← Product Code [HSB](/productcode/HSB) · K141023

# PRECICE INTRAMEDULLARY LIMB LENGTHENING SYSTEM (K141023)

_Ellipse Technologies, Inc. · HSB · May 20, 2014 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K141023

## Device Facts

- **Applicant:** Ellipse Technologies, Inc.
- **Product Code:** [HSB](/productcode/HSB.md)
- **Decision Date:** May 20, 2014
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3020
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Ellipse PRECICE System is indicated for limb lengthening of the tibia and femur.

## Device Story

The PRECICE Intramedullary Limb Lengthening System is an orthopedic implant used for lengthening the tibia or femur. The system consists of an intramedullary nail, locking screws, surgical instruments, and an external remote controller (ERC). The nail contains an internal rare earth magnet, a telescoping lead screw/nut assembly, and planetary gearing. During surgery, the nail is inserted into the intramedullary canal and secured with locking screws. Post-operatively, the patient or clinician uses the ERC to non-invasively adjust the nail length via magnetic coupling. This controlled, incremental lengthening benefits patients requiring limb reconstruction or correction of limb length discrepancies. The device is used in a clinical or home setting under physician guidance.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing included static and dynamic four-point bend testing and static torque-to-failure testing per ASTM F1264-03, as well as magnet holding torque and device functionality/verification testing.

## Technological Characteristics

Intramedullary fixation rod; rare earth magnet, telescoping lead screw/nut assembly, planetary gearing. Materials: standard orthopedic implant materials. Standards: ASTM F1264-03 for mechanical testing. Sterilization: Gamma radiation for nail; non-sterile screws/accessories. Connectivity: External remote controller (ERC) for non-invasive magnetic adjustment.

## Regulatory Identification

An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

## Predicate Devices

- Ellipse PRECICE System ([K131677](/device/K131677.md))
- Ellipse PRECICE System ([K133289](/device/K133289.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### MAY 2 0 2014

Ellipse Technologies, Inc. PRECICE Intramedullary Limb Lengthening System Special 510(k) Application

April 2014 Product Code: HSB

## PRECICE® Intramedullary Limb Lengthening System 510(k) Summary - K TBD April 2014

1. Company: Ellipse Technologies. Incorporated 13900 Alton Parkway, Suite 123 Irvine, CA 92618

> Rebecca Shelburne Contact: Regulatory Affairs Specialist Phone: (949) 837-3600 x227 Fax: (949) 837-3664 Date Summary Prepared: April 17, 2014

Proprietary Trade Name: PRECICE Intramedullary Limb Lengthening System 2.

Common Name: Intramedullary Nail 3.

4. Classification Name: Intramedullary Fixation Rod (21 CFR 888.3020)

న. Product Code: HSB (Rod, Fixation, Intramedullary and Accessories)

Product Description: The Ellipse PRECICE System is composed of the PRECICE nail 6. (supplied sterile), locking screws, surgical instruments and an external remote controller (ERC). The nail is available in tibia or femur models with various diameters. Iengths and screwhole configurations to accommodate a variety of patient anatomies. The locking screws are also available in a variety of diameters and lengths. The PRECICE nail is supplied sterile by gamma radiation while the locking screws and PRECICE specific accessories are supplied non-sterile and must be sterilized prior to use. The nail contains an enclosed rare earth magnet. telescoping lead screw/nut assembly, and planetary gearing.

7. Indications: The Ellipse PRECICE System is indicated for limb lengthening of the tibia and femur.

Substantial equivalence: Documentation that includes mechanical test results and 8. detailed comparison to the predicate devices demonstrates that the Ellipse PRECICE System is substantially equivalent to the following 510(k) cleared device:

Ellipse PRECICE System (K131677 and K133289) .

Substantial equivalence is based on similar indications for use, designs, and on in vitro testing performed. The Risk Management file was updated to include the design modifications and no new risks were identified.

The modified Ellipse PRECICE System and the predicate device have the same intended use. Specifically, to lengthen the femur or tibia. The modified PRECICE nail and the predicate are available in the same application, screwhole configurations, stroke lengths, and overall lengths. The PRECICE Nail has the same patient contacting materials, technological characteristics and principles of operation as that of the predicate. Both devices are inserted into the intramedullary

{1}------------------------------------------------

canal of the femur or tibia and secured with locking screws. Both devices are adjusted noninvasively by the Ellipse external remote controller (ERC). The differences between the modified PRECICE System and the predicate device are as follows:

- The magnet stabilizer feature is being modified and re-located to the proximal end of the . housing tube for ease of manufacturing, and
- The anti-rotation lug feature is being modified to be press-fit into the internal diameter of . the housing tube, eliminating the tack welding process.

Data relied upon to determine substantial equivalence of the PRECICE System with the device modifications described in this submission to the cleared PRECICE System include the following:

- . Mechanical testing
- Design functionality and verification testing .

Conclusions can be drawn from the tests that the modifications to the PRECICE System are safe and effective and meet the performance specifications. The following specific tests have been performed in order to establish equivalence to the predicate devices:

| Test Description                      | Applicable Test Standard |
|---------------------------------------|--------------------------|
| Static Four Point Bend                | ASTM F1264-03            |
| Dynamic Four Point Bend               | ASTM F1264-03            |
| Static Torque to Failure              | ASTM F1264-03            |
| Magnet Holding Torque                 | None                     |
| Device functionality and verification | None                     |

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three tail feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 20, 2014

Ellipse Technologies, Incorporated Ms. Rebecca Shelburne Regulatory Affairs Specialist 13900 Alton Parkway, Suite 123 Irvine. California 92618

Re: K141023

Trade/Device Name: Ellipse PRECICE® System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: II Product Code: HSB Dated: April 17, 2014 Received: April 21. 2014

Dear Ms. Shelburne:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please notc: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

### Page 2 - Ms. Rebecca Shelburne

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

## Lori A. Wiggins

- for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

#### 510(k) Number (if known) K141023

Device Name Ellipse PRECICE® System

Indications for Use (Describe)

The Ellipse PRECICE System is indicated for limb lengthening of the tibia and femur.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

> Casey antev-Division of Orthopedle D vices

Page 1/1

E

---

**Source:** [https://fda.innolitics.com/device/K141023](https://fda.innolitics.com/device/K141023)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
