K141474 · Becton, Dickinson & CO · PHC · Dec 18, 2014 · General Hospital
Device Facts
Record ID
K141474
Device Name
BECTON DICKINSON INTELLIPORT SYSTEM
Applicant
Becton, Dickinson & CO
Product Code
PHC · General Hospital
Decision Date
Dec 18, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Indications for Use
The BD Intelliport System is an automated record keeping system that incorporates patient safety features that are aligned with hospital patient records and protocols. The system is comprised of an injection port and software that enables the identification, measurement, alerting and documentation of the administration of medications to patients. The BD Intelliport System allows the clinician to record anesthesia related events in the pre-op, intra-op, and PACU areas. The system is indicated for use by healthcare professionals in a hospital or medical center setting with patients who are receiving manually administered bolus intravenous injections as part of their care to facilitate documentation of the medications.
Device Story
System automates documentation of IV bolus medication administration; comprises intelligent injection port and software. Clinician connects port to patient fluid-delivery line; injects medication via syringe. Port reads 2-D barcode on syringe (drug name/concentration); uses ultrasonic sensor to measure volume and time of injection. Data transmitted to electronic record system. Used in pre-op, intra-op, and PACU by healthcare professionals. Facilitates accurate medication logging; replaces manual entry; improves documentation efficiency and patient safety.
Clinical Evidence
Bench testing only. Includes software unit testing, code reviews (verification), system validation testing, and electrical/electromagnetic safety testing. Validation covered user interface, usability, wireless communication, and volume delivery accuracy (±5% for >1.0-55 mL; ±20% for 0.4-1.0 mL).
Technological Characteristics
System includes intelligent injection port and software. Uses ultrasonic sensor for volume measurement. Connectivity via wireless communication to hospital information system. Accuracy: ±5% for bolus >1.0 mL to 55 mL; ±20% for 0.4 to 1.0 mL. Complies with IEC 60601-2-24 for bolus delivery.
Indications for Use
Indicated for healthcare professionals in hospital/medical center settings to document manually administered bolus intravenous injections for patients in pre-op, intra-op, and PACU areas.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
DocuSys Anesthesia Information and Digital-Drug Management System (K062388)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 18,2014
Becton, Dickinson and Company C/O Mr. Daniel Olivier Certified Compliance Solutions, Inc. President 11665 Avena Place, Suite 203 San Diego, CA 92128
Re: K141474
Trade/Device Name: BD Intelliport System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: PHC, FPA Dated: December 15, 2014 Received: December 16, 2014
Dear Mr. Olivier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Olivier
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runno DDS, MA
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141474
Device Name BD Intelliport System
### Indications for Use (Describe)
The BD Intelliport System is an automated record keeping system that incorporates patient safety features that are aligned with hospital patient records and protocols. The system is comprised of an injection port and software that enables the identification, measurement, alerting and documentation of medications to patients.
The BD Intelliport System allows the clinician to record anesthesia related medication events in the pre-op, intra-op, and PACU areas. The system is indicated for use by healthcare professionals in a hospital or medical center setting with patients who are receiving manually administered bolus intravenous injections as part of their care to facilitate documentation of the medications.
### Type of Use (Select one or both, as applicable)
- Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 5. 510(k) Summary
| Submitter: | Becton, Dickinson and Company |
|--------------------------|-------------------------------------------------------------|
| Address: | 1 Becton Drive<br>Franklin Lakes, NJ 07417 |
| Phone number: | (201) 847-4496 |
| Fax number: | (201) 847-4845 |
| Contact person: | Daniel Olivier |
| Phone number: | (858) 675-8200 |
| Fax number: | (858) 675-8201 |
| Date prepared: | September 14, 2014 |
| Trade name: | BD Intelliport System |
| Common name: | Intelliport System |
| Product Code, Primary: | PHC, Infusion Safety Management Software |
| Regulation: | 21 CFR 880.5725, Infusion Pump, Class II |
| Product Code, Secondary: | FPA, Set, Administration, Intravascular |
| Regulation: | 21 CFR 880.5440, Intravascular Administration Set, Class II |
Substantial equivalence claimed to: DocuSys Anesthesia Information and Digital-Drug Management System
## Description:
The BD Intelliport System is used for automated documentation of medication, concentration, dose, volume and time of each IV injection for when intravenous bolus injections of medication are given to a patient.
During treatment, the clinician connects an intelligent injection port, called the Intelliport device, to a patient's fluid-delivery line and performs standard drug-delivery activities. The health practitioner injects the drug; the Intelliport reads the 2-D barcode adhered to the syringe containing the drug such as Becton Dickinson syringe K110771. This barcode contains the drug name and concentration. As the drug is injected, the Intelliport measures the volume of the injected drug and the time the drug was administered. Once the drug has been administered, the Intelliport transmits all of the information to an electronic record maintained by the Computer.
{4}------------------------------------------------
# Indications for Use:
The BD Intelliport System is an automated record keeping system that incorporates patient safety features that are aligned with hospital patient records and protocols. The system is comprised of an injection port and software that enables the identification, measurement, alerting and documentation of the administration of medications to patients.
The BD Intelliport System allows the clinician to record anesthesia related events in the pre-op, intra-op, and PACU areas. The system is indicated for use by healthcare professionals in a hospital or medical center setting with patients who are receiving manually administered bolus intravenous injections as part of their care to facilitate documentation of the medications.
## Technological Characteristics
The BD Intelliport System is substantially equivalent to the predicate Device, the DocuSys Anesthesia Information and Digital-Drug Management System.
| | BD<br>Intelliport System | DocuSys Anesthesia Information &<br>Digital-Drug Management System |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Becton Dickinson | DocuSys |
| 510(k) | NA | K062388 |
| Indications for<br>use | The BD Intelliport System is an<br>automated record keeping system<br>that incorporates patient safety<br>features that are aligned with<br>hospital patient records and<br>protocols. The system is comprised<br>of an injection port and software that<br>enables the identification,<br>measurement, alerting and<br>documentation of the administration<br>of medications to patients.<br><br>The BD Intelliport System allows the<br>clinician to record anesthesia<br>related events in the pre-op, intra-<br>op, and PACU areas. The system is<br>indicated for use by healthcare<br>professionals in a hospital or<br>medical center setting with patients<br>who are receiving manually<br>administered bolus intravenous<br>injections as part of their care to<br>facilitate documentation of the<br>medications. | The DocuSys Anesthesia Information<br>and Digital-Drug Management System<br>offers anesthesia a comprehensive<br>record keeping system with advanced<br>patient safety features. The clinician<br>begins interacting with the system<br>through the electronic anesthesia pre-<br>operative assessment in which patient<br>data relative to anesthesia is<br>recorded. Electronic documentation of<br>the patient's allergies and home<br>medications allows for proactive<br>adverse drug event screening when<br>the patient is in the operating room<br>prior to anesthesia delivering drugs.<br>The DocuSafe electronic anesthesia<br>information system allows the clinician<br>to record anesthesia related events in<br>the pre-op, intra-op, and PACU areas.<br>The DocuSafe software receives<br>physiologic data from various patient<br>monitoring devices. Clinicians enter<br>drug information onto the anesthesia<br>record in one of two methods: a)<br>manually selecting the drug from a list<br>of drugs in the formulary, or b)<br>through the use of DocuSys' optional<br>digital-Drug Management System. If<br>the optional d-DMS is used, the<br>clinician may scan a barcode affixed<br>to a Syringe Label Cradle or use |
| | BD<br>Intelliport System | DocuSys Anesthesia Information &<br>Digital-Drug Management System |
| | | Formulary Reference module, managed by pharmacy, contains drug specific information used for documenting drugs on the anesthesia record, performing ADE checks, and submitting medication utilization information for billing purposes. The DocuRx pharmacy component of the digital-Drug Management System provides for pre-screening of medications that are ordered by anesthesia to check for any potential adverse drug events. It also provides a methodology for narcotics tracking through its comprehensive medication tracking and wasting feature. Each component of the Anesthesia Information and Digital-Drug Management System plays an integral part in standardizing medication administration techniques and documentation. |
| Allergy data | Provides notice of patient allergies | Provides notice of patient allergies |
| Drug<br>interactions | NA | Adverse drug event screening |
| Medication<br>history record | Provides a record of drug delivery related events | Provides a record of drug delivery related events |
| Drug formulary | Provides drug formulary to select drugs to be delivered | Provides drug formulary to select drugs to be delivered |
| Drug delivery | Pre-filled syringe that has a barcode attached to register drug delivered | Pre-filled syringe that has a barcode to register drug delivered and volume |
| Measuring<br>drug delivery | Measures volume of the injected drug using an ultrasonic sensor | Measures the syringe plunger displacement |
| Accuracy of<br>drug delivery | $\pm$ 5% for bolus volumes > 1.0 mL to 55 mL<br>$\pm$ 20% for bolus volumes of 0.4 to 1.0 mL<br>(at nominal conditions) | Calculated as a percent of total syringe volume not volume of drug delivered so not used for comparison. |
| Formulary<br>database | Provides ability to edit drug formulary | Provides a Formulary Reference module |
| Narcotics<br>tracking | Provides database for tracking narcotics | Provides database for tracking narcotics |
| External data<br>interface | Provides interface to hospital information system | Provides interface to hospital information system |
{5}------------------------------------------------
# Substantial Equivalence Discussion
The BD Intelliport System is substantially equivalent to the DocuSys Anesthesia Information and Digital-Drug Management System based on indications for use and comparison of functional
{6}------------------------------------------------
capabilities. Both devices are intended to provide the healthcare professional with an electronic record of "anesthesia related events." The BD Intelliport System provides an automated record of the name, concentration, dose, volume of the drug injected into the patient, as well as the time of the injection. Features common to both systems include:
- · Medication history record: name, concentration, dose, volume of the drug injected into the patient as well as the time of the injection
- · Allergy Alerts
- Drug Formularv
The minor differences between the BD Intelliport System and the DocuSys Anesthesia Information and Digital-Drug Management System do not raise new questions of safety or effectiveness. These differences will be discussed in turn.
The DocuSys Anesthesia Information and Digital-Drug Management System provides adverse drug event screening such as notification of potential drug - drug interactions. This feature is not provided by the BD Intelliport System. Failure to provide this feature does not raise safety or efficacy concerns as this screening is currently performed in the hospital setting.
The drug-delivery accuracy of the DocuSys Anesthesia Information and Digital-Drug Management System is ±3%. However, this accuracy measurement is always calculated based on the total syringe volume not the volume of drug delivered to the patient. As stated in the DocuSys Summary Data for DocuJet II Accuracy, "If a user delivers a bolus injection of 2ml using a 20ml syringe and the DocuJet records 2.2ml on the record, that is not considered a discrepancy of 10%. The discrepancy is based on the Nominal Syringe Volume and therefore would be calculated as 0.2ml/20ml, a 1% discrepancy." The term discrepancy, in this case, is tantamount to accuracy. The accuracy of the BD Intelliport System is ± 5% for bolus volumes from 1.0 mL to 55 mL and ± 20% for bolus volumes from 0.4 to 1.0 mL at nominal conditions. This measurement is based on the volume of the drug delivered.
A performance test comparison of the accuracy of the DocuSys Anesthesia Information and Digital-Druq Management System could not be performed as this product is no longer marketed and not available.
Testing of the BD Intelliport System syringe delivery is conducted in accordance with the bolus delivery requirements as defined in IEC 60601-2-24 (Particular requirements for the safety of infusion pumps and controllers). The accuracy of drug delivery for the BD Intelliport System does not raise safety or efficacy concerns as this accuracy is in line with the accuracy currently achieved through manual activation of a syringe.
## Test Summary
The performance testing for the BD Intelliport System includes software unit testing and code reviews (verification), system validation testing, and testing to compliance standards for electrical and electromagnetic safety. Traceability has been documented between the system specification to validation test protocols. Validation test procedures also address the user interface, user manual descriptions, usability, wireless communication and general performance including volume delivery accuracy.
{7}------------------------------------------------
# Conclusion:
The performance of the BD Intelliport System is substantially equivalent to that of the DocuSys Anesthesia Information and Digital-Drug Management System and raises no safety or effectiveness issues and performs as well or better than the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.