Software as a Medical Device, Pediatric, 3rd-Party Reviewed
Indications for Use
Q-Station is application software intended to manage, view, analyze, and report qualitative and quantitative image data from ultrasound exams. It is designed to host optional advanced analysis applications via QLAB integration and provides integrated tools that allow users to manually assess and score cardiac wall motion and export images and/or exams and reports. Q-Station can view DICOM images of non-ultrasound images such as CT, MR, NM, CR, MG, XA, PET, RT, and X-Ray modalities for reference viewing. It supports connectivity to ultrasound systems, PACS and other DICOM storage repositories.
Device Story
Q-Station is PC-based application software for post-acquisition management of ultrasound data; integrates with QLAB for advanced analysis. Inputs include ultrasound images/measurements from acquisition devices and non-ultrasound DICOM images (CT, MR, etc.) for reference. Clinicians use software to review exams, perform manual cardiac wall motion scoring, create/edit measurements, and generate reports. Supports import/export of exams and anonymization. Facilitates clinical decision-making by providing tools for quantitative/qualitative assessment of cardiac and vascular data. Used in clinical settings by healthcare professionals; enables sharing of interpretive findings and reports across clinical teams.
Clinical Evidence
No clinical studies were required to support substantial equivalence. Evidence consists of bench testing, including system-level performance tests, risk analysis, and verification and validation of software functionality.
Technological Characteristics
Application software for PC; DICOM compatible. Features multi-modality viewing (1-up/n-up) and integrated analysis packages (Adult/Pediatric Echo, Vascular). Connectivity via DICOM to ultrasound systems and PACS. Manual measurement tools for labels, groups, and collections. Software-based; no specific hardware requirements listed beyond PC platform.
Indications for Use
Indicated for clinicians requiring management, viewing, analysis, and reporting of qualitative and quantitative ultrasound image data. Supports adult echo, pediatric echo, and vascular applications. Used for reference viewing of non-ultrasound DICOM images (CT, MR, NM, CR, MG, XA, PET, RT, X-Ray). Prescription use only.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
# K140808 APR 1 7 2014
# 5. 510(k) Summary
This summary of safety and effectiveness is provided as part of the Premarket Notification in compliance with 21CFR. Part 807, Subpart E, Section 807.92
1) Submitter's name, address, telephone number, contact person Penny Greco, Regulatory Affairs Specialist Philips Healthcare, Ultrasound 3000 Minuteman Road Andover, MA Telephone: (978) 659-4615 Facsimile: (978) 975-7324 E-mail: penny.greco@philips.com
Date prepared: March 14, 2014
2) Name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known:
Common/Usual Name: Picture Archiving and Communications Systems Workstation
Proprietary Name: Q-Station
CFR 892.2050, system, image processing, radiological, Product Classification Name: code LLZ, Class II
#### 3) Substantially Equivalent Devices
Philips Ultrasound believes that modifications introduced with Q-Station 3.0 are substantially equivalent to other commercially available devices including Philips Xcelera (K061995) and Philips EPIQ (K132304). The Q-Station analysis packages and multi-modality viewing are the same as those found on Philips ultrasound systems, including EPIQ.
#### 4) Device Description
Q-Station is designed to manage post-acquisition ultrasound images and other data, for the purposes of diagnosing the patient's condition. This includes using Q-Station on a PC to review
{2}------------------------------------------------
images and measurements sent from an ultrasound acquisition device, analyze 3D and other data with QLAB. Q-Station is used to review various ultrasound exam types, including Adult echo, General Imaging, Stress echo, Vascular, and TEE. In addition, Q-Station can be used for reference viewing of non-ultrasound DICOM images. Q-Station can be used to add interpretive findings, key images, measurements and calculations and other comments that create reports that can be shared with other clinicians. During this review, users may also use Q-Station to import and export exams, print reports, and anonymize images for export. Q-Station supports QLAB Q-Apps for advanced analysis (K132165).
#### 5) Intended Use
O-Station is application software intended to manage, view, analyze, and report qualitative and quantitative image data from ultrasound exams. It is designed to host optional advanced analysis applications via QLAB integration and provides integrated tools that allow users to manually assess and score cardiac wall motion and export images and/or exams and reports. Q-Station can view DICOM images of non-ultrasound images such as CT, MR, NM, CR, MG, XA, PET, RT, and X-Ray modalities for reference viewing. It supports connectivity to ultrasound systems, PACS and other DICOM storage repositories.
#### 6) Technological comparison to predicate devices
Philips Q-Station 3.0 is application software similar to Q-Station as submitted in K103815. It is designed to manage, view and report image data acquired by Ultrasound systems, but with Q-Station 3.0 now includes analysis packages and multi-modality viewing similar to Philips Xcelera and Philips EPIQ.
Both Q-Station and Xcelera are software applications for off-cart image analysis and patient report preparation. Both are DICOM compatible.
The integrated analysis packages included in Q-Station 3.0 are essentially the same as those included with the EPIQ ultrasound system (K132304), except that the Q-Station analysis packages are limited to Adult Echo, Pediatric Echo and Vascular applications. The Analysis packages are also similar to Philips Xcelera analysis packages (K061995).
The Q-Station 3.0 multi-modality viewing allows users to view non ultrasound images such as CT, MR, NM, CR, MG, XA, PET, RT, and X-Ray modalities for reference viewing. These images can be viewed in 1-up or in n-up for comparison. In cases where the images are present in stack format, O-Station viewer will allow the user to navigate the frames/slices of these stacks and analyze them one by one. No modification is allowed on multi-modality images/studies, nor is reporting allowed on such studies. The Q-Station Multi-modality viewing is similar to that of both EPIQ and Xcelera.
{3}------------------------------------------------
| Model | Philips<br>Q-Station | Philips<br>Q-Station 3.0 | Philips<br>Xcelera | Philips<br>EPIQ<br>Ultrasound System |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)<br>Reference | K103815 | Subject submission | K061995 | K132304 |
| Product Code | LLZ | LLZ | LLZ | IYN, IYO, ITX |
| Intended Use | Q-Station software is<br>a software<br>application package.<br>It is designed to<br>manage, view and<br>report image data<br>acquired by<br>Ultrasound systems<br>and cardiac<br>waveformn data<br>from Philips<br>StressVue<br>EGG systems. Q-<br>Station offers<br>support for QLAB<br>plug-ins for analysis,<br>quantification and<br>reporting of data<br>from ultrasound<br>systems. | Q-Station is application<br>software intended to manage,<br>view, analyze, and report<br>qualitative and quantitative<br>image data from ultrasound<br>exams. It is designed to host<br>optional advanced analysis<br>applications via QLAB<br>integration and provides<br>integrated tools that allow<br>users to manually assess and<br>score cardiac wall motion<br>and export images and/or<br>exams and reports. Q-<br>Station can view DICOM<br>images of non-ultrasound<br>images such as CT, MR, NM,<br>CR, MG, XA, PET, RT, and<br>X-Ray modalities for<br>reference viewing. It supports<br>connectivity to ultrasound<br>systems, PACS and other<br>DICOM storage repositories. | An integrated<br>multimodality image and<br>information system,<br>designed to perform the<br>necessary functions<br>required for<br>import/export/storage/<br>archiving/review/<br>analysis/quantification/<br>reporting and database<br>management of digital<br>cardiovascular images.<br>waveforms and data related<br>to cardiology. | The EPIQ Diagnostic<br>Ultrasound System is a<br>general purpose,<br>software controlled<br>diagnostic ultrasound<br>system. Its function is<br>to acquire ultrasound<br>data and to display the<br>data in a various modes<br>of operation. |
| Analysis<br>Packages | No | Adult Echo<br>Pediatric Echo<br>Vascular | Xcelera packages include:<br>Adult Echo<br>Pediatric Echo<br>Vascular | EPIQ packages include:<br>Adult Echo<br>Pediatric Echo<br>Vascular |
| Measurement<br>Tool | No | View, copy, and edit system-<br>defined measurement labels,<br>groups, and collections.<br>Create, edit, and delete<br>customized measurement<br>labels, groups, and<br>collections. | Create, edit, delete<br>measurements &<br>calculations and re-label<br>existing measurements &<br>calculations | Same as Q-Station 3.0 |
| Viewer for<br>other<br>Modalities | No | CT, MR, NM, CR, MG, XA,<br>PET, RT, and X-Ray images | Same + US | Same as Q-Station 3.0 |
.
{4}------------------------------------------------
#### 7) Non-clinical performance data
No performance standards for PACS systems or components have been issued under the authority of Section 514. O-Station 3.0 was tested in accordance with Philips verification and validation processes. Quality assurance measures were applied to the system design and development, including:
- . Risk Analysis
- . Product Specifications
- Design Reviews .
- . Verification & Validation
#### Summary of Clinical Tests:
The subject of this premarket submission, Q-Station 3.0 software did not require clinical studies to support substantial equivalence.
#### Conclusion:
Verification and Validation activities required to establish the performance, functionality, and reliability characteristics of Q-Station 3.0 were performed. Testing involved system level tests, performance tests, and safety testing from risk analysis. Testing performed demonstrated that the O-Station 3.0 meets all defined reliability requirements and performance claims.
#### 8) General Safety and Effectiveness Concerns
The device labeling contains operating instructions for the safe and effective use of the Q-Station 3.0 software.
#### 9) Conclusions
The Q-Station software is designed and manufactured to meet United States and international standards for the display and reporting of qualitative and quantitative image data acquired on Phillips Ultrasound devices. The system is designed to incorporate components common to all image viewing systems for the display, manipulation and quantification tasks within a clinical setting. The Q-Station software incorporates features of predicate devices cleared through premarket notification and no new issues of safety or effectiveness are raised.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the department's name around the perimeter. Inside the circle is an abstract symbol that represents the department's mission related to health and human services. The symbol is composed of three curved lines that form a stylized human figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 17, 2014
Philips Ultrasound, Inc. % Mr. Mark Job Responsible Third Party Official 1394 25th Street NW Buffalo, MN 55313
Re: K140808
Trade/Device Name: Q-station Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: March 31, 2014 Received: April 1, 2014
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2-Mr. Job
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Sm.h.D
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration |
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
#### 510(k) Number (if known)
#### K140808
Device Name
# Q-Station
Indications for Use (Describe)
Q-Station is application software intended to manage, view, and report qualitative and quantitative image data from ultrasound < rans. It is designed to host optional advanced analysis applications via QLAB integration and provides integrated tools that allow users to manually assess and score cardiae wall motion and export images and/or exams and reports. Q-Station can view DICOM images of nonwhansound images such as CT, MR, NM, XA, PET, RT, and X-Ray modalities for reference viewing. It supports connectivity to ultrasound systems, PACS and other DICOM storage repositories.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
# FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
FORM FDA 3881 (1/14)
Page 1 of 1
EF PSC Publishing Services (301) 443-6740
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.