← Product Code [MQB](/productcode/MQB) · K141271

# AERODR SYSTEM 2 (K141271)

_Konica Minolta, Inc. · MQB · Sep 26, 2014 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K141271

## Device Facts

- **Applicant:** Konica Minolta, Inc.
- **Product Code:** [MQB](/productcode/MQB.md)
- **Decision Date:** Sep 26, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1680
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The AeroDR SYSTEM 2 is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen system in general-purpse diagnostic procedures. The AeroDR SYSTEM 2 is not in mammography. fluoroscopy, tomography and angiography applications.

## Device Story

AeroDR SYSTEM 2 is a digital radiography system designed to replace film/screen systems. It utilizes a flat panel digital detector (P-51) to capture X-ray signals, which are then transmitted via wired or wireless connection to the Console CS-7 operator console. The system includes a new Generator Interface Unit2 and is designed for compact clinical layouts. Operated by healthcare professionals in clinical settings, the device processes X-ray inputs into digital images for diagnostic review. The system integrates with legally marketed X-ray generators. Benefits include improved workflow, lighter detector weight, extended battery life, and IPX6 waterproof rating for the detector. Clinical decision-making is supported by the high-quality digital images produced, which are equivalent to those of the predicate AeroDR systems.

## Clinical Evidence

Bench testing only. Non-clinical and clinical performance testing was conducted in accordance with the FDA Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices. Results demonstrated equivalent performance outcomes to the predicate devices, confirming no safety or effectiveness issues.

## Technological Characteristics

Digital radiography system using a flat panel detector (P-51). Materials evaluated per EN ISO 10993-1. Electrical safety: AAMI/ANSI ES60601-1:2005/(R)2012, C1:2009/(R)2012, A2:2010/(R)2012. EMC: IEC 60601-1-2. Connectivity: Wired or wireless (Radio Frequency Wireless Technology in Medical Device Guidance). Features: IPX6 waterproof detector, updated battery. Software/hardware verification and validation completed.

## Regulatory Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- AeroDR SYSTEMS ([K102349](/device/K102349.md))
- AeroDR SYSTEMS ([K113248](/device/K113248.md))
- AeroDR SYSTEMS ([K120477](/device/K120477.md))
- AeroDR SYSTEMS ([K130936](/device/K130936.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human profile.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 26, 2014

Konica Minolta, Inc. % Mr. Russell D. Munves Storch, Amini & Munves, PC 140 East 45th Street, 25th Floor NEW YORK NY 10017

Re: K141271 Trade/Device Name: AeroDR SYSTEM 2

Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MOB, LLZ Dated: August 8, 2014 Received: August 13, 2014

Dear Mr. Munves:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Smh.7)

for

Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K141271

Device Name

AeroDR SYSTEM 2

Indications for Use (Describe)

The AeroDR SYSTEM 2 is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen system in general-purpse diagnostic procedures.

The AeroDR SYSTEM 2 is not in mammography. fluoroscopy, tomography and angiography applications.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary

as required by 807.92

## 1. Company Identification

Konica Minolta, Inc. No.1 Sakura-machi, Hino-shi, Tokyo 191-8511, Japan Establishment Registration Number: 3004485675

## 2. Submitter's Name and Address

Shigeyuki Kojima Manager of Quality Assurance Operations Regulations and Standards Section, Quality Assurance Center No. 1 Sakura-machi, Hino-shi, Tokyo 191-8511, Japan Telephone: 81-42-589-8429 Fax: 81-42-589-8053

# 3. Date of Submission

May 9, 2014

#### 4. Device Trade Name AeroDR SYSTEM 2

## 5. Common Name

Digital Radiography

## 6. Classification

Class II

## 7. Product Code

Primary Product Code 90MQB: (Solid State X-ray Imager) Subsequent Product Code: 90LLZ (Picture archiving and communications system)

## 8. Corresponding regulation

21CFR 892.1680: Stationary x-ray system

## 9. Predicate Device

AeroDR SYSTEMS 510(k) number K102349, K113248, K120477, K130936

{4}------------------------------------------------

#### 10. Description of Device

The AeroDR SYSTEM 2 is a digital imaging system to be used with diagnostic x-ray systems. A new AeroDR Detector (flat panel digital detector: hereafter P-51) and AeroDR Generator Interface Unit2 has been just added to AeroDR SYSTEMS (The predicate devices:K102349, K113248, K120477, K130936) to function together such as with Console CS-7 (operator console), AeroDR Interface Unit, AeroDR Interface Unit2, AeroDR Generator Interface Unit, AeroDR Access Point and AeroDR Battery Charger, AeroDR Battery Charger2 and perform fundamentally same as Aero DR SYSTEMS do in physical and performance characteristics such as in device design, material safety and physical properties. Therefore, images captured with the flat panel digital detector in the AeroDR SYSTEM 2 can be communicated to the operator console via wired connection or wireless, depend on user's choice. The AeroDR SYSTEM 2 is just developed to meet user's compact layout needs without changing fundamental functions of the predicate devices.

AeroDR SYSTEM 2 is only connected with X-ray devices which are regally marketed in the United States of America and are compatible with XGIF, UEC, XIF Board along with certain electronic requirement, Specific signal controls for hardware and software and accessories described in Operation manual and Installation manual which is also fulfilled how to compatibility test at the time of installation also. In addition, for the use of pediatric, X-ray control system for pediatric are required.

## 11. Indications for Use

The AeroDR SYSTEM2 is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in general-purpose diagnostic procedures. The AeroDR SYSTEM 2 is not indicated for use in mammography, fluoroscopy, tomography and angiography applications.

## 12. Substantial Equivalence to Predicate Device

Although the new AeroDR Detector (flat panel digital detector whose material has been changed) and AeroDR Generator Interface Unit2 has been designed in the AeroDR SYSTEM 2 (the proposed device) to be used with diagnostic x-ray systems and with AeroDR SYTEMS' components (the predicate device). AeroDR SYSTEM 2 had been evaluated substantial equivalencies to the predicate device by the following points:

The Indications for use of proposed device and predicate devices are identical.

{5}------------------------------------------------

As a part of new outer materials of the new panel (p-51) which intact of human skin has been evaluated with the latest EN ISO 10993-1 and assured the safety as same as the predicate devices has. In addition, Inner material change for a righter weight, battery change for an attainment of its longer time/duration and IPX6 waterproof has been designed for the P-51 panel. So the electrical safety (AAMI / ANSI ES60601-1:2005/(R) 2012 and C1:2009/(R) 2012 and A2:2010/(R) 2012) and the electromagnetic compatibility testing (IEC 60601-1-2) had been conducted and assured as the predicate devices as well. In technical characteristics, hardware and software of the new panel and an accessory's verification and validation, (for the Wireless function, Radio Frequency Wireless Technology in Medical Device Guidance issued on August 14, 2013 was referred), Risk management based on ISO14971 had been completed without problem, performance testing (Bench Testing) including Non clinical and clinical testing referring to the FDA Guidance of the Submission of 510K (k')s for Solid State X-ray imaging Device had been concluded and showed equivalent evaluation outcome, which has supported a fact that no impacts in technological characteristic such as design, material chemical composition energy sauce and other factors of the proposed device were recognized. The all evaluation results can assure that there are no safety and effectiveness and performance issue or no differences were found in further than the predicate devices have which have been regally marketed in the United Sates. Therefore, we confirmed that AeroDR SYSTEM 2 has the same substantial equivalency to the predicate device, AeroDR SYSTEMs have.

#### 13. Conclusion

Comprehensively, we judged that the AeroDR SYSTEM 2 has fundamentally the same technological characteristics as the predicate devices have. Therefore, this 510(k) has been able to conclude as substantial equivalence as the predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K141271](https://fda.innolitics.com/device/K141271)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
