KLASSIC KNEE TIBIAL STEM EXTENSION

K140942 · Total Joint Orthopedics, Inc. · JWH · May 16, 2014 · Orthopedic

Device Facts

Record IDK140942
Device NameKLASSIC KNEE TIBIAL STEM EXTENSION
ApplicantTotal Joint Orthopedics, Inc.
Product CodeJWH · Orthopedic
Decision DateMay 16, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3560
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Klassic Knee Tibial Stem Extension is intended for use with the Klassic Knee System. The Klassic Knee System is intended for prosthetic replacement with the use of bone cement in treatment of the following: Patient conditions of non-inflammatory degenerative joint disease (NIDJD): avascular necrosis and osteoarthritis Patient conditions of inflammatory joint disease (IJD): rheumatoid arthritis Patients with failed previous surgery where pain, deformity, or dysfunction persists Correctable varus-valgus deformity and moderate flexion contracture Revision of a previously failed knee arthroplasty Patients who require a total knee replacement

Device Story

Klassic Knee Tibial Stem Extension is a modular titanium alloy component designed for attachment to the Klassic Knee System Tibial Baseplate via a threaded connection. It provides additional distal fixation for the tibial baseplate in total knee arthroplasty. Used by orthopedic surgeons in clinical settings during knee replacement or revision procedures. The device is implanted to stabilize the tibial component, potentially reducing pain and restoring joint function for patients with degenerative or inflammatory joint disease. It is compatible with the existing Klassic Knee System and offered in lengths up to 150mm.

Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical bench testing, including cantilever bending fatigue testing, post-fatigue fretting-corrosion analysis, and modular connection integrity analysis, conducted in accordance with ASTM F2083-12, ASTM F1814-97(2009), and ASTM F1800-12.

Technological Characteristics

Modular tibial stem extension; material: titanium alloy; connection: threaded; length: up to 150mm; energy source: none (mechanical); sterilization: not specified; standards: ASTM F2083-12, ASTM F1814-97(2009), ASTM F1800-12.

Indications for Use

Indicated for patients requiring total knee replacement due to non-inflammatory degenerative joint disease (avascular necrosis, osteoarthritis), inflammatory joint disease (rheumatoid arthritis), failed previous surgery with persistent pain/deformity/dysfunction, correctable varus-valgus deformity, moderate flexion contracture, or revision of failed knee arthroplasty.

Regulatory Classification

Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## MAY 1-S 2014 ### 510(k) SUMMARY #### 140942 510(k) Notification K__ #### GENERAL INFORMATION ## Applicant: Total Joint Orthopedics 1567 E. Stratford Avenue Salt Lake City, UT 84106 United States Phone: 801-486-6070 FAX: 801-486-6117 ## Contact Person: Jean M. Wheeler Product Development Engineer Total Joint Orthopedics 1567 E. Stratford Avenue Salt Lake City, UT 84106 United States Phone: 801-486-6070 Fax: 801-486-6117 Date Prepared: April 11, 2014 ## DEVICE INFORMATION ## Trade/Proprietary Name: Klassic™ Knee Tibial Stem Extension #### Generic/Common Name: Knee prosthesis, tibial stem extension #### Classification: 21 CFR §888.3560 Knee joint patellofemorotibial polymer/metal/polymer semiconstrained cemented prosthesis Product Code: IWH {1}------------------------------------------------ ## 510(k) SUMMARY #### PREDICATE DEVICE(S) The Klassic Knee Tibial Stem Extension is substantially equivalent in intended use, design, function, performance testing to the following predicate devices: - Klassic Knee System Total Joint Orthopedics K112906 . - Universal Stem Extensions DePuy K063633 . - Triathlon TS Knee System Stryker Orthopedics K070095 . #### P #### INTENDED USE The Klassic Knee Tibial Stem Extension is intended for use with the Klassic Knee System. The Klassic Knee System is intended for prosthetic replacement with the use of bone cement in treatment of the following: - Patient conditions of non-inflammatory degenerative joint disease (NIDJD): . avascular necrosis and osteoarthritis - Patient conditions of inflammatory joint disease (IJD): rheumatoid arthritis . - Patients with failed previous surgery where pain, deformity, or dysfunction . . persists - Correctable varus-valgus deformity and moderate flexion contracture . - Revision of a previously failed knee arthroplasty . - . Patients who require a total knee replacement #### PRODUCT DESCRIPTION The Klassic Knee Tibial Stem Extension is a modular component designed for use with the Klassic Knec System Tibial Baseplate to provide the user with options for stemming the Tibial Baseplate where additional distal fixation is desired. The Stem Extension is fully compatible with the predicate Klassic Knee System and offered in lengths up to 150mm. The Klassic Knee Tibial Stem Extension is manufactured from titanium alloy and features a threaded connection which allows for attachment to the Klassic Knee System Tibial Baseplate. ## TECHNOLOGICAL CHARACTERISTICS The technological characteristics of the Klassic Knee Tibial Stem Extension were compared to predicate devices to serve as a basis for determining substantial equivalence. The Klassic Knee Tibial Stem Extension is similar in device design, material of manufacture, sterilization and size offerings. {2}------------------------------------------------ ## 510(k) SUMMARY ## TESTING IN SUPPORT OF SUBSTANTIAL EQUIVALENCE DETERMINATION Non-clinical bench testing was conducted on the Klassic Knee Tibial Stem Extension to support a determination of substantial equivalence to the predicate devices. Non-clinical bench testing was based off of guidance from: - ASTM F2083-12 Standard Specification for Knee Replacement Prosthesis . - ASTM F1814-97(2009) Standard Guide for Evaluating Modular Hip and Knee . Joint Components - ASTM F1800-12 Standard Practice for Cyclic Fatigue Testing of Metal Tibial . Tray Components of Total Knee Joint Replacements The performance bench testing for the Klassic Knee Tibial Stem Extension evaluated the use of the worst case geometry of the subject device with the Klassic Knee System in fatigue and the impact of fatigue loading on the modular connection mechanism of the device. Specifically, the following tests were conducted: - Cantilever Bending Fatigue Testing t - . Post Fatigue Fretting-Corrosion Analysis - Analysis of the Integrity of the Modular Connection . All bench testing demonstrates the Klassic Knee Tibial Stem Extension is equivalent in regards to safety and efficacy, is suitable for use in Total Knee Replacement and is substantially equivalent to predicate devices. #### SUBSTANTIAL EQUIVALENCE The indications for use for the predicate devices are substantially equivalent to the proposed indications for use for the Klassic Knee Tibial Stem Extension. The Klassic Knee Tibial Stem Extension is similar to the predicate devices based on technological characteristics, design, material, non-clinical performance testing, sterilization and intended use. Any differences in the technological characteristics between the devices do not raise any new issues of safety or effectiveness. Thus, the Klassic Knee Tibial Stem Extension is substantially equivalent to the predicate device. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 16, 2014 Total Joint Orthopedics, Incorporated Ms. Jean M. Wheeler Product Development Engineer 1567 East Stratford Avenue Salt Lake City, Utah 84106 Re: K140942 Trade/Device Name: Klassic 1 Knee Tibial Stem Extension Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH Dated: April 11, 2014 Received: April 14, 2014 Dear Ms. Wheeler: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {4}------------------------------------------------ Page 2 - Ms. Jean M. Wheeler forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also. please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. # Ronald P. Jean -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ TOTAL JOINT ORTHOPEDICS KLASSIC KNEE TIBIAI. STEM EXTENSION 5 10(k) PREMARKET NOTIFICATION ## SECTION 4 INDICATIONS FOR USE STATEMENT K140942 510(k) Number (if known): Klassic" Knee Tibial Stem Extension: ## Indications For Use: The Klassic" Knee Tibial Stem Extension is intended for use with the Klassic" Knee System. The Klassic" Knee System is intended for prosthetic replacement with the use of bone cement in treatment of the following: - Patient conditions of non-inflammatory degenerative joint disease (NIDJD): . avascular necrosis and osteoarthritis - Patient conditions of inflammatory joint disease (IJD): rheumatoid arthritis . - . Patients with failed previous surgery where pain. deformity, or dysfunction persists - Correctable varus-valgus deformity and moderate flexion contracture . - Revision of a previously failed knee arthroplasty . - Patients who require a total knee replacement . Prescription Use X (21 CFR Part 801 Subpart D) And/Or Over the Counter Use (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE: CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Casey L. Hanley, Ph.D. --- Division of Orthopedic Devices
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...