POLARIS PRESSURE ADJUSTABLE VALVE SYSTEM/POLARIS ADJUSTABLE VALVE WITH SIPHONX
Applicant
Sophysa SA
Product Code
JXG · Neurology
Decision Date
Apr 2, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5550
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Polaris® Adjustable Pressure Valve is designed for the treatment of hydrocephalus by shunting the Cerebrospinal Fluid (CSF) to the abdominal cavity or right atrium of the heart.
Device Story
The Polaris Pressure Adjustable Valve System is a cerebrospinal fluid (CSF) shunt system used for the treatment of hydrocephalus. The device functions by shunting excess CSF from the ventricles to the abdominal cavity or right atrium. It features an adjustable pressure valve mechanism, allowing clinicians to modify the valve's opening pressure settings non-invasively to manage patient-specific drainage requirements. The system includes the Polaris Adjustable Valve, the SiphonX anti-siphon device, and an adjustment kit. It is intended for use by neurosurgeons in a clinical or surgical setting. The device helps manage intracranial pressure by regulating CSF flow, thereby mitigating symptoms associated with hydrocephalus.
Clinical Evidence
No clinical data provided; substantial equivalence is based on bench testing and design characteristics.
Technological Characteristics
The system consists of an adjustable pressure valve and an anti-siphon device (SiphonX). It is a mechanical CNS fluid shunt system. No software, electronics, or energy sources are utilized. Materials are standard for implantable shunts. Sterilization is performed per standard medical device manufacturing practices.
Indications for Use
Indicated for the treatment of hydrocephalus in patients requiring CSF shunting to the abdominal cavity or right atrium.
Regulatory Classification
Identification
A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.
Related Devices
K122118 — HAKIM PROGRAMMABLE AND PRECISION VALVE SHUNT SYSTEM · Codman & Shurtleff, Inc. · Aug 3, 2012
K212641 — StrataMR II Valves and Shunts · Medtronic Neurosurgery · Sep 16, 2021
K031097 — POLARIS PRESSURE ADJUSTABLE VALVE SYSTEM MODES SPV, SPVA, SPVB · Sophysa SA · Feb 26, 2004
K141687 — AESCULAP - MEITHKE PROGAV 2.0 ADJUSTABLE SHUNT SYSTEM · Aesculap, Inc. · Apr 2, 2015
K053350 — MODIFICATION TO CODMAN HAKIM SHUNT SYSTEMS · Codman & Shurtleff, Inc. · Jan 19, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human figure in profile, composed of three overlapping silhouettes. The silhouettes are arranged in a way that suggests a sense of community and support. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 2, 2015
Sophysa SA Mr. Gustavo Kobrin Regulatory Consultant 25108 Marguerite Parkway, Suite A-354 Mission Viejo, California 92692
Re: K141227
Trade/Device Name: Polaris Pressure Adiustable Valve System / Polaris Adjustable Valve with SiphonX / Polaris Adjustment Kit Regulation Number: 21 CFR 882.5550 Regulation Name: Central Nervous System Fluid Shunt and Components Regulatory Class: Class II Product Code: JXG Dated: February 4, 2015 Received: February 5, 2015
Dear Mr. Gustavo Kobrin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -550/Δ
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K141227
Device Name
Polaris Pressure Adjustable Valve System / Polaris Adjustable Valve with SiphonX / Polaris Adjustment Kit
Indications for Use (Describe)
The Polaris® Adjustable Pressure Valve is designed for the treatment of hydrocephalus by shunting the Cerebrospinal Fluid (CSF) to the abdominal cavity or right atrium of the heart.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches