← Product Code MTJ · K141502

# HBP4 HARDENING, RESORBABLE HEMOSTATIC BONE PUTTY (K141502)

_Orthocon, Inc. · MTJ · Oct 3, 2014 · SU · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K141502

## Device Facts

- **Applicant:** Orthocon, Inc.
- **Product Code:** MTJ
- **Decision Date:** Oct 3, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Intended Use

HBP4 Hardening, Resorbable Hemostatic Bone Putty is indicated for the control of bleeding from cut or damaged bone by acting as a mechanical barrier or tamponade.

## Device Story

HBP4 is a sterile, resorbable, putty-like hemostatic material for bone surfaces. It consists of two components—granular calcium phosphate, calcium stearate, vitamin E acetate, triglyceride, polyalcohol, and lactide-diester/polyester-based polymers—that are manually mixed immediately prior to use. The mixture forms a hardening, radiopaque putty applied directly to bleeding bone by a surgeon. It functions via mechanical tamponade, occluding vascular openings in damaged bone. The material is resorbable, with components degrading via dissolution and hydrolysis. It is intended for single-use in surgical settings to achieve local hemostasis, providing a benefit by stopping bleeding from bone while being resorbed over time, unlike traditional non-resorbable bone waxes.

## Clinical Evidence

No human clinical data provided. Evidence consists of bench testing (stiffness, spreadability, stickiness, temperature sensitivity, electrocautery compatibility, dissolution, swelling), biocompatibility testing (ISO 10993: irritation, sensitization, toxicity, genotoxicity, implantation, hemolysis, pyrogenicity), and animal studies (in vivo hemostasis, resistance to irrigation, resorption time).

## Technological Characteristics

Two-part putty system; components: calcium phosphate, calcium stearate, vitamin E acetate, triglyceride, polyalcohol, lactide-diester/polyester-based polymers. Radiopaque. Resorbable. Manual mixing required. Sterilized via gamma irradiation. Single-use. Mechanical tamponade principle.

## Predicate Devices

- Skeletal Kinetics CAAP (Calcium Apatite) Bone Wax ([K111538](/device/K111538.md))
- US Surgical Auto Suture Bone Wax ([K971680](/device/K971680.md))
- CP Medical Bone Wax ([K024372](/device/K024372.md))
- Ceremed Ostene® CT Bone Hemostasis Implant ([K102071](/device/K102071.md))
- Orthocon Hemostatic Bone Putty 3 ([K123243](/device/K123243.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of a stylized eagle with three human profiles incorporated into its design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 3, 2014

Orthocon Incorporated Mr. Howard Schrayer 1 Bridge Street, Suite 121 Irvington, New York 10533

Re: K141502

Trade/Device Name: HBP4™ Hardening, Resorbable Hemostatic Bone Putty Regulatory Class: Unclassified Product Code: MJT Dated: July 7, 2014 Received: July 8, 2014

Dear Mr. Schrayer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Binita S. Ashar -S 2014.10.03 15:55:00 -04'00'

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K141502

Device Name

HBP4TM Hardening, Resorbable Hemostatic Bone Putty

Indications for Use (Describe)

HBP4 Hardening, Resorbable Hemostatic Bone Putty is indicated for the control of bleeding from cut or damaged bone by acting as a mechanical barrier or tamponade.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

| Contact:               | Howard Schrayer<br>Orthocon, Inc.<br>1 Bridge Street, Suite 121<br>Irvington, NY 10533<br>Telephone: 914-357-2600<br>Fax: 914-231-7884<br>hs.ss@verizon.net |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared:         | September 24, 2014                                                                                                                                          |
| Device Trade Name:     | HBP4™ Hardening, Resorbable Hemostatic Bone Putty                                                                                                           |
| Manufacturer:          | Orthocon, Inc.<br>1 Bridge Street, Suite 121<br>Irvington, NY 10533                                                                                         |
| Common Name:           | Calcium phosphate bone hemostasis material                                                                                                                  |
| Classification:        | Unclassified                                                                                                                                                |
| Product Code:          | MTJ                                                                                                                                                         |
| Primary Predicate:     | Skeletal Kinetics CAAP (Calcium Apatite) Bone Wax<br>510(k) K111538                                                                                         |
| Additional predicates: | US Surgical Auto Suture Bone Wax<br>510(k) K971680                                                                                                          |
|                        | CP Medical Bone Wax<br>510(k) K024372                                                                                                                       |
|                        | Ceremed Ostene® CT Bone Hemostasis Implant<br>510(k) K102071                                                                                                |
|                        | Orthocon Hemostatic Bone Putty 3<br>510(k) K123243                                                                                                          |

{4}------------------------------------------------

### Indications for Use:

HBP4 Hardening, Resorbable Hemostatic Bone Putty is indicated for the control of bleeding from cut or damaged bone by acting as a mechanical barrier or tamponade.

### Device Description:

HBP4 Hardening, Resorbable Hemostatic Bone Putty is a sterile, biocompatible, resorbable material of putty-like consistency for use in the control of bleeding from bone surfaces. The single use HBP4 device contains two separate components of putty-like consistency comprised of granular calcium phosphate, calcium stearate, vitamin E acetate, a triglyceride, a polyalcohol and a mixture of a lactide-diester and polyester-based polymers. When mixed together, the components of the HBP4 device form a resorbable putty-like material that can be applied directly to bleeding bone. The resulting hardening material is primarily comprised of calcium phosphate. HBP4 must be mixed immediately prior to use.

When applied to surgically cut or traumatically damaged bone. HBP4 Hardening, Resorbable Hemostatic Bone Putty achieves local control of bleeding by acting as a mechanical barrier (tamponade).

### Substantial Equivalence and Predicate Devices:

The device was shown to be substantially equivalent to previously cleared bone wax devices including Skeletal Kinetics CAAP (Calcium Apatite) Bone Wax (K111538), US Surgical Auto Suture Bone Wax (K971680), CP Medical Bone Wax (K024372), Ceremed Ostene CT Bone Hemostasis Implant (K102071) and Orthocon Hemostatic Bone Putty 3 Resorbable Hemostatic Bone Putty (K123243). Additional bone wax devices are available in a variety of forms (e.g., waxes, putties, and hardening materials) and some are permanent implants while others are resorbable.

### Technological Characteristics:

The tables below provide comparisons of HBP4 Hardening, Resorbable Hemostatic Bone Putty with the predicate devices.

{5}------------------------------------------------

# Predicate Comparison Table

| Manufacturer                    | Orthocon, Inc.                                                                                                                                                                                          | Skeletal Kinetics                                                                                                                                                                      | US Surgical                                                                                      |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| Trade Name                      | HBP4 Hardening, Resorbable<br>Hemostatic Bone Putty                                                                                                                                                     | CAAP (Calcium Apatite) Bone<br>Wax                                                                                                                                                     | Auto Suture Bone Wax                                                                             |
| 510(k) Number                   | Subject Device                                                                                                                                                                                          | K111538                                                                                                                                                                                | K971680                                                                                          |
| Type of Device/<br>Product Code | Bone hemostat / MTJ                                                                                                                                                                                     | Bone hemostat / MTJ                                                                                                                                                                    | Bone hemostat / MTJ                                                                              |
| Indications for Use             | HBP4 Hardening, Resorbable<br>Hemostatic Bone Putty is<br>indicated in the control of<br>bleeding from cut or damaged<br>bone by acting as a<br>mechanical barrier or<br>tamponade                      | CAAP (Calcium Apatite) Bone<br>Wax is indicated to control<br>bleeding from cut or damaged<br>bone by acting as a<br>mechanical barrier or<br>tamponade                                | Auto Suture Bone Wax is<br>indicated for use in the control<br>of bleeding from bone<br>surfaces |
| Intended Use                    | Bone hemostasis                                                                                                                                                                                         | Bone hemostasis                                                                                                                                                                        | Bone hemostasis                                                                                  |
| Mechanism of<br>Action          | Mechanical tamponade that<br>occludes vascular openings in<br>damaged bone                                                                                                                              | Mechanical tamponade that<br>occludes vascular openings in<br>damaged bone                                                                                                             | Mechanical tamponade that<br>occludes vascular openings in<br>damaged bone                       |
| Form of Device                  | HBP4 Hardening, Resorbable<br>Hemostatic Bone Putty is<br>formulated as a two-part<br>putty/putty device that forms a<br>"settable" (hardening) putty<br>when manually mixed at the<br>time of surgery. | CAAP (Calcium Apatite) Bone<br>Wax is formulated as a two-<br>part powder/liquid system that<br>forms a "settable" (hardening)<br>putty when manually mixed at<br>the time of surgery. | Paste-like material.                                                                             |

{6}------------------------------------------------

| Radiopacity | Radiopaque – Contains calcium phosphate                                                                                                                                                                                                                                                                                                                                                                                                         | Radiopaque – Contains calcium apatite                                                                                                                                                                                                                                                                                                        | Radiopaque – Contains β-tricalcium phosphate                                                                                                                                                               |
|-------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Materials   | Sterile mixture of two separate components of putty-like consistency comprised of calcium phosphate, calcium stearate, vitamin E acetate, triglyceride, polyalcohol and a mixture of a lactide-diester and polyester-based absorbable polymers. HBP4 is to be mixed immediately prior to use. Resulting hardening material from the two putties is primarily comprised of calcium phosphate similar to the mineral phase of native bone tissue. | A sterile kit containing calcium phosphate powder, dilute sodium silicate liquid, and a mixing system (mixing bowl, pestle and spatula). CAAP Bone wax is to be mixed immediately prior to use. Resulting hardening material from the paste is primarily comprised of calcium phosphate, similar to the mineral phase of native bone tissue. | A sterile mixture of glycolide, caprolactone, mannitol and β-tricalcium phosphate. The copolymer derived from glycolide and caprolactone is the same copolymer used to coat US Surgical's POLYSORB Suture. |
| Resorbable  | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                             | Yes                                                                                                                                                                                                                                                                                                                                          | Yes                                                                                                                                                                                                        |

{7}------------------------------------------------

| Resorption Time       | Greater than 30 days primarily due to presence of calcium phosphate.                                                                                                                      | Greater than 30 days primarily due to presence of calcium phosphate        | Greater than 30 days primarily due to presence of calcium phosphate.                                                       |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Method of Application | Manually applied and spread onto bone tissue                                                                                                                                              | Manually applied and spread onto bone tissue                               | Manually applied and spread onto bone tissue                                                                               |
| Degradation Process   | The non-calcium salt and non-polymeric components degrade via dissolution; the polymer degrades via hydrolysis and calcium salts degrade via chemical dissolution and/or cellular removal | Believed to be combination of chemical dissolution and/or cellular removal | Copolymer degrades via hydrolysis; calcium phosphate degrades via combination of chemical dissolution and cellular removal |
| Sterility             | Provided sterile for single use by gamma irradiation                                                                                                                                      | Provided sterile for single use by gamma irradiation                       | Provided sterile for single use by gamma irradiation                                                                       |

{8}------------------------------------------------

# Predicate Comparison Table (cont'd)

| Manufacturer                    | CP Medical                                                                                           | Ceremed                                                                                                                                                                                                               | Orthocon, Inc.                                                                                                                                                                          |
|---------------------------------|------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name                      | CP Medical Bone Wax                                                                                  | Ostene® CT Bone<br>Hemostasis Implant                                                                                                                                                                                 | Hemostatic Bone Putty 3<br>Resorbable Hemostatic Bone<br>Putty                                                                                                                          |
| 510(k) Number                   | K024372                                                                                              | K102071                                                                                                                                                                                                               | K123243                                                                                                                                                                                 |
| Type of Device/<br>Product Code | Bone hemostat / MTJ                                                                                  | Bone hemostat / MTJ                                                                                                                                                                                                   | Bone hemostat / MTJ                                                                                                                                                                     |
| Indications for Use             | The CP Medical Bone Wax is<br>indicated for use in the control<br>of bleeding from bone<br>surfaces. | Ostene® CT is indicated for<br>use as a water-soluble<br>implant material and for use<br>in the control of bleeding from<br>bone surfaces in<br>cardiothoracic surgery<br>following sternotomy                        | Hemostatic Bone Putty 3 is<br>indicated for the<br>control of bleeding from cut or<br>damaged bone by acting as a<br>mechanical barrier or<br>tamponade.                                |
| Intended Use                    | Bone hemostasis                                                                                      | Bone hemostasis                                                                                                                                                                                                       | Bone hemostasis                                                                                                                                                                         |
| Mechanism of<br>Action          | Mechanical tamponade that<br>occludes vascular openings in<br>damaged bone                           | Mechanical tamponade that<br>occludes vascular openings in<br>damaged bone                                                                                                                                            | Mechanical tamponade that<br>occludes vascular openings in<br>damaged bone                                                                                                              |
| Form of Device                  | A hard wax-like material that<br>must be warmed by kneading<br>prior to use,                         | An odorless, opaque wax-like<br>material designed to be<br>utilized directly out of the<br>package. It can be softened<br>and increased in stickiness by<br>warming and by additional<br>handling and<br>manipulation | Putty-like material that does<br>not require kneading prior to<br>application.                                                                                                          |
| Radiopacity                     | Not Radiopaque                                                                                       | Not Radiopaque                                                                                                                                                                                                        | Radiopaque - Contains β-tricalcium phosphate                                                                                                                                            |
| Materials                       | CP Medical bone wax is a<br>sterile mixture of bees wax<br>and paraffin.                             | Ostene® CT is a sterile<br>mixture of water-soluble<br>alkylene oxide copolymers                                                                                                                                      | Hemostatic Bone Putty 3 is a<br>mixture of alkylene oxide<br>polymer-based materials,<br>vitamin E acetate, granular<br>calcium phosphate and<br>carboxymethylcellulose<br>sodium salt. |
| Resorbable                      | Non-resorbable                                                                                       | Yes                                                                                                                                                                                                                   | Yes                                                                                                                                                                                     |
| Resorption Time                 | Permanent implant                                                                                    | Less than 30 days                                                                                                                                                                                                     | Greater than 30 days<br>primarily due to presence of<br>calcium phosphate.                                                                                                              |
| Method of<br>Application        | Manually applied and spread<br>onto bone tissue                                                      | Manually applied and spread<br>onto bone tissue                                                                                                                                                                       | Manually applied and spread<br>onto bone tissue                                                                                                                                         |
| Degradation<br>Process          | Does not degrade - has been<br>found to be present years<br>after implantation                       | Degrades via dissolution                                                                                                                                                                                              | The non-calcium salt degrade<br>via dissolution; the calcium<br>salts degrade via chemical<br>dissolution and/or cellular<br>removal                                                    |
| Sterility                       | Provided sterile for single use<br>by gamma irradiation                                              | Provided sterile for single use<br>by irradiation                                                                                                                                                                     | Provided sterile for single use<br>by gamma irradiation                                                                                                                                 |

{9}------------------------------------------------

{10}------------------------------------------------

## Performance Testing:

Bench testing, biocompatibility and animal functionality testing performed on HBP4 ™ Hardening, Resorbable Hemostatic Bone Putty demonstrate that the device is substantially equivalent to predicate devices in intended use, technological characteristics, and performance. This testing included the following:

Bench Testing was conducted to verify the device's handling properties, to characterize the device's performance over a range of temperatures and to evaluate the device's dissolution properties. The following bench studies were completed: relative stiffness, spreadability, stickiness, temperature sensitivity, electrocautery compatibility, dissolution and swelling.

Biocompatibility Testing was conducted to evaluate the device's biocompatibility in accordance with the recommendations of ISO 10993. The following biocompatibility studies were conducted on the final, finished, gamma-irradiated sterile device in accordance with the GLP requirements: irritation, sensitization, acute systemic toxicity, genotoxicity, implantation, subacute systemic toxicity, chronic systemic toxicity, hemolysis, endotoxicity and pyrogenicity.

Animal Testing included animal studies to demonstrate intraoperative in vivo hemostasis, resistance to irrigation, and to characterize resorption time.

## Conclusion

HBP4 is substantially equivalent to previously cleared bone wax devices with respect to intended use, general technological characteristics and performance.

---

**Source:** [https://fda.innolitics.com/device/K141502](https://fda.innolitics.com/device/K141502)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
