

| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)         </span> </div> | <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)         </span> </div> |
|--------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## NanOss™ Bone Void Filler 510(k) Summary Pursuant to 21 CFR 807.92

| Sponsor:                       | Pioneer Surgical Technology, Inc.<br>(DBA RTI Surgical, Inc.)<br>375 River Park Circle<br>Marquette, MI 49855 USA<br>Contact: Sarah McIntyre<br>Ph: (906) 225-5861<br>Fx: (906) 226-4459                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                | Prepared: September 16, 2014                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Name:                          | NanOss™ Bone Void Filler                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Common Name:                   | Bone Void Filler                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Trade/Proprietary              | NanOss™, NanOss™ Loaded, NanOss™ Loaded Kit                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Classification<br>Regulations: | Filler, Bone Void, Calcium Compound (Product Code MQV)<br>21 CFR 888.3045, Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Classification Panel:          | Orthopaedic and Rehabilitation Devices                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Predicate:                     | FortrOss Bone Void Filler (K110561)<br>NanOss™ BVF-E (K100361)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Intended Use:                  | NanOss™ is intended for bony voids or gaps that are not intrinsic to the<br>stability of bony structure. The product is indicated to be gently packed into<br>bony voids or gaps in the skeletal system (i.e., extremities, posterolateral<br>spine, and pelvis). NanOss must be mixed with autogenous blood or sterile<br>saline for use in the extremities or pelvis. NanOss must be mixed with bone<br>marrow aspirate and autograft bone as a bone graft extender in the<br>posterolateral spine . These defects may be surgically created osseous<br>defects or defects created from traumatic injury to the bone. The product<br>provides a bone void filler that resorbs and is replaced with bone during the<br>healing process. |
| Description:                   | NanOss™ is a resorbable bone void filler (BVF) consisting of calcium<br>phosphate and a porcine gelatin carrier provided in granular form. The<br>BVF is radiopaque, provided sterile and is intended for single use only. The<br>product is provided pre-filled in a mixing container as NanOss™ or in a<br>mixing chamber/syringe as NanOss™ Loaded.                                                                                                                                                                                                                                                                                                                                                                                   |
|                                | The purpose of this submission is to obtain clearance for a merged<br>indications for use statement and new packaging configuration (NanOssTM<br>Loaded Kit). In addition, instrument accessories previously cleared for MIS<br>graft placement in the posterolateral spine through a small incision are now<br>available for graft placement in the skeletal system (i.e., extremities, pelvis).                                                                                                                                                                                                                                                                                                                                        |
| Testing:                       | The following non-clinical tests were performed to support a determination<br>of substantial equivalence:<br>Packaging and Shelf-Life Validations per ASTM D4169-09: DC 13<br>& 17 and ASTM F1980 Sterilization Validation per ISO 11137-1 Instrument Validation Testing                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Substantial<br>Equivalence:    | The bone void filler has the following similarities to the previously<br>cleared bone void filler product (K110561/K100361):  has the same intended uses, uses the same operating principles, incorporates the same basic design, incorporates the same materials, is provided sterile and for single-use.                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Conclusion:                    | The comparisons and testing conducted demonstrate the subject bone<br>void filler product described in this submission is substantially equivalent<br>to the predicate devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |

{4}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/device/K141600](https://fda.innolitics.com/device/K141600)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
