K140825 · Del Medical, Inc. · MQB · Jul 17, 2014 · Radiology
Device Facts
Record ID
K140825
Device Name
DELWORKS DR SYSTEM
Applicant
Del Medical, Inc.
Product Code
MQB · Radiology
Decision Date
Jul 17, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
DelWorks DR System is intended for digital image capture use in general radiographic examinations, wherever conventional screen-film systems may be used, excluding fluoroscopy, angiography and mammography. DelWorks DR System allows imaging of the skull, chest, shoulders, spine, abdomen, pelvis, and extremities.
Device Story
Digital radiography system integrating flat panel detectors (FPD) with acquisition software; replaces conventional film systems. Inputs: X-ray signals captured by FPDs. Processing: Digital Radiography Operator Console (DROC) software controls technique factors via serial interface; provides image enhancement and stitching (combining 4 views into one). Output: DICOM images transmitted to PACS/radiology information systems. Used in clinical settings by radiology staff. Features include auto-exposure detection and optional synchronization modules for high-voltage generators. Wireless connectivity available for specific detectors. Benefits: filmless environment, improved cost-effectiveness, and PACS integration.
Clinical Evidence
Clinical images were acquired and evaluated by a board-certified radiologist. The study concluded that the images were of good diagnostic quality and demonstrated equal or better image quality compared to the predicate device.
Technological Characteristics
Digital radiography system using flat panel detectors (FPD). Connectivity via Ethernet or Wi-Fi. Power source: AC line. Software: DROC. Compliance: IEC 60601-1 (Electrical Safety), IEC 60601-1-2 (EMC). Features auto-exposure detection and image stitching. DICOM output.
Indications for Use
Indicated for digital image capture in general radiographic examinations (skull, chest, shoulders, spine, abdomen, pelvis, extremities) where conventional screen-film systems are used. Excludes fluoroscopy, angiography, and mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Sedecal Digital Radiographic Upgrade, Model: SDRU-T (K130883)
Submission Summary (Full Text)
{1}------------------------------------------------
'Page 1 of 2
JUL 1 7 2014
# 510(k) Summary K140825 Submitted by Del Medical, Inc. 241 Covington Drive Bloomingale, IL 60108 (Tel) 800.800.6006 (Fax) 847.288.7011 Prepared by: Marc Lorenzo, Vice President Date Prepared: June 4, 2014
- 1. Identification of the Device: Proprietary-Trade Name: DelWorks DR System Classification Name: Stationary X-Ray System, Product Code MQB, Regulation 892.1680 Common/Usual Name: Digital X-Ray Receptor Panel
- 2. Equivalent legally marketed device: Sedecal Digital Radiographic Upgrade, Model: SDRU-T, K130883
- 3. Indications for Use (intended use) DelWorks DR System is intended for digital image capture use in general radiographic examinations, wherever conventional screen-film systems may be used, excluding fluoroscopy, angiography and mammography. DelWorks DR System allows imaging of the skull, chest, shoulders, spine, abdomen, pelvis, and extremities.
- 4. Description of the Device: The DelWorks DR System represents the straightforward integration of a number of digital x-ray receptor panels using digital acquisition software. The DelWorks DR System is compatible with all modern HF generators. DelWorks DR System is a Digital Radiography system, featuring an integrated flat panel digital detector (FPD) The DelWorks DR System is designed to perform digital radiographic examinations as a replacement for conventional film. This integrated platform provides the benefits of PACS with the advantages of digital radiography for a filmless environment and improves cost effectiveness. Many of the compatible panels provide Auto-exposure detection. A synchronization module is available where exposure sync is required. This module connects between the digital panel and the x-ray high voltage generator. It provides "ready" synchronization and optical isolation between the two devices. Some of the detectors feature wireless Wi-Fi connection via a wireless access point. The software features Image Stitching (4 views can be combined into one single view) and other image enhancement features. The system software interacts with and controls technique factors of compatible generators via serial interfaces. DICOM output allows for transmission of images to the hospital radiology information system. The system employs Digital Radiography Operator Console (DROC) software, outputs a DICOM image. A USB "dongle" is provided to assure the authenticity of the software.
- 5. Safety and Effectiveness, comparison to predicate device. The results of clinical image inspection, bench, and test laboratory results indicates that the new device is as safe and effective as the predicate devices. Clinical images collected demonstrate equal or better image quality as compared to our predicates.
{2}------------------------------------------------
#### Substantial Equivalence Chart 6.
| Characteristic | Sedecal Digital Radiographic Upgrade, Model: SDRU-T, K130883 | DelWorks DR System K140825 |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: | Sedecal Digital Radiographic Upgrade Model SDRU-T is intended for digital image capture use in general radiographic examinations, wherever conventional screen film systems may be used, excluding fluoroscopy, angiography and mammography. The kit allows imaging of the skull, chest, shoulders, spine, abdomen, pelvis, and extremities. | DelWorks DR System is intended for digital image capture use in general radiographic examinations, wherever conventional screen-film systems may be used, excluding fluoroscopy, angiography and mammography. DelWorks DR System allows imaging of the skull, chest, shoulders, spine, abdomen, pelvis, and extremities. |
| Configuration | This submission is for the Digital Panel and Software only, no generator or stand provided. | This submission is for the Digital Panel and Software only, no generator or stand provided. |
| Digital Panels | Toshiba FDX4343R Detector<br>Toshiba FDX3543RP Detector | Toshiba FDX4343R<br>Toshiba FDX3543RP<br>Thales Pixium Portable 2430EZ*<br>Thales Pixium Portable 3543EZ*<br>Konica Aero DR 14x17 Detector*<br>Konica Aero DR 17x17 Detector*<br>Fuji FDR D-EVO G35s Detector<br>Fuji FDR D-EVO G35i Detector*<br>Fuji FDR D-EVO G43i Detector*<br>Fuji FDR D-EVO C24i Detector*<br>Fuji FDR D-EVO C35i Detector*<br>Fuji FDR D-EVO C43i Detector*<br>Varian PaxScan 4336R<br>Varian PaxScan 4343R<br>Thales Pixium RAD 4143<br>Thales Pixium RAD 4343<br>*Wireless Wi-Fi |
| Software | Digital Radiography Operator Console (DROC) software, outputs a DICOM image | SAME as K130883,<br>outputs a DICOM image. |
| DICOM | Yes | Yes |
| Interface | Ethernet | Ethernet or Wi-Fi |
| Power source | AC Line | AC Line |
| Electrical safety and EMC | Electrical Safety per IEC 60601-1 and EMC per IEC 60601-1-2. | Electrical Safety per IEC 60601-1 and EMC per IEC 60601-1-2. |
- Summary of Bench Testing Conducted: IEC Standards were employed for: Electrical Safety 7. and Electromagnetic Compatibility. MTF and DQE measurements were supplied by the panel . manufacturer in accordance with the FDA guidance document. Risk Analysis was conducted in accordance with FDA guidance documents. Software validation was performed using all panels. All of the digital panels have had previous successful 510(k) reviews.
- Summary of Clinical Testing: Clinical images were acquired and evaluated by a board certified 8. radiologist who concluded the images were of good diagnostic quality.
- ರು. Conclusion: After analyzing bench, clinical image, and external laboratory testing to applicable standards, it is the conclusion of Del Medical Inc that the DelWorks DR System is as safe and effective as the predicate device, have few technological differences, and has the same indications for use, thus rendering it substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 17, 2014
Del Medical, Inc. % Mr. Daniel Kamm Principle Engineer 8870 Ravello Court NAPLES FL 34114
Re: K140825
Trade/Device Name: DelWorks DR System Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: June 12, 2014 Received: June 16, 2014
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
ff your device is classified (see above) into cither class 11 (Special Controls) or class 111 (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2-Mr. Kamm
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.goy/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Sm.7)
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known) K140825
#### Device Name Del Works DR System
#### Indications for Use (Describe)
The DelWorks DR System is intended for digital image capture use in general radiographic examinations, wherever conventional screen-film systems may be used, excluding fluoroscopy, angiography and mammography. DelWorks DR System allows imaging of the skull, chest, shoulders, spine, abdomen, pelvis. and extremities.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FOA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{6}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.