K140929 · Human Design Medical, LLC · BZD · Oct 17, 2014 · Anesthesiology
Device Facts
Record ID
K140929
Device Name
Z1 CPAP SYSTEM
Applicant
Human Design Medical, LLC
Product Code
BZD · Anesthesiology
Decision Date
Oct 17, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Provides continuous positive airway pressure (CPAP) to support treatment of adults with obstructive sleep apnea.
Device Story
Z1 CPAP System is a single-patient, reusable, non-critical care device for home use; provides continuous positive airway pressure (CPAP) to treat obstructive sleep apnea in adults >66 lbs. Input: pneumatic air flow signals from patient breathing. Operation: pneumatic pump delivers pressure (4-20 cmH2O); includes APAP mode with auto-pressure adjustment and 'Z-Breathe' dynamic pressure control (3 settings). Output: pressurized air delivered via tube/mask; therapy data logged for management. Operated by patient; includes optional mobile app for interaction. Benefits: provides respiratory support for sleep apnea; portable design with optional battery. Healthcare providers use logged data to monitor therapy effectiveness.
Clinical Evidence
No clinical data generated. Bench testing only. Performance verified via side-by-side comparison with predicate (iSleep 20i) using test lung air flow signals representing diverse respiratory patterns. Metrics included pressure accuracy, apnea/hypopnea event detection, and software-level output agreement. All V&V acceptance criteria met.
Technological Characteristics
Pneumatic pump-based CPAP; 4-20 cmH2O pressure range (0.5 cmH2O increments). Materials comply with ISO 10993. Standards: ISO 5356-1, ASTM F 1246-91, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11, ISO 23328-2, ISO 17510-1. Connectivity: Bluetooth, USB, SD card. Power: AC adapter or 14.4V 4-cell Li-ion battery. Software: APAP algorithm ported from K063476.
Indications for Use
Indicated for adults weighing over 66 lbs (30 kg) with obstructive sleep apnea requiring continuous positive airway pressure (CPAP) therapy.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of three faces in profile, stacked on top of each other, resembling a stylized human figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 17, 2014
Human Design Medical, LLC C/O Ms. Maureen O'Connell President O'Connell Regulatory Consultants, Inc. 5 Timber Lane North Reading, MA 01864
Re: K140929
Trade/Device Name: Z1 CPAP System Regulation Number: 21 CFR 868.5905 Regulation Name: Ventilator, Non-Continuous Regulatory Class: II Product Code: BZD Dated: September 16, 2014 Received: September 17, 2014
Dear Ms. O'Connell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Ms. O'Connell
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/5 description: The image shows the name "Tejashri Purohit-Sheth, M.D." in a stylized font. The text is black and the background is white. The text appears to be a name and title, possibly for a medical professional.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K140929
Device Name Z1 CPAP System
Indications for Use (Describe)
The Z1 CPAP System is a single patient reusable device that provides continuous positive airway pressure (CPAP) to support treatment of adults weighing over 66 lbs (30 kg) with obstructive sleep apnea.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| <div> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
Prepared in accordance with 21 CFR § 807.92
| Date Summary Prepared: | September 16, 2014 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Information: | |
| Company Name: | Human Design Medical, LLC |
| Company Address: | 119 Braintree Street, Unit 703<br>Boston, MA 02134 |
| Contact Person: | Maureen O'Connell<br>O'Connell Regulatory Consultants, Inc.<br>5 Timber Lane<br>North Reading, MA 01864<br>Phone: 978-207-1245<br>Fax: 978-824-2541 |
| Device Information: | |
| Trade Name: | Z1 CPAP System |
| Common Name: | CPAP System |
| Classification Name: | Non-Continuous (Respirator) Ventilator, 21 CFR § 868.5905 |
| Device Class: | Class II |
| Predicate Devices: | 510(k) Number: K121374<br>Manufacturer: Human Design Medical, LLC<br>Product Name: Z1 Blower |
| | 510(k) Number: K063476<br>Manufacturer: Breas Medical AB<br>Product Name: iSleep 20i |
| | 510(k) Number: K100121<br>Manufacturer: AeioMed, Inc.<br>Product Name: Model 300157 |
| Device Description: | The Z1 CPAP System provides continuous positive airway pressure to support the treatment of adults over 66 lbs (30 kg) with obstructive sleep apnea. The device is a non-critical care, reusable, single patient device for use in the |
{4}------------------------------------------------
| | home setting, and it is intended to be a prescription use<br>only device. The Z1 CPAP System is a modified version of<br>the Z1 Blower cleared in K121374. The modifications<br>include: addition of an optional battery, addition of an<br>APAP mode, addition of Bluetooth and a mobile app,<br>addition of new user settings, and changes to firmware. |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: | Provides continuous positive airway pressure (CPAP) to<br>support treatment of adults with obstructive sleep apnea. |
| Indications for Use: | The Z1 CPAP System is a single patient reusable device<br>That provides continuous positive airway pressure (CPAP)<br>To support treatment of adults weighing over 66 lbs (30 kg)<br>with obstructive sleep apnea. |
### Technological Characteristics
| Compared to Predicate: | The Z1 CPAP System consists of the main flow generator<br>and accessories. The Z1 CPAP System and the identified<br>predicate devices have the same fundamental technological<br>characteristic: Use of a pneumatic pump to deliver<br>continuous positive airway pressure within a clinically<br>indicated therapeutic pressure range. |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
All four of the systems include a flow generator and a power adaptor. The Z1 CPAP System, the Breas iSleep 20i System and the AeioMed Model 300157 also include an air tube and battery while the Z1 Blower cleared in K121374 can only be operated using the power adaptor and the air tube must be purchased separately. All four of the systems have buttons and a status display that allow the user to control the system. Additionally, the Z1 CPAP System allows the user to interact with the system via a mobile application. This is an optional user convenience component that does not impact the safety or effectiveness of the treatment delivered.
All four of the systems include a CPAP treatment mode with a ramp feature. The Z1 CPAP System, the Z1 Blower and the AeioMed Model 300157 include a ramp period up to 45 minutes while the Breas iSleep 20i System has a ramp period up to 60 minutes. This minor difference between the Z1 CPAP System and the Breas iSleep 20i System does not impact the substantial equivalence. All ramp features are identical between the Z1 Blower and the Z1 CPAP System.
The Z1 CPAP System and the Breas iSleep 20i System
{5}------------------------------------------------
both include an APAP mode with an auto pressure setting algorithm. The Z1 employs the exact same software code for the APAP mode auto pressure setting algorithm as was cleared for the iSleep20i in K063476. Performance data is provided which shows that the performance of the Z1 CPAP System in the APAP mode is substantially equivalent to the predicate device.
All four of the systems include dynamic pressure control. The predicate systems have one fixed setting. The Z1 CPAP System has three user selectable settings (called Z-Breathe 1, 2 and 3). Performance data is provided which shows that all Z-Breathe settings are equivalent between the Z1 CPAP System and one of the identified predicate devices.
The pressure range for all four of the systems is 4-20 cmH20 in 0.5 cm H2O increments. The Z1 CPAP System, the Z1 Blower and the AeioMed Model 300157 include an automatic altitude adjustment for conditions up to 8,000 feet. The operating range and storage range for the Z1 CPAP System and the Z1 Blower are identical. All four systems allow data management functionality through an SD data card or memory card or through a USB port.
Both the Z1 CPAP System and one of the identified predicate devices use a lithium ion battery pack with 4 cells and a voltage of 14.4 volts. The battery longevity was measured to be 5-8 hours for the Z1 CPAP System depending upon the exact use conditions while it was measured to be 5-7 hours for the predicate. The environment of use for all four systems is the home while the Breas iSleep 20i System can also be used in a clinical setting.
Human Design Medical has performed the following Performance Testing: performance testing to support the safety and effectiveness of the Z1 CPAP System:
> Maximum temperature at the patient connection port under normal and single fault conditions was < 43 degrees C at a pressure setting of 20 cmH20;
Pressure stability under static long-term conditions was within ± 1.0 cmH2O of the set pressure of 10 cmH2O;
{6}------------------------------------------------
Short term static pressure accuracy: the actual pressure was within the specified tolerance of the set pressure for all pressure settings;
Dynamic pressure stability: the average peak-to-peak pressure deviation was < the maximum limit established by Human Design Medical, for each pressure and breathe rate setting and was compared to a predicate device;
Maximum flow rate was ≥ minimum limit established by Human Design Medical for each pressure setting.
Acoustic noise testing found that the sound pressure level was 26dBA:
Testing of the battery was performed to verify the safety, performance and longevity of the battery accessory.
Ability of the system to detect air leaks: air leaks were detected within +5 mmH20 of the specified air leak detection thresholds:
Volatile organic compounds: no volatile organic compounds were observed in the air output of the device above what was found in the ambient air;
Particulate matter: no particulate matter was observed in the output air of the device above what was found in the ambient air;
The Z1 Blower added no carbon monoxide or carbon dioxide to the output air and the ozone output of the device met the 21 CFR 801.415 requirement of 0.05 ppm or less.
Altitude testing showed that the Z1 maintained actual output pressure within the specified tolerance limits throughout the entire specified altitude pressure range of operation.
The pressure waveforms for the three z-Breathe settings have been compared with the pressure waveforms for the predicate device and it was determined that the performance was substantially equivalent.
The Z1 Blower complies with the following performance
{7}------------------------------------------------
and safety standards: ISO 5356-1:2004 Anesthetic and Respiratory Equipment - Conical Connectors: Part 1: Cones and Sockets. ASTM F 1246-91 standard specification for Electrically Powered Home Care Ventilators, Part 1 - Positive-Pressure Ventilators and Ventilator Circuits. IEC 60601-1:2005 Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performance. IEC 60601-1-2:2007 Medical Electrical Equipment – Part 1-2: General requirements for basic safety and essential performance - collateral standard: Electromagnetic compatibility - requirements and tests. IEC 60601-1-6: Medical Electrical Equipment-Part 1-6: General requirements for basic safety and essential performance-collateral standard: Usability. IEC 60601-1-11 Medical Electrical Equipment-Part 1-11: General requirements for basic safety and essential performancecollateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment (including Technical Corridgendum 1).
ISO 23328-2:2002 Breathing system filters for anesthetic and respiratory use - part 2: non filtration aspects. ISO 17510-1:2007 Sleep apnoea breathing therapy- Part 1: Sleep apnoea breathing therapy equipment.
Biocompatibility testing was performed which showed that the patient contacting components of the system comply with ISO 10993-1 Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing; ISO 10993-3:2003 Biological Evaluation of Medical Devices - Part 3: Tests for Genotoxicity, Carcinogenicity, and Reproductive Toxicity; ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro Cytotoxicity; ISO 10993-6:2007 Biological Evaluation of Medical Devices – Part 6: Tests for Local Effects After Implantation: ISO 10993-10 Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization; and ISO 10993-12:2012 Biological Evaluation of Medical Devices – Part 12: Sample Preparation and Reference Materials.
Testing of the CPAP Tube was performed which validated the recommended cleaning procedures and replacement interval.
{8}------------------------------------------------
The APAP software of the Z1 CPAP System is identical to the APAP software of the iSleep20i System cleared by FDA in K063476. The iSleep20i software source code, as well as the entire iSleep20i Design History File and Regulatory Master File were acquired by Human Design Medical as part of a business acquisition. Human Design Medical ported the iSleep20i APAP software into the Z1 CPAP System without modification. Therefore, the Z1 CPAP System employs the exact same software code as was cleared for the iSleep20i in K063476.
Since the APAP software has already been cleared in K063476 for the iSleep20i, it was only necessary to verify and validate that the Z1 CPAP System, when running the same software and presented with the same clinically diverse dataset on the bench, produces the same results as the iSleep20i System. Therefore Human Design Medical has conducted extensive bench testing of the Z1 CPAP System side-by-side with the iSleep20i System to verify and validate that the two devices produce the same respiratory event detections and pressure adjustment outputs for the same input air flow data. This testing is described in the following paragraph.
The same pneumatic air flow signals from a test lung were presented to the iSleep20i and the Z1. The test lung air flow signals comprised a clinically diverse data set of respiratory patterns, breathing rates, pressure settings, recovery to normal breathing and non-recovery patterns, etc. The therapy pressures generated by each device over time were compared. The pressure output of the Z1 matched the pressure output of the iSleep20i within the specified accuracy across all air flow signals within each sample time interval and met the acceptance criterion of matching within the specified accuracy. The outputs of ancillary algorithms for detecting apnea and hypopnea events were also compared. The number of apneahypopnea events detected by the two devices was the same, and met the acceptance criterion for an exact match. Additionally, software-level testing was performed in which the same digitized flow samples were presented to the iSleep20i and the Z1. The digitized flow data comprised a clinically diverse data set. The outputs of the software in each device were compared on a sample by sample basis and met all acceptance criteria for level of
{9}------------------------------------------------
| | agreement. The results of the APAP testing satisfied each<br>and every acceptance criterion of the V&V Plan without<br>exception. |
|-------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | No clinical data was generated to support a substantial<br>equivalence determination. A determination was made that<br>bench performance testing was appropriate versus clinical<br>performance testing because there are accepted bench<br>performance testing methodologies available which allow a<br>direct comparison of the performance of the Z1 CPAP<br>System compared to the predicate devices under a wide<br>variety of operating conditions. |
| Conclusion: | The Z1 CPAP System is substantially equivalent to the<br>predicate devices, as the devices share a common intended<br>use, and technological differences between the Z1 CPAP<br>System and the predicate do not raise new questions of<br>safety or effectiveness. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.