K141093 · Ge Vingmed Ultrasound AS · ITX · May 15, 2014 · Radiology
Device Facts
Record ID
K141093
Device Name
6VT-D ULTRASOUND TRANSDUCER
Applicant
Ge Vingmed Ultrasound AS
Product Code
ITX · Radiology
Decision Date
May 15, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1570
Device Class
Class 2
Indications for Use
The device is intended for use by a qualified physician for use with GE Diagnostic Ultrasound Systems for ultrasound evaluation of Cardiac anatomy of adults in 2D modes as well as in real-time 3D / 4D mode.
Device Story
6VT-D is a transesophageal ultrasound transducer for adult cardiac imaging; connects to GE ultrasound systems. Device functions as passive transducer; does not control energy delivery; contains no software. Modification involves replacement of one patient-contact material in endoscope. Used by physicians in clinical settings for diagnostic imaging. Output displayed on GE ultrasound system for clinical decision-making regarding cardiac anatomy. Benefits include non-invasive visualization of cardiac structures.
Clinical Evidence
No clinical studies required. Substantial equivalence supported by bench testing, including acoustic output, biocompatibility, cleaning/disinfection effectiveness, and thermal/electrical/mechanical safety per IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, and NEMA UD 2/UD 3 standards.
Technological Characteristics
Transesophageal ultrasound transducer. Materials evaluated per ISO 10993-1. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, NEMA UD 2, and NEMA UD 3. No software or internal energy control circuitry. Standalone transducer component for GE ultrasound systems.
Indications for Use
Indicated for ultrasound evaluation of adult cardiac anatomy via transesophageal access. Supports 2D, 3D, and 4D imaging modes.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
Predicate Devices
GE Vivid E9 Diagnostic Ultrasound System including 6VT-D transducer (K131514)
{0}------------------------------------------------
MAY 1 5 2014
K141093
Page 1 of 3
GE Healthcare 510(k) Premarket Notification Submission
## 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
April 25, 2014 Date:
Submitter: GE Healthcare, GE Vingmed Ultrasound AS Strandpromenaden 45 N-3191. Horten. Norway
6VT-D Ultrasound Transducer
6VT-D Ultrasound Transducer
Bryan Behn Contact Person:
Regulatory Affairs Manager GE Healthcare. GE Medical Systems Ultrasound and Primary Care Diagnostics, LLC. Phone: 414-721-4214 Fax: 414-918-8275
Trade Name: Device: Common/Usual Name:
Diagnostic Ultrasound Transducer, 21 CFR 892.1570 Classification Names:
Product Code: 90-ITX
Predicate Device(s):
K131514 GE Vivid E9 Diagnostic Ultrasound System including 6VT-D transducer.
K060907 Karl Storz Video Gastroscope System
Device Description:
The 6VT-D is an ultrasound-imaging device that is attached to a GE ultrasound imaging system and used for diagnostic imaging. 6VT-D does not directly control energy delivered to the patient nor does it contain any software. The 6VT-D is a transducer for imaging cardiac anatomy in adults with transesophageal means of access.
The objective of this 510(k) submission is to present the proposed modifications to the 6VT-D Transducer (K131514), and to demonstrate substantial equivalence. The 6VT-D transducer modification is an incremental improvement where one patient contact material of the endoscope is replaced by another. Both the current endoscope material, and the proposed new material, were evaluated for biocompatibility according to the same international standards, and both were found to be acceptable for the transducer's intended use and relevant patient contact. No design inputs are affected by the change of material.
Image /page/0/Picture/18 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular shape. The logo is black and white and appears to be a vector graphic.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The letters and the border are black, while the background is white.
# GE Healthcare
510(k) Premarket Notification Submission
Intended Use:
Determination of
Substantial Equivalence:
The device is intended for use by a qualified physician for use with GE Diagnostic Ultrasound Systems for ultrasound evaluation of Cardiac anatomy of adults in 2D modes as well as in real-time 3D / 4D mode.
The 6VT-D Transducer employs the same fundamental scientific Technology: technology as its predicate device(s).
# Comparison to Predicate Devices
The 6VT-D Diagnostic Ultrasound Transducer is substantially equivalent to the predicate devices with regard to intended use, imaging capabilities, technological characteristics and safety and effectiveness.
- The 6VT-D is identical to the predicate 6VT-D transducer . with respect to imaging and applications.
- . The transducers are manufactured with materials which have been evaluated and found to be safe for the intended use of the device.
- The endoscope of 6VT-D uses identical patient contact . material as the Karl Storz Video Gastroscope System (K060907)
## Summary of Non-Clinical Tests:
The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic and mechanical safety, and has been found to conform with applicable medical device safety standards.
The 6VT--D Transducer and its applications comply with voluntary standards:
- 1. IEC60601-1, Medical Electrical Equipment Part 1: General Requirements for Safety
- 2. IEC60601-1-2,Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility Requirements and Tests
- 3. IEC60601-2-37, Medical Electrical Equipment Part 2-37: Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment
- 4. NEMA UD 3. Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment when connected to a GE Ultrasound System
- 5. ISO10993-1. Biological Evaluation of Medical Devices-
ﺗﺮ ﺗ
{2}------------------------------------------------
K141093
Page 3 of 3
GE Healthcare
510(k) Premarket Notification Submission Part 1: Evaluation and Testing
- 6. NEMA UD 2, Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment when connected to a GE Ultrasound System
- 7. ISO14971, Application of risk management to medical devices
The following quality assurance measures were applied to the development of the complete medical system:
- Risk Analysis .
- Requirements Reviews .
- Design Reviews .
- Testing on unit level (Module verification) .
- . Integration testing (System verification)
- Performance testing (Verification) .
- Safety testing (Verification) .
- Final Acceptance testing (Validation) .
## Summary of Clinical Tests:
The subject of this premarket submission, 6VT-D Transducer, did not require clinical studies to support substantial equivalence. 100
GE Healthcare considers the 6VT-D Transducer to be as safe, as Conclusion: effective, and performance is substantially equivalent to the predicate device(s).
Image /page/2/Picture/17 description: The image contains the number 22. The number is written in a simple, sans-serif font. The number is black and is set against a white background.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 15, 2014
GE VINGMED ULTRASOUND AS % Mr. Bryan Behn Regulatory Affairs Manager GE Healthcare 9900 W. Innovation Drive, RP-2138 WAUWATOSA WI 53226
Re: K141093
Trade/Device Name: 6VT-D Diagnostic Ultrasound Transducer Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: ITX Dated: April 25, 2014 Received: April 28, 2014
Dear Mr. Behn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Image /page/3/Picture/11 description: The image shows a circular seal with text around the perimeter and a symbol in the center. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The symbol in the center appears to be an eagle or similar bird with outstretched wings, stylized with thick, curved lines.
{4}------------------------------------------------
Page 2-Mr. Behn
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misoranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.

for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K141093
### Device Name
6VT-D Diagnostic Ultrasound Transducer
#### Indications for Use (Describe)
6VT-D is intended for use by a qualified physician for use with GE Diagnostic Ultrasound Systems for ultrasound evaluation of Cardiac anatomy of adults through Transesophageal means of access, having imaging capabilities in 2D as well as in real-time 3D / 4D mode.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017 See PRA Statement below.
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Smh.7)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The letters and the border are in black, contrasting with the white background. The logo is a classic and recognizable symbol of the General Electric company.
# GE Healthcare 510(k) Premarket Notification Submission
### Diagnostic Ultrasound Indications for Use Form
## 6VT-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|-----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|---------------|
| Clinical Application<br>Anatomy/ Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | RT3D<br>Mode* |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | P | P | P | P | P | P | P | P | P | P | P |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other (specify) | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | P | P | P | P | P | P | P | P | P | P | P |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(*) RT3D is Realtime 3D / 4D volume tissue scan acquisition (with or w/o color flow);
(PLEASE DO NOT WRITE BELOW THIS LINE · CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription User (Per 21 CFR 801.109)
.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.