QUILL MONODERM KNOTLESS TISSUE-CLOSURE DEVICE

K141558 · Surgical Specialties Corp · GAM · Jul 11, 2014 · General, Plastic Surgery

Device Facts

Record IDK141558
Device NameQUILL MONODERM KNOTLESS TISSUE-CLOSURE DEVICE
ApplicantSurgical Specialties Corp
Product CodeGAM · General, Plastic Surgery
Decision DateJul 11, 2014
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4493
Device ClassClass 2
AttributesTherapeutic

Intended Use

Quill™ MONODERM™ device is indicated for use in soft tissue approximation where the use of absorbable sutures is appropriate.

Device Story

Quill™ Monoderm™ is a sterile, synthetic, absorbable tissue approximation device. It consists of a monofilament copolymer of glycolide and e-caprolactone (PGA-PCL). The suture features small, bi-directional barbs along its long axis, allowing for knotless tissue closure. It is available in various diameters (Sizes 0 to 5-0) and lengths, affixed to various needle types. The device is used by clinicians for soft tissue approximation. By eliminating the need for knots, the device simplifies the suturing process and provides secure tissue approximation, potentially reducing operative time and improving patient outcomes in surgical settings.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical laboratory performance testing, including USP monograph conformance for absorbable sutures, in vitro post-hydrolysis tensile testing, and in vivo resorption testing.

Technological Characteristics

Synthetic, absorbable monofilament suture composed of a copolymer of glycolide and e-caprolactone (PGA-PCL). Features bi-directional barbs along the long axis for knotless closure. Available in sizes 0 to 5-0. Conforms to USP monograph for absorbable sutures. Sterilized device.

Indications for Use

Indicated for soft tissue approximation in patients requiring absorbable sutures.

Regulatory Classification

Identification

An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K141558 Page 1/2 # 510(k) Summary . JUL 1 1 2014 | Date Prepared: | July 1, 2014 | JUL 1 1 2014 | |--------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------| | Company: | Surgical Specialties Corporation<br>100 Dennis Dr.<br>Reading, PA 19606 | | | Contact: | Kirsten Stowell Franco<br>Sr. Regulatory Affairs Manager<br>Phone: 610-404-3367<br>Fax: 610-404-3924<br>Email: kstowell@surgicalspecialties.com | | | Device trade name: | Quill™ Monoderm™ Knotless Tissue-Closure Device (PGA-PCL) | | | Device Common Name: | Absorbable Poly(glycolide/l-lactide) Surgical Suture | | | Device classification: | Suture, Absorbable, Synthetic, Polyglycolic Acid<br>Product code, GAM<br>21 CFR 878.4493<br>Class II | | | Legally marketed device<br>to which the device is<br>substantially equivalent: | K072028: | Quill™ Monoderm™ (PGA-PCL) | | | K052437: | Monoderm™ (PGA-PCL) Surgical Suture | | Description of the<br>device: | The Quill™ Monoderm™ Knotless Tissue-Closure Device is a<br>sterile, synthetic absorbable tissue approximation device that is<br>comprised of a copolymer of glycolide and e-caprolactone,<br>undyed, or dyed with D&C Violet No. 2. The device is designed<br>with small bi-directional barbs along the long axis of the suture<br>monofilament. It is available in diameter Sizes 0 to 5-0, in various<br>lengths affixed to various needle types. | | | Indications for Use: | Quill™ MONODERM™ device is indicated for use in soft tissue<br>approximation where the use of absorbable sutures is appropriate. | | {1}------------------------------------------------ ## K141558 Page 2/2 Substantial Equivalence: Performance tests: The proposed additional diameter sizes, Size 4-0 and 5-0, of the Quill™ Monoderm™ Knotless Tissue-Closure Device product line have the same material, design, intended use and technological characteristics as the predicate device. The only difference between the proposed and predicate device is the suture diameter. Non-clinical laboratory performance testing was conducted to confirm that the Quill™ Monoderm™ Knotless Tissue-Closure Device conforms to the USP monograph for absorbable sutures (as applicable). This testing was performed in accordance with FDA's Class II Special Controls Guidance Document: Surgical Sutures, Issued June 3, 2003. Additional performance testing was conducted in order to demonstrate substantial equivalence to the predicate device including in vitro post-hydrolysis tensile testing and in vivo resorption testing. The results of this testing demonstrates that the Quill™ Monoderm™ Knotless Tissue-Closure Device is substantially equivalent to the predicate device. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 11, 2014 Surgical Specialties Corporation % Angiotech Puerto. Inc. Kirsten Stowell Franco Sr. Regulatory Affairs Manager 100 Dennis Drive Reading, Pennsylvania 19606 Re: K141558 Trade/Device Name: Quill" Monoderm™ Knotless Tissue-Closure Device Regulation Number: 21 CFR 878.4493 Regulatory Name: Absorbable poly(glycolide/L-lactide) surgical suture Regulatory Class: Class II Product Code: GAM Dated: June 10, 2014 Received: June 12, 2014 Dear Ms. Franco: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {3}------------------------------------------------ Page 2 - Kirsten Stowell Franco device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, # David Krause -S - for Binita Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health cc: DMC 510(k) Staff Division D.O. {4}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug AdminIstration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page. ## 510(k) Number (if known) 441558 Device Name Quill™ MONODERM™ Knotless Tissue-Closture Device (PGA-PCL) Indications for Use (Describe) Quill™ MONODERM™ Knoiless Tissue-Closure Device is indicated for use in soft tissue approximation where the use of absorbable sutures is appropriate. Type of Use (Select one or both, as applicable) [ሬ] Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) # PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure) Peter L. Hudson -S FORM FDA 3881 (1/14) QuillTM Monoderm™ "* Device Special 510(k) {5}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%