K141256 · Okamoto USA, Inc. · HIS · Jul 23, 2014 · Obstetrics/Gynecology
Device Facts
Record ID
K141256
Device Name
OKAMOTO STUDDED CONDOM
Applicant
Okamoto USA, Inc.
Product Code
HIS · Obstetrics/Gynecology
Decision Date
Jul 23, 2014
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 884.5300
Device Class
Class 2
Indications for Use
This Okamoto Studded condom is used for contraceptive and prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).
Device Story
Okamoto Studded Condom is a natural rubber latex sheath designed to cover the penis. Features include a tapered shape, reservoir tip, and an exterior dotted texture created during the molding process using an engraved glass mold. The device is lubricated with silicone oil and uses cornstarch/MgO as a dusting agent. It is intended for over-the-counter use by individuals for contraception and the prevention of sexually transmitted infections. The device functions as a physical barrier. Safety and effectiveness are supported by conformance to ASTM D3492:2008 standards and biocompatibility testing (ISO 10993) for cytotoxicity, irritation, sensitization, and acute systemic toxicity.
Clinical Evidence
Bench testing only. Conformance to ASTM D3492:2008 for air burst pressure (≥ 1 kPa) and volume (≥ 17 dm3). Biocompatibility testing per ISO 10993 (cytotoxicity, irritation, sensitization, acute systemic toxicity) confirmed the device is non-toxic, non-sensitizing, and non-irritating. Stability studies support a 5-year shelf life.
Technological Characteristics
Natural rubber latex sheath; tapered shape with reservoir tip; dotted surface texture; silicone lubricant (200 cps); cornstarch and MgO dusting agent. Nominal length 180 mm; nominal flat width 54 mm; thickness 0.070 mm. Conforms to ASTM D3492:2008. No electronic components or software.
Indications for Use
Indicated for contraceptive and prophylactic use to prevent pregnancy and sexually transmitted infections in the general population.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
{0}------------------------------------------------
K141256 page 1 of 3
# JUL 2 3 2014
#### II. 510(k) SUMMARY
| Submitted by: | OKAMOTO U.S.A., INC.<br>3130 West Monroe Street<br>Sandusky, OH 44870<br>Phone: 419-626-1633 |
|----------------------|-------------------------------------------------------------------------------------------------|
| Contact Person: | Joseph W. Cormier, JD, PhD / Hyman, Phelps & McNamara<br>Mr. Hirofumi Chiba, Okamoto USA., Inc. |
| Date Prepared: | July 21, 2014 |
| Proprietary Name: | Okamoto Studded Condom |
| Common Name: | Male Latex Condom |
| Classification Name: | Condom (21 C.F.R. § 884.5300) |
| Predicate Device: | |
| Brand Name: | BillyBoy Dotted (Beaded) condom |
| 1) | |
Company Name: MAPA GmbH
510(k) Document Control Number: K111093
### Description of the Device:
This condom is made of a natural rubber latex sheath, which completely covers the penis with a closely fitted membrane. This condom is textured surface (dotted), tapered shape with reservoir tip, silicone lubricated condom with nominal length 180 mm, nominal flat width 54 mm measured at 30 mm from the open end, a maximum width of 58 mm at the closed end, and nominal thickness of 0.070 mm. The air burst pressure is ≥ 1 kPa and air burst volume is ≥ 17 dm3. It is lubricated with silicone (viscosity 200 cps) and cornstarch is used as a dressing This condom conforms to current established national standard ASTM D3492: 2008. material.
The term Studded refers to the irregular surface of the irregular surface is not applied externally but is achieved in the molding process. The glass mold has an engraved pattern of semispherical pits replicated as a pattern of small round tips on the surface of the
{1}------------------------------------------------
This results in an exterior surface that has a pattern of small raised areas best condom. described as a dotted pattern. The new condom is defined as a condom of increased thickness relative to some of other Okamoto condoms (e.g., Crown condom, K893039),
### Intended Use of the Device:
This Okamoto Studded condom is used for contraceptive and prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).
#### Technological Characteristics:
The subject condom has nearly identical technological characteristics to the predicate condom. It is identical in terms of additives and manufacturing process. The subject and predicate condoms have the same indications for use. As indicated in the table below, the only differences are the shape, nominal width and length, the quantity of lubrication, and the latex formulation. These differences do not affect the safety or effectiveness or affect performance of the subject condom as compared to the predicate. Testing on the subject condom demonstrates conformance with ISO 10993, Biological Evaluation of Medical Devices for a device in contact for 24 hours or less for cytotoxicity, irritation and sensitization, acute systemic toxicity testing, and sample preparation and reference material. Testing shows these condoms are non-toxic, non-sensitizing, and non-irritating. Stability studies conducted on the similar condoms established a shelf life of the device at 5 years. The differences in dimension, latex formulation, and quantity of lubrication do not affect cytotoxicity, irritation, sensitization, or shelf life. Furthermore, because the differences in characteristics are common for natural rubber latex male condoms, the differences do not raise new questions of safety or effectiveness.
The results of biocompatibility testing for cytotoxicity, irritation and sensitization, and acute systemic toxicity as well as airburst testing demonstrate that the subject device is substantially equivalent to the predicate device in term of safety and effectiveness.
The similarities and differences of the features and technological characteristics of the condom as compared to the predicate condom are shown in the table below.
{2}------------------------------------------------
.
| | | Table 1- Characteristics of Submission Device and Predicate Device |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|
| | Submission device:<br>Okamoto Studded Male Latex<br>Condom | Predicate device:<br>Billy Boy Male Latex Condom<br>K111093 |
| FDA Classification | Class II; 884.5300 | Identical |
| Classification code | HIS | Identical |
| Intended Use | For contraceptive and<br>prophylactic purposes (to help<br>prevent pregnancy and the<br>transmission of sexually<br>transmitted infections). | Identical |
| Material | Natural rubber latex | Identical |
| Performance | Complied with ASTM<br>D3492:2008 | Identical |
| Type | Dotted | Dotted (beaded) |
| Shape | Tapered shape with reservoir tip;<br>dotted | Straight walled with reservoir<br>tip; cylindrical; dotted/beaded |
| Length | 170-190 mm | > 175 mm |
| Width | 54 ± 2 mm measured 30 mm<br>from open end,<br>Maximum width of 58 ± 2 mm<br>at the closed end | 52 mm |
| Thickness | 0.070 ± 0.010 mm | 0.100 ± 0.010 mm |
| Lubricant | 0.550 g silicone oil | 0.49 g silicone oil |
| Color | None | Identical |
| Scent | None | Identical |
| Latex Formulation | Natural Rubber Latex | Comparable |
| Dusting Agent | Cornstarch and MgO | Identical |
.
.
and the comments of the comments of
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake entwined around it, and three horizontal bars above it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 23, 2014
Okamoto USA, Inc. % Joseph W. Cormier Law Offices of Hyman, Phelps & McNamara, P.C. 700 Thirteenth Street N.W., Suite 1200 Washington, DC 20005-5929
Re: K141256 Trade Name: Okamoto Studded Condom Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: HIS Dated: May 14, 2014 Received: May 14, 2014
Dear Joseph W. Cormier,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of general bonders profisions practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 - Joseph W. Cormier
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301)796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Image /page/4/Picture/8 description: The image shows the name "Benjamin R Fisher -S" in a bold, sans-serif font. The letters are black against a white background, making them easily readable. The text appears to be a signature or a nameplate.
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K 141256
Device Name Okamoto Studded Condom
Indications for Use (Describe)
The Okamoto Studded Condom is used for contraceptive and prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
2 Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY -------------------------------------------------------------------------------------------------------------------------------------------------------------
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
11
. : ... .
# Benjamin R. Fisher -S 2014.07.23 15:57:22 -04'00'
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the The builden fine for this oclicens of mornizatia sources, gather and maintain the data needed and complete line to review the collection. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.