POWDER FREE NITRILE SURGICAL GLOVES, WHITE COLOR (BRAND NAME: TITANFINE)
K140989 · Hebei Hongsen Plastics Technology Co, Ltd. · KGO · Dec 22, 2014 · General, Plastic Surgery
Device Facts
Record ID
K140989
Device Name
POWDER FREE NITRILE SURGICAL GLOVES, WHITE COLOR (BRAND NAME: TITANFINE)
Applicant
Hebei Hongsen Plastics Technology Co, Ltd.
Product Code
KGO · General, Plastic Surgery
Decision Date
Dec 22, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4460
Device Class
Class 1
Indications for Use
The powder free nitrile surgical glove, white color (Brand Name: Titanfine), is a sterile and single use device made of synthetic rubber intended to be worn by operation room personnel to protect a surgical wound from contamination. The gloves do not contain lubricating or dusting powder.
Device Story
Disposable sterile nitrile rubber surgical glove; provides physical barrier between surgical personnel hands and patient open wound; prevents contamination during surgical procedures. Used in operating rooms by surgical personnel. Device is single-use; white color; powder-free. Benefits patient by reducing risk of surgical site contamination. Performance verified via bench testing for dimensions, tensile strength, elongation, pinhole resistance, and biocompatibility.
Clinical Evidence
No clinical data. Bench testing only. Compliance with ASTM D3577-09 (physical properties), ASTM D6124-06 (powder residue), ASTM D5151-06 (pinholes), ASTM D5712-10 (protein), and ISO 10993-10 (biocompatibility/irritation/sensitization).
Technological Characteristics
Material: Synthetic nitrile rubber. Form factor: Hand/wrist covering with finger sheaths. Sterilization: Radiation (ISO 11137-2:2006). Standards: ASTM D3577-2009, ISO 10993-10. Dimensions: Length ≥265mm, thickness ≥0.10mm. Tensile strength ≥17 MPa. Elongation ≥650%.
Indications for Use
Indicated for use by operating room personnel as a sterile, single-use physical barrier to protect surgical wounds from contamination. No specific patient population age or gender restrictions.
Regulatory Classification
Identification
A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Motex Powder Free Nitrile Surgical Gloves, White and Green Color (K122557)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of a human figure, with three profiles overlapping to create a sense of depth and connection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the figure in a circular pattern.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 22, 2014
Hebei Hongsen Plastics Technology Co., Ltd. C/O Mr. Charles Shen Manager Manton Business and Technology Services 853 Dorchester LN, Unit-B New Milford, NJ 07646
Re: K140989
Trade/Device Name: Powder Free Sterile Nitrile Surgical Gloves; White Color (Brand Name: Titanfine) Regulation Number: 21 CFR 878.4460 Regulation Name: Surgeon's Glove Regulatory Class: I Product Code: KGO Dated: November 23, 2014 Received: November 26, 2014
Dear Mr. Shen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Shen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Teiashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K140989
Device Name
Powder Free Nitrile Surgical Gloves, White Color (Brand Name: Titanfine)
Indications for Use (Describe)
The powder free nitrile surgical glove, white color (Brand Name: Titanfine), is a sterile and single use device made of synthetic rubber intended to be worn by operation room personnel to protect a surgical wound from contamination. The gloves do not contain lubricating or dusting powder.
Type of Use (Select one or both, as applicable)
- Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Hebei HongSen Plastics Technology Co., Ltd 15 Eastern Industrial Accumulation Area, Nangong Town, Xingtai City, Hebei Province, China Tel: (086) 311-85656588
# Section 5: 510(k) Summary: K140989
This summary of 510k is being submitted in accordance with the requirements of 21CFR 807.92
#### 5.1 Submission Sponsor
Hebei HongSen Plastics Technology Co., Ltd Contact Person: Vincent Lu 15 Eastern Industrial Accumulation Area Nangong Town, Xingtai City, Hebei Province, China Tel: (086) 311-85656588 Submitter's FDA Registration Number: 3010582952
#### 5.2 Submission Correspondent
Manton Business and Technology Services Contact Person: Charles Shen 853 Dorchester LN, Unit-B New Milford, NJ 08534 Tel: 608-217-9358 Email: cyshen@aol.com
#### 5.3 Date of Summary: December 16, 2014
#### 5.4 Device Name:
Proprietary Name: Powder Free Nitrile Surgical Gloves, White Color (Brand Name: Titanfine) Common Name: Surgeon's glove Classification Name: Surgeon's glove Device Classification: -Regulation Number: 21 CFR 878.4460 Panel: General General & Plastic Surgery Product Code: KGO
#### ર્સ્ટ Predicate Device Information:
K122557, "Motex Powder Free Nitrile Surgical Gloves, White and Green Color", (1) manufactured by "Shanghai Motex Healthcare Co., Ltd."
{4}------------------------------------------------
15 Eastern Industrial Accumulation Area, Nangong Town, Xingtai City, Hebei Province, China Tel: (086) 311-85656588
#### 5.6 Device description:
Surgical glove is a disposable device worn on the surgical room personnel's hand or finger to prevent contamination during surgical procedures. The principle of operation is to provide a physical barrier between the surgical personnel's hands and the patient's open wound.
The powder free nitrile surgical gloves described in this submission are made of synthetic nitrile rubber, and are sterile that meets all of the requirements of ASTM standard D 3577-2009. They have white color and are single use.
The powder free nitrile surgical gloves described in this submission are tested for dimension, percent elongation, tensile, tensile at 500% elongation, free of pinholes, residue powder, water soluble proteins. The test results conform to related requirements in ASTM 3577-2009.
The powder free nitrile surgical gloves described in this submission are also tested biocompatibility per ISO 10993-10. The test results conform to related requirements in the standard.
The product sterilization is performed using radiation and the procedure is validated per ISO 11137-2: 2006.
#### 5.7 Indications for Use:
The powder free nitrile surgical glove, white color (Brand Name: Titanfine), is a sterile and single use device made of synthetic rubber intended to be worn by operation room personnel to protect a surgical wound from contamination. The gloves do not contain lubricating or dusting powder.
#### Comparison to Predicate Devices 5.8
The powder free nitrile surgical gloves, white, sterile are compared with the following Predicate Devices in terms of intended use, design, material, specifications, and performance.
K122557, "Motex Powder Free Patient Nitrile Surgical Gloves", manufactured by (1) "Shanghai Motex Healthcare Co., Ltd."
Substantial equivalence is established with respect to intended use, Labeling, performance, design, materials, and other applicable characteristics. Side-by-side comparison tables that include the following four areas are provided below:
- A: Indication for Use, design and materials
- B: Labeling (labels, instructions for use, promotional material) for the legally marketed device to which substantial equivalence is claimed.
{5}------------------------------------------------
15 Eastern Industrial Accumulation Area, Nangong Town, Xingtai City, Hebei Province,China Tel: (086) 311-85656588
- C: Performance data supporting substantial equivalence
- A: Indications for Use. Design, and Material:
The following table shows similarities and differences of use, design, and material between subject device and the predicate devices.
| Description | Subject Device (K140989) | Predicate Device (K122557) |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for<br>Use | The powder free nitrile surgical glove<br>(Brand Name: Titanfine) is a sterile and<br>single use device made of synthetic<br>rubber intended to be worn by operation<br>room personnel to protect a surgical<br>wound from contamination. The gloves<br>do not contain lubricating or dusting<br>powder. | A Surgeon's Glove is a sterile<br>and single use device made of<br>synthetic rubber intended to be<br>worn by operation room<br>personnel to protect a surgical<br>wound from contamination. The<br>lubricating or dusting powder<br>used in the glove is excluded. |
| Basic Design | A device covering the hand and wrist<br>area. Gloves have separate sheaths or<br>openings for each finger and the thumb. | Same |
| Materials | Nitrile rubber | Same |
| Size | 6, 6.5, 7, 7.5, 8, 8.5, 9 | 6, 6.5, 7, 7.5, 8, 8.5 |
| Single Use | Yes | Yes |
| Color | White | White and green |
| Sterile | Sterile (Irradiation) | Sterile (Irradiation) |
## Table 5.1: Comparison of Intended Use, Design, and Material
The subject device is essentially identical to the predicate device in terms of indications for use, design, and material between subject device and the predicate devices. The only minor difference is that the predicate device has both green and white color, while subject device in submission has only white color. This minor difference in color does not impact the efficacy and safety of the gloves, demonstrated by satisfactory performance and biocompatibility test results, and the fact that the white pigment of TiO2 has a well established safety profile.
#### B: Labeling:
The following table shows similarities and differences of key elements of the labeling between subject device and the predicate devices.
{6}------------------------------------------------
15 Eastern Industrial Accumulation Area, Nangong Town, Xingtai City, Hebei Province, China Tel: (086) 311-85656588
| Description | Subject Device<br>(K140989) | Predicate Device<br>(K122557) |
|--------------------------------------------------|-----------------------------|-------------------------------|
| Product Name | Yes | Yes |
| Indication for Use | Yes | No |
| Technical Characteristics (Powder, Nitrile, etc) | Yes | Yes |
| Manufacturer Name and Address | Yes | Yes |
| Product Model, Size, and Lot Number | Yes | Yes |
| Single Use Statement | Yes | Yes |
| Sterility Statement | Yes | Yes |
| Expiration Date | No | Yes |
| Information on Wrapper | Size, and<br>Left/Right | Size |
| | Table 5.2: Comparison of Key Elements in Labeling | |
|--|---------------------------------------------------|--|
| | | |
The label for this subject device is identical to K122557, except for two minor differences. The first one is that the label for subject device has Indications for Use on it, while the second one being that the predicate device claims expiration date, while the subject device does not.
#### ﻥ Performance and Specifications
The following table shows similarities and differences of the specification/performance between subject device and the predicate devices. Tests were conducted following the recommended procedures outlined in the respective consensus standards, and results for Powder Free Nitrile Surgical Gloves, White Color (Brand Name: Titanfine) met all relevant requirements in the specifications and test standards. The subject device/gloves are equivalent with the predicate device in terms of performance and technological characteristics.
| Description | Subject Device (K140989) | Predicate Device (K122557) |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Dimension | Length: ≥265 mm<br>Width: within ±6 nm of the<br>specified value in ASTM 3577:<br>2009<br>Palm, Finger, Cuff Thickness:<br>≥0.10 mm | Length: ≥265 mm<br>Width: within ±6 nm of the<br>specified value in ASTM 3577:<br>2009<br>Palm, Finger, Cuff Thickness:<br>≥0.10 mm |
| Tensile Strength (MPa) | ≥17 | ≥17 |
{7}------------------------------------------------
15 Eastern Industrial Accumulation Area, Nangong Town, Xingtai City, Hebei Province, China Tel: (086) 311-85656588
| Tensile at 500 %<br>Elongation (MPa) | ≤ 7.0 | ≤ 7.0 |
|-------------------------------------------------|---------------------------------------------------------------------------|-------------------------------|
| Percent Elongation | ≥650 | ≥650 |
| Free of Pinhole | Free of pinholes at AQL = 1.5 | Free of pinholes at AQL = 1.5 |
| Residue Powder | ≤ 2.0 mg | ≤ 2.0 mg |
| Water Soluble Protein | ≤ 200 µg/dm² | ≤ 200 µg/dm² |
| Primary Skin Irritation<br>(ISO 10993-10: 2010) | Not a primary skin irritant (Grade<br><1.0) under the conditions of study | Not a primary skin irritant |
| Dermal sensitization<br>(ISO 10993-10: 2010) | Not a dermal sensitizer (Grade 0 or<br>1) under the conditions of study | Not a dermal sensitizer |
### 5.9 A brief discussion of the non-clinical testing data that was submitted, referenced or relied on to demonstrate that the Subject Device is safe and effective, and whose performance meets the requirements of its user-defined acceptance criteria and intended uses:
Powder Free Nitrile Surgical Gloves, White Color (Brand Name: Titanfine) meet requirements per ASTM D3577-09, ASTM D6124-06, ASTM D 5151-06, ASTM D5712-10 and ISO 10993-10. Its performance meets the requirements of its pre-defined acceptance criteria and intended uses.
### A brief discussion of the clinical submitted, reference, or relied on in the 5.10 premarket notification submission for a determination of substantial equivalence.
Clinical data is not needed for gloves cleared by the 510(k) process.
#### 5.11 Substantial Equivalent Conclusions
Based on the comparison of intended use, design, materials, and performance, the subject device is as safe and effective and substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.