AEROSURE MEDIC

K140772 · Actegy , Ltd. · BWF · Dec 19, 2014 · Anesthesiology

Device Facts

Record IDK140772
Device NameAEROSURE MEDIC
ApplicantActegy , Ltd.
Product CodeBWF · Anesthesiology
Decision DateDec 19, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5690
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Aerosure is indicated for use as a Positive Expiratory Pressure (PEP) Device. - The use of Aerosure improves clearance of secretions - The use of Aerosure may reduce the need for postural drainage - Aerosure facilitates opening of airways in patients with Cystic Fibrosis, COPD, asthma, and lung diseases with secretory problems - Aerosure may be used to prevent or reverse atelectasis - Aerosure may also be useful in the removal of mucus from the lungs of patients who have chronic bronchitis or bronchiectasis Aerosure is intended for patients age 21 and above with Cystic Fibrosis, COPD, asthma, and lung diseases with secretory problems, and patients with atelectasis. All patients must be capable of following instructions for Positive Expiratory Pressure Therapy. Aerosure may be used in hospital as well as the home after a period of training.

Device Story

Aerosure Medic is a small, battery-powered, handheld device for airway clearance; utilizes an electrically powered rotary valve mechanism to create alternating airflow occlusion; generates positive expiratory pressure (PEP) and vibration of the respiratory tree; used in hospital or home settings by patients (age 21+) after training; device restricts inhalation/exhalation efforts to loosen mucus; facilitates secretion removal; reduces need for postural drainage; helps prevent/reverse atelectasis; provides therapeutic benefit through improved airway patency.

Clinical Evidence

Bench testing only. Functional testing verified design specifications. Comparative bench testing against predicates demonstrated substantially similar output. Usability testing (summative protocol per FDA human factors guidance) confirmed safe and effective use by target population.

Technological Characteristics

Electrically powered rotary valve mechanism; 3-cell NiMH rechargeable battery (3.6V); plastic housing; dimensions 45x96x149mm; weight 195g; IP45 ingress protection; standards: ISO 10993 (biocompatibility), IEC 60601-1 (safety), IEC 60601-1-2 (EMC).

Indications for Use

Indicated for patients age 21+ with Cystic Fibrosis, COPD, asthma, lung diseases with secretory problems, and atelectasis; aids in mucus clearance and airway opening; requires patient ability to follow PEP therapy instructions.

Regulatory Classification

Identification

An incentive spirometer is a device that indicates a patient's breathing volume or flow and that provides an incentive to the patient to improve his or her ventilation.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 19,2014 Actegy, Ltd. C/O Dr. John J. Smith Hogan Lovells US LLP Regulatory Counsel Columbia Square 555 13th Street, NW Washington, DC 20004 Re: K140772 Trade/Device Name: Aerosure Medic Regulation Number: 21 CFR 868.5690 Regulation Name: Incentive spirometer Regulatory Class: II Product Code: BWF Dated: December 2, 2014 Received: December 2, 2014 # Dear Dr. Smith: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Page 2 - Dr. Smith Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Susan Runno DDS, MA Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {2}------------------------------------------------ 510(k) Number (if known) Device Name Aerosure Medic Indications for Use (Describe) Aerosure is indicated for use as a Positive Expiratory Pressure (PEP) Device. - · The use of Aerosure improves clearance of secretions - · The use of Aerosure may reduce the need for postural drainage - Aerosure facilitates opening of airways in patients with Cystic Fibrosis, COPD, asthma, and lung diseases with secretory problems - · Aerosure may be used to prevent or reverse atelectasis · Aerosure may also be useful in the removal of mucus from the lungs of patients who have chronic bronchitis or bronchiectasis Aerosure is intended for patients age 21 and above with Cystic Fibrosis. COPD, asthma, and lung diseases with secretory problems, and patients with atelectasis. All patients must be capable of following instructions for Positive Expiratory Pressure Therapy. Aerosure may be used in hospital as well as the home after a period of training. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number. {3}------------------------------------------------ # 510(k) SUMMARY ## Actegy's Aerosure Medic # Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared Actegy Ltd. 8 Queen Square, Ascot Business Park Lyndhurst Road Ascot Berkshire SL5 9FE UK | Phone: | +44(0) 1344 636 940 | |------------|---------------------| | Facsimile: | +44(0) 8452 255 612 | Contact Person: Angie Glover Date Prepared: December 18, 2014 ## Name of Device and Name/Address of Sponsor Aerosure Medic Actegy Ltd. 8 Queen Square, Ascot Business Park Lyndhurst Road Ascot Berkshire SL5 9FE UK #### Common or Usual Name Oscillatory positive expiratory pressure devices (OPEP) #### Classification Name Incentive spirometer, 21 C.F.R. §868.5690, Product code BWF Powered percussor, 21 C.F.R. § 868.5665, Product code BYI #### Predicate Devices Flutter D (Clement Clarke), K972859; Acapella (Smiths Medical), K002768; Roadrunner (DHD), K991561. {4}------------------------------------------------ # Intended Use / Indications for Use Aerosure is indicated for use as a Positive Expiratory Pressure (PEP) Device. - · The use of Aerosure improves clearance of secretions - · The use of Aerosure may reduce the need for postural drainage · Aerosure facilitates opening of airways in patients with Cystic Fibrosis, COPD, asthma, and lung diseases with secretory problems - · Aerosure may be used to prevent or reverse atelectasis • Aerosure may also be useful in the removal of mucus from the lungs of patients who have chronic bronchitis or bronchiectasis Aerosure is intended for patients age 21 and above with Cystic Fibrosis, COPD, asthma, and lung diseases with secretory problems, and patients with atelectasis. All patients must be capable of following instructions for Positive Expiratory Pressure Therapy. Aerosure may be used in hospital as well as the home after a period of training. ## Technological Characteristics The Aerosure Medic consists of an electrically powered rotary valve mechanism which acts as a means of restricting the user's inhalation and exhalation efforts and generating vibration of the respiratory tree, thus facilitating the removal of mucus from the lungs. ### Performance Data Biocompatibility (ISO 10993), electromagnetic compatibility (IEC 60601-1-2), medical device safety testing (IEC 60601-1), and other performance testing were conducted. In all instances, the Aerosure Medic functioned as intended and the results observed were as expected. To support the cleaning instructions, cleaning and ingress protection testing was conducted to demonstrate that reasonably expected cleaning of the device would not have an effect of performance or safety. The results verified that there was no depreciation in performance or safety. In addition, in order to verify the Aerosure Medic ingress protection rating of IP45 [IEC60529], testing was conducted to investigate the effects of expected in-use exposure to water. The results verified that there was no water ingress into the device and that the device function was normal after exposure. Actegy conducted functional testing as part of the design verification process which demonstrated the Aerosure Medic device met its specifications. In addition, the company conducted bench testing comparing the performance characteristics for the Aerosure Medic and the legally marketed predicates. The results showed that all of the devices have directly comparable basic characteristics and substantially similar output specifications. Finally, Actegy conducted usability testing to demonstrate the safe use of the device in the target user population, i.e., lay persons. The Aerosure Medic summative protocol design, conduct, and analysis followed FDA's human factors draft guidance for industry entitled, Applying Human Factors and Usability Engineering to Optimize Medical Device Design (June 22, 2011). Based on the findings of this study, it was concluded that Aerosure Medic is reasonably safe and effective for the intended users, uses and use environments. {5}------------------------------------------------ # Substantial Equivalence The Aerosure Medic is substantially equivalent to the predicated devices. The Aerosure Medic has substantially equivalent intended uses, indications, technological characteristics and principles of operation as its predicate devices. The Aerosure Medic differs from the predicate devices in terms of certain technological characteristics (minor differences in construction, battery and motor-driven vs. manual power, minor differences in occlusion mechanism, weight and dimensions). Performance bench test data demonstrate that the Aerosure Medic is substantially equivalent to the predicate devices. Thus, the Aerosure Medic is substantially equivalent. # Conclusion The Aerosure Medic is substantially equivalent to the predicate devices Flutter D (K.972859), Acapella (K002768) and Roadrunner (K991561). {6}------------------------------------------------ | Basic Unit characteristics | | | | | |------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Aerosure Medic | Roadrunner | Flutter D | Acapella | | 510k number | TBD | K991561 | K972859 | K002768, | | Manufacturer | Actegy Ltd. | DHD Healthcare | Clement Clarke<br>(Axcan Scandipharm) | Smiths Medical | | Intended use | Aerosure is indicated for use as a Positive<br>Expiratory Pressure (PEP) Device.<br><br>• The use of Aerosure improves clearance of<br>secretions<br>• The use of Aerosure may reduce the need for<br>postural drainage<br>• Aerosure facilitates opening of airways in<br>patients with Cystic Fibrosis, COPD, asthma,<br>and lung diseases with secretory problems<br>• Aerosure may be used to prevent or reverse<br>atelectasis<br>• Aerosure may also be useful in the removal of<br>mucus from the lungs of patients who have<br>chronic bronchitis or bronchiectasis<br><br>Aerosure is intended for patients age 21 and<br>above with Cystic Fibrosis, COPD, asthma, and<br>lung diseases with secretory problems, and<br>patients with atelectasis. All patients must be<br>capable of following instructions for Positive<br>Expiratory Pressure Therapy.<br><br>Aerosure may be used in hospital as well as the<br>home after a period of training. | 1 - Purpose:<br>The DHD Roadrunner is indicated for use as a<br>Positive Expiratory Pressure (PEP) Device.<br><br>2 - Claims:<br>• The use of DHD Roadrunner improves<br>clearance of secretions<br>• The use of DHD Roadrunner may reduce<br>the need for postural drainage<br>• DHD Roadrunner facilitates opening of<br>airways in patients with Cystic Fibrosis,<br>COPD, asthma, and lung diseases with<br>secretory problems<br>• The DHD Roadrunner may be used to<br>prevent or reverse atelectasis<br><br>3 - Target Patient Population<br>Patients with Cystic Fibrosis, COPD, asthma,<br>and lung diseases with secretory problems, and<br>patients with atelectasis. All patients must be<br>capable of following instructions for Positive<br>Expiratory Pressure Therapy.<br><br>4 - Intended Environment for Use<br>• Labeling reflects the statement: "Federal<br>(USA) Law restricts this device to sale by<br>or on the order of a physician."<br>• May be used in hospital as well as the<br>home after a period of training. | The efficacy of the Flutter 4D as a<br>mucus clearance device for cystic<br>fibrosis patients is based on its<br>ability to<br>1) vibrate the airways (which<br>loosens mucus from the airway<br>walls),<br>2) intermittently increase<br>endobronchial pressure (to<br>maintain the patency of airways<br>during exhalation, so that mucus<br>does not become trapped as it<br>moves up the airways), and<br>3) accelerate expiratory airflow (to<br>facilitate the upward movement of<br>mucus through the airways so<br>that it can be more easily<br>coughed out).<br><br>FlutterD may also be useful in the<br>removal of mucus from the lungs<br>of patients who have chronic<br>bronchitis or bronchiectasis and<br>in conjunction with a medical<br>need for Positive Expiratory<br>Pressure (PEP) Therapy. | Acapella is intended for<br>use as a Positive<br>Expiratory Pressure<br>(PEP) device. It may also<br>be used simultaneously<br>with nebulized aerosol<br>drug delivery. | | Device | Aerosure is a small, battery powered device. It<br>produces a rapidly alternating occlusion of airflow as | Positive Expiratory Pressure and vibration<br>feature | The Flutter®D is a small, single<br>use, hand held device expiration | Acapella is a small, single<br>use, hand held secretion | | Description | a user breathes through the replaceable valve head<br>(containing the powered rotating vane mechanism)<br>which creates a general increase in resistance to<br>breathing and a vibration of the respiratory tree and<br>the associated structures, including the musculature.<br>Frequency is adjusted by switching modes. | Use when both inhale and exhale<br>Adjust frequency with knob<br>Manually powered device. | resistance device. consisting of a<br>hardened plastic mouthpiece at<br>one end, a plastic cover with an<br>opening at the other end, and a<br>valve on the inside created by a<br>high density steel ball resting in a<br>plastic circular cone. Exhalation through the device causes<br>vibrations to loosen mucus from<br>the airways. The resistance is<br>increased by tilting the device | clearance and lung<br>expansion plastic<br>mechanical device that<br>creates vibrating positive<br>expiratory pressure when<br>a patient exhales through<br>the device via a vibrating orifice mechanism. The<br>resistance is adjusted via<br>the rotating dial. | | Software/<br>Firmware/<br>Microprocess<br>or control | Yes | No | No | No | | Power source | 3 cell NiMH rechargeable battery (nominal 3.6V) to<br>drive the motor. The charger operates normally with a<br>power supply range of 110~240 V / 50Hz | Inhalation & Exhalation - manual, mechanical<br>powered device | Exhalation - manual, mechanical<br>powered device | Inhalation & Exhalation -<br>manual, mechanical<br>powered device | | Compliance<br>with<br>Voluntary<br>standards | ISO 10993; IEC 60601-1-2; IEC 60601-1 | Unknown | Unknown | Unknown | | Weight | 195g | Unknown | 86g | 89g | | Dimensions | 45x96x149 | Unknown | 90x56x44 | 172x60x58 | | Housing<br>materials and<br>construction | Plastic mouldings | Unknown | Plastic mouldings | Plastic mouldings | {7}------------------------------------------------
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...