← Product Code [DWF](/productcode/DWF) · K140999

# TANDEMHEART VENO-VENOUS CANNULA (K140999)

_Cardiacassist, Inc. · DWF · Aug 6, 2014 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K140999

## Device Facts

- **Applicant:** Cardiacassist, Inc.
- **Product Code:** [DWF](/productcode/DWF.md)
- **Decision Date:** Aug 6, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4210
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The TandemHeart Veno-Venous Cannula Set is intended for use as a single cannula for both venous drainage and reinfusion of blood via an internal jugular vein during extracorporeal life support procedures.

## Device Story

TandemHeart Veno-Venous Cannula Set facilitates extracorporeal life support; provides single-cannula venous drainage and reinfusion via internal jugular vein. Device comprises 29 Fr dual-lumen wire-reinforced cannula and 13 Fr introducer. Introducer enables placement over 0.038-inch guidewire under fluoroscopy. Cannula features co-axial lumen design; distal section has side holes; proximal sections allow visualization and clamping. Barbed connectors link to standard 3/8-inch blood circuit tubing. Suture wing secures device to patient. Used in clinical settings by healthcare providers. Tapered design aids vessel dilation during insertion. Output is blood flow path for extracorporeal support; enables patient stabilization during life support procedures.

## Clinical Evidence

Bench testing only. No clinical data. Performance testing included in-vitro system capacity and flow vs. pressure drop (HQ) testing to compare flow characteristics against the Avalon Elite predicate. Additional testing included flexibility, strength, biocompatibility, in-vitro hemolysis, leak testing, sterilization, and shelf life.

## Technological Characteristics

Wire-reinforced dual-lumen co-axial cannula; 29 Fr outer diameter; 13 Fr introducer. Features tantalum markers and radiopaque introducer for fluoroscopic visualization. Includes suture wing and barbed connectors for 3/8-inch tubing. Single-use, sterile. No software or electronic components.

## Regulatory Identification

A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.

## Predicate Devices

- Avalon (Maquet) Elite Bi-Caval Dual Lumen Catheter ([K081820](/device/K081820.md))
- TandemHeart Venous Cannula ([K133236](/device/K133236.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 6, 2014

CardiacAssist, Inc. Greg Johnson Director of Regulatory Affairs 240 Alpha Drive Pittsburgh, PA 15238

Re: K140999

Trade/Device Name: TandemHeart Veno-Venous Cannula Set Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary bypass vascular catheter, cannula or tubing Regulatory Class: Class II Product Code: DWF Dated: May 15, 2014 Received: May 16, 2014

Dear Mr. Johnson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Melissa A. Torres -S

For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

#### K140999

Device Name TandemHeart Veno-Venous Cannula

Indications for Use (Describe)

The TandemHeart Veno-Venous Cannula Set is intended for use as a single cannula for both venous drainage and reinfusion of blood via an internal jugular vein during extracorporeal life support procedures.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

# Melissa A. Torres -S

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

# Section 5 510(k) Summary

Date: 7/22/2014

## Applicant

CardiacAssist, Inc. 240 Alpha Drive Pittsburgh, PA 15238 Telephone: 412-963-7770 Fax: 412-963-0800

#### Contact person

Greg Johnson, PhD Title: Director of Regulatory Affairs Phone: 412-963-7770 x266 e-mail: gjohnson@tandemheart.com

## Device

| Trade/Proprietary Name: | TandemHeart Veno-Venous Cannula Set                                                                  |
|-------------------------|------------------------------------------------------------------------------------------------------|
| Common Name:            | Veno-Venous Cannula and Introducer                                                                   |
| Classification Name:    | Cardiopulmonary bypass vascular catheter, cannula, or<br>tubing. (21 CFR 870.4210, Product Code DWF) |

#### Predicate Devices

Avalon (Maquet) Elite Bi-Caval Dual Lumen Catheter (K081820) TandemHeart Venous Cannula (K133236)

#### Device Description

The TandemHeart Veno-Venous Cannula Set consists of two components: a 29 Fr. Dual lumen Veno-Venous Cannula and a 13 Fr. Introducer. The Introducer is designed to accept a standard 0.038 inch guidewire. The TandemHeart Veno-Venous Cannula Set is intended as a single patient, single use, sterile device.

{4}------------------------------------------------

Image /page/4/Figure/0 description: The image is a diagram of the TandemHeart Veno-Venous Cannula. The diagram labels the barbed connectors, clamp areas, insertion depth markers, and tantalum markers. The insertion depth markers are labeled with the numbers 44, 42, and 40. The cannula also has a wire-reinforced area.

The TandemHeart Veno-Venous Cannula (Figure 1) consists of two distinct lumens with a wire-reinforced cannula body. The inner lumen is located entirely within the outer lumen forming two concentric channels

The distal section (inner lumen/cannula) of the cannula body has six side holes near the distal tip opening. The proximal sections of each lumen are clear and not wire-reinforced to allow visualization of blood and to enable clamping to prevent blood flow during setup and removal of the cannula from the extracorporeal circulatory support equipment (see Figure 1). A non-vented barbed connector is affixed to both proximal ends (inner/distal and outer/proximal lumens) of the cannula and allow for connection of standard 3/8 inch blood circuit tubing for subsequent connection to extracorporeal circulatory support equipment. The cannula has printed insertion depth markings every 10 centimeters from 10 to 30 cm followed by every 2 centimeters for the remainder of the insert-able length, measured from the distal end. The cannula also has a suture wing that can be used for securing the cannula in place to the patient.

The introducer (Figure 2) consists of a tube with a hub. The introducer fits inside the inner lumen of the cannula during insertion of the cannula/introducer assembly. The introducer is used to advance the cannula over a guidewire and facilitate cannula placement within the target vessel. The introducer has a hub at its proximal end to manage introducer insertion and removal from the cannula. The hemostasis cap minimizes blood loss when the cannula/introducer assembly is inserted into the target vessel. The introducer body is constructed from radiopaque material for visualization under fluoroscopy.

Image /page/4/Figure/4 description: The image shows a medical device, specifically an introducer needle, which is used to insert catheters or other devices into the body. The introducer hub is located at the end of the needle, and the hemostasis cap is located near the hub. The needle is long and thin, with a sharp point at the end.

Figure 2: 13 Fr. Introducer

{5}------------------------------------------------

#### Intended Use

The TandemHeart Veno-Venous Cannula Set is intended for use as a single cannula for both venous drainage and reinfusion of blood via an internal jugular vein during extracorporeal life support procedures.

#### Comparison of Technological Characteristics

The TandemHeart Veno-Venous Cannula Set and the predicate Avalon (Maquet) Elite Bi-Caval Dual Lumen Catheter (K081820) are both wire-reinforced dual lumen venous cannula. Both kits include an introducer to facilitate placement using a guidewire and fluoroscopy. There are a number of technological differences: 1) the TandemHeart Veno-Venous Cannula Set contains an additional component located on the exterior proximal portion of the cannula: a suture wing to provide a means of securing the cannula to the patient; 2) the TandemHeart Veno-Venous cannula is co-axial whereas the Avalon Elite predicate is bi-caval; 3) the TandemHeart Veno-Venous Cannula has a 29 Fr outer diameter at the proximal end compared to 27 Fr for the comparable Avalon Elite; 4) the TandemHeart Veno-Venous Cannula is longer than the predicate with greater distance between inlet and outlet to reduce recirculation and to allow improved flexibility in positioning within the venous system; and 5) the TandemHeart Veno-Venous Cannula features a taper from the narrow distal, single lumen portion to the larger diameter, proximal, dual lumen portion that is designed to aid with insertion by gradually dilating the vessel.

#### Performance Data

Non-clinical performance testing was conducted to demonstrate substantial equivalence of flow characteristics between the TandemHeart Veno-Venous Cannula and the predicate, Medtronic Avalon Elite The performance testing included in-vitro system capacity testing and flow vs. pressure drop (HQ). This testing verified that despite the longer length of the TandemHeart Veno-Venous Cannula relative to the Avalon Elite Cannula, it was able to sustain flows comparable to the predicate.

Flexibility, strength, biocompatibility, in-vitro hemolysis, leak testing, sterilization, and shelf life of the TandemHeart Veno-Venous Cannula are addressed by comparison with both the Avalon elite and a second predicate, the TandemHeart Venous Cannula (K133236). Based on the performance test results and data from the predicate devices, the TandemHeart Veno-Venous Cannula was found to meet established design input requirements and thus to be substantially equivalent to the predicate Avalon Elite Dual Lumen Veno-Venous Cannula.

---

**Source:** [https://fda.innolitics.com/device/K140999](https://fda.innolitics.com/device/K140999)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
