K140927 · Medos International SARL · OLO · Oct 24, 2014 · Neurology
Device Facts
Record ID
K140927
Device Name
UNIVERSAL NAVIGATION INSTRUMENTS
Applicant
Medos International SARL
Product Code
OLO · Neurology
Decision Date
Oct 24, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The Universal Navigation Instruments for EXPEDIUM® and VIPER® MIS Spine Systems are intended to assist the surgeon in precisely locating anatomical structures in either open, minimally invasive, or percutaneous procedures. These are indicated for use in surgical spinal procedures, in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the pelvis or a vertebrae can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy using tracking arrays provided by the navigation manufacturer. These procedures include but are not limited to spinal fusion. These devices can be pre-calibrated and/or manually calibrated with Brainlab Navigation system, where other navigation systems require manual calibration.
Device Story
Manual surgical instruments designed to interface with previously cleared surgical navigation systems; used in open, minimally invasive, or percutaneous spinal procedures. Instruments feature universal adaptors and clamps to attach to tracking arrays provided by navigation manufacturers. Surgeon uses instruments to precisely locate anatomical structures relative to patient images (CT, MR, fluoroscopy) or anatomical models. System supports pre-calibration or manual calibration with Brainlab navigation systems; manual calibration for other systems. Output is visual guidance on navigation system display, assisting surgeon in instrument positioning during spinal fusion or other orthopedic procedures. Benefits include improved precision in anatomical localization during surgery.
Clinical Evidence
Bench testing only. Verification and validation activities compared accuracy and connection rigidity against predicate instruments in a controlled environment. Testing confirmed successful calibration with third-party navigation software and verified that connection performance between universal clamps/arrays and subject devices was equal to or better than predicates.
Technological Characteristics
Manual surgical instruments. Materials: Stainless steel (17-4PH, custom 455, custom 465, 18-8, 316, 316L, 420, 302), aluminum (6061-T6), plastic (RADEL R5500), and titanium alloy. Interfaces with third-party surgical navigation systems via universal adaptors and clamps. No internal energy source; passive mechanical instruments.
Indications for Use
Indicated for patients undergoing surgical spinal procedures (e.g., spinal fusion) where stereotactic surgery is appropriate and rigid anatomical structures (pelvis or vertebrae) can be referenced against acquired images (CT, MR, 2D/3D fluoroscopy) or anatomical models.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
Predicate Devices
EXPEDIUM and VIPER Navigated Instruments (K120867)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white seal for the Department of Health and Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of three faces in profile, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 24, 2014
DePuy Spine, Incorporated Ms. Kirsten Lehmuller Regulatory Affairs Specialist II 325 Paramount Drive Raynham, Massachusetts 02767
Re: K140927
Trade/Device Name: Universal Navigation Instruments for EXPEDIUM® and VIPER® MIS Spine Systems
Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: Class II Product Code: OLO Dated: September 15, 2014 Received: September 16, 2014
Dear Ms. Lehmuller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K140927
### Device Name
Universal Navigation Instruments for EXPEDIUM® and VIPER® MIS Spine Systems
### Indications for Use (Describe)
The Universal Navigation Instruments for EXPEDIUM® and VIPER® MIS Spine Systems are intended to assist the surgeon in precisely locating anatomical structures in either open, minimally invasive, or percutaneous procedures. These are indicated for use in surgical spinal procedures, in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the pelvis or a vertebrac can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy using tracking arrays provided by the navigation manufacturer. These procedures include but are not limited to spinal fusion. These devices can be pre-calibrated and/or manually calibrated with Brainlab Navigation system, where other navigation systems require manual calibration.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY
B.
C.
#### Submitter Information A.
| Manufacturer: | Medos International Sárl<br>Chemin-Blanc 38<br>2400 Le Locle, Switzerland |
|-------------------------|--------------------------------------------------------------------------------|
| Submitter: | DePuy Spine, Inc.<br>325 Paramount Drive<br>Raynham, MA 02767 |
| Contact Person: | Kirsten Lehmuller<br>325 Paramount Drive<br>Raynham, MA 02767 |
| Telephone number: | 508-828-3291 |
| Fax number: | 508-828-3797 |
| Email: | klehmull@its.jnj.com |
| Date Prepared | April 10, 2014 |
| Trade/Proprietary Name: | Universal Navigation Instruments for EXPEDIUM® and<br>VIPER® MIS Spine Systems |
Common/Usual Name: Stereotaxic Instrument
- Classification Name: Class II, per 21 CFR §882.4560 OLO; Orthopedic
#### D. Predicate Device Name
Trade name: EXPEDIUM and VIPER Navigated Instruments (K120867) Synthes Navigable Pedicle Preparation Instruments (K122211) Brainlab VectorVision Fluoro 3D System (K070106) Medtronic StealthStation (K050438)
{4}------------------------------------------------
#### Device Description E.
The Universal Navigation Instruments for EXPEDIUM® and VIPER® MIS Spine Systems are manual surgical instruments which are designed to interface with previously cleared surgical navigation systems. The instruments in this system are manually calibrated to previously cleared surgical navigation systems using manufacturers' instructions. These instruments are intended to be used in spine applications to perform general manual functions within the orthopedic surgical environment. The Universal Navigation Instruments for EXPEDIUM® and VIPER® MIS Spine Systems are intended for use with the EXPEDIUM 5.5 Spine System and VIPER 2 MIS Spine Systems components.
#### F. Intended Use
The Universal Navigation Instruments for EXPEDIUM® and VIPER® MIS Spine Systems are intended to assist the surgeon in precisely locating anatomical structures in either open, minimally invasive, or percutaneous procedures. These are indicated for use in surgical spinal procedures, in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the pelvis or a vertebrae can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy using tracking arrays provided by the navigation manufacturer. These procedures include but are not limited to spinal fusion. These devices can be pre-calibrated and/or manually calibrated with Brainlab Navigation system, where other navigation systems require manual calibration.
### G. Summary of Similarities and Differences in Technological Characteristics, Performance and Intended Use
The design features, materials, and indications for use of the subject Universal Navigation Instruments are substantially equivalent to the predicate devices identified.
#### H. Materials
The Universal Navigation Instruments are manufactured from stainless steel: 17-4PH, custom 455, custom 465, 18-8, 316, 316L, 420, 302, aluminum: 6061-T6, plastic: RADEL R5500, and titanium alloy.
#### l. Performance Data
DePuy Spine performed verification and validation activities for the Universal Navigation Instruments for EXPEDIUM and VIPER MIS Spine Systems. The Universal Navigation Instruments for EXPEDIUM and VIPER MIS Spine Systems were tested in a controlled
{5}------------------------------------------------
bench top environment in order to compare the overall accuracy to predicate instruments. Testing with third party navigation software was completed to validate calibration with the use of the universal adaptor and the universal clamp attachment. All the subject instruments were successfully calibrated using the third party software according to the instructions provided by the software manufacturer.
In order to assess the rigidity of the connection between the universal clamps and arrays and the subject devices, DePuy Spine performed verification testing in a controlled bench-top environment. For each third-party manufacturer, all combinations of connections were evaluated under expected loading conditions and compared to predicate instruments. All connections performed equal to or better than the predicate instrument.
#### J. Conclusion
The Universal Navigation Instruments for EXPEDIUM® and VIPER® MIS Spine Systems intended use, principles of operation, and technological characteristics are substantially equivalent to those predicates identified.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.