The Alliance Spine RAAS Cranial Plating System is indicated for osteotomies, cranial fractures, and reconstruction of non-load bearing cranial areas.
Device Story
The RAAS Cranial Plating System provides internal fixation for cranial fractures and bone flaps, and stabilization for burr holes and cranial defects. The system comprises malleable low-profile plates (straight, box, gap), burr hole covers, malleable mesh, and fasteners (screws). Components are used by surgeons in clinical settings to secure bone segments. The device is mechanical; it does not involve software or algorithms. Benefits include stable fixation and reconstruction of non-load bearing cranial areas.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including static axial compression, insertion/removal torque, ultimate torque, and axial pullout testing per ASTM standards.
Technological Characteristics
Materials: Plates/Screws (ASTM F136 Ti alloy), Mesh (ASTM F67 CP Ti). Surface treatment: Type III Anodized mesh and screws. Dimensions: Plates/covers 0.5mm thick; mesh 0.3mm/0.6mm thick. Screws: 1.5mm self-drilling/tapping and 1.7mm rescue screws. Sterilization: End-user steam sterilization.
Indications for Use
Indicated for patients requiring osteotomies, craniotomy, cranial fractures, and reconstruction of non-load bearing cranial areas.
Regulatory Classification
Identification
A preformed alterable cranioplasty plate is a device that is implanted into a patient to repair a skull defect. It is constructed of a material, e.g., tantalum, that can be altered or reshaped at the time of surgery without changing the chemical behavior of the material.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of two main elements: a circular text element and a stylized symbol. The text element reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion. The symbol is a stylized representation of a human figure, with three profiles facing right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 27, 2015
Alliance Partners, LLC % Mr. Kellen Hills Quality & Regulatory Consultant Orchid Design. A Division of Orchid Orthopedic Solutions 4600 East Shelby Drive, Suite 1 Memphis. TN 38118
Re: K141123
> Trade/Device Name: Alliance Spine RAAS Cranial Plating System Regulation Number: 21 CFR 882.5320 Regulation Name: Preformed Alterable Cranioplasty Plate Regulatory Class: Class II Product Code: GWO, GXR, HBW Dated: February 20, 2015 Received: February 25, 2015
Dear Mr. Kellen Hills,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Carlos L. Pena -S
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141123
Device Name RAAS Cranial Plating System
Indications for Use (Describe)
The Alliance Spine RAAS Cranial Plating System is indicated for osteotomies, cranial fractures, and reconstruction of non-load bearing cranial areas.
Type of Use (Select one or both, as applicable)
| <label><input checked="true" type="checkbox"/> For Activities Under CEE 301.5(c)</label> | <label><input type="checkbox"/> For Site Controls Under CEE 301.5(d)</label> |
|------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|
X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| (a)(1) | Submitted By: | Alliance Partners, LLC<br>(dba Alliance Spine™)<br>14206 Northbrook Dr.<br>San Antonio, TX 78232 | |
|-------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Date: | February 20, 2015 | |
| | Contact Person: | Tim M. Lohnes<br>Senior Regulatory Consultant<br>Orchid Design | |
| | Phone:<br>Fax: | 203-922-0105<br>203-922-0130 | |
| (a)(2) | Proprietary Name: | RAAS Cranial Plating System | |
| | Common Name: | Bone Plates, Bone Fixation Fasteners | |
| | Classification Names and References: | 882.5320 - Preformed alterable<br>cranioplasty plate<br>882.5250 - Burr hole cover<br>882.3560 - Cranioplasty plate fastener | |
| | Product Codes: | GWO, GXR, HBW | |
| (a)(3) | Predicate Devices: | Stryker-NEURO II (K031659)<br>OsteoMed-Low Profile Neuro (K111176) | |
| | SUBJECT DEVICE | PREDICATE DEVICES | |
| Manufacturer | Alliance Spine | Stryker | OsteoMed |
| System Name | RAAS Cranial<br>Plating System | Neuro 2 Cranial Plate<br>System | Low Profile Neuro<br>Fixation System |
| 510(k) | Subject | K031659 | K111176 |
| Product Codes | GWO, GXR, HBW | JEY, HWC | GWO, GXR |
| Indications for Use | The RAAS Cranial<br>Plating System is<br>indicated for<br>osteotomy,<br>craniotomy,<br>cranial fractures,<br>and<br>reconstruction of<br>non-load bearing<br>cranial areas. | The Stryker Leibinger<br>Universal Neuro<br>System is a low-<br>profile plate and<br>screw system<br>intended for<br>osteotomy,<br>craniotomy,<br>stabilization and rigid<br>fixation of craniofacial<br>fractures and<br>reconstruction of<br>non-load bearing<br>areas. | The OsteoMed<br>LowProfile Neuro<br>Fixation System is<br>indicated for use in<br>ostemtomies,<br>fractures or<br>reconstruction of<br>cranial bones.<br>Implants and drills<br>are single use only. |
| Sterilization | End User (steam) | Non-sterile and sterile | Not specified |
| Shelf Life | n/a | Not specified | Not specified |
| Implant Material: | | | |
| Plates and<br>Burr Hole Covers - | ASTM F136 | CP Ti | ASTM F136/F67 |
| Mesh – | ASTM F67 | CP Ti | ASTM F136/F67 |
| Screws - | ASTM F136 | Ti Alloy | ASTM F136 |
| Thickness: | | | |
| Plates & Burr Hole<br>Covers - | 0.5mm | 0.6 and 0.5mm | 0.25 to 1.0mm |
| Mesh - | 0.3 and 0.6mm | 0.1, 0.2, 0.3 and<br>0.6mm | 0.25 to 1.0mm |
| Surface<br>Treatment:<br>Plates & Burr Hole<br>Covers | None | Type III Anodized | Not specified |
| Surface<br>Treatment: Mesh | Type III Anodized | Type III Anodized | Not specified |
| Screws - type | Self-tapping,<br>drilling, and<br>rescue | Self-tapping, drilling,<br>and rescue | "Auto Drive" and<br>rescue |
| Screws:<br>diameter/length | 1.5 x 4 & 5mm,<br>1.7 x 5 & 6mm | 1.5mm x 4, 5 & 6mm<br>1.7mm x 4mm | 1.6mm x 4, 5, 6 & 8mm, 1.9mm x 4, 5, 6 & 8mm |
| Surface<br>Treatment: Screws | Type III Anodized | Type III Anodized | Not specified |
### Device Description: (a)(4)
The Alliance Spine RAAS Cranial Plating System consists of various sizes and geometries of malleable low profile cranial plates including straight, box, and gap plates, burr hole covers, malleable mesh as well as fasteners (screws) that provide stable internal fixation of cranial fractures and bone flaps, as well as coverage and stabilization of burr holes and other cranial defects.
The RAAS plates and burr hole covers are 0.5mm thick, while the RAAS malleable Mesh is supplied in 0.3mm and 0.6mm thickness. The fasteners (screws) are supplied in 1.5mm diameter Self-Drilling and 1.5mm diameter Self-Tapping configurations in 4mm and 5mm lengths, while the RAAS Rescue Screws are 1.7mm in diameter in 5mm and 6mm lengths. The Screws and Mesh are color anodized for identification.
{4}------------------------------------------------
The Alliance Spine™ RAAS Cranial Plating System is indicated for osteotomies, craniotomy, cranial fractures, and reconstruction of non-load bearing cranial areas.
- Technological Characteristics: (a)(6)
{5}------------------------------------------------
### (b)(1) Non-clinical testing:
| Test | Test Method Summary | Results |
|--------------------------|---------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Static Axial Compression | Worst case Burr Hole plates<br>were subjected to loading as<br>described in ASTM F382. | Peak load and stiffness values<br>were acceptable based on<br>intended use and predicate data.<br>Subject device expected to<br>perform similarly to predicate. |
| Insertion/Removal Torque | Worst case screws were<br>inserted into and then<br>removed from a test block as<br>described in ASTM F543. | Driving torque values were<br>acceptable based on intended<br>use and predicate data. Subject<br>device expected to perform<br>similarly to predicate. |
| Ultimate Torque | Worst case screws were fixed<br>below the head and subjected<br>to loading as described in<br>ASTM F543. | Peak torque values were<br>acceptable based on intended<br>use and predicate data. Subject<br>device expected to perform<br>similarly to predicate. |
| Axial Pullout | Worst case screws were<br>inserted to a fixed depth and<br>subjected to loading as<br>described in ASTM F543. | Peak load values were acceptable<br>based on intended use and<br>predicate data. Subject device<br>expected to perform similarly to<br>predicate. |
## Clinical testing: (b)(2)
Clinical testing was not required to demonstrate substantial equivalence in this premarket notification.
## (b)(3) Conclusions:
Based on the information provided in this premarket notification, we believe that the subject RAAS Cranial Plating System demonstrates substantial equivalence and has a safety and effectiveness profile that is similar to the identified predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.