The DeVilbiss DV5MB SmartLink II System can only be used in conjunction with the DeVilbiss DV51, DV53, DV54, DV55, DV56, and DV57 Series CPAP Systems for follow up of obstructive sleep apnea in patients weighing above 30 kg on nasal CPAP therapy. The system is to be used in home and healthcare environments.
Device Story
Accessory module for DeVilbiss IntelliPAP DV5x series CPAP devices; collects usage/performance data (compliance, daily performance, Rx changes, oximetry, fault logs) from CPAP; enables data download to PC software for storage, viewing, and reporting. Physically/electrically attaches to CPAP; powered by CPAP (12 VDC). Adds Bluetooth v2.1 + EDR wireless communication (max range 10m) to existing wired RS232/memory card data transfer methods. Used in home or healthcare settings by patients or clinicians. Healthcare providers use PC software to review data for therapy management. Benefits include remote/convenient access to CPAP performance data to monitor therapy adherence and efficacy.
Clinical Evidence
No clinical testing or clinical performance evaluations were conducted. Substantial equivalence is supported by bench testing and compliance with applicable safety and performance standards.
Technological Characteristics
Enclosure: GE Plastics Cycoloy CX2244ME PC/ABS (UL94 V-0). Connectivity: Bluetooth v2.1 + EDR (max +2.5 dBm, 10m range), wired RS232, memory card. Power: 12 VDC from host CPAP. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, ISO 14971, EN55011, FCC Part 15.
Indications for Use
Indicated for follow-up of obstructive sleep apnea in patients >30 kg on nasal CPAP therapy, using DeVilbiss DV51, DV53, DV54, DV55, DV56, or DV57 Series CPAP systems.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure with outstretched arms, possibly representing care and protection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 4, 2014
DeVilbiss Healthcare LLC Ms. Betty Miller Regulatory/Compliance Manager 100 DeVilbiss Drive Somerset, PA 15501
Re: K140880
Trade/Device Name: DeVilbiss DV5MB Smartlink II System Regulation Number: 21 CFR 868.5905 Regulation Name: Ventilator, Non-Continuous Respirator Regulatory Class: II Product Code: BZD Dated: July 1, 2014 Received: July 3, 2014
Dear Ms. Miller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
Indications for Use
510(k) Number: (if known):_
Device Name: DeVilbiss DV5MB Smartlink II Svstem
The DeVilbiss DV5MB SmartLink II System can only be used in conjunction with the DeVilbiss DV51, DV53, DV54, DV55, DV56, and DV57 Series CPAP Systems for follow up of obstructive sleep apnea in patients weighing above 30 kg on nasal CPAP therapy. The system is to be used in home and healthcare environments.
| | Prescription Use _____________________________________________________________________________________________________________________________________________________________ |
|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Image /page/2/Picture/7 description: The image shows the FDA logo. The logo is a stylized version of the letters "FDA". The letters are white and are outlined in a slightly darker shade of white. The logo is simple and modern.
Anya C. Harry -S 2014.08.01 14:18:20 -04'00'
Page of of a
{3}------------------------------------------------
## Section 5.0 510(k) Summary
# Administrative Information and Device Identification
| Name and address of the manufacturer and<br>sponsor of the 510(k) submission: | DeVilbiss Healthcare LLC<br>100 DeVilbiss Drive<br>Somerset, PA 15501 |
|-------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FDA registration number of the manufacturer<br>of the new device: | 2515872 |
| Official contact person for all<br>correspondence: | Betty Miller<br>Regulatory/Compliance Manager<br>DeVilbiss Healthcare LLC<br>100 DeVilbiss Drive<br>Somerset, PA 15501<br>Phone: 814-443-7606<br>Fax: 814-443-7575<br>Email: betty.miller@devilbisshc.com |
| Date Prepared: | 03/21/2014 |
| Device Name: | DeVilbiss DV5MB SmartLink II |
| Proprietary name of new device: | DeVilbiss DV5MB SmartLink II System |
| Common or usual name of the device: | SmartLink II |
| DeVilbiss Model Number | DV5MB |
| Classification of the predicate device: | Class II |
| Classification of new device: | Class II |
| Classification Panel: | Anesthesiology |
| Panel Code: | BZD |
| CFR Regulation Number: | 21 CFR 868.5905 Ventilator, non-<br>continuous respirator |
| Predicate Device Name(s) and 510(k)<br>number(s): | DeVilbiss DV5M SmartLink System<br>(K082209) |
{4}------------------------------------------------
#### Description of Device:
The proposed DeVilbiss DV5MB SmartLink II System is an accessory to DeVilbiss IntelliPAP DV5x Series CPAPs, models DV51, DV53, DV54, DV55, DV56 and DV57. The proposed DeVilbiss DV5MB SmartLink II System connects physically and electrically to a DV5x Series CPAP for the purpose of collecting usage and performance data from the CPAP for subsequent download to PC software for storage, viewing and reporting.
The proposed DeVilbiss DV5MB SmartLink II System is identical to the predicate device with the addition of a Bluetooth Wireless connection. The Bluetooth radio used is a complete Bluetooth module with built in antenna and Bluetooth v2.1 + EDR. The maximum output power of the Bluetooth radio is +2.5 dBm, has a maximum range of 10 meters. The DeVilbiss DV5MB SmartLink II System wireless feature can be used in the home or in a healthcare setting.
#### Comparison of Device Technological Characteristics to Predicate Devices:
The proposed DeVilbiss DV5MB SmartLink II Module has the following similarities to those which previously received 510(k) concurrence:
- Has the same intended use
- Uses the same operating principle
- Incorporates similar materials ●
The only modifications that were made are:
- . Added Bluetooth Wireless communication interface
| Feature | Predicate Device:<br>DeVilbiss DV5M SmartLink<br>System (K082209) | Modified Device:<br>DeVilbiss DV5MB SmartLink<br>II System | Nature of<br>Change |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use<br>Statement | The DeVilbiss DV5M<br>SmartLink System in<br>conjunction with a DeVilbiss<br>CPAP Unit is intended for use<br>in treating obstructive sleep<br>apnea in patients 30 Kg and<br>above. The system is to be<br>used in home and healthcare<br>environments. | The DeVilbiss DV5MB<br>SmartLink II System in<br>conjunction with a DeVilbiss<br>CPAP Unit is intended for use<br>in treating obstructive sleep<br>apnea in patients 30 Kg and<br>above. The system is to be<br>used in home and healthcare<br>environments. | Same as<br>predicate. |
| Feature | Predicate Device:<br>DeVilbiss DV5M SmartLink<br>System (K082209) | Modified Device:<br>DeVilbiss DV5MB SmartLink<br>II System | Nature of<br>Change |
| Indications for<br>Use Statement | The DeVilbiss DV5M<br>SmartLink System can only<br>be used in conjunction with<br>the DeVilbiss DV51, and<br>DV54 Series CPAP Systems<br>for follow up of obstructive<br>sleep apnea in patients<br>weighing above 30 kg on<br>nasal CPAP therapy | The DeVilbiss DV5MB<br>SmartLink II System can<br>only be used in conjunction<br>with the DeVilbiss DV51,<br>DV53, DV54, DV55, DV56,<br>and DV57 Series CPAP<br>Systems for follow up of<br>obstructive sleep apnea in<br>patients weighing above 30<br>kg on nasal CPAP therapy.<br>The system is to be used in<br>home and healthcare<br>environments. | Add<br>compatibility<br>with additional<br>DV5x Series<br>CPAP models.<br>No impact on<br>safety or<br>effectiveness. |
| Compliance<br>recording | Records 3 years of "while<br>breathing" ON/OFF<br>compliance, 120 days of Daily<br>Performance, Rx change,<br>approximately 72 hours of<br>Oximetry, and Fault logs | Records 3 years of "while<br>breathing" ON/OFF<br>compliance, 120 days of Daily<br>Performance, Rx change,<br>approximately 72 hours of<br>Oximetry, and Fault logs | Same as<br>predicate. |
| User interface | The DeVilbiss DV5M<br>SmartLink Module is<br>designed to mechanically and<br>electrically attach to a DV5x<br>series CPAP. The module<br>uses the onboard Keypad/<br>Display of the DV5x CPAP as<br>the user interface. The user<br>interface is located on the top<br>of the DV5x CPAP and<br>consists of a 6 button keypad<br>and a 2x16 LCD Display. | The DeVilbiss DV5MB<br>SmartLink II System is<br>designed to mechanically and<br>electrically attach to a DV5x<br>series CPAP. The module<br>uses the onboard Keypad/<br>Display of the DV5x CPAP as<br>the user interface. The user<br>interface is located on the top<br>of the DV5x CPAP and<br>consists of a 6 button keypad<br>and a 2x16 LCD Display. | Same as<br>predicate. |
| Oximeter<br>Function | The DeVilbiss DV5M<br>SmartLink Module uses an<br>optional Nonin oximeter to log<br>SpO2 and Pulse Rate at 4-<br>second intervals. | The DeVilbiss DV5MB<br>SmartLink II System uses an<br>optional Nonin oximeter to log<br>SpO2 and Pulse Rate at 4-<br>second intervals. | Same as<br>predicate. |
| Feature | Predicate Device:<br>DeVilbiss DV5M SmartLink<br>System (K082209) | Modified Device:<br>DeVilbiss DV5MB SmartLink<br>II System | Nature of<br>Change |
| Software<br>Function | The SmartLink software is<br>used to display data as it is<br>recorded from the module.<br>The software stores data in a<br>data based structure and has<br>querying and reporting<br>functionality. | The SmartLink software is<br>used to display data as it is<br>recorded from the module.<br>The software stores data in a<br>data based structure and has<br>querying and reporting<br>functionality. (See Section<br>16.0 software section) | Improvements<br>to user interface<br>and<br>performance.<br>No impact on<br>safety or<br>effectiveness. |
| Remote settings<br>change function | CPAP Settings can be<br>changed through an SD card,<br>settings are programmed<br>through SmartLink Desktop<br>software. | CPAP Settings can be<br>changed through an SD card,<br>settings are programmed<br>through SmartLink Desktop<br>software. | Same as<br>predicate. |
| Remote control<br>function | Wired RS232 connection to a<br>distance of up to 200 feet. | Wired RS232 connection to a<br>distance of up to 200 feet.<br>Wireless Bluetooth<br>connection to a distance of up<br>to 33 feet. | Added<br>Bluetooth<br>wireless<br>communication,<br>operates same<br>as wired RS232<br>connection.<br>No impact on<br>safety or<br>effectiveness. |
| Means of Data<br>Transfer | Memory card or<br>wired RS232 connection | Memory card or<br>wired RS232 connection | Same as<br>predicate. |
| Environmental<br>Specifications | Operating temperature:<br>5 to +40 °C.<br>Operating humidity range: 0<br>to 95% R.H. non condensing<br>Storage temperature range:<br>-40 to +70 °C<br>Storage humidity range: 0 to<br>95% R.H. non-condensing | Operating temperature:<br>5 to +40 °C.<br>Operating humidity range: 0<br>to 95% R.H. non condensing<br>Storage temperature range:<br>-40 to +70 °C<br>Storage humidity range: 0 to<br>95% R.H. non-condensing | Same as<br>predicate. |
| Power<br>Requirements | The DeVilbiss DV5M<br>SmartLink Module is not<br>equipped with an on-board<br>power supply. It only<br>receives its power (12 VDC)<br>from the DV5x Series CPAP<br>device during operation. | The DeVilbiss DV5MB<br>SmartLink II System is not<br>equipped with an on-board<br>power supply. It only<br>receives its power (12 VDC)<br>from the DV5x Series CPAP<br>device during operation. | Same as<br>predicate. |
| Feature | Predicate Device:<br>DeVilbiss DV5M SmartLink<br>System (K082209) | Modified Device:<br>DeVilbiss DV5MB SmartLink<br>II System | Nature of<br>Change |
| 510(k) Product<br>Code | BZD (Ventilator, non-<br>continuous (respirator)) | BZD (Ventilator, non-<br>continuous (respirator)) | Same as<br>predicate. |
| Materials<br>comparison | | | |
| Enclosure | GE Plastics Cycoloy<br>CX2244ME PC/ABS Blend<br>FR30U - Meets UL94 V-0<br>flame retardant spec. (P/N:<br>420-0200-004) | GE Plastics Cycoloy<br>CX2244ME PC/ABS Blend<br>FR30U - Meets UL94 V-0<br>flame retardant spec. (P/N:<br>420-0200-004) | Same as<br>predicate. |
Comparison of Similarities and Differences:
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
Statement of Intended Use:
The DeVilbiss DV5MB SmartLink II System can only be used in conjunction with the DeVilbiss DV51, DV53, DV54, DV55, DV56, and DV57 Series CPAP Systems for follow up of obstructive sleep apnea in patients weighing above 30 kg on nasal CPAP therapy. The system is to be used in home and healthcare environments.
### Non-Clinical Testing:
This device has been tested to appropriate ISO and IEC standards and other applicable requirements passing all test protocols. The proposed DeVilbiss DV5MB SmartLink II has been fully tested to demonstrate compliance with the applicable sections of the following standards:
- 1. IEC 60601-1-1:2000 Medical Electrical Equipment—Part 1: General Requirements for Safety; Safety Requirements for Medical Electrical Systems
- 2. IEC 60601–1–2:2007 Ed 3.0 Medical Electrical Equipment—Part 1: General Requirements for Safety; Electromagnetic Compatibility Requirements and Tests
- 3. IEC 60601-1-4:2000 Consol. Ed. 1.1. Medical electrical equipment Part 1-4: General requirements for safety -- Collateral standard: Programmable electrical medical systems, edition 1.1.
- 4. ISO 14971:2012 Medical devices Application of risk management to medical devices
- 5. EN55011:2007 Industrial, scientific and medical (ISM) radio-frequency equipment -Radio disturbance characteristics - Limits and methods of measurements
- 6. IEC 529 (1989): Classification of Degrees of Protection Provided by Enclosures
- 7. IEC 68 (1988): Environmental Testing
{8}------------------------------------------------
- 8. ISTA Procedure 3A Package –Products for Parcel Delivery System Shipments 70kg (150lb) or less
- 9. FCC Requlation 47 CFR Part 15 Radio Frequency Devices, Class B
Additional design considerations were applied based on published quidance and draft quidance documents from FDA:
- 1. FDA Reviewer Guidance for Premarket Notification Submissions (Anesthesiology and Respiratory Device Branches, November 1993)
- 2. DRAFT FDA Guidance Design Considerations for Devices Intended for Home Use Dec 12- 2012
- 3. DRAFT FDA Guidance Cybersecurity in Medical Devices 6-13-13
- 4. FDA Guidance Radio Frequency Wireless Technology in Medical Devices Aug 14-2013
The proposed DeVilbiss DV5MB SmartLink II System in conjunction with a DeVilbiss CPAP Unit meets the required performance criteria and functioned as intended. See Section 12.0 and Section 9.0.
#### Clinical Testing:
No clinical testing or clinical performance evaluations were conducted on the proposed DeVilbiss DV5MB SmartLink II relating to clinical studies to demonstrate substantial equivalence to the predicate device.
#### Conclusion:
The proposed DeVilbiss DV5MB SmartLink II System is substantially equivalent to the predicate DV5M SmartLink Compliance Module (K082209) and the device, as changed, does not raise any new issues of safety and effectiveness.
Analysis of comparison of design, function and features of the proposed DeVilbiss DV5MB SmartLink II System to the predicate DV5M SmartLink Compliance Module (K082209), together with the results of testing demonstrate the device to be substantially equivalent to the predicate devices in terms of meeting performance criteria and functioning as intended.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.