MCC Para-Fix CARY BLAIR MEDIUM

K141506 · Medical Chemical Corporation · JSM · Feb 25, 2015 · Microbiology

Device Facts

Record IDK141506
Device NameMCC Para-Fix CARY BLAIR MEDIUM
ApplicantMedical Chemical Corporation
Product CodeJSM · Microbiology
Decision DateFeb 25, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.2390
Device ClassClass 1

Indications for Use

Para-FixTM Cary Blair Medium provides a method for collecting and preserving fecal specimens for the culture of intestinal enteric bacteria. Because the medium is capable of maintaining the bacteria for 96 hours, immediate transportation and processing of the specimen is not necessary.

Device Story

MCC Para-Fix Cary Blair Medium is a non-nutritive, buffered, isotonic transport medium designed for fecal specimen collection and preservation. The device consists of a 30 mL vial containing 15 mL of solution, a built-in sample collection spoon, and a phenol red pH indicator. The formulation includes agar and sodium thioglycolate to maintain low oxygen tension, supporting the viability of anaerobic and other enteric bacteria for up to 96 hours. Used in clinical settings, the device allows for delayed transport and processing of samples. Healthcare providers collect fecal specimens into the vial; the medium maintains bacterial counts within ±2 log10 of the original inoculum, facilitating accurate downstream culture and identification of pathogens. This preservation capability eliminates the requirement for immediate laboratory processing, potentially improving diagnostic workflow and patient access to testing.

Clinical Evidence

Bench testing only. Performance was evaluated using the plate count technique and Roll-Plate method to assess the maintenance of 10 representative enteric bacterial species (e.g., C. jejuni, C. difficile, S. enterica, E. coli) in both pure suspension and spiked human fecal matrix. Testing was conducted at 2-8°C and 20-25°C over 96 to 120 hours. Results demonstrated that viable organism counts remained within ±2 log10 of the original inoculum, meeting CLSI M40-A2 performance criteria. Bioburden testing confirmed no viable organisms at manufacture or expiration.

Technological Characteristics

Non-nutritive, buffered, isotonic solution containing potassium phosphate, disodium phosphate, phenol red, 0.15% (w/v) agar, sodium thioglycolate, calcium chloride, and distilled water. Form factor: 30 mL vial with integrated spoon. Single-use. Manufacturing includes 30-minute steam chamber treatment to reduce bioburden. Shelf life: 18 months.

Indications for Use

Indicated for the collection and preservation of fecal specimens for the culture of intestinal enteric bacteria in patients requiring diagnostic testing for enteric pathogens.

Regulatory Classification

Identification

A transport culture medium is a device that consists of a semisolid, usually non-nutrient, medium that maintains the viability of suspected pathogens contained in patient specimens while in transit from the specimen collection area to the laboratory. The device aids in the diagnosis of disease caused by pathogenic microorganisms and also provides epidemiological information on these diseases.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: K141506 B. Purpose for Submission: To make a substantial equivalence determination for the Medical Chemical Corporation (MCC) Para-Fix Cary Blair Medium transport system for the collection and transport of clinical specimens containing enteric bacteria from the clinic to the laboratory for bacteriological examination and culture. C. Measurand: Not applicable D. Type of Test: Collection and transport culture media system E. Applicant: Medical Chemical Corporation F. Proprietary and Established Names: MCC Para-Fix Cary Blair Medium G. Regulatory Information: 1. Regulation section: 21 CFR 866.2390 – Transport culture medium. 2. Classification: Class I (general controls) 3. Product code: JSM - Culture Media, non-propagating transport media LIO - Device, Specimen Collection {1} 4. Panel: Microbiology (83) H. Intended Use: 1. Intended use(s): Para-Fix Cary Blair Medium provides a method for collecting and preserving fecal specimens for the culture of intestinal enteric bacteria. Because the medium is capable of maintaining the bacteria for 96 hours, immediate transportation and processing of the specimen is not necessary. 2. Indication(s) for use: Para-Fix Cary Blair Medium provides a method for collecting and preserving fecal specimens for the culture of intestinal enteric bacteria. Because the medium is capable of maintaining the bacteria for 96 hours, immediate transportation and processing of the specimen is not necessary. 3. Special conditions for use statement(s): For in vitro diagnostic use only For prescription use only 4. Special instrument requirements: None I. Device Description: Para-Fix Cary Blair medium is supplied in a 30 mL vial that contains 15 ml of medium. The lid has a built in sample collection spoon. The Cary Blair medium is a non-nutritive, buffered, isotonic solution with a pH indicator added. The medium also contains agar and sodium thioglycolate to maintain a low oxygen tension for the preservation of anaerobic species. The phenol red indicator will turn yellow when the solution is acidic and the conditions are not optimal for recovery of the intended organisms. The transport vial is intended to transport common enteric parasites found in the stool of infected patients. The kit does not include a swab. {2} Para-Fix Cary Blair medium contains the following components: Potassium phosphate Disodium phosphate Phenol Red Agar Sodium Thioglycolate Calcium Chloride Distilled water J. Substantial Equivalence Information: 1. Predicate device name(s): Meridian Para-Pak (C&S) 2. Predicate 510(k) number(s): K792712 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | MCC Para-Fix Cary Blair K141506 | Meridian Para-Pak C&S K792712 | | Intended use | Para-Fix Cary Blair Medium provides a method for collecting and preserving fecal specimens for the culture of intestinal enteric bacteria. Because the medium is capable of maintaining the bacteria for 96 hours, immediate transportation and processing of the specimen is not necessary. | The Para-Pak (C&S) System provides a standardized procedure for the routine collection, transportation, preservation, and culture of stool specimens for bacterial enteric pathogens. It is designed for easy collection of specimens by individuals not trained in microbiological techniques and affords an excellent means of minimizing the adverse effects of delay in specimen transportation. | | Sample types | Fecal collection | Same | | Sample transport | Vial with built in spoon | Same | | Differences | | | | --- | --- | --- | | Item | MCC Para-Fix Cary Blair K141506 | Meridian Para-Pak C&S K792712 | | Medium composition | No glycerol in the medium, medium contains agar at 0.15% (w/v) | C&S medium contains glycerol and not agar | {3} | Differences | | | | --- | --- | --- | | Item | MCC Para-Fix Cary Blair K141506 | Meridian Para-Pak C&S K792712 | | Sample medium | Cary Blair | C&S | | Specimen stability | 2-8°C and 20-25°C, up to 96 hrs | Specimens must not be refrigerated, up to 72 hrs | ## K. Standard/Guidance Document Referenced (if applicable): Quality Control of Microbiological Transport Systems; Approved Standard, M40-A2 (2014), Clinical and Laboratory Standards Institute (CLSI), Wayne, PA. ## L. Test Principle: Not applicable. ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: The performance characteristics of Para-Fix Cary Blair transport vials were determined by spiking enteric organisms prepared either in Para-Fix Cary Blair transport medium or in clinically negative fecal matrix/Para-Fix Cary Blair transport medium. Swab Elution and Roll-Plate methods were used in accordance with the CLSI M40-A2 document to determine recovery of viable enteric organisms. Initially, ten (10) enteric organisms in Para-Fix Cary Blair transport medium were tested using only the Swab Elution method. Subsequently, as an additional challenge, further evaluation of the transport device was also performed using three (3) representative organisms, which were tested in fecal matrix/Para-Fix Cary Blair transport medium using both the Swab Elution and Roll-Plate methods. The 10 organisms listed below were added directly to Para-Fix Cary Blair Medium and tested by the Swab Elution Method: Enterococcus faecalis ATCC 29212 Escherichia coli ATCC 8739 Shigella dysenteriae ATCC 9361 Pseudomonas aeruginosa ATCC 9027 Salmonella enterica ATCC 10708 Vibrio parahaemolyticus ATCC 17802 Clostridium difficile ATCC 9689 Campylobacter jejuni ATCC 33291 Staphylococcus aureus ATCC 6538 Bacillus subtilis ATCC 6633 {4} The following 3 challenge organisms were spiked into clinically negative fecal matrix and Para-Fix Cary Blair Medium and tested by the Roll-Plate and Swab Elution Methods: Escherichia coli O157 ATCC 43894 Salmonella enterica ATCC 10708 Vibrio parahaemolyticus ATCC 17802 Acceptance criteria for recovery of bacteria as recommended in the CLSI document M40-A2 were applied. No over growth limit is defined by CLSI M40-A2 at room temperature (20-25°C) for the Swab Elution or Roll-Plate Methods because most commercial transport media cannot control for it. Manual colony counts were conducted for the Roll-Plate Method and the Swab Elution Method at the four time intervals (0 hr, 72 hr, 96 hr and 120 hr) for the representative organisms. The averages from 3 vials plated in duplicate, (n=6) were used to generate plate counts and assess recovery and viability of organisms. ## Swab Elution Method: The results of the Swab Elution Method studies are presented in Table 1 for all enteric organisms in Para-Fix Cary Blair transport medium and in Table 2 for the three (3) representative challenge enteric organisms in a mixture of Para-Fix Cary Blair transport medium and fecal matrix. The results demonstrate that the Para-Fix Cary Blair Medium is able to support and maintain acceptable viability of all organisms up to 96 hours when taken from pure culture and up to 120 hours when spiked into fecal matrix, at both room temperature (20-25°C) and under refrigerated (2-8°C) conditions. Acceptability is based on the acceptance criteria set forth in CLSI M40-A2. For viability in the Swab Elution Method to be considered acceptable, there shall be no more than a 3 log₁₀ (1 × 10³ +/- 10%) decline in CFU between the zero-time CFU count and the CFU counts for the predetermined end time point. This was demonstrated for all organisms tested. ## Roll-Plate Method for Challenge Organisms: The results of the Roll-Plate Method study are presented in Table 3 below for the three (3) representative challenge enteric organisms. The results demonstrate that the Para-Fix Cary Blair Medium is able to maintain acceptable viability for all organisms up to 120 hours at room temperature (20-25°C) and under refrigerated (2-8°C) conditions based on the acceptance criteria set forth in CLSI M40-A2. For the transport system to demonstrate acceptable viability by the Roll-Plate method, ≥5 CFU should be recovered following the specified maintenance time from the specific dilution that yielded time-zero plate counts closest to 300 CFU. 5 {5} Table 1. Enteric Organism Recovery Results for Para-Fix Cary Blair Medium Using Swab Elution Method. | Organism* | Hold Temperature | Average CFU's Recovered: Time 0 hrs | Average CFU's Recovered: Time 96 hrs | T=96 hrs Log reduction (-) or Log increase (+) | | --- | --- | --- | --- | --- | | Escherichia coli | 2-8°C | 2.2x107 | 2.7x107 | +0.09 | | | 20-25°C | 2.1x107 | 2.8x108 | +1.12 | | Staphylococcus aureus | 2-8°C | 1.6x107 | 1.3 x107 | -0.09 | | | 20-25°C | 1.5 x107 | 1.6 x107 | +0.03 | | Pseudomonas aeruginosa | 2-8°C | 7.6x106 | 8.2x106 | +0.03 | | | 20-25°C | 8.1x106 | 2.9x108 | +1.55 | | Salmonella enterica | 2-8°C | 4.2x107 | 4.5x107 | +0.03 | | | 20-25°C | 5.8x107 | 2.8x108 | +0.68 | | Bacillus subtilis | 2-8°C | 4.0x106 | 4.4x106 | +0.04 | | | 20-25°C | 4.7x106 | 1.5x107 | +0.50 | | Vibrio parahaemolyticus | 2-8°C | 9.9x105 | 9.1x105 | -0.04 | | | 20-25°C | 1.1x106 | 9.8x105 | -0.05 | | Clostridium difficile | 2-8°C | 1.4x106 | 1.2x106 | -0.07 | | | 20-25°C | 1.4x106 | 1.2x106 | -0.07 | | Campylobacter jejuni | 2-8°C | 4.1x106 | 3.7x106 | -0.04 | | | 20-25°C | 5.4x106 | 4.8x106 | -0.05 | | Enterococcus faecalis | 2-8°C | 2.1x107 | 1.9x107 | -0.04 | | | 20-25°C | 1.8x107 | 1.6x107 | -0.05 | | Shigella dysenteriae | 2-8°C | 2.4x107 | 2.2x107 | -0.04 | | | 20-25°C | 2.1x107 | 1.4x107 | -0.18 | *0.5 McFarland microorganism suspension diluted with Cary Blair medium at 1:15 {6} Table 2. Representative Challenge Enteric Organism Recovery Results for Para-Fix Cary Blair Medium Using the Swab Elution Method. | Organism* | Hold Temperature | Average CFU's Recovered: Time 0 hrs | Average CFU's Recovered: Time 72 hrs | Average CFU's Recovered: Time 96 hrs | Average CFU's Recovered: Time 120 hrs | T=120 hrs Log reduction (-) or Log increase (+) | | --- | --- | --- | --- | --- | --- | --- | | Salmonella enterica | 2-8°C | 2.4x102 | 2.5 x102 | 2.8 x102 | 2.1 x102 | -0.06 | | | 20-25°C | 8.0 x102 | ** | ** | ** | N/A | | Vibrio parahaemolyticus | 2-8°C | 2.9 x102 | 1.6x102 | 8.3x102 | 1.4x102 | -0.32 | | | 20-25°C | 2.1 x102 | ** | ** | ** | N/A | | Escherichia coli | 2-8°C | 2.9 x102 | 1.6x102 | 5.2x102 | 6.4x101 | -0.66 | | | 20-25°C | 1.2 x102 | 7.0 x102 | 5.2 x102 | 2.3 x102 | +0.28 | *0.5 McFarland microorganism suspension diluted with fecal matrix and Cary Blair medium at 1:2000 ** Too numerous to count Table 3. Representative Enteric Challenge Organism Recovery Results for Para-Fix Cary Blair Medium using the Roll-Plate Method. | Organism* | Hold Temperature | Average CFU's Recovered: Time 0 hrs | Average CFU's Recovered: Time 72 hrs | Average CFU's Recovered: Time 96 hrs | Average CFU's Recovered: Time 120 hrs | T=120 hrs Log reduction (-) or Log increase (+) | | --- | --- | --- | --- | --- | --- | --- | | Salmonella enterica | 2-8°C | 220 | 140 | 160 | 130 | -0.23 | | | 20-25°C4 | 39 | ** | ** | 250 | +0.81 | | Vibrio parahaemolyticus | 2-8°C | 52 | 160 | 190 | 100 | +0.28 | | | 20-25°C | 110 | ** | ** | ** | N/A | | Escherichia coli | 2-8°C | 79 | 48 | 55 | 64 | -0.09 | | | 20-25°C | 75 | ** | ** | 150 | +0.30 | * 0.5 McFarland microorganism suspension diluted with fecal matrix and Cary Blair medium to $2^{4}$ unless noted, diluted $2^{5}$ ** Too numerous to count {7} 8 # Overgrowth Evaluation: No over growth limit is defined by CLSI M40-A2 at room temperature (20-25°C) for the Swab Elution or Roll-Plate Methods because most commercial transport media cannot control for it. Para-Fix Cary Blair is no different; overgrowth at room temperature (20-25°C) was recorded as Too Numerous To Count (TNTC) for the studies conducted with fecal matrix. a. Precision/Reproducibility: Not applicable. b. Linearity/assay reportable range: Not applicable. c. Traceability, Stability, Expected values (controls, calibrators, or methods): **Shelf life**: The shelf life for Para-Fix Cary Blair vials is 18 months after the date of manufacture. The stability tests were performed on 3 lots of Para-Fix Cary Blair Medium up to 18 months of storage. **pH Stability**: The pH value of Para-Fix Cary Blair Medium up to 18 months of storage was 6.88 to 7.05. **Sterilization**: Para-Fix Cary Blair Medium vials are not sold as sterile, nor are they intended to be sterilized by the user. These vials are single use devices that do not require cleaning by the operator. Although there is no sterility claim, the manufacturing process includes treatment of finished vials in a steam chamber for 30 minutes to eliminate bioburden. d. Detection limit: Not applicable. e. Analytical specificity: Not applicable. f. Assay cut-off: Not applicable. 2. Comparison studies: a. Method comparison with predicate device: {8} Method comparison is not applicable for a bacterial transport medium. The device itself does not provide a result that can be used in making a clinical decision. Bench testing studies were done to determine the ability of Para-Fix Cary Blair Medium to maintain viability of different strains of intestinal enteric bacteria with and without fecal matrix. Para-Fix Cary Blair Medium showed recovery of representative bacteria from fecal matrix within the acceptance criteria recommended in CLSI-M40-A2 and similar to the predicate device. (See above under Performance Characteristics). b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Not applicable N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...