K141504 · Therozone USA, Inc. · EIA · Feb 24, 2015 · Dental
Device Facts
Record ID
K141504
Device Name
T-8000 THEROZONE UNIT
Applicant
Therozone USA, Inc.
Product Code
EIA · Dental
Decision Date
Feb 24, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6640
Device Class
Class 1
Attributes
Therapeutic
Intended Use
The T-8000 TherOzone Unit is intended for the reduction of microorganisms in dental unit water lines.
Device Story
T-8000 TherOzone Unit cleans dental unit water lines; utilizes ozone generator to create ozone from ambient air; ozonated gas applied to water lines; mechanism involves oxidation leading to cell lysis; operated by dental professionals in clinical settings; daily flush cycle; device consists of ozone generator and dispensing bottles; output is ozonated water for microbial reduction; benefits patient by reducing microbial contamination in dental water systems.
Clinical Evidence
Bench testing only. Performance testing included microbial challenge, software validation, environmental evaluation, functional verification, material compatibility analysis, usability evaluation, and pressure verification. No clinical data provided.
Technological Characteristics
Materials: Polyethylene, Fluoropolymers. Principle: Ozone generation from ambient air via electrical power (100-240VAC). Mechanism: Oxidation/cell lysis. Dimensions: 8"x 9.5" x 18.25". Weight: 12 lbs. Connectivity: Standalone. Sterilization: Non-sterile, reusable.
Indications for Use
Indicated for the reduction of microorganisms in dental unit water lines for use by dental professionals.
Regulatory Classification
Identification
A dental operative unit and accessories is an AC-powered device that is intended to supply power to and serve as a base for other dental devices, such as a dental handpiece, a dental operating light, an air or water syringe unit, and oral cavity evacuator, a suction operative unit, and other dental devices and accessories. The device may be attached to a dental chair.
K964271 — IGN 500 TM WATER CLEANING UNIT WITH CALBENIUM LIQUID · Airel-West · Dec 19, 1997
K983847 — MODIFICATION OF: ODYSSEY I DENTAL WATER UNIT · Germiphene Corp. · Dec 1, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, with three overlapping faces suggesting a sense of community and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 24, 2015
TherOzone USA. Inc. Ms. Rebecca K Pine Official Correspondent 2701 Ocean Park Blvd, Suite 108 Santa Monica, CA 90405
Re: K141504
Trade/Device Name: T-8000 TherOzone Unit Regulation Number: 21 CFR 872.6640 Regulation Name: Dental Operative Unit (accessory) Regulatory Class: II Product Code: EIA Dated: January 27, 2015 Received: January 28, 2015
#### Dear Ms. Pine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Ms. Pine
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141504
Device Name T-8000 TherOzone Unit
Indications for Use (Describe)
The T-8000 TherOzone Unit is intended for the reduction of microorganisms in dental unit water lines.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 6. 510(k) Summary
This 510(k) [K141504] summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| APPLICANT: | TherOzone USA, Inc. |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| DATE PREPARED: | February 24, 2015 |
| CONTACT PERSON: | Rebecca K Pine<br>Official Correspondent<br>TherOzone USA, Inc.<br>2701 Ocean Park Blvd, Suite 108<br>Santa Monica, CA 90405<br>Phone: 760.809.5178<br>FAX: 760.290.3216<br>Email: beky@cox.net |
| TRADE NAME: | T-8000 TherOzone Unit |
| COMMON NAME: | Accessory, Dental Unit |
| CLASSIFICATION<br>NAME: | Dental Operative Unit (accessory) |
| DEVICE<br>CLASSIFICATION: | Class II, per 21 CFR 872.6640 |
| PRODUCT CODE | EIA |
| PREDICATE DEVICES: | UltraKleen(K991946)<br>Odyssey Dental Water Unit (K964796) |
### Substantially Equivalent To:
The T-8000 TherOzone Unit is substantially equivalent in intended use, principal of operation and technological characteristics to the existing UltraKleen and the Odyssey Dental Water Unit devices.
### Description of the Device Subject to Premarket Notification:
The T-8000 TherOzone Unit is a device intended to clean dental unit water lines. The device consists of an ozone generator and dispensing bottles.
### Indication for Use:
The T-8000 TherOzone Unit is intended for the reduction of microorganisms in dental unit water lines.
{4}------------------------------------------------
### Technical Characteristics:
The T-8000 TherOzone Unit has similar physical and technical characteristics to the predicate devices, as illustrated in the table below.
| | T-8000<br>TherOzone Unit | Odyssey I Dental<br>Water Unit<br>Germiphene Corp.<br>(K964796) | UltraKleen,<br>Sterilex (K991946) | Substantial<br>Equivalence |
|--------------------------------------|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | For the reduction of<br>microorganisms in<br>dental unit water<br>lines. | Indicated to be used<br>as an in-line water<br>disinfecting system<br>to reduce<br>microorganisms in<br>the dental water<br>lines | The product has<br>been specially<br>formulated and<br>clinically proven to<br>clean deposits and<br>control bacterial<br>contamination in<br>Dental Unit Water<br>Lines. | Although minor<br>grammatical<br>difference exist, the<br>intended use of the<br>subject device is<br>substantially<br>equivalent to the<br>predicate devices |
| Function | Dental unit water<br>line cleaner | Dental unit water<br>line cleaner | Dental unit water<br>line cleaner | Same. The subject<br>device is<br>substantially<br>equivalent to the<br>predicate devices. |
| Principle of<br>Operation | Creates ozone from<br>air. Ozonated gas<br>(in water) applied<br>to dental unit water<br>lines | Creates ozone from<br>air. Ozonated gas<br>(in water) applied<br>to dental unit water<br>lines | Anti-microbial<br>chemical agent | Same. The subject<br>device is<br>substantially<br>equivalent to the<br>predicate device. |
| Mechanism of<br>Action | Oxidation, leading<br>to cell lysis | Oxidation, leading<br>to cell lysis | Oxidation<br>Hydrolysis<br>Microbubbling | Same. The subject<br>device is<br>substantially<br>equivalent to the<br>predicate device. |
| Patient contact | None | In-line use | None | Same. The subject<br>device is<br>substantially<br>equivalent to the<br>predicate device. |
| Intended User | Dental professional | Dental professional | Dental professional | Same. The subject<br>device is<br>substantially<br>equivalent to the<br>predicate devices. |
| Delivery to Site | Direct application<br>to dental water<br>lines | Direct application<br>to dental water<br>lines | Direct application<br>to dental water<br>lines | Same. The subject<br>device is<br>substantially<br>equivalent to the<br>predicate devices. |
| Microorganis<br>m reduction<br>cycle | Daily flush through<br>water lines | Continuous flush | Flush with<br>overnight soak | Similar. The<br>subject device cycle<br>varies slightly, but<br>all device cycles |
| | T-8000<br>TherOzone Unit | Odyssey I Dental<br>Water Unit<br>Germiphene Corp.<br>(K964796) | UltraKleen,<br>Sterilex (K991946) | Substantial<br>Equivalence |
| Materials<br>(wetted) | Polyethylene<br>Fluoropolymers | Polycarbonate,<br>304V stainless<br>steel, Kynar,<br>polyethylene | Sodium carbonate<br>Sodium<br>percarbonate<br>Benzenemethanaminium, N, N-<br>dimethyl-N-<br>tetradecyl-.<br>Chloride<br>Tetrasodium EDTA | effective microorganism<br>reduction. The<br>minor differences<br>do not pose a<br>functional<br>difference,<br>therefore the<br>subject device is<br>substantially<br>equivalent to the<br>predicate devices.<br>Similar, subject<br>device and the<br>predicate device are<br>both fabricated<br>from common<br>medical device<br>materials, therefore<br>the subject device<br>and predicate<br>device are<br>substantially<br>equivalent |
| Air Supply | Ambient,<br>compressed to 5-<br>25psi | 60 PSIG, min | N/A | Same. The subject<br>device and<br>predicate device<br>both use a<br>pressurized air<br>source. The<br>differences in<br>operating<br>parameters are<br>minor and do not<br>affect the<br>fundamental<br>technology<br>therefore the<br>subject device and<br>the predicate device<br>are substantially<br>equivalent. |
| Electrical | 100-240VAC,50-<br>60Hz,1.5 65 watts | 100-130 VAC, 50-<br>60 Hz, 20 watts,<br>grounded | N/A | Same. The subject<br>device and<br>predicate device<br>both use electrical<br>power. The<br>differences in<br>electrical<br>specifications are |
| | T-8000<br>TherOzone Unit | Odyssey I Dental<br>Water Unit<br>Germiphene Corp.<br>(K964796) | UltraKleen,<br>Sterilex (K991946) | Substantial<br>Equivalence |
| Water<br>capacity | ~ 600mL | 1 liter (1,000 mL) | N/A | minor and do not<br>affect the<br>fundamental<br>technology<br>therefore the<br>subject device and<br>the predicate device<br>are substantially<br>equivalent.<br>Same. The minor<br>differences in<br>volume do not<br>affect the<br>fundamental<br>technology<br>therefore the<br>subject device and<br>the predicate device<br>are substantially<br>equivalent. |
| Water<br>requirements | Distilled only | Distilled only | Distilled only | Same. The subject<br>device is<br>substantially<br>equivalent to the<br>predicate device. |
| Equipment<br>Weight (dry) | 12 lbs | 7 lbs | N/A | Same. The minor<br>differences in<br>equipment weight<br>do not affect the<br>fundamental<br>technology<br>therefore the<br>subject device and<br>the predicate device<br>are substantially<br>equivalent. |
| Dimensions | 8"x 9.5" X 18 1/4" | 6" x 6" x 12 ½ " | N/A | Same. The minor<br>differences in<br>equipment<br>dimension do not<br>affect the<br>fundamental<br>technology<br>therefore the<br>subject device and<br>the predicate device<br>are substantially<br>equivalent. |
| How provided | Non-sterile,<br>reusable | Non-sterile,<br>reusable | Non-sterile, single<br>use | Same. The subject<br>device is |
| T-8000<br>TherOzone Unit | Odyssey I Dental<br>Water Unit<br>Germiphene Corp.<br>(K964796) | UltraKleen,<br>Sterilex (K991946) | Substantial<br>Equivalence | |
| | | | substantially<br>equivalent to the<br>predicate device. | |
TherOzone USA, Inc. T-8000 TherOzone Unit
Page 12 of 56 Premarket Notification
{5}------------------------------------------------
TherOzone USA, Inc.
T-8000 TherOzone Unit
Page 13 of 56
Premarket Notification
{6}------------------------------------------------
TherOzone USA, Inc.
T-8000 TherOzone Unit
Page 14 of 56
Premarket Notification
{7}------------------------------------------------
## Performance Data:
All necessary verification and validation testing has been performed for the T-8000 TherOzone Unit to assure substantial equivalence to the predicate devices and demonstrate the device performs as intended. All testing was performed on test units representative of finished devices. Performance Testing included:
- Microbial challenge
- Software Validation ●
- Environmental Evaluation ●
- Functional Verification
- Material Compatibility Analysis
- Usability Evaluation
- Pressure Verification ●
- Environmental Condition Verification
The performance testing conducted demonstrates that the T-8000 TherOzone Unit is substantially equivalent to the predicate devices.
### Basis for Determination of Substantial Equivalence:
Based on the intended use, indications for use, technological characteristics, performance data and nonclinical tests performed, the subject T-8000 TherOzone Unit is substantially equivalent and is as safe and as effective as the legally marketed predicate devices, UltraKleen(K991946) and Odyssey Dental Water Unit (K964796).
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.