SPHYGMOMANOMETER

K141392 · Suzhou JJ Meter Co., Ltd. · DXQ · Jul 2, 2014 · Cardiovascular

Device Facts

Record IDK141392
Device NameSPHYGMOMANOMETER
ApplicantSuzhou JJ Meter Co., Ltd.
Product CodeDXQ · Cardiovascular
Decision DateJul 2, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1120
Device ClassClass 2

Intended Use

The device is intended to be used by medical professionals or in the home for the measurement of systolic and diastolic pressure on adults. The device is intended to be manually attached to a patient and manually inflated along with a manual method for detecting Korotkoff sounds.

Device Story

The device is a manual sphygmomanometer consisting of an inflatable bladder within a soft, inelastic sleeve, connected via tubing to a non-invasive pressure measurement unit. It is used by medical professionals or patients in home settings. The device is manually wrapped around the patient's limb and manually inflated. The user employs a manual method (e.g., stethoscope) to detect Korotkoff sounds to determine systolic and diastolic blood pressure. It provides no automated processing or electronic output.

Clinical Evidence

Bench testing only. The device was tested for conformance to ISO 81060-1:2007 (Non-invasive sphygmomanometers - Part 1: Requirements and test methods for non-automated measurement type) and biocompatibility standards ISO 10993-5 and ISO 10993-10.

Technological Characteristics

Manual sphygmomanometer; soft inelastic sleeve with integrated inflatable bladder; hook and loop closure; manual inflation; non-automated measurement. Conforms to ISO 81060-1:2007, ISO 10993-5, and ISO 10993-10.

Indications for Use

Indicated for adult patients for the measurement of systolic and diastolic blood pressure. Intended for use by medical professionals or in the home setting.

Regulatory Classification

Identification

A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # K141392; page 1 of 3 # 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR §807.92. Date: May 27, 2014 1. Submitter: Name: SUZHOU JJ METER CO., LTD. Add: No.156 Xuqing Rd, Xushuguan Town, Suzhou New District, P.R. China Tel: 0086-512-66168979 Fax: 0086-512-66168979 2. Contact Person: Long Yang (COO.) Shenzhen Hlongmed Biotech Co., Ltd. R1508, East Building, Yihai Plaza, Chuangye Road, Nanshan District,Shenzhen, P.R. China Tel: 0086-755-86664986 Fax: 0086-755-86664933 E-mail: yanglong@hlongmed.com {1}------------------------------------------------ ### K141392 l # 3. Device Information: | Trade name: | Sphygmomanometer | |----------------------|----------------------------| | Model: | JHT-1110, JHT-1500, JHT-16 | | Common name: | Blood Pressure Cuff | | Classification name: | Blood Pressure Cuff | | Review Panel: | Cardiovascular | | Product Code: | DXQ | | Regulation Class: | II | | Regulation Number: | 870.1120 | # 4. Predicate Device Information: Company Name: Wenzhou Bokang instruments Co., Ltd. Device Name: Aneroid Sphygmomanometer BK2002 510(k) number: K043286. # 5. Device Description: The device comprises tubing attached to a soft inelastic sleeve with an integrated inflatable bladder that is wrapped around the patient's limb and secured by hook and loop closure . The device tubing is connected to a non-invasive sphygmomanometer. ## 6. Indication For Use: The device is intended to be used by medical professionals or in the home for the measurement of systolic and diastolic pressure on adults. The device is intended to be {2}------------------------------------------------ K141392 ### Page 3 of 3 manually attached to a patient and manually inflated along with a manual method for detecting Korotkoff sounds. ### 7. Technological Characteristics: The JJ METER's Sphygmomanometer(model: JHT-1110, JHT-1500, JHT-1611) are virtually the same as Bokang's Aneroid Sphygmomanometer BK2002. #### 8. Safety and Performance Data: The JJ METER's Sphygmomanometer(model: JHT-1110, JHT-1500, JHT-1611) have been tested conform to the ISO 81060-1:2007 Non-invasive sphygmomanometers - Part 1: Requirements and test methods for non-automated measurement type, and ISO 10993-5 and ISO 10993-10. #### 9. Conclusions: In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this pre-market notification. SUZHOU JJ METER CO., LTD concludes that the Sphygmomanometer( models JHT-1110, JHT-1500, JHT-1611) are safe and effective, and substantially equivalent to the predicate device described herein. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird symbol, with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the bird symbol. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G Silver Spring, MD 20993-0002 July 2, 2014 Suzhou Jj Meter Co., Ltd. c/o Ms. Long Yang Official Correspondent R1508, East Bldg, Yihai Plaza Chuangye Rd. Nanshan District Shenzhen, Guangdong, 518054 CH K141392 Re: > Trade/Device Name: Sphygmomanometer (Models JHT-1110, JHT-1500, JHT-1611) Regulation Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II Product Code: DXQ Dated: May 20, 2014 Received: May 27, 2014 Dear Ms. Yang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 IVD and Part 809); medical device reporting (reporting of {4}------------------------------------------------ Page 2 - Ms. Long Yang medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 IVD and Part 809), please contact the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/dcfault.htm. Sincerely yours, Image /page/4/Picture/6 description: The image shows a logo with the letters FDA in a stylized font. The letters are interconnected and have a textured appearance. There is a signature or some other form of writing across the letters. forbram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ K141392 Page _________________________________________________________________________________________________________________________________________________________________________ 510(k) Number (if known): K141392 Device Name: __ Sphygmomanometer (Model: JHT-1110, JHT-1500, JHT-1611) _______________________________________________________________________________________________________ Indications For Use: í The device is intended to be used by medical professionals or in the home for the measurement of systolic and diastolic pressure on adults. The device is intended to be manua attached to a patient and manually inflated along with a manual method for detecting Korotkoff sounds. (BLEASE DO NOT WRITE BETON THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Image /page/5/Picture/7 description: The image shows a document with the title "Concurrence of CDRH, Office of Device Evaluation (ODE)". There is a signature and the date "2014.07.02 14:34:45-04'00'". The document indicates "Over-The-Counter Use" with an "X" marked, and also mentions "Prescription Use (Per 21 CFR 801.109)" along with an "OR" separator.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...