K141488 · Careray Digital Medical System Co., Ltd. · MQB · Sep 8, 2015 · Radiology
Device Facts
Record ID
K141488
Device Name
CAREVIEW 1800R
Applicant
Careray Digital Medical System Co., Ltd.
Product Code
MQB · Radiology
Decision Date
Sep 8, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
CareView 1800R X-ray Flat Panel Detectors is indicated for digital imaging solution designed for providing general radiographic system in all general purpose diagnostic procedures. CareView 1800R X-ray Flat Panel Detectors is a component of a digital imaging system. Properly integrated into a completed X-ray system, the detector enables the digital X-ray imaging intended for medical application. This product is not intended for mammography applications and dental X-ray applications.
Device Story
CareView 1800R is a stationary X-ray flat panel detector (FPD) component for digital radiography systems. It uses a TFT/PD image sensor with a direct-deposited CsI scintillator to convert X-ray photons into visible light, then into electron signals via a diode capacitor array. The device captures single-frame images, which are processed and transmitted to a PC via a user interface. Used in clinical settings by healthcare professionals, the detector integrates into a completed X-ray system to enable digital imaging. It supports external and manual synchronization modes with X-ray generators. The output is a digital image used by clinicians for diagnostic purposes, facilitating medical decision-making through high-resolution radiographic visualization.
Clinical Evidence
Clinical evidence included a concurrence study of 30 clinical images comparing the CareView 1800R to the predicate device. Results showed no significant difference between the images produced by the subject and predicate devices. Primary evidence for equivalence was based on laboratory performance data (MTF, DQE, NPS, sensitivity, lag, uniformity, etc.) and established scintillator material usage.
Technological Characteristics
Active area 43cm x 43cm; CsI scintillator; TFT readout mechanism; 16-bit ADC; 154µm pixel pitch; 2816x2816 pixel array. Stationary installation. Power consumption ~36W. Connectivity via wired communication. Tested to IEC/ES 60601-1 (electrical/mechanical/environmental safety) and IEC/EN 60601-1-2 (EMC).
Indications for Use
Indicated for digital imaging in general purpose diagnostic radiographic procedures. Not for mammography or dental applications.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 8, 2015
CareRay Digital Medical System Co., Ltd. % Jamie Ku Official Correspondent Compass Innovations, Inc. 3001 Winchester Blvd, Suite 3 CAMPBELL CA 95008
Re: K141488
Trade/Device Name: CareView 1800R X-ray Flat Panel Detector Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: August 13, 2015 Received: August 17, 2015
Dear Jamie Ku:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
For
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K141488
Device Name CareView 1800R X-ray Flat Panel Detectors
Indications for Use (Describe)
CareView 1800R X-ray Flat Panel Detectors is indicated for digital imaging solution designed for providing general radiographic system in all general purpose diagnostic procedures. CareView 1800R X-ray Flat Panel Detectors is a component of a digital imaging system. Properly integrated into a completed X-ray system, the digital X-ray imaging intended for medical application.
This product is not intended for mammography applications and dental X-ray applications.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> </span> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <input type="checkbox"/> </span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
[As required by 21 CFR 807.92]
## 1. Date Prepared [21 CFR807.92 (a) (1)]
August 13, 2015
#### 2. Submitter's Information [21 CFR807.92 (a) (1)]
| Company Name: | CareRay Digital Medical System Co., Ltd. |
|------------------|---------------------------------------------------------------------------------------------------------|
| Company Address: | Suite 201, Building A2, BIOBAY, 218 Xing Hu Street,<br>SuZhou Industrial Park, SuZhou 215123, P.R.China |
| Contact Person: | Ms. Ding |
| Phone Number: | (86) 512-86860288 |
| Fax Number: | (86) 512-86860388 |
| E-mail: | yl.ding@careray.com |
### 3. Trade Name, Common Name, Classification [21 CFR807.92(a)(2)]
| Trade Name: | X-ray Flat Panel Detectors |
|----------------------|----------------------------|
| Model Name: | CareView 1800R |
| Classification Name: | Stationary X-ray system |
| Regulation Number: | 21 CFR 892.1680 |
| Regulatory Class: | Class II |
| Product Code: | MQB |
#### 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)[
The identified predicates within this submission are as follows:
| Manufacturer: | Canon Inc. |
|----------------------|-------------------------|
| Trade Name: | Canon |
| Model Name: | CXDI-40EG |
| Classification Name: | Stationary X-ray system |
| Regulation Number: | 21 CFR 892.1680 |
| Regulatory Class: | Class II |
{4}------------------------------------------------
| Product Code: | MQB |
|---------------|---------|
| FDA 510(k) #: | K050987 |
### 5. Description of the Device [21 CFR 807.92(a)(4)]
CareView 1800R X-ray Flat Panel Detectors supports the single frame mode, with the key component of TFT/PD image sensor flat panel of active area: 43cm×43cm. The sensor plate of CareView 1800R X-ray Flat Panel Detectors is direct-deposited with Csl scintillator to achieve the conversion from X-ray to visible photon. The visible photons are transformed to electron signals by diode capacitor array within TFT panel, which are composed and processed by connecting to scanning and readout electronics, consequently to form a panel image by transmitting to PC through the user interface.
The major function of the CareView 1800R X-ray Flat Panel Detectors detector is to convert the X-ray to digital image, with the application of high resolution X-ray imaging. This detector is the key component of DR system, enables to complete the digitalization of the medical X-ray imaging with the DR system software.
# 6. Intended Use [21 CFR 807.92(a)(5)]
CareView 1800R X-ray Flat Panel Detectors is indicated for digital imaging solution designed for providing general radiographic system in all general purpose diagnostic procedures. CareView 1800R X-ray Flat Panel Detectors is a component of a digital imaging system. Properly integrated into a completed X-ray system, the detector enables the digital X-ray imaging intended for medical application.
This product is not intended for mammography applications and dental X-ray applications.
| Item | Proposed Device:<br>CareView 1800R X-ray Flat Panel<br>Detectors | Predicate Device: CXDI-40EG |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(K)<br>Number | To be assigned | K050987 |
| Intended Use | The CareView 1800R X-ray Flat<br>Panel Detectors is indicated for<br>digital imaging solution designed<br>for providing general radiographic<br>system in all general purpose<br>diagnostic procedures. CareView<br>1800R X-ray Flat Panel Detectors<br>is a component of a digital imaging<br>system. Properly integrated into a | DIGITAL RADIOGRAPHY<br>CXDI-40EG provides digital<br>image capture for conventional<br>film/screen radiographic<br>examinations. The device is<br>intended to replace radiographic<br>film/screen systems in all general<br>purpose diagnostic procedures. |
| | completed X-ray system, the<br>detector enables the digital X-ray<br>imaging intended for medical<br>application.<br>This product is not intended for<br>mammography applications and<br>dental X-ray applications. | |
| Classification<br>Name | Stationary X-ray system | Stationary X-ray system |
| Product<br>Code | MQB | MQB |
| Regulation<br>Number | 21 CFR 892.1680 | 21 CFR 892.1680 |
| Panel | Radiology | Radiology |
| Class | II | II |
| X-ray<br>Absorber | CsI Scintillator | GOS |
| Installation<br>Type | Stationary, permanently installed | Stationary, permanently installed |
| Readout<br>Mechanism | Thin Film Transistor | Thin Film Transistor |
| Detector<br>Size | 492 x 492 x 33.5 mm | 550 x 550 x 67.5 mm |
| Detector<br>Weight | 13 kg | 20 kg |
| Pixel Array | 2816x2816 | 2688 x2688 |
| Active Area | 43cm×43cm | 43cm×43cm |
| Pixel Pitch | 154µm | 160µm |
| ADC<br>Digitization | 16 bit | 14 bit |
| Power<br>Consumption | ~36 W | Max.200 VA |
| Modulation<br>Transfer<br>Function<br>(MTF) | 1 lp/mm (70±3)%<br>2 lp/mm (42±3)%<br>3 lp/mm (23±3)% | 1 lp/mm 68%<br>2 lp/mm 38%<br>3 lp/mm 18% |
| Detective<br>Quantum<br>Efficiency<br>(DQE) | DQE at 5 µGy<br>0 lp/mm (63±3)%<br>1 lp/mm (45±3)%<br>2 lp/mm (33±3)%<br>3 lp/mm (20±3)% | DQE at 5 mR<br>0 lp/mm 37%<br>I lp/mm 34%<br>2 lp/mm 24%<br>3 lp/mm 12% |
| Power<br>Supply | Voltage: 100-250 VAC<br>Frequency: 50/60 Hz | Voltage: 100V, 120V, 230/240 V<br>Frequency: 50/60 Hz |
| Operating<br>Condition | Temperature: 5 ~ 35°C<br>Relative humidity: 30~75% RH<br>Air pressure: 700hPa~1060hPa | Temperature: 5~35°C<br>Relative humidity: 30~75%<br>Air pressure: 700hPa ~1060hPa |
| Storage<br>Condition | Temperature: -20~55 °C<br>Relative humidity: 10~90%RH<br>Air pressure: 700hPa~1060hPa | Temperature: -30~60 °C<br>Relative humidity: 10~60%<br>Air pressure: 700hPa ~1060hPa |
# 7. Technological Characteristics [21 CFR 807.92(a)(6)]
{5}------------------------------------------------
{6}------------------------------------------------
# 8. System requirements to operate with other radiographic system components
- 1) Recommended Generator Specification:
Energy range: 40~150kVp
mA range: 10~1000mA (depending on the generator power)
ms range: 10~6300ms to produce 0.1~1000mAs (depending on the generator power) Note: To our best knowledge, the detector is compatible with the X-ray generators with the specifications described above. If you have questions regarding the compatibility issue for other generators, please contact your distributor or CareRay.
- 2) Application Program Interface (API) for system integration manufacturer
Peripheral hardware: CareView detector connected via wired communication.
CPU: Intel (R) Core (TM) 2 Duo, 2.93GHz or above
RAM: 2 GB or higher
Hard disk: 160 GB or higher
Monitor: 1280 x 1024 or higher
OS: Windows XP or Windows 7
Development environment: MS Visual Studio 2005
- 3) X-ray exposure mode
The synchronous connection mode is the signal transfer mode between the X-ray generator which sends the X-ray and the detector which receives the X-ray.
CareView 1800R supported typical sync mode contains external sync, manual sync mode.
The detector can't provide feedback to the generator to terminate the x-ray exposure.
## 9. Substantial Equivalence [21 CFR 807.92(b) (1) and 807.92(b)(2)]
- > Electrical safety and EMC testing
Electrical, mechanical, environmental safety and performance testing according to IEC/ES 60601-1 was performed, and EMC testing was also conducted in accordance with IEC/EN 60601-1-2. All test results are satisfactory. Section 201.7.2.7, 201.7.9.1, 201.7.9.2.1.103 and 203.4.1 of IEC 60601-2-54 were also conducted.
- A Nonclinical considerations
The following non-clinical studies have been performed and the results have shown that the CareView 1800R X-ray flat panel detector is substantially equivalent to the predicate devices on the Market (CXDI-40EG).
{7}------------------------------------------------
-Detective quantum efficiency (DQE), Quantum limited performance, Modulation transfer function (MTF), Effects of aliasing, Sensitivity linearity, Lag(Erasure thoroughness), Change in detection sensitivity, Dose requirement and reciprocity changes, Stability of device characteristics with time, Uniformity of device characteristic, Noise power spectrum(NPS), Spatial resolution, Image Acquisition time, & Black level
- A Clinical considerations
A concurrence study of 30 clinical images was conducted to compare the performance of the CareView 1800R to that of the predicate device (K050987).
Clinical images were provided; these images were not necessary to establish substantial equivalence based on the superior performance in comparison to the predicate based on laboratory testing and based on the use of a detector scintillator material that is already on the marketplace; however, they provide further evidence in addition to the laboratory performance data to show that the complete system works as intended.
There was no significant difference between the images of the CareView 1800R and those of the predicate device.
#### 10. Conclusion [21 CFR 807.92(b) (3)]
In accordance with the Federal Food, Druq and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, the CareView 1800R X-ray flat panel detector is substantially equivalent to predicate device CXDI-40EG (K050987). Both propose and predicate devices are same or very similar in the intended use, the design principle, the applicable standards and specification. Some characteristics, for example, the detector size, pixel pitch, weight and ADC digitization, X-ray absorber, storage condition are different. However the test reports in this submission documents provide demonstration that these differences do not raise any new questions of safety and effectiveness. Therefore, CareRay Digital Medical System Co., Ltd. concludes the CareView 1800R X-ray flat panel detector is substantially equivalent with the predicate device CXDI-40EG (K050987) of Canon Inc.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.